Here we present a protocol for characterizing monocyte subsets by whole blood flow cytometry. This includes outlining how to gate the subsets and assess their expression of surface markers and giving an example of the assessment of the expression of M1 (inflammatory) and M2 markers (anti-inflammatory).
Monocytes are key contributors in various inflammatory disorders and alterations to these cells, including their subset proportions and functions, can have pathological significance. An ideal method for examining alterations to monocytes is whole blood flow cytometry as the minimal handling of samples by this method limits artifactual cell activation. However, many different approaches are taken to gate the monocyte subsets leading to inconsistent identification of the subsets between studies. Here we demonstrate a method using whole blood flow cytometry to identify and characterize human monocyte subsets (classical, intermediate, and non-classical). We outline how to prepare the blood samples for flow cytometry, gate the subsets (ensure contaminating cells have been removed), and determine monocyte subset expression of surface markers — in this example M1 and M2 markers. This protocol can be extended to other studies that require a standard gating method for assessing monocyte subset proportions and monocyte subset expression of other functional markers.
Monocytes are a type of white blood cells which play a major role in promoting and resolving inflammation. There are three main subsets of monocytes recognized, classical (~85%), intermediate (~5%), and non-classical (~10%) monocytes, which are characterized by their level of cluster of differentiation (CD)14 and CD16 expression1. The proportions of monocyte subsets can differ with the presence of disease, such as an increased proportion of intermediates in various inflammatory states2,3 including cardiovascular disease, where the level of intermediates is associated with clinical events4,5. Furthermore, in disease conditions, monocytes can also undergo functional changes, with many changes detectable by a difference in surface marker expression6,7. One such example is monocyte M1-skewing, an increase in markers associated with M1 macrophages, which has been observed in cardiovascular disease, diabetes, obesity, and metabolic syndrome7,8,9,10.
Despite the popularity of flow cytometry to assess monocyte subset proportion and function, there is a considerable variability in sample preparation and subset gating between studies which makes it difficult to compare findings between such studies. Importantly, there is no consensus in the demarcation of monocyte subsets, yet a standardized approach is essential given the clinical significance of changes in subset proportions in several diseases. Part of the difficulty in gating arises from the fact that monocytes differentiate from the classical through the intermediate to the non-classical subset11 and as such, monocytes exist as a continuous spectrum rather than distinct populations12. Interestingly, Zawada et al. showed that using either a rectangular or trapezoid gating of the intermediate subset, both resulted in a higher intermediate subset that predicted a cardiovascular endpoint13. This highlights that, at least for calculating proportions, the key issue is applying a consistent gating strategy between different samples (and studies), rather than attempting to definitively discriminate between subsets. While definitive gating may be more important when assessing function, the change in marker expression between subsets is incremental12,14, and thus again, consistency in gating is perhaps key. As such, an objective gating method that reproducibly apportions the monocyte subsets between different samples is needed. The purpose of the method presented here is to gate monocyte subsets with a clear explanation and justification for the gating technique employed and assess the subsets for surface marker expression, thus providing a method which allows researchers to have confidence in the use of this technique when assessing different samples.
This study has been approved by the WSLHD Human Research Ethics committee (HREC) (approval AU RED HREC/15/WMEAD/289).
1. Sample Preparation for Whole Blood Flow Cytometry
NOTE: As human blood is potentially infectious, the sample set-up should be performed in a biohazard hood.
2. Flow Cytometry
3. Monocyte Gating
4. Validation of Gating Method
5. Phenotypic Monocyte Marker Expression
The monocyte gating strategy and flow cytometry analysis used here (Figure 1) successfully gated the monocyte subsets and revealed their relative proportions. The proportions (for this sample) were calculated as 88.1% classicals, 4.33% intermediates, and 7.49% non-classicals. These subset gates were not contaminated with B cells, T cells, neutrophils or NK cells, which was confirmed with markers CD19, CD3, CD56, and CD66b, respectively. By assessing the relative position of other populations, it is clear that the T cells and neutrophils fall well outside the monocyte "┐" shape on a CD16/CD14 plot (Figure 2A and 2D). However, both the NK cells and B cell populations overlapped with the non-classical monocyte population (Figure 2B and 2C). The steps of the gating strategy (Figure 1D and 1E) were confirmed to exclude the NK cells (Figure 3A) and B cells (Figure 3B). Although the B cell population included a small portion of non-classical monocytes, the amount was negligible.
Having gated the subsets, the degree to which they expressed different surface markers, M1 (CD64, CD86, and CD120b) and M2 (CD163, CD11b, and CD93) was assessed. The markers showed positive expression compared to their corresponding isotype controls, as seen by the shift of the histograms (Figure 4). The median of the markers was greater than that of the isotype controls.
Application of this gating strategy to a blood sample, which was split into four tubes, stained and analyzed separately, yields comparable results between tubes (Table 4).
Figure 1: Representative monocyte gating strategy in whole human blood. (A) FSC(A) vs. FSC(H) plot: Gating the cells that have an equal area and height, thus removing clumps (greater FSC(A) relative to FSC(H)) and debris (very low FSC) K= 1000. (B) FSC(A) vs. SSC(A) plot: Broad selection of monocytes based on their SSC/FSC properties. (C) CD16 vs. CD14 plot: Gating to select monocytes based on their characteristic "┐" shape. (D) CD16 vs. HLA-DR plot: Gating to select HLA-DR positive cells and remove NK cells. (E) CD14 vs. HLA-DR: Gating to exclude the B cells (HLA-DR high/CD14 low) from the monocytes. (F) Selected monocytes redisplayed on CD16 vs. CD14 plot to gate the monocyte subsets. For A-F color represents cell density with blue and green indicating low density, red and orange indicating high density, and orange indicating mid-range density. Please click here to view a larger version of this figure.
Figure 2: Validation of gating strategy by identification of potentially contaminating cells. Potentially contaminating cells are identified by markers (A) T cells (CD3), (B) B cells (CD19), (C) NK cells (CD56), and (D) neutrophils (CD66b). The left side panels show identification of each population after gating as per Figure 1A and 1B, with color representing cell density from high (red) to low (blue). The right side panels show each cell population (blue) superimposed on the final monocyte CD16/CD14 plot to reveal proximity of these populations to monocytes (red). Please click here to view a larger version of this figure.
Figure 3: Confirmation that gating steps remove contaminating cells. Heat maps showing the degree of expression from high (red) to low (blue) of (A) CD56 and (B) CD19. Gating steps successfully exclude cells with high CD56 (NK cells) and high CD19 (B cells). Please click here to view a larger version of this figure.
Figure 4: Monocyte expression of M1 (CD120b) and M2 (CD93) markers. Smoothed histograms of monocyte M1 and M2 marker expression (blue) showing clear shift from the isotype (red) for (A) classicals, (B) intermediates, and (C) non-classicals. Please click here to view a larger version of this figure.
Antibody | Volume (50 µL blood) | Volume (16x) 800 µL blood |
CD14 – V450 | 0.75 µL | 12 µL |
CD16 – APC | 0.5 µL | 8 µL |
HLA-DR-PerCP | 0.625 µL | 10 µL |
Table 1: Antibodies for whole blood flow and master mix.
Tube | PE Antibodies | Code | Isotype Match | Stock Concentration | Working Volume |
1 | IgG1 | BD (555749) | NA | 1.0 µg/20 µL | 0.625 µL |
2 | CD163 | BD (556018) | BD IgG1 | 0.125 µg/20 µL | 5 µL |
3 | CD64 | BD (558592) | BD IgG1 | 0.06 µg/20µL | 10 µL |
4 | IgG1 | BD (555749) | NA | 1.0 µg/20 µL | 1.25 µL |
5 | CD86 | BD (555658) | BD IgG1 | 1.0 µg/20 µL | 1.25 µL |
6 | CD11b | BD (561001) | BD IgG1 | 1.0 µg/20 µL | 1.25 µL |
7 | IgG2a | R &D (IC003P) | NA | 25 µg/mL | 5 µL |
8 | CD120b (TNFRII) | R &D (FAB226P) | R &D IgG2a | 25 µg/mL | 5 µL |
9 | IgG1 | BL (400112) | NA | 0.2 mg/mL | 2.5 µL |
10 | CD93 | BL (336108) | BL IgG1 | 50 µg/mL | 1.25 µL |
Table 2: M1/M2-PE Fluorochrome labeled monoclonal antibodies for whole blood flow.
Tube | Antibody | Working Volume |
11 | CD3 | 5 µL |
12 | CD19 | 5 µL |
13 | CD66b | 1.25 µL |
14 | CD56 | 5 µL |
Table 3: PE Fluorochrome labeled monoclonal antibodies for lymphocytes (T cells, B cells), neutrophils, and NK cells.
Tube | % Classical | % Intermediate | % Non-classical |
1 | 84.3 | 5.11 | 10.1 |
2 | 84.2 | 5.05 | 10.5 |
3 | 84.3 | 5.03 | 10.7 |
4 | 81.6 | 5.03 | 12.1 |
Table 4: Monocyte subset proportions from one blood sample stained and analyzed separately.
Whole blood flow cytometry is an ideal approach to study monocytes as the cells are examined in conditions close to their physiological microenvironment providing an insight into their roles in infection and inflammatory conditions. Furthermore, the use of fresh (i.e., unprocessed) blood samples minimizes the alterations or cell transformations that can occur due to storage or handling18,19, such as those known to occur with freeze-thawed monocytes20. Prompt sample preparation is recommended as some markers are upregulated if samples are kept at room temperature prior to processing19. The optimal concentrations of M1 and M2 markers were determined by titration, and this should be done for any new antibody to limit non-specific binding whilst ensuring that the degree of shift is due to antigen expression and not restricted by lack of antibody. The removal of red blood cells and fixation of white blood cells with lysis solution is an important step in this protocol as the presence of red blood cells can interfere with flow cytometry21,22. Note that while some lysis solutions are compatible with no-wash staining, clearer populations are evident in our hands when a wash step is used.
Correct setup of the flow cytometer is also critical when comparing expression of monocyte markers. We recommend that researchers maintain consistent target fluorescence intensities of control beads and perform quality control on the instrument to be used to provide consistent results across different samples run on different days. In addition to this, isotype controls are used to assist in interpreting any non-specific background signal generated by non-specific antibody binding. Monocytes have high levels of Fc receptors11 and therefore are prone to non-specific binding. Of note, the level of non-specific binding differs for the different subsets, and thus the use of an isotype control becomes important when comparing the degree of marker expression between subsets.
Another important criterion to be considered is the gating steps employed. Some studies suggest that it is crucial to draw a tight gate around the monocyte population in FSC(A)/SSC(A) plot to get rid of most of the non-monocytic CD16 positive cells23,24,25, but this may lead to loss of some monocytes as non-monocyte cells can overlap with monocytes on FSC/SSC plots26. Rather, to exclude any other blood cells that may contaminate the monocytes, the inclusion of a third monocyte marker in addition to CD14 and CD16, is essential26,27. For this reason, HLA-DR is often used and is ideal as it is not expressed by NK cells or neutrophils17,28. Although lymphocytes (B cells and T cells) may express HLA-DR, they differ from monocytes in respect to CD14 expression. While HLA-DR is an ideal third marker, CD86 has also been recommended5,27,29 but was not used here as it is also an M1 marker and thus its degree of expression on monocyte subsets was assessed.
Validation of the gating strategy used is of crucial importance. While NK cells are known to overlap with non-classicals if they are not gated out28, we notice that B cells can also overlap with the non-classicals (Figure 2B); whether this is the case in other studies may depend on the fluorochrome choice, instrument configuration, detector sensitivity, or even the disease condition being examined. Here, the overlapping B cells showed high expression of HLA-DR and were not gated out by selection of HLA-DR positive cells in Figure 1D. Rather, to remove B cells we used an additional plot of CD14/ HLA-DR, where the B cells separate out from non-classicals due to their higher HLA-DR and low CD14 expression.
There are also many different ways in which the gates for the monocytes themselves have been drawn in the literature; these include quadrants (the subsets are separated by quadrant markers), and rectangular or trapezoid boxes13 (with separate boxes drawn for each subset) which furthermore differ in their placement delineating where one subset ends and another starts. These differences likely reflect the fact that monocytes exist as a continuum of cells, differentiating from classical to non-classical, rather than as clearly distinct populations. However, because variations in techniques for identifying the subsets themselves can lead to differences in the calculated monocyte subset proportions, it becomes important that the gating method is reasonably objective, rather than subjective, as this will make the method more robust and reproducible. Some studies use an isotype control for CD16 to determine the border between the classical and intermediate subsets30. On the other hand, to define the separation between intermediates and non-classicals, it has been proposed that the cut-off line may be vertical or oblique, with the choice up to the investigators, the proviso being it should be reproducible, but a rectangular gate has been recommended to facilitate comparisons between studies13,30,31. Here, increased objectiveness was obtained by plotting the data on a zebra plot and applying objective visual rules, because zebra plots provide an additional visualization cue by mixing color gradient to each equal probability bin over a traditional contour plot. The right border of the classical subset was drawn such that the population was evenly distributed around the population median. The division between the intermediates and non-classicals were also standardized by having the base of the intermediates align with the bottom of the concentric circles within the classical population (i.e., the intermediate population clearly expresses high levels of CD14, as per the standard nomenclature1).
While some studies have suggested the use of additional markers, such as C-C chemokine receptor type 2 (CCR2) or 6-sulfo LacNAc (SLAN) for obtaining successful enumeration of monocytes and to reveal their clinical significance32,33, in our hands the level of expression of many monocyte functional markers varies widely between individuals14. Such variation may limit the usefulness of these markers to define subsets based on their expression. Automated computational approaches have also been used to visualize and cluster monocyte subsets including visual interactive stochastic neighbor embedding (viSNE), t-distributed stochastic neighbor embedding (tSNE), or Spanning-tree Progression Analysis of Density-normalized Events (SPADE)34,35, which can provide visual representation of the cells based on the set of multiple markers used. While this has been shown to increase the accuracy of the gating strategy in monocyte subset classification, it is recognized that a drawback is the number of antibodies (and corresponding fluorophore channels) required. Its usefulness will depend on the questions being asked; the extra complexity may not be warranted, for example, in enumeration studies.
Monocytes gated with our technique show proportions in line with the literature and the expression of surface markers by the three subsets can be readily determined. Overall, the technique and methodology explained here provides a standardized and straightforward method of enumerating monocyte subset proportions and assessing surface marker expression, which can be extended to include other markers as well, thereby validating their functional roles in various other conditions.
The authors have nothing to disclose.
Flow cytometry was performed in the Flow Cytometry Core Facility that is supported by the Westmead Institute for Medical Research, Westmead Research Hub, Cancer Institute New South Wales, and National Health and Medical Research Council. This study was supported by Clinical Chemistry Research and Education Fund.
BD vacutainer blood collection set | Becton Dickinson | 367286 | |
BD vacuitainer K2E 5.5 mg plus blood collection tubes -3.0 ml | Becton Dickinson | 7128959 | |
Phospahate Buffered Saline (PBS) | Lonza | 17-516F | |
5ml polystyrene round bottom FACS tube (12 x 75 mm style) | Invitro technologies | 352054 | |
V450 Mouse Anti-Human CD14 Antibody [MφP9] | BD Biosciences | 560349 | |
APC Mouse monoclonal Anti-CD16 Antibody [3G8] | Abcam Australia | ab140477 | |
Per CP Anti-Human HLA-DR Antibody [L243] | BioLegend | 307628 | |
PE Mouse IgG1 κ Isotype Control [MOPC-21] | BD Biosciences | 555749 | Lot # 4283901 |
PE Mouse Anti-Human CD163 Antibody [GHI/61] | BD Biosciences | 556018 | Lot # 2335626 |
PE Mouse Anti-Human CD64 Antibody [10.1] | BD Biosciences | 558592 | Lot # 36768 |
PE Mouse Anti-Human CD86 Antibody [2331 (FUN-1)] | BD Biosciences | 555658 | Lot # 3109766 |
PE Mouse Anti-Human CD11b/Mac-1 Antibody [ ICRF44] | BD Biosciences | 561001 | Lot # 3228959 |
PE Mouse IgG2A Isotype control [20102] | R&D Systems | IC003P | Lot # 1212031 |
PE Human TNF RII/TNFRSF1B Antibody [22235] | R&D Systems | FAB226P | Lot # 612051 |
PE Mouse IgG1, κ Isotype Control [MOPC-21] | BioLegend | 400112 | Lot # B220359 |
PE Anti-human CD93 Antibody [VIMD2] | BioLegend | 336107 | Lot # B143544 |
PE Mouse Anti-Human CD3 Antibody [SK7] | BD Biosciences | 347347 | Lot # 3010929 |
PE Mouse Anti-Human CD19 Antibody [HIB19] | BD Biosciences | 561741 | Lot # 2307721 |
PE Mouse Anti-Human CD56 Antibody [B159] | BD Biosciences | 561903 | Lot # 3011796 |
PE Anti-human CD66b Antibody [G10F5] | BioLegend | 305105 | Lot # B154037 |
OptiLyse C | Beckman Coulter | A11895 | |
Formaldehyde solution 37% | Sigma | F1635 | |
BD FACSCanto II Flow Cytometer | BD Biosciences | ||
Automatic haematology analyser, XT-1800i | Sysmex | ||
Centrifuge GS-6R | Beckman | ||
Flowjo software v10.0.7 | Tree Star Inc |