A protocol for determining the vitreous phase densities of micro- to pico-liter size drops of aqueous mixtures at cryogenic temperatures is described.
We demonstrate a method for determining the vitreous phase cryogenic temperature densities of aqueous mixtures, and other samples that require rapid cooling, to prepare the desired cryogenic temperature phase. Microliter to picoliter size drops are cooled by projection into a liquid nitrogen-argon (N2-Ar) mixture. The cryogenic temperature phase of the drop is evaluated using a visual assay that correlates with X-ray diffraction measurements. The density of the liquid N2-Ar mixture is adjusted by adding N2 or Ar until the drop becomes neutrally buoyant. The density of this mixture and thus of the drop is determined using a test mass and Archimedes principle. With appropriate care in drop preparation, management of gas above the liquid cryogen mixture to minimize icing, and regular mixing of the cryogenic mixture to prevent density stratification and phase separation, densities accurate to <0.5% of drops as small as 50 pL can readily be determined. Measurements on aqueous cryoprotectant mixtures provide insight into cryoprotectant action, and provide quantitative data to facilitate thermal contraction matching in biological cryopreservation.
The physical properties of water and aqueous mixtures in their various phases are of fundamental interest, and are important to in vivo and in vitro understanding of biological systems. In contemporary cryobiology and biological cryopreservation, the vitreous or amorphous phases of aqueous cryoprotectant mixtures are of particular interest1,2. Nucleation and growth of ice crystals can disrupt cells and tissues, and promote protein denaturation and aggregation, so cryopreservation protocols that vitrify the solvent have become increasingly popular. In biomolecular crystallography, the crystallization of solvent in the channels between biomolecules disrupts crystal lattices and degrades diffraction properties. Vitrification is achieved through a combination of rapid cooling, dehydration, and addition of cryoprotective solutes such as glycerol, ethylene glycol, polyethylene glycols (PEGs), alcohols, and salts.
Vitrification limits ice crystallization and growth, but does not eliminate all cooling-related sample damage. For example, crystal mosaicity (a measure of the distribution of crystal plane orientations) routinely increases by a factor of 10 to 100 when protein crystals are cooled into a vitrified state3, and post-thaw survival rates of vitrified sperm cells and oocytes vary widely.
One damage mechanism is differential contraction of solvent and surrounding material during cooling3,4,5. The equilibrium solvent and solute concentrations within a crystal, cell or tissue are temperature dependent, and the solvent plus the solute and surrounding material may contract by different amounts. Rapid cooling may prevent solvent and solute redistribution prior to vitrification, and differential contraction may lead to large, inhomogeneous, nonequilibrium stresses that cause sample damage.
Rational approaches to reducing cooling-induced damage could thus benefit from the knowledge of temperature-dependent densities of liquid and vitrified aqueous mixtures. At solute concentrations above 50% weight of solute to weight of solution (w/w), most aqueous cryoprotectant mixtures can be vitrified with modest cooling rates of 10 K/s or less, allowing production of and density measurements using large vitreous samples6. Density can then be determined using Archimedes' principle, by measuring the apparent weight of the sample when suspended in a liquid cryogen like nitrogen. However, as solute concentration decreases, cooling rates required for vitrification increase rapidly: Cooling rates for aqueous glycerol mixtures increase from <10 K/s at 50% weight of solute in g to volume of solution in mL (w/v) to >1,000 K/s at 25% w/v7. Heat transfer becomes boundary-layer limited, so that achieving larger cooling rates requires smaller and smaller samples8.
Measurements of the density of pure vitreous water and ice have been achieved by depositing micrometer-diameter (femtoliter volume) drops in a vacuum onto a cryogenically-cooled surface so as to build up a macroscopic (gram mass) sample. The density of this sample was determined by cryoflotation in a liquid nitrogen-argon mixture, in which the density of the cryogenic liquid was adjusted until the sample became neutrally buoyant9. However, generating large samples from a large number of small drops in a way that minimizes void volumes – an important source of error in previous vitreous phase density measurements – is non-trivial. For aqueous mixtures, differential evaporation of solution components during aerosolization and deposition in a vacuum can lead to substantial uncertainties in deposited concentrations.
We have developed a method, based on cryoflotation, that allows accurate density determination of aqueous mixtures using individual drops as small as 50 pL10. These drops can be rapidly cooled while retaining their original concentrations, and their cryogenic temperature state (vitrified or crystalline) can be assessed using a simple visual assay that correlates with X-ray diffraction measurements. This method is broadly applicable to aqueous and non-aqueous mixtures, and can be extended to a variety of biological samples including cells (e.g., stem and egg), tissue samples, and protein crystals having low-temperature densities between 0.8 and 1.4 g/mL.
CAUTION: Please consult all relevant material safety data sheets (MSDS) before use. Please use all appropriate safety practices when using compressed gases, including appropriate calibrated gas handling regulators and valves, and approved gas tubing. Contact with liquid cryogens can cause severe frostbite and necrosis. Use appropriate personal protective equipment (face shield, gloves, lab coat, full length pants, closed-toe shoes), all of which must be impermeable to liquid nitrogen. Remain standing and ensure an unobstructed exit path from the apparatus when using liquid cryogens. Be aware of asphyxiation hazards when using compressed gases and liquid cryogens, and work in a well-ventilated area with adequate make-up air (a fume hood or a high air turnover rate room.)
1. Preparation of Aqueous Solutions for Density Measurements
NOTE: Because weights are more easily measured to a high accuracy than volumes, solution concentrations are measured in w/w units. All densities and melting or boiling temperatures assume an atmospheric pressure of ~100 kPa. The following steps describe the preparation of a 35% w/w glycerol solution. The same procedure can be used for other concentrations and solutes.
2. Preparation of the Sample Cooling Chamber
3. Determination of the Volume and Density of the Test Mass at T = 298 K and T = 77 K
4. Preparation of the Initial Liquid N2-Ar Mixture
5. Measuring and Adjusting the Density of the Initial N2-Ar Mixture
6. Cooling Drops of Sample Solution
7. Assessment of the State of the Sample
8. Determination of the Density of the Sample
Density measurements at T = 77 K for vitrified drops of aqueous glycerol and ethylene glycol versus cryoprotectant concentration are shown in Figure 1A and Figure 1B respectively, and the corresponding change in specific volume between T = 298 K and 77 K, calculated using previously determined T = 298 K densities, is shown in Figure 2. At high cryoprotectant concentrations, the solutions contract on cooling to the vitrified state, while pure water expands. Near 20-25% w/w solutions of both cryoprotectants are predicted to show no net expansion or contraction. The slope of the volume change versus concentration has the largest magnitude below 40% w/w, where the effects of additional cryoprotectant on water's tetrahedral low-temperature structure are most pronounced.
Figure 1: Vitreous Phase T =77 K Density Versus Cryoprotectant Concentration. T =77 K density versus concentration for vitrified aqueous drops containing (A) glycerol and (B) ethylene glycol. Data is presented as mean ± SEM of three individual drops. Please click here to view a larger version of this figure.
Figure 2: Change in Specific Volume on Cooling from Liquid at 298 K to the Vitreous Phase at 77 K. Percent volume change on cooling from 298 K to 77 K for aqueous solutions of glycerol and ethylene glycol. The T = 298 K solution densities are obtained from previous measurements14,15. Data is presented as mean ± SEM of three individual drops. Please click here to view a larger version of this figure.
The present apparatus and methods, developed primarily by undergraduates with limited access to instrument-building tools and machinery, nevertheless delivers highly accurate density measurements for individual liquid drops as small as 50 pL. In the concentration range near and above 50% w/w, where small cooling rates are sufficient to obtain vitrified samples, the densities agree with those obtained in previous measurements on bulk samples. Extrapolations of the present densities to 0% concentration – pure water – also agree quite well with the accepted density of low-density amorphous ice at 77 K9.
Generating the N2-Ar mixtures required between 30 min and 5 h of Ar flow, depending upon the final N2-Ar density needed to render a given drop composition neutrally buoyant. This time may be reduced by using a diffuser or multiple gas tubes to increase the surface area for Ar mixing in the cryogenic liquid. Adjusting the N2-Ar density until a drop is ascertained to be neutrally buoyant can also be time consuming, especially for small radius (r) drops that have small terminal speeds (r2) and so require more careful observations. The N2-Ar mixture tends to develop a vertical density/composition gradient, and so must be regularly mixed. Consequently, determining a single vitreous phase density point for a given solute type and concentration, requiring measurements on at least 3-5 drops, can take several h.
At each concentration, the densities of two or three drops are typically measured. The "density" of each drop is estimated as the average of the measured upper and lower bounds on the density, given by the largest measured N2-Ar density that made the drop sink and the smallest density that made it float. Since both a tight upper bound and a tight lower bound – where tightness is assessed by the speed of ascent or descent of the drop – are not always obtained in measurements on a given drop (e.g., the drop may be lost during mixing), measurements on drops of the same concentration and size were sometimes combined into a single density estimate.
To reduce experiment times, attempts were made to prepare and store high-density N2-Ar mixtures in cryogenic storage containers for use one to three days later. In all cases, Ar crystallized out of the solution and the liquid density decreased with storage time. Ar solidification and liquid density decreases also occurred during drop density measurements if the liquid N2-Ar was not regularly mixed.
A major challenge in these measurements is minimizing frosting and ice formation. Water vapor condensation, ice formation, and ice accumulation on the sample cooling chamber, on other cold surfaces, in the cold gas above the N2-Ar mixture, and in the N2-Ar mixture itself can contaminate samples used in density measurements, promote ice nucleation within them, and change their apparent density. Ice on the sample and floating on and in the liquid N2-Ar mixture can make assessing the sample's low temperature state (vitreous or polycrystalline) difficult. To minimize ice formation, regularly inspect all cold surfaces for ice. Carefully remove any ice mechanically or using warm dry N2 gas. If ice accumulates in the copper chamber, remove it using a fine mesh screen, or else remove, empty, dry, and refill the chamber.
The minimum (critical) cooling rate required to obtain a vitrified sample increases with decreasing solute concentration, approaching 106 K/s for pure water7. Sample cooling rates depend on drop shape and size (increasing with decreasing diameter), the speed with which the drop is projected into the liquid cryogen, the presence of cold gas above the liquid cryogen (which usually decreases cooling rates), and the properties of the liquid cryogen. Generally, cooling rates larger than 1,000 K/s require drops with volumes (diameters) smaller than ~1 nL (~100 μm).
The lowest solute or cryoprotectant concentration for which vitreous densities can be measured is set by maximum drop cooling rates, and by the smallest size drops for which the visual assay for vitrification can be reliably used. Cooling rates could be increased by a factor of ~5 by cooling samples in liquid propane or a liquid propane-ethane mixture. Unlike liquid N2, these cryogenic liquids have a large separation between boiling and melting temperatures and so can absorb much more heat without heat-transfer-limiting surface boiling. Cooled drops could then be transferred to the N2-Ar mixture for density measurements. The transition from clear drops to hazy or cloudy drops is abrupt, occurring over a narrow range of solute concentration (roughly 2% w/w) and cooling rate, and has been correlated with the appearance ice rings in X-ray diffraction patterns16,17. However, accurate visual clarity assessment becomes more difficult as drop volumes decrease toward 10 pL.
The accessible range of sample densities using N2-Ar mixtures is set by the densities of the pure liquids, 0.81 g/mL and 1.40 g/mL, respectively. Liquid Ar-Kr mixtures are susceptible to Kr crystallization, but might be used to extend this density range provided the liquids were constantly mixed.
The methods described here are broadly applicable to determining densities of aqueous mixtures, cells, cell aggregates, other biological materials, and other systems where small samples and large cooling rates are required to achieve the desired low-temperature phase. These densities will be useful in understanding and minimizing sample damage in cryopreservation, and in understanding the behavior of water in aqueous solutions and in confined and crowded environments.
The authors have nothing to disclose.
This work was supported by the NSF under award MCB-1330685. DWM acknowledges partial support from Cornell University's Molecular Biophysics Training Grant (NIH T32GM0082567).
centrifuge tube | Falcon | 6029236 | 15 mL conical centrifuge tube |
glycerol, >99.5% | Sigma | G9012-100 mL | |
ethylene glycol, >99.8% | Sigma | 324558-100 mL | |
analytical microbalance | Mettler | AE240 | Analytical balance, 0.01 mg resolution, has hook on bottom for weighing below the balance |
vortexer | Scientific Industries | SI-0236 | Vortex-Genie 2 |
Apparatus enclosure framing | Unistrut | 1-5/8" metal framing | 48" wide x 24" deep x 40" tall |
Apparatus enclosure air barrier | any clear plastic sheeting | ||
neoprene rubber disk | 4" diameter, 1/8" thick | ||
dewar flask | Scilogix Dilvac | SS333 | 4.5 liter dewar flask with steel case and clamp lid |
copper chamber | This fabricated part is comprised of a 1.43" diameter, 0.017" wall thickness copper tube with a solid cylindrical copper base soldered to seal one end. The copper base is 0.87" tall and the overall chamber height is 7". | ||
nitrogen gas | Airgas | NI HP300 | 99.998% pure N2 gas |
argon gas | Airgas | AR HP300 | 99.998% pure Ar gas |
rotameter | Omega | FL3692ST | 2.52 l/min max flow rate |
foam insulating lid | This part is fabricated from 4 lb/ft3 crosslinked polyethylene foam (supplied by Technifab, 1355 Chester Industrial Parkway, Avon, OH), and has an OD of 2.42", and ID of 1.52", and a thickness of 0.79". | ||
PTFE test mass | This fabricated part is a 0.246" diameter, 0.580" tall cylinder with a 0.060" diameter hole running perpendicular to and intersecting the cylinder axis ~0.10" from one end. | ||
microbalance platform | Unistrut | 1-5/8" metal framing | 11" wide x 24" long x 24" high rectangular frame with an top aluminum sheet containing a hole for the monofilament and hanging test mass |
2 mil (50 um) monofilament line | Berkley | NF1502-CM | Nanofil fishing line |
Argon precooling coil tubing | VWR | 60985-512 | 1/8" ID x 1/4" OD PVC tubing |
perforated copper foil mixer | 1.4" diameter, 35 micron thick copper disk, cut from 1 ounce/ft2 copper sheet and perforated with holes using an awl or other sharp pointed tool. Insert 1-2 mm diameter rigid thermally insulating (plastic or wood) rod into the center and fix using epoxy as needed. | ||
syringe | BD | 309628 | 1 ml Luer-Lok tip syringe |
vacuum generator | Gast | VG-015-00-00 | compressed air venturi single stage vacuum generator |
hydrophobic coating spray | RainX | 620036 | plastic water repellent |
long focal length stereo microscope | Bausch and Lomb Stereozoom 6 | 0.67-4 x zoom pod with 20x eyepieces, 10 cm working distance | |
LED ring illuminator | Amscope | LED144S | |
LED spot illuminator | Newhouse Lighting | NHCLP-LED | 3W LED gooseneck clamp lamp |
1.8 ml cryo vial | Nunc | V7634-500EA | Any 1.8 or 2 ml cryovial is adequate |