Here, we present a standardized method for measurement of elbow passive kinesthesia using the threshold to detection of passive movement (TDPM) that is appropriate for a research setting.
Proprioception is an important component of controlled movement. The threshold to detection of passive movement (TDPM) is a commonly used method for quantifying the proprioceptive submodality of kinesthesia in research settings. The TDPM paradigm has been found to be valid and reliable; however, the equipment and methods used for TDPM vary between studies. In particular, the research laboratory apparatuses for producing passive movement of an extremity are often custom designed by individual laboratories or inaccessible due to high cost. There is a need for a standardized, valid, and reliable method for measuring TDPM using readily available equipment. The purpose of this protocol is to provide a standardized method for measurement of TDPM at the elbow that is economical, easy to administer, and that produces quantitative results for measurement purposes in research-based settings. This method was tested on 20 healthy adults without neurological impairment, and eight adults with chronic stroke. The results obtained suggest this method is a reliable way to quantify elbow TDPM in healthy adults, and provides initial support for validity. Researchers seeking a balance between equipment affordability and measurement precision are most likely to find this protocol of benefit.
Proprioceptive information is an important contributor to the control of human movement. Proprioceptive deficits accompany a wide range of neurologic conditions such as stroke1,2,3,4,5,6, Parkinson’s disease7, and sensory neuropathies8. Orthopedic injuries such as ligament and muscle tears have also been shown to reduce proprioceptive function9. The construct of proprioception is often tested in clinical outcome measures via detection of provider-applied small alterations in finger or toe position10,11,12,13,14. Such measures produce relatively coarse measurements: “absent”, “impaired”, “normal”12. While sufficient for detection of gross proprioceptive impairments, laboratory mechanical testing methods are required to precisely measure subtle proprioceptive impairments14,15,16.
Researchers and clinicians often divide proprioception into submodalities for measurement. The most commonly investigated submodalities of proprioception are joint position sense (JPS) and kinesthesia, typically defined as the sense of movement3,16,17. Joint position sense is often tested via active matching tasks, where individuals replicate a reference joint angle18,19. Kinesthesia is commonly measured using the threshold to detection of passive movement (TDPM), whereby a participant’s limb is passively moved slowly, with the participant indicating the point at which movement is first detected16,17,19. Measurement of TDPM typically requires use of specialized equipment to provide the slow passive movement and denote the point of detection17.
Valid and reliable results have been found at different joints using TDPM methods9,16,19,20,21,22. However, there is considerable variation in TDPM equipment and methods, creating a challenge for comparison of findings across studies16,17. Laboratories often develop their own limb movement and measurement devices, or use expensive commercial devices and software16. Passive movement speeds also vary; movement speed is known to affect detection thresholds7,16,23. A standardized, easily reproducible method capable of quantifying TDPM across a range of impairment levels is needed. Because the anatomy and physiology of each joint differs, protocols should be joint specific19. The protocol outlined here is specific to the elbow joint. However, the methods of this protocol may be useful to establish protocols for other joints.
To increase generalizability across sensorimotor research laboratories, the preferred apparatus for providing the passive movement for elbow TDPM testing would be commercially available at an affordable cost. To this end, an elbow continuous passive movement (CPM) machine (available speed range 0.23°/s – 2.83°/s) was chosen to produce the motorized, consistent motion. CPM machines are commonly found in rehabilitation hospitals and medical supply stores and can be rented or borrowed to reduce research costs. Additional equipment requirements include items commonly found in sensorimotor laboratories (i.e., electrogoniometer and electromyography (EMG) sensors), and hardware stores (e.g., PVC pipe, string and tape).
Two different groups were tested to explore the measurement properties of this TDPM protocol: healthy adults and adults with chronic stroke. For the adults with chronic stroke, the ipsilesional (i.e., less affected) arm was tested. Kinesthetic sense in the ipsilesional elbow in adults with chronic stroke may appear normal with clinical testing, but impaired when evaluated using quantitative laboratory methods5,15. This example illustrates the importance of developing and using sensitive and precise measures of somatosensory impairment and makes this a useful population for testing purposes. For validation of this protocol, we used the known groups method24. We compared TDPM to another quantitative measure of kinesthesia, the Brief Kinesthesia Test (BKT). The BKT has been shown to be sensitive to ipsilesional upper limb impairment post stroke25. The tablet-based version (tBKT) was used in this study because it is the same test as the BKT, administered on a tablet with more trials. The tBKT has been shown to be stable in one-week test-retest measurement and sensitive to proprioceptive knockdown26. It was hypothesized that the elbow TDPM and tBKT outcomes would be correlated as sensorimotor control of the elbow contributes to BKT performance26.
The purpose of this paper is to outline a standardized method of measuring elbow TDPM that is reproducible using common equipment. Data is presented regarding reliability and initial validity testing of the method, as well as feasibility of use for persons with no known pathology, and those who were hypothesized to have mild somatosensory impairment.
The Institutional Review Board at The College of St. Scholastica has approved the study under which this protocol was developed and tested.
1. Fabrication of the visual screen
2. Preparation of the testing equipment
3. Preparation of the participant for TDPM testing
4. Administration of the TDPM test
Trial Number | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
Delay (s) | 1 | Catch | 3 | 1 | 2 | Catch | 3 | 1 |
Table 1: Standardized time delays and catch trial locations. Varied trial start time delays are included to prevent participant attempts to guess when movement will begin. Catch trials are included to test whether or not participant is actually detecting movement19,31.
5. Calculation of participant’s TDPM score
Figure 2: Example electrogoniometer tracing with detection point. The electrogoniometer line tracing (green line), start point of the continuous passive motion (CPM) machine movement, and the point at which participant indicated movement was detected (first blue peak) are shown. The difference between electrogoniometer readings at the start of the trial (pink circle) and at detection point (orange circle) determines the TDPM value for that trial. Please click here to view a larger version of this figure.
Participants:
Using the protocol presented here, elbow TDPM was measured in an academic research laboratory for two different groups of individuals: 20 healthy adults, and eight adults with chronic stroke. Participants for both groups were recruited from the community using fliers, emails, and word-of-mouth. The healthy adults (14 females, six males; mean age (SD) = 28 (7.9) years; 19 right- and one left-handed) were tested to generate representative results for an unimpaired population. Inclusion criteria were: age of 18 to 85 years; ability to follow two-step directions as determined by screening at initial meeting. Exclusion criteria were: history of disease or conditions affecting neuromuscular function of the upper limbs based upon self-report; reported allergy to metal or latex. Handedness was assessed using the Edinburgh Handedness Inventory30. Half of the healthy adult participants had TDPM of their right elbow tested, and half had their left elbow tested (block randomization). To determine the test-retest reliability of this protocol, healthy adult participant elbow TDPM was measured twice, one week apart. The tBKT was completed on Day 1 following TDPM testing. No adverse events occurred for any participant in the healthy participants group.
The elbow of the ipsilesional (i.e., less affected) upper limb of the individuals with chronic stroke (five males, three females; mean age (SD) = 69 (11.3) years; five right hemisphere stroke, three left hemisphere stroke) was tested to represent the protocol’s capability for detecting and quantitatively discriminating TDPM in individuals with suspected mild proprioceptive impairment. Inclusion criteria for this group were the same as for the healthy adult group, with the addition of: history of stroke occurring more than six months prior that impacted upper extremity function. Exclusion criteria were: any history of ipsilesional upper extremity pain or musculoskeletal injury; reported allergy to metal or latex. Participants with chronic stroke completed one elbow TDPM testing session. The tBKT was completed following elbow TDPM testing. One participant with stroke reported mild irritation from the EMG sensor adhesive; no other adverse events occurred.
Results:
No statistical difference was found between right and left elbow TDPM scores for healthy adults (p = 0.86, two-tailed); the data was combined for subsequent analyses. The average elbow TDPM for healthy adult participants (n = 20) was 1.19 (±1.02) degrees. The Spearman correlation and intraclass correlation coefficient (ICC) were calculated to evaluate test-retest reliability of the TDPM; a positive and statistically significant relationship was found (rs = 0.72, p < 0.001), (ICC 2,4 = 0.84), suggesting moderate to good reliability of the measure among healthy adult participants24 (Figure 3).
The average ipsilesional elbow TDPM for participants with chronic stroke (n = 8) was 8.24 (±4.53) degrees (Table 2). Participants with chronic stroke were more variable than healthy adult participants (Figure 4A). Using a two-tailed t-test, the TDPM of the healthy adult and chronic stroke groups were found to be statistically different, with the adults with chronic stroke requiring a greater elbow extension excursion prior to movement being detected (t = 4.4, p = 0.003, two-tailed) (Table 2). Spearman correlation between elbow TDPM and error in targeted reaching as measured by the tBKT showed a moderate relationship between these two measures (rs = 0.63, p < 0.001) (Figure 5). Participant tBKT scores are shown in Figure 4B.
Figure 1: Participant setup for elbow threshold to detection of passive movement (TDPM) testing. The continuous passive motion (CPM) machine extended the participant’s elbow at a constant speed of 0.23°/s. Note the visual screen placed to occlude vision of the testing arm. Not visible are hearing occlusion headphones, and a trigger switch for participant indication of movement detection. Please click here to view a larger version of this figure.
Figure 3: Test-retest reliability of elbow threshold to detection of passive movement (TDPM) method in healthy adults. Spearman correlation and intraclass correlation coefficient (ICC) of Day 1 and Day 2 (tested one week apart) were used to compare TDPM scores. Figures show line of fit with 95% confidence interval (shaded area) and a density ellipse. A positive and statistically significant relationship was found (rs = 0.72, p < 0.001). Please click here to view a larger version of this figure.
Figure 4: Representative results for elbow threshold to detection of passive movement (TDPM) (A) and the tablet version of Brief Kinesthesia Test (B) for healthy adult control subjects versus participants with chronic stroke. Note that one individual with chronic stroke was unable to detect movement on any trial; the maximum specified TDPM value of 15° was assigned. This same individual had the greatest amount of error during tBKT testing. Please click here to view a larger version of this figure.
Figure 5: Elbow threshold to detection of passive movement (TDPM) scores compared to tablet version of the Brief Kinesthesia Test (tBKT) scores in healthy adults and adults with chronic stroke. Spearman correlation between elbow TDPM and error in targeted reaching as measured by the tablet version of the Brief Kinesthesia Test (tBKT) is shown. There was a moderate positive relationship (rs = 0.63, p < 0.001). Please click here to view a larger version of this figure.
Age | Sex | Stroke Chronicity | Handedness | Fugl-Meyer | TDPM | tBKT Error | |
Mean(SD) in years | Mean(SD) in months | Mean (SD) of shoulder-elbow subscore/36 | Mean (SD) in degrees | Mean (SD) in cm | |||
Healthy adults (control) n = 20 | 28(7.9) | 14 F; 6 M | NA | 19 – R 1- L | NA | 1.19 (1.02) | 1.12 (0.26) |
Adults with chronic stroke n = 8 | 69(11.3) | 3 F; 5 M | 33(19) | 7 – R 1- L | 23.9(8.5) 5 – R CVA 3 – L CVA | 8.24 (4.53) | 2.85 (1.16) |
SD = Standard deviation; F = Female; M = Male; R = Right; L = Left; CVA = Cerebrovascular Accident; NA = Not applicable; cm = centimeter | t = 4.4, p = 0.003 (two-tailed) | t = 4.15, p = 0.004 |
Table 2: Participant description, average elbow threshold to detection of passive movement (TDPM) scores (degrees), and average tablet version of the Brief Kinesthesia Test (tBKT) scores. A significant difference was found in average elbow TDPM between healthy controls and adults with chronic stroke, as well as in the average tBKT scores.
The presented protocol describes how to measure elbow TDPM in a standardized fashion using a common CPM machine to provide the passive movement. Across 20 healthy participants the average elbow TDPM measurement was found to be similar to the average value identified in previous studies using other TDPM measuring setups7,19,32, and produced reliable results across testing sessions. TDPM of the ipsilesional elbow among the eight participants with chronic stroke on average differed significantly, and perhaps clinically meaningfully, from the healthy adult population as has been previously shown5,15. It is likely a portion of the difference in TDPM between groups can be attributed to age differences21,33,34,35 and to potential reaction time differences1. Regardless, the findings indicate this method is able to discriminate between groups that have subtle differences in performance.
Selecting the CPM machine movement speed is a critical protocol step that will affect TDPM scores (protocol step 2.2). Previous studies have shown that TDPM increases with decreasing passive motion velocity7,16,23. The speed selected for this protocol, 0.23°/s, is similar to values tested in prior studies7,22,28, and is near the inflection point where TDPM exponentially increases in difficulty for healthy subjects7. As noted in the representative results, one participant with chronic stroke was unable to feel movement in any trial, suggesting the CPM machine movement speed of 0.23°/s has a potential floor effect and may need to be increased for testing of individuals with more severe kinesthetic impairments. The range of available speeds differs across CPM machine manufacturers; researchers should select a model that will meet their study needs. Providing clear participant instruction with verification of understanding is also a critical protocol element to support accurate performance of the TDPM task.
All participants with chronic stroke were able to depress the trigger switch with their more affected upper extremity; alternative methods of indicating when movement is felt may be needed for participants who are unable to do so. It is possible a larger style of switch could be used. Additional modifications to the protocol may include elimination of the biceps and triceps EMG sensors. EMG use was incorporated into the protocol to confirm muscle contraction did not occur during trials, as active muscle contraction and muscle contraction history have been shown to impact proprioceptive thresholds due to the thixotropic properties of muscle fibers and spindles27,28. However, muscle activation was not noted during any trial for any participant, suggesting EMG monitoring may be unnecessary.
A possible limitation of this protocol is the testing position of 90° of shoulder and elbow flexion, as some individuals may be unable to achieve or tolerate this position. Modification of the testing position is known to change kinesthesia36. An aspect of the TDPM paradigm that is not unique to this protocol is the high attentional demand of the task, which limits the appropriateness of this measurement method for individuals with attention deficits. To reduce error due to inattention or fatigue19, we intentionally designed this protocol to take no more than 15 minutes per limb. This protocol does not control for potential differences in reaction time between participants, which is a potential limitation. The slow passive movement speed used in this protocol decreases the proportional contribution of reaction time error to a participant’s TDPM score.
This detailed elbow TDPM protocol provides sensorimotor researchers a sensitive and precise measure of kinesthesia. The data suggest the TDPM’s resolution is high affording the possibility of detecting mild impairment or perhaps being sensitive to change if used in a study of recovery of function. Future research could be conducted to determine the minimal clinically important difference in TDPM. Adaptation of this protocol to other joints may also be appropriate.
The authors have nothing to disclose.
The authors would like to thank Dr. Jon Nelson for technical support of EMG and electrogoniometer equipment used here.
3/4 inch diameter PVC pipe | Charlotte Pipe | Pipe to be cut into lengths of: 30 inches/76.2 cm (x2); 8 inches/20.3 cm (x2); 44 inches/111.8 cm (x1); 32 inches/81.3 cm (x1). | |
3/4 inch diameter PVC pipe end caps (x3) | Charlotte Pipe | ||
45° PVC elbow (x1) | Charlotte Pipe | ||
90° PVC elbows (x2) | Charlotte Pipe | ||
Athletic tape | 3M | ||
Delsys acquisition software (EMGworks) | Delsys | ||
Double-sided tape | 3M | ||
Duct tape | 3M | Used to assist in removal of dead skin cells on participant's skin prior to EMG sensor placement. | |
Elbow Continuous Passive Motion (CPM) Machine | Artromot | Chattanooga Artromot E2 Compact Elbow CPM; Model 2038 | |
Electrogoniometer | Biometrics, Ltd | ||
Flour sack dishcloths (x2) | Room Essentials | Fabric used for creation of visual screen. | |
Handheld external trigger switch | Qualisys | Trigger switch used for electrogoniometer event marking. | |
Hearing occlusion headphones | Coby | ||
Isopropyl alcohol | Mountain Falls | ||
Paper tape | 3M | ||
Ruler with inch markings | Westcott | ||
Standard height chair | KI | ||
String | Quality Park | Approximately 15 inches of string needed. String used for standardization of electrogoniometer placement. | |
Trigno Goniometer Adapter | Delsys | ||
Trigno Wireless Electromyography Sensors | Delsys | ||
Washable marker | Crayola | ||
Washcloth | Aramark | Used in combination with isopropyl alcohol for cleaning participant's skin prior to EMG sensor placement. |