Here we present a protocol used to perform traction force microscopy experiments on B cells. We describe the preparation of soft polyacrylamide gels and their functionalization, as well as data acquisition at the microscope and a summary of data analysis.
Traction force microscopy (TFM) enables the measurement of forces produced by a cell on a substrate. This technique infers traction force measurements from an experimentally observed displacement field produced by a cell pulling on an elastic substrate. Here, we adapted TFM to investigate the spatial and temporal structure of the force field exerted by B cells when activated by antigen engagement of the B cell receptor. Gel rigidity, bead density, and protein functionalization must be optimized for the study of relatively small cells (~ 6 µm) that interact with, and respond specifically to ligands for cell surface receptors.
B cells are the antibody-producing cells of the immune system. To activate the adaptive immune response, they first acquire the antigen in a native form (i.e., non-processed) through a specific receptor called B cell receptor (BCR)1. This process occurs in the lymph node B cell zone. Even if some antigens can reach the B cell through lymphatic fluids, most antigens, especially with high molecular weight (>70 kDa, which is the limit size for lymphatic conduits) are indeed presented in their native form on the surface of an antigen presenting cell (APC), typically a subcapsular sinus macrophage or follicular dendritic cell, through lectin or Fc receptors (non-specific). The contact with this cell leads to the formation of an immune synapse where the BCR exerts force on the APC-associated antigens. The binding of an antigen to the BCR initiates BCR signaling, which may activate force-generating mechanisms. These forces could be important for amplifying BCR signaling, but are also essential for B cells to extract and then internalize the antigen.
Recent studies have shown that the BCR is indeed mechanosensitive2. For example, stiffer substrates elicit enhanced BCR signaling3. Moreover, force generated at the immune synapse pulls on single BCRs to probe its affinity to antigen and thereby ensure affinity discrimination4. It is therefore interesting to investigate the mechanical response of B cells to antigen presentation and to dissect this response in terms of type of receptors implicated (IgG/IgM)5, adhesion molecules (integrin ligands) or in pharmacologically and genetically modified cells (i.e., silencing of a protein downstream of BCR signaling or cytoskeleton dynamics)6.
A simple method to observe the response of a cell to a substrate of physiological rigidity and, at the same time, study forces exerted on the substrate is Traction Force Microscopy (TFM). TFM consists of observing the displacement field produced by the cell pulling on an elastic substrate. Originally the deformation of the gel was observed through wrinkles of the elastomer itself by phase-contrast microscopy7, but the insertion of fluorescence microbeads as fiducial markers allowed for better resolution and has since become the standard8. This method has been used to investigate the traction force exerted by adherent cells, tissues, and even organoids embedded in gels. Several variations of TFM have been developed9 including, combination with superresolution microscopy (i.e., STED10 or SRRF11), modification of the refractive index of the gel to allow for TIRF microscopy12, replacing beads by nano-printed patterns13, and using nanopillars instead of flat surface14. For a complete review of these variations, see Colin-York et al.15.
The protocol presented here describes a procedure to measure forces exerted by B cells on an antigen-coated substrate. These forces are applied on the ligands (antigen) in order to cluster them and subsequently extract them from the antigen-presenting substrate. We have adapted the standard TFM protocol to mimic the rigidity of physiological antigen-presenting substrates, the size and the relevant coating for the B cells. This protocol allows for the study of several cells simultaneously and can be used in conjunction with fluorescence microscopy techniques and chemical treatments. However, it does not aim to probe single molecule force measurements, for which optical tweezers16, molecular tension probes17,18, biomembrane force probes19, and atomic force microscopy20 are more suitable techniques. Compared to other single cell force measurement methods (e.g., micropipettes21 or microplates22) TFM allows for the reconstruction of a complete map of the forces exerted at the synapse with a resolution of ~300 nm. This is useful to identify spatio-temporal patterns in the forces exerted on the surface and, as the gel is compatible with confocal imaging, to correlate them with the recruitment of specific proteins (for example, cytoskeleton and signaling proteins).
Although 3D TFM is possible, it is not compatible with the rigidity and the setup we used. Deformations in 3D are achievable by other more complex setups such as protrusion force microscopy (AFM scanning a deformable membrane where the cells are plated)23,24 and elastic resonator interference stress microscopy (ERISM, a gel acting as resonating cavity for light and highlighting deformations of the substrate with accuracy of a few nanometers)25. Although these techniques are very promising, they have not yet been employed in B cells. Other types of TFM, such as on nanopillars14, could be used to have more reproducible substrates. However, this geometry is not adapted to soft cells as the cell interpenetrates the pillars, which complicates the analysis. This approach has indeed been used in T cells to observe the capability of the cell to build structures around the pillars26.
Despite its simplicity, TFM using polyacrylamide gels allows for the simultaneous observation of many cells and can be easily and inexpensively implemented in any lab equipped with a bench and an epifluorescence microscope (although we recommend confocal/spinning disk).
To mimic the physiological rigidity of an APC, we used polyacrylamide gels with a rigidity of ~500 Pa27 and functionalized the gel with activating antigens. In this protocol, we functionalized the surface of the polyacrylamide gel with hen egg lysozyme (HEL). This allows for the measurement of forces generated by stimulation of the BCR through engagement of the antigen binding site. The use of this antigen and the HEL-specific B cells from MD4 mice ensures relatively uniform force generation in response to antigen ligation28. However, other molecules (such as anti-IgM for B6 mice) can be grafted onto the gel, but the forces generated in these cases could be more heterogeneous and less intense. Because B cells are small cells (diameter ~6 µm), the number of beads has been optimized to be maximal but still trackable. For large cells that exert ~kPa forces on their substrates, one can achieve satisfactory results using relatively sparse beads or performing simple particle image velocimetry (PIV) to reconstruct the deformation field. However, for small cells such as B lymphocytes that exert stress as small as ~50 Pa, the use of single particle tracking is required (particle tracking velocimetry, PTV) to achieve the desired accuracy when reconstructing the deformation field. In order to reliably track beads individually, the magnification of the objective lens needs to be at least 60x and its numerical aperture around 1.3. Thus, the gels must be relatively thin (<50 µm), otherwise the beads are not visible as they are above the working distance of the objective.
The main protocol consists of three sections: gel preparation, gel functionalization and imaging; two more sections are optional and are dedicated to the antigen extraction quantification and imaging of fluorescent cells.
1. Gel preparation
2. Gel functionalization
3. Cell loading and imaging
4. Fluorescent HEL extraction experiment
5. Fluorescence imaging
6. Analysis
NOTE: Data analysis is in general performed by first correcting the whole stack for drift, finding the beads in each frame, tracking their movements with respect to a reference frame (taken in absence of cells), interpolating the displacement field and inverting the problem to obtain the stress using Fourier transform29. To this end, we suggest using a combination of ImageJ Macro and MATLAB programs downloadable from an online repository30.
Given the size of the cells, algorithms that extract the displacement map of the beads via correlative techniques (such as particle image velocimetry) are in general not very precise. However depending on the degree of resolution required, one can easily obtain qualitative results using a free Fiji/ImageJ plugin31,32. While this approach is sufficient to compare stimulating versus non-stimulating conditions, for a thorough analysis we recommend using our software downloadable from an online repository30, that tracks the beads individually and provides the displacement field map at a given time point as the interpolation of the individual bead displacements33. Several quantifications are possible at this point. For example (by assuming the displacement is caused only by stress tangential to the gel surface) the software also provides the stress at each point causing that specific displacement map. This is a type of “inversion problem”: the displacement at a certain point depends on the sum of all the forces applied all over the other points. The “inversion algorithm” takes into account the physical parameters of the substrate: its rigidity (Young modulus) and Poisson ratio. Direct algorithms are typically very accurate but computationally expensive. Algorithms based on Fourier transform, like ours, perform essentially a deconvolution in Fourier space and are more efficient but prone to some errors (mainly due to the interpolation step). These algorithms generally require the tuning of a parameter that prevents small local (and potentially artifactual) displacements to become too relevant in the computation of the stress field (Tikhonov regularization parameter8,29; “Regularization” variable in the dialog window; here we typically set equal to 5 x 10-19). For more advanced interpretation and analysis, such as spatio-temporal correlations, local movements, correlations with fluorescent channels, we recommend collaborating with experts in the field. For a review on computational methods see Schwarz et al.9.
As mentioned above, correct bead images look like a “starry sky”, a uniform and random distribution of bright spots (Figure 2A). Data and analysis are not reliable when the number of beads is too low (Figure 2B) or the image is out of focus (Figure 2C). Once B cells have settled onto the surface of the gel, the beads underneath the cells start to move due to the traction force exerted by the cell on the gel. Frames for which the beads are not trackable should be discarded.
As a check, it is possible to observe by eye the movement of beads comparing the “reference frame”, typically the one preceding the first contact of the cell with the substrate. Approximate results can be obtained from the single particle tracking (e.g., Trackmate, Fiji 34) as done in Figure 3A. The analysis provides a segmentation of the beads in the reference image (“FILENAME.fig”) as a control.
With the software we propose, one can obtain the displacement (Figure 3B) and stress field (the vector of the local stress at each pixel and each time point obtained by inversion from the displacement field, Figure 3C). Scalar product of the displacement and force fields integrated on the area of the cell provides total work exerted by the cell on the substrate (Figure 4A). This computation requires the mask of the cell introduced in step 6.2 of the protocol.
To compare two biological conditions (as activating HEL versus non-activating substrate BSA, or wild type versus knock-out) it is useful to compute the average curve (Figure 4B) or, even more synthetically, an average value over the last time points (20 min) where the energy reaches a plateau (Figure 4C). When the spatial information of the forces is relevant it is possible to compare single time points of each condition (Figure 4D). Refer to Kumari et al.6 for deeper analysis.
An example of fluorescence antigen extraction time lapse is shown in Figure 5A: the progressive appearance of fluorescence signals at the synapse indicated antigen detachment from the gel. The average extraction curve with its confidence interval (standard error of the mean) over 15 cells is shown in Figure 5B.
Figure 1: Schematic showing of the preparation of the gel and its functionalization. Steps are described in the protocol. Please click here to view a larger version of this figure.
Figure 2: Three examples of bead images of different qualities. (A) Example of bead image with the correct signal to noise ratio and the correct density. (B) Examples of images with a too insufficient number of beads and (C) out of focus plane. Please click here to view a larger version of this figure.
Figure 3: Processing of the images to extract the force field. (A) Example of an image of the beads (outline of the cell in white, extracted from the transmission image), bead tracking at time t = 5 min (red overlay) and displacement (arrows) relative to time t = 0 min (scale bar 5 µm). (B) Interpolated displacement field (represented as vector quiver and magnitude map, arrows are proportional to the displacement [nm]; see the color bar on the right); bottom: a smoother image of the magnitude (obtained by interpolation with a bicubic function). (C) Stress field from displacement field in panel B (represented as vector quiver and magnitude map; arrows are proportional to the shear stress [Pa]; see the color bar on the right); bottom: a smoother image of the magnitude (obtained by interpolation with a bicubic function). Please click here to view a larger version of this figure.
Figure 4: Example of information that can be extracted from force and displacement fields. (A) Example of evolution of energy in time for a single cell: a plateau phase (highlighted in gray) shows up after about 10 min. (B) Comparison of the average energy curves and (C) of the relative plateau levels for 65 cells plated on HEL (activating) coated gel and 35 cells on BSA (non-activating) coated gel (median ± interquartile ranges are shown, Mann-Whitney test was used for statistical significance). (D) Time-lapse color maps of stress for HEL and control BSA condition; both magnitude and quiver plots are shown. These images have been adapted from Kumari et al.6. Please click here to view a larger version of this figure.
Figure 5: Example of experiments with fluorescent antigen. (A) Time lapse of the extraction of fluorescent HEL (below: percentage of the maximum, scale bar = 3µm). (B) Antigen gathering over time (Mean ± SEM, n = 15). These images have been adapted from Kumari et al.6. Please click here to view a larger version of this figure.
The TFM method described here allows for the systematic study of the active mechanical capabilities of B cells. In the context of B cells, this is related to the ability to extract and internalize the antigen. Compared to other TFM methods, the protocol presented here is simple and rather reproducible: the rigidity, measured by indentation of a glass microsphere and using Hertz model, is between 400 and 600 Pa. Similar protocols have been successfully used not only for B cells35 but also for T cells36. In comparison to nanopillars (also used for T lymphocytes37) it provides a flat homogeneous surface, hence the results are easier to interpret as the interaction of the gel are mainly constrained to be tangential to the surface.
The protocol we described gives access to the spatiotemporal dynamics of the forces exerted by B cells on antigen-presenting substrates. On the spatial level this provides information of the localization of forces, and in combination with fluorescence microscopy, enables the experimenter to correlate local forces with the presence of specific molecules (i.e., components of the cytoskeleton or BCR signaling cascade). At the temporal level, it is possible to integrate quantities (such as total energy or total stress) to provide one value per time point and reduce the noise. This allows for observation of the evolution of the traction force in time (growth and plateau) and the presence of pulsatile patterns.
Critical experimental aspects for the analysis are described as following. (i) Cell density: to perform a correct analysis, cells should be sufficiently separated. We consider a cell to be analyzable if it has an empty region of its own size around it. (ii) Transmission image: it is advisable to collect at least a transmission image of the cells during the experiment to be used as a mask in the analysis. (iii) Number of beads in the image: we suggest analyzing only images where the number of beads in the synapse is between 30 and 200 (i.e., 1–8 beads/µm²). Lower densities do not allow for adequate map displacement reconstruction. High bead densities make single particle tracking unreliable. (iv) Number of beads should be constant during the experiment; however, fluctuations can occur due to small variability in the imaging conditions (especially in beads that are too close to each other). Focus drift, if occurring, must be corrected and problematic frames should be discarded. (v) Gel quality: gels with too many cracks, variability in beads distribution or gels that are too thick should be discarded. (vi) Depending on the cell type, after repeated exposures, cells at late time points (>300 frames) may suffer phototoxic effects. It is advisable to run the program on a mask devoid of cells as a “baseline” to be compared with the data. This provides a magnitude of the noise level solely due to the experimental conditions.
Gels used to measure traction force in classical adhesion allow for the investigation of processes that occur at the focal adhesion (actin flows and recruitment of signaling molecules)—the points where forces are applied38,39. However, forces at the synapse are not applied through focal adhesions. The spatiotemporal pattern of force generation at the B cell immune synapse has not been quantitatively investigated using this method until recently. Using TFM, we observed for the first time, force patterning at the B cell immune synapse, as presented in our recent study6, opening encouraging perspectives in the study of lymphocytes.
Notably, this method employs an image taken before the arrival of the cells on the gel as a reference image for the force computation. Usual TFM protocols suggest taking the reference image at the end of the experiment, after detaching the cells with trypsin; this allows the experimenter to look for a region rich in cells. Although this is possible here too, trypsin is rather inefficient at detaching B cells from antigen-coated gel, one needs to wait long for detachment and the risk of gel modification and movements (that make the whole data set unexploitable) is higher.
The method presented here is flexible and can be applied to study the effect of other signals at the immune synapse as it allows for grafting other proteins onto the gel surface (e.g., integrin ligands and immunoglobulins have been tested) and even fluorescent antigen (see section 4). Moreover, cells remain accessible to the experimenter for drug treatment and local perturbations. Finally, the method is also compatible with imaging fixed cells. For these observations, it is recommended to make the gel on a coverslip, stain the cells, glue the coverslip on a slide and only then add mounting media and another coverslip. Observation will then be done with the gel on top to avoid the degradation of the image through the gel.
Possible pitfalls are the variability in gel in polymerization and coating. Polymerization problems are mainly due to the quality of initiator/catalyst. Also, the gel can inflate, especially if not used right after assembly. This problem does not seem to dramatically affect the mechanical properties of the gel, but it can make the bead layer unreachable for the objective, effectively making the gel useless. We recommend preparing extra gels for each condition when this problem appears. There might be also a certain variability in the coating, and it is crucial to have freshly diluted Sulfo SANPAH.
In conclusion, we have described a simple, cheap and reproducible method to measure the forces exerted by B cells at the immunological synapse when activated by BCR ligand. It can be adapted to study the reaction to other ligands and other kinds of lymphocytes (memory B cells, T cells, etc.) with the use of the proper receptor ligand.
The authors have nothing to disclose.
The authors thank M. Bolger-Munro for critical reading and acknowledge the Nikon Imaging Center@CNRS-InstitutCurie and PICT-IBiSA, Institut Curie, Paris, member of the France-BioImaging national research infrastructure, for support in image acquisition and the Curie Animal Facility. PP was supported by CNRS. AK and JP were supported by Paris Descartes PhD fellowship and Ecole Doctorale FIRE—Programme Bettencourt. This project was funded by grants to PP (ANR-10-JCJC-1504-Immuphy) and AMLD (ANR-PoLyBex-12-BSV3-0014-001, ERC-Strapacemi-GA 243103).
3-aminopropyltrimethoxysilane (APTMS) | Sigma-Aldrich | 281778 | Store aliquoted, protected from humidity |
40% Acrylamide Solution | Biorad | 1610140 | |
Alexa555 microscale protein labeling kit | Molecular Probes | A30007 | |
Ammonium Persulfate (APS) | Sigma-Aldrich | A3678 | |
B cell Isolation Kit, Mouse | Miltenyi Biotec | 130-090-862 | |
B-mercaptoethanol | Gibco | 31350-010 | |
2% Bis Solution | Biorad | 161-0142 | |
Bovine Serum Albumin (BSA) | Euromedex | 04-100-812-C | |
Coverslip 18mm | VWR | 631-1580 | |
Fetal calf serum | PAA | A15-151 | Decomplemented (40min @56°C) |
Fluorodishes FD35 | World Precision Instruments, Inc | FD35100 | |
Fluosphere: carboxylate-modified, 0.2um, dark red | Molecular Probes | F8807 | |
Hen Egg Lysozyme | Sigma-Aldrich | L6876 | Stocked in aliquote 100mg/ml |
MEM Non-Essential Amino Acids Solution (100X) | Thermofisher/Gibco | 11140035 | |
N,N,N',N'-tetrametiletilendiammine (TEMED) | Euromedex | 50406-B | |
PBS (Phosfate Buffer Saline) | Gibco | 10010-015 | |
Penicillin–streptomycin | Gibco | 15140-010 | |
RMPI 1640 – Glutamax I | Thermofisher | 61870-010 | |
Sigmacote | Sigma-Aldrich | SL2 | |
Sodium pyruvate | Gibco | 11360-039 | |
sulfosuccinimidyl 6-(4'-azido-2'-nitrophenylamino)hexanoate (Sulfo-SANPAH) | Thermo Scientific | 22589 |