A comunicação inter-celular é crítico para controlar várias actividades fisiológicas dentro e fora da célula. Este documento descreve um protocolo para avaliar a natureza espaço-temporal de secreção das células individuais. Para conseguir isso, uma abordagem multidisciplinar é usado que integra livre-label nanoplasmonic sentindo com imagens de células vivas.
Inter-cellular communication is an integral part of a complex system that helps in maintaining basic cellular activities. As a result, the malfunctioning of such signaling can lead to many disorders. To understand cell-to-cell signaling, it is essential to study the spatial and temporal nature of the secreted molecules from the cell without disturbing the local environment. Various assays have been developed to study protein secretion, however, these methods are typically based on fluorescent probes which disrupt the relevant signaling pathways. To overcome this limitation, a label-free technique is required.
In this paper, we describe the fabrication and application of a label-free localized surface plasmon resonance imaging (LSPRi) technology capable of detecting protein secretions from a single cell. The plasmonic nanostructures are lithographically patterned onto a standard glass coverslip and can be excited using visible light on commercially available light microscopes. Only a small fraction of the coverslip is covered by the nanostructures and hence this technique is well suited for combining common techniques such as fluorescence and bright-field imaging.
A multidisciplinary approach is used in this protocol which incorporates sensor nanofabrication and subsequent biofunctionalization, binding kinetics characterization of ligand and analyte, the integration of the chip and live cells, and the analysis of the measured signal. As a whole, this technology enables a general label-free approach towards mapping cellular secretions and correlating them with the responses of nearby cells.
Comunicação inter-celular é crucial para a regulação de muitas atividades fisiológicas dentro e fora da célula. Uma variedade de proteínas e vesículas pode ser secretado que posteriormente desencadear processos celulares complexos, tais como a diferenciação, a cicatrização de feridas, a resposta imune, migração, proliferação e. 1-5 O mau funcionamento de vias de sinalização inter-celulares têm sido implicados em numerosas doenças, incluindo o cancro, aterosclerose , e diabetes, para citar alguns.
O ensaio de secreção celular óptima deve ser capaz de espacialmente e temporalmente o mapeamento da proteína segregada de interesse sem interromper as vias de sinalização relevantes. Desta forma relações causais entre os perfis de concentração e a resposta das células receptoras pode ser inferida. Infelizmente, muitas das técnicas baseadas em fluorescência mais geralmente utilizados não satisfazem estes critérios. Proteínas de fusão fluorescentes podem ser utilizadas para etiquetar o analito Within da célula, mas podem interromper a via secretora, ou, se segregado, resulta em uma luz difusa no exterior da célula que é difícil de quantificar. Ensaios baseados em immunosandwich fluorescentes são as técnicas mais vulgarmente utilizadas para a detecção de secreções celulares, mas tipicamente requerem o isolamento de células individuais. 6-11 Além disso, a introdução do anticorpo de detecção, tipicamente interrompe ou termina a experiência e o tamanho das etiquetas de anticorpos, 150 kDa para IgG, é um impedimento para sinalização a jusante.
Devido a estas barreiras é preferível que uma técnica livre rótulo ser utilizado para secreção de proteínas e entre tecnologias de imagem livre de etiquetas existentes, a superfície de ressonância de plasmão (SPR) e sensores localizados de ressonância de plasmon de superfície (LSPR) são excelentes candidatos. 12-17 Estes sensores têm sido amplamente utilizados para estudos de ligação de analisável, exossomas de proteínas e outros biomarcadores. 18-24 Em caso de LSPR, o nanostr plasmonicuctures pode ser modelado lithographically em lamelas de vidro e animado usando luz visível através de configurações padrão de microscopia de campo amplo. Devido à sua presença em nanoescala, a maioria do substrato de vidro está disponível para imagens técnicas comuns, tais como de campo brilhante e microscopia de fluorescência tornando assim estas sondas adequados para integração com microscopia de células vivas. 25-28 Demonstrámos a medição em tempo real de secreção de anticorpos a partir de células de hibridoma utilizando nanoestruturas plasmonic ouro funcionalizado com resoluções espaciais e temporais de 225 mseg e 10 uM, respectivamente. A configuração básica do chip é ilustrado na Figura 1. 28 O percurso de luz de saída do microscópio é dividido entre uma câmara CCD usada para aparência e um espectrómetro de fibra-opticamente acoplada para a determinação quantitativa de ocupação fraccionada de uma dada matriz de nanoestruturas (Figura 2 ).
O PROTOCol apresentada neste documento descreve a concepção experimental para a medição em tempo real das secreções celulares individuais enquanto monitorando simultaneamente a resposta das células utilizando a microscopia de campo brilhante padrão. A abordagem multidisciplinar inclui a fabricação de nanoestruturas, funcionalização das nanoestruturas para a ligação de analitos de elevada afinidade, a optimização da superfície tanto para minimizar a ligação não específica e caracterização das constantes cinéticas usando uma superfície comercial Plasmon Resonance (SPR) instrumento, a integração de linhas celulares sobre o substrato, e a análise de imagens e de dados espectrais. Prevemos que esta técnica seja uma tecnologia capacitadora para o mapeamento espaço-temporal das secreções celulares e suas relações causais com células receptoras.
The LSPR imaging technique described in this work has numerous advantages over more traditional methodologies for detecting cell secretions. First, the time resolution of our technique is on the order of seconds whereas the commercial alternative, an immunosandwhich assay known as EliSpot, has a typical time resolution of 2 to 3 days.7,32 As a result we were able to resolve sudden changes in the rate of protein secretion, such as that shown in Figure 6. Second, having arrays distributed over the chip allows for the secreted signal to be tracked in space and time which enables more rigorous comparisons to diffusion-based models of cell secretion. In addition, arrays like the control array shown in Figure 6 can be used to subtract out global changes in the image that typically arise from instrumental factors such as focus drift. Third, our technique requires no modification of the cells. If desired, the experiment can incorporate commonly used tags such as fluorescent proteins, but if there is concern that such tags may negatively affect cell viability or homeostasis the label-free nature of our approach does not require them. Fourth, using the spectroscopic data we have demonstrated that quantitative information regarding the fractional occupancy of surface bound ligands can be calculated.
There are numerous alternative methods to EBL for fabricating metallic nanoparticles. However, we have found that the EBL provides considerable flexibility for optimizing nanostructure and array dimensions to best suit the optics and the cells under investigation. Also critical is the fact that the chips can be readily regenerated by plasma ashing. In this way, a typical chip can be used dozens of times. Biofunctionalization details must be modified for the specific application. The protocol presented here conjugated the surface with relatively small c-myc peptide ligands. Larger ligands such as whole antibodies typically require more spacing and thus a higher SPO to SPN/SPC ratio. Regardless, a well formed SAM layer is essential for preventing non-specific binding in live-cell experiments. In general, larger molecular weight analytes are more readily detected by LSPR. Thus, in its single-cell manifestation, this technique may not be appropriate for detecting the secretion of small proteins, such as cytokines.
The current setup has been used for studying individual non-adherent cells. There are significant number of secreted signaling proteins and vesicles to which the results reported in this work are directly applicable. For example carcinoembryonic antigen (CEA) which for decades now has been a diagnostic marker for cancer. Colon cancer cells are known to secrete CEA at the rates of thousands of molecules/cell/hr and the molecular weight is 180 kDa which exceeds that of IgG antibodies. CEA is believed to be involved in autocrine and paracrine signaling pathways but the spatio-temporal nature of these secretions have never been measured. Our technique can directly address these signaling questions. An extension of this work will be to measure the spatio-temporal nature of CEA secretion from single cells.33 Future work will also focus on integrating LSPRi with two and three dimensional cell cultures of adherent cells. By incorporating multiplexed arrays capable of detecting a number of secreted proteins in parallel, this technique has the potential to open a new window into cell secretions and how they influence neighboring cells.
The authors have nothing to disclose.
The authors have nothing to disclose.
25mm diameter glass coverslips | Bioscience Tools | CSHP-No1.5-25 | 170±5 µm is optimal |
Poly-methyl methacrylate | Microchem | PMMA 950 A4 | |
Ethyl lactate methyl metacrylate | Microchem | MMA EL6 | |
Electron beam evaporator | Temescal | FC-2000 | |
Electron beam lithography | Raith | Series 150 | |
Ethanol | Sigma-Aldrich | 459836 | |
Acetone | Sigma-Aldrich | 320110 | |
CR-7 chromium etchant | Cyantek | CR-7 | |
Scanning electron microscope | Zeiss | Ultra 55 | |
Atomic force microscope | Veeco | Nanoscope III | |
Plasma ashing system | Technics | Series 85 RIE | |
SH-(CH2)8-EG3-OH (SPO) | Prochimia | TH 001-m8.n3-0.2 | |
SH-(CH2)11-EG3-COOH (SPC) | Prochimia | TH 003m11n3-0.1 | |
SH-(CH2)11-EG3-NH2 (SPN) | Prochimia | TH 002-m11.n3-0.2 | |
Surface plasmon resonance system | Biorad | XPR36 | |
Bare gold chip | Biorad | GLC chip | Plasma ashed to remove the monolayer |
1-Ethyl-3-(3-dimethylaminopropyl) carbodiimide | Thermo | 22980 | |
N-hydroxysuccinimide (NHS) | Thermo | 24510 | |
Pentylamine-Biotin | Thermo | 21345 | |
Ethanolamine | Sigma-Aldrich | E9508 | |
Neutraavidin | Thermo | 31000 | |
Phosphate buffered saline | Thermo | 28374 | |
Tween 20 | Sigma-Aldrich | P2287 | |
Inverted microscope | Zeiss | Axio Observer | Microscope is equipped with 40X oil immersion objective; CO2 and humidity incubation from Pecon GmbH |
CCD camera | Hamamatsu | Orca R2 | Thermoelectrically cooled (16 bit) |
Spectrometer | Ocean Optics | QE65Pro | |
Spectrasuite | Ocean Optics | version1.4 | |
c-myc peptide HyNic Tag | Solulink | SP-E003 | |
monoclonal anti-c-myc antibody | Sigma-Aldrich | M4439 | |
Hybridoma cell line | ATCC | CRL-1729 | |
Antibiotic Antimycotic Solution (100×) | Sigma-Aldrich | A5955 | |
Serum free media RPMI 1640 | Invitrogen | 11835-030 | |
Fetal bovine serum | ATCC | 30-2020 | |
Rhodamine DHPE | Life Technologies | L-1392 |