Summary

Multiplexed Fluorescent Microarray for Human Salivary Protein Analysis Using Polymer Microspheres and Fiber-optic Bundles

Published: October 10, 2013
doi:

Summary

We describe a procedure for profiling salivary proteins using multiplexed microsphere-based antibody arrays. Monoclonal antibodies were covalently linked to fluorescent dye-encoded 4.5 μm polymer microspheres using carbodiimide chemistry. The modified microspheres were deposited in fiber-optic microwells to measure protein levels in saliva using fluorescence sandwich immunoassays.

Abstract

Herein, we describe a protocol for simultaneously measuring six proteins in saliva using a fiber-optic microsphere-based antibody array. The immuno-array technology employed combines the advantages of microsphere-based suspension array fabrication with the use of fluorescence microscopy. As described in the video protocol, commercially available 4.5 μm polymer microspheres were encoded into seven different types, differentiated by the concentration of two fluorescent dyes physically trapped inside the microspheres. The encoded microspheres containing surface carboxyl groups were modified with monoclonal capture antibodies through EDC/NHS coupling chemistry. To assemble the protein microarray, the different types of encoded and functionalized microspheres were mixed and randomly deposited in 4.5 μm microwells, which were chemically etched at the proximal end of a fiber-optic bundle. The fiber-optic bundle was used as both a carrier and for imaging the microspheres. Once assembled, the microarray was used to capture proteins in the saliva supernatant collected from the clinic. The detection was based on a sandwich immunoassay using a mixture of biotinylated detection antibodies for different analytes with a streptavidin-conjugated fluorescent probe, R-phycoerythrin. The microarray was imaged by fluorescence microscopy in three different channels, two for microsphere registration and one for the assay signal. The fluorescence micrographs were then decoded and analyzed using a homemade algorithm in MATLAB.

Introduction

Since the first microarray reported by Mark Schena and coworkers in the mid-1990s, this powerful tool has been utilized in many fields of biological research1. Antibody microarrays capable of simultaneously detecting multiple proteins in diagnostic fluids, such as blood, have important applications in clinical diagnostics and biomarker screening2-10. Saliva, containing many of the same analytes as blood, has been considered as a preferable alternative to blood because saliva collection is safe, noninvasive, and can be carried out by minimally-trained medical personnel11-13. Currently, multiplexed protein analysis using saliva samples is limited by several important factors, including the low concentration of target analyte14 and the wide concentration range of different biomarkers15.

.

Herein, we demonstrate the analysis of six proteins: human vascular endothelial growth factor (VEGF), interferon gamma-induced protein 10 (IP-10), interleukin-8 (IL-8), epidermal growth factor (EGF), matrix metallopeptidase 9 (MMP-9), and interleukin-1 beta (IL-1β). The performance of the method was initially verified using standard solutions constituting recombinant analyte proteins and blocking buffer. Real saliva samples collected from patients of different chronic respiratory diseases as well as healthy controls were also tested with satisfactory performance. The protocol should be applicable to other protein analytes and other microsphere-based assays. This platform offers considerable advantages to the Analytical Chemistry field as it enables fast, accurate, and reproducible simultaneous analysis of low concentrations of several proteins with a broad dynamic range, minimal non-specific interactions, reduced sample consumption, and low cost in comparison to an analogous Enzyme-Linked Immunosorbent Assay (ELISA).

Protocol

Figure 1
Figure 1. Workflow for applying fiber-optic microsphere antibody array to saliva profiling. (1) Microspheres are internally encoded with two fluorescent dyes; (2) the encoded microspheres are externally modified with protein-specific monoclonal antibodies; (3) the multiplexed microspheres are mixed, and (4) randomly deposited in microwells etched at the proximal end of a fiber-optic bundle; (5) salivary proteins are captured by microspheres through sandwich immunoassay, and (6) quantified using fluorescence microscopy.

1. Microspheres Encoding

  1. Weigh sodium europium (III) thenoyltrifluoroacetonate trihydrate (Eu-TTA, MW = 869.54 g/mol) in an amber glass vial and prepare a 200 mM stock solution in tetrahydrofuran (THF). Mix gently by pipetting; visually verify the dye is completely dissolved.
  2. Weigh coumarin 30 (C30, MW = 347.41 g/mol) in an amber glass vial and prepare a 12 mM stock solution in THF. Mix gently by pipetting; visually verify the dye is completely dissolved.
  3. Prepare 700 μl of working solution for each microsphere type using the stock solutions and THF to reach final concentration listed in Table 1.
  4. Ensure microspheres (10% w/v) are well suspended by vortexing, and then transfer 60 μl suspension to a 1.5 ml microcentrifuge tube.
  5. Add 600 μl PBS to microsphere suspension and mix by pipetting 20x. Centrifuge the tube at 10,000 rpm for 3 min and carefully remove the supernatant. Repeat this process another 2x to wash the microspheres.
  6. Wash the microsphere pellet 3x with THF using a similar procedure as step 1.5.
  7. Centrifuge the microsphere/THF suspension at 10,000 rpm for 3 min, remove the supernatant, and re-suspend the pellet in 600 μl working solution listed in Table 1. Transfer the mixture to a new 1.5 ml microcentrifuge tube.
  8. Seal the microcentrifuge tube with Parafilm and place it on a shaker. Shake at 3,000 rpm and incubate for 24 hr, protect from light by covering the setup using a box covered with aluminum foil.
  9. Centrifuge the microsphere suspension at 10,000 rpm for 3 min to form a pellet.
  10. Remove the supernatant dye solution, wash the microspheres 6x with 600 μl methanol and then with 600 μl PBS containing 0.01% Tween-20 another 6x, as described in step 1.5.
  11. Store the encoded microspheres in 600 μl PBS containing 0.01% Tween-20 at 4 °C until use, protect from light.
Microsphere type name: 1 2 3 4 5 6 7
Eu-TTA (mM) 100 100 10 100 10
C30 (mM) 1 1 6 6 1 6

Table 1. Concentrations of Eu-TTA and C30 working solutions.

2. Preparation of Protein-capture Microspheres

  1. Prepare MES buffer [2-(N-morpholino)ethanesulfonic acid (MES) 0.1 M, NaCl 0.9%, SDS 0.01%, pH 5.7] and PBS/SDS buffer (Sodium Phosphate 0.01 M, NaCl 0.154 M, SDS 0.01%, pH 7.4) in advance at room temperature.
  2. Add 200 μl encoded microsphere suspension (containing ~2 mg microspheres) into a Safe-Lock 1.5 ml microcentrifuge tube. It is important to use this type of microcentrifuge tube to avoid spillage during the incubation in step 2.6.
  3. Wash the microspheres with 600 μl MES buffer 3x as described in step 1.5. Add 700 μl MES to the microsphere pellet and mix well by pipetting.
  4. Prepare 0.105 g/ml 1-ethyl-3-[3-dimethylaminopropyl] carbodiimide hydrochloride (EDC, MW = 191.7 g/mol) and 0.163 g/ml sulfo-N-hydroxysulfosuccinimide (sulfo-NHS, MW = 217.13 g/mol) solutions in MES buffer in separate tubes.
  5. Add 150 μl freshly prepared EDC solution drop by drop into the microsphere suspension. The EDC hydrolyzes very quickly, thus it is essential to use freshly made solution to ensure the immobilization efficiency. Immediately after the addition of the EDC, add 150 μl sulfo-NHS solution to the microspheres, mix well by pipetting and verify the suspension is homogenous.
  6. Cap the tube, seal it with Parafilm, and place it on a shaker. Shake at 1,500 rpm for 4 hr, protect from light using a box covered with aluminum foil.
  7. Centrifuge the microsphere suspension at 10,000 rpm for 3 min to form a pellet and remove the supernatant. Wash the microspheres with 500 μl PBS/SDS buffer 3x as described in the step 1.5.
  8. Add 200 μl PBS/SDS buffer to the pellet, mix the microspheres well and transfer the solution to 300 μl PBS/SDS buffer containing 60 μg capture antibodies. It is important to suspend the microsphere pellet first and then add the microsphere suspension into the antibody solution.
  9. Incubate the antibody-microspheres mixture by shaking at 1,500 rpm for 4 hr on a shaker, protect from light using a box covered with aluminum foil.
  10. Centrifuge the microsphere suspension at 10,000 rpm for 3 min to form a pellet and remove the supernatant. Wash the microspheres with 500 μl StartingBlock (TBS) blocking buffer 3x, as described in step 1.5.
  11. Mix microspheres with 1 ml of TBS blocking buffer and incubate microspheres at 1,500 rpm for 1 hr on a shaker, protect from light using a box covered with aluminum foil.
  12. Centrifuge the microsphere suspension at 7,000 rpm for 3 min to form a pellet and remove the supernatant. Wash the pellet 3x with 500 μl TBS blocking buffer, as described in step 1.5.
  13. Centrifuge the microsphere solution at 7,000 rpm for 3 min to form a pellet and remove the supernatant. Add 500 μl TBS blocking buffer and mix by pipetting.
  14. Store the modified microsphere suspension at 4 °C until use, protect from light.
  15. Repeat this procedure to make other types of microspheres using corresponding antibodies.

3. Fiber-optic Microarray Assembly

  1. Mix 50 μl of each type of protein-capture microspheres into a new autoclaved microcentrifuge tube. Centrifuge the stock microspheres pool at 7,000 rpm for 3 min, and remove the supernatant so that the remaining volume is approximately 100 μl.
  2. Hand-cut fiber-optic bundles to approximately 5 cm in length and sequentially polish both ends on a polishing machine using 30, 15, 9, 6, 3, 1, 0.5, and 0.05 μm-sized diamond lapping films. Sonicate the polished bundle in deionized water for 2 min to remove any particles.
  3. Put a small magnetic stirring bar into a 0.5 ml microcentrifuge tube, add 400 μl of protein-free PBS buffer and stir on a magnetic stirring plate for 30 min to remove air bubbles.
  4. Etch one end of the polished fiber bundle in a 0.025 N HCl solution for 150 sec10,16. Immediately submerge and sonicate the etched end in deionized water for 1 min. Dry the fiber bundles with compressed air.
  5. Mount the fiber bundle on a fiber holder and block the etched end in the bubble-free, Protein-free PBS buffer for 1 hr.
  6. Deposit 1 μl aliquot of the stored microsphere suspension on the etched end of the fiber-optic bundle. Protect the setup from light with a box with aluminum foil, and wait for 15 min. The volume of the suspension will decrease by evaporation and force the microspheres into the etched microwells. Repeat this loading step one more time.
  7. Use a swab saturated with sample solution to remove microspheres that are not trapped in the microwells. The protein microarray is now ready to use. Go immediately to step 4.1 for saliva analysis.

4. Saliva Sample Analysis Using Microspheres Microarray

  1. Add 200 μl sample solution (100 μl saliva supernatant diluted with equal volume of StartingBlock T20 (PBS) blocking buffer. The collection protocol for saliva supernatant has been published previously10). into a 0.5 ml autoclaved microcentrifuge tube. Dip the protein microarray loaded on the fiber into the solution and incubate on the shaker at 600 rpm for 2 hr, protect from light using a box covered with aluminum foil.
  2. Prepare 20 ml wash buffer by adding 200 μl 10 % BSA solution and 200 μl 10% Tween-20 into 19.6 ml PBS, mix well by gently shaking.
  3. Dip the protein microarray into a new autoclaved microcentrifuge tube containing 200 μl wash buffer and shake at 600 rpm for 2 min. Repeat this wash 3x.
  4. Incubate the microarray in 100 μl solution of a detection antibody cocktail containing 5 μg/ml of anti-VEGF, IP-10, IL-8, EGF, MMP-9, IL-1β biotinylated detection antibodies in StartingBlock T20 (PBS) blocking buffer for 30 min at room temperature on the shaker at 600 rpm, protect from light using a box covered with aluminum foil.
  5. Wash the microarray 3x with wash buffer as described in step 4.3.
  6. Incubate the microarray in 200 μl of 20 μg/ml streptavidin R-phycoerythrin conjugate (SAPE) solution in PBS at 600 rpm on the shaker for 10 min, protect from light using a box covered with aluminum foil.
  7. Wash the microarray 5x with wash buffer as described in step 4.3.
  8. Clean the distal end of the fiber (i.e., the end away from the microspheres) bundle with a swab saturated with absolute ethanol.
  9. Dry both ends of the fiber with compressed air. Mount the fiber on an epifluorescent microscope and image through the distal end of the fiber.
  10. Acquire microsphere images corresponding to Eu-TTA, C30, and SAPE fluorescence emission intensities. Optical filters setup and exposure times for fluorescence imaging of Eu-TTA, C30, and SAPE are depicted in Table 2.
Channel Eu-TTA C30 SAPE
Exciter 365/10x 350/50x 546/10x
Beamsplitter 525DCLP 400DCLP 560LP
Emitter 620/60m 460/50m 580/30m
Exposure time 1 sec 0.3 sec 1 sec

Table 2. Parameters for the three fluorescent images of the microarray.

Representative Results

Fluorescence images from three channels showing a small section of the fiber-optic bundle are shown in Figures 2A-C. These images were analyzed using an algorithm written in MATLAB (as described in more detail in the Discussion section). The analysis employs both information from the Eu-TTA encoding image (Figure 2A) and the C30 encoding image (Figure 2B) to decode the microspheres, and the fluorescence intensities of different microspheres in the signal image (Figure 2C) were calculated. With the calibration curves obtained from correlated protein standards, the concentrations of proteins in saliva samples were calculated.

Figure 2
Figure 2. (A) Eu-TTA encoding image, (B) C30 encoding image, (C) SAPE signal image, (D) decoding results overlapped on the signal image; different types of microspheres were labeled with different colors and shapes.

Discussion

Researchers should pay extra attention to the following steps: for better decoding accuracy, it is necessary to verify the microspheres were homogeneously suspended in all incubation and wash steps during the microspheres encoding procedure. In addition, the encoded microspheres need to be protected from light throughout the entire experiment. Following proper encoding and storage procedures, we found that overall decoding accuracy was above 99%. The encoded microspheres should be stored at 4 °C. Avoid freezing and protect from light. Under proper storage conditions, the encoded microspheres are stable for more than six months. Extreme care is required during the encoding and modification steps to reduce the loss of microspheres. For example, do not disturb the microsphere pellet when removing supernatant. In addition, including Tween-20 and SDS in the buffers is essential for better microsphere pelleting.

Some of the common troubleshooting procedures are: (1) use only freshly prepared EDC solution and the EDC solution should be added dropwise, (2) autoclaved tubes should be used and the microsphere solution should be transferred into a new tube before each incubation step, (3) during each wash step, try to remove the supernatant as much as possible, (4) the modified microspheres should be stored at 4 °C, avoid freezing and protect from light. Under proper storage conditions, the modified microspheres may be stored for more than 3 months with no detectable signal loss.

The MATLAB algorithm used to analyze the fluorescence images is briefly described below. The relevant code can be found in supplemental materials. Pixels within a user-defined intensity range for the Eu-TTA image are filtered first. Different areas were checked with a size-filter, areas containing a "hollow" center caused by light attenuation were removed and areas where microspheres are present were sorted into a watch list for future processing. All microspheres from the watch list were further filtered with responses in the corresponding C30 image. The signals for all the pixels of a particular microsphere were averaged. The signal strength of this microsphere of interest was calculated by averaging the intensities of all microspheres in the watch list. To make the results statistically more representative, a minimum of 25 microspheres of each type was required. To minimize the nonspecific signal and decrease the variation between different experiments, signals for all microsphere types were normalized by subtracting the signal from the control microspheres.

The signal response of the microspheres can be adjusted by the amount of antibodies immobilized on the surface of microspheres, which is ideal for multiplexed detection of proteins with wide ranges of concentrations. Based on different diameters of the microspheres, the amount of antibody required to form a monolayer on the microspheres can be estimated by the equation from the manufacturer17. For microspheres with a diameter of 4.5 μm, 3 μg of antibody is needed per 1 mg of microspheres. We have tested different antibody amounts in the coupling solution in step 2.8 from 3 μg (0.5x) to 60 μg (10x) of antibody. Increased signals were observed when more antibody was used in the solution. Optimal antibody amounts for different antibodies and applications must be experimentally determined. The "control microspheres" were coupled with the mouse isotype IgG antibody, which is supplied by the manufacturer as the negative control in direct ELISA and has shown no cross-reactivity with up to 40 recombinant human proteins18. To evaluate the reproducibility of the coupling method, two batches of MMP-9 microspheres were coupled on different days. The performance of the microspheres was tested by using multiplexed detection of 10 ng/ml MMP-9. The results showed no significant differences among different batches of microspheres (detailed results are shown in Table 3).

No Date Test 1 Test 2 Test 3 Average Std
1 01/12/2011 773.8 690.1 756.8 740.2 44.2
2 01/26/2011 791.9 721.8 867.0 793.6 72.6

Table 3. Reproducibility of the coupling methods (results shown in arbitrary units).

The multiplexed method presented here enables fast, accurate, and reproducible analysis of different proteins over a wide concentration range (results listed in Table 4). The potential cross-reactivity for the six antibody pairs we used in this study was also tested. All signals from cross-reactivity were found to be negligible (less than 3%). Other advantages of this method, when compared to a conventional Enzyme-Linked Immunosorbent Assay (ELISA), include a broad dynamic range, minimal non-specific interactions, reduced sample consumption and low cost10. The required volume of saliva sample is as low as 100 μl and the assay can be completed within 3.5 hr. When compared with other suspension arrays, one limitation of this approach is that once the microarray was assembled, the detection needs to be started in less than 15 min. The method described here has been used in our lab to analyze human saliva samples collected from patients with inflammatory diseases and healthy controls (unpublished data). The method should be applicable for protein analysis of other complex fluid specimens. A similar microsphere-based protein array can be used on other platforms such as microfluidic devices. The encoding and immobilized methods could also be applied to microspheres made with other materials9.

VEGF IP-10 IL-8 EGF MMP-9 IL-1β
Test 1 5365.9 2578.4 6508.7 2584.0 515.5 1147.4
Test 2 5787.8 2577.9 7238.9 2919.2 375.7 1099.2
Test 3 4944.6 2883.3 6726.3 3619.3 406.8 1084.1
Average 5366.1 2679.9 6824.6 3040.8 432.7 1110.2
Std. Dev 421.6 176.2 374.9 528.3 73.4 33.1

Table 4. Results for the three independent tests on the same saliva sample (results shown in arbitrary units).

Disclosures

The authors have nothing to disclose.

Acknowledgements

This work was supported by the National Institutes of Health (grant 08UDE017788-05). E.B.P. also acknowledges support from the Spanish Foundation for Science and Technology (FECYT). The authors thank Shonda T. Gaylord and Pratyusha Mogalisetti for critical reading of the manuscript.

Materials

Name of Reagent Company Catalog Number Comments
Eu-TTA dye Fisher Scientific AC42319-0010
THF Sigma-Aldrich 34865-100ML
Amber glass vial Fisher Scientific 03-339-23B
Coumarin 30 dye Sigma-Aldrich 546127-100MG
Microspheres Bangslabs PC05N/6698
1.5 ml microcentrifuge tubes Fisher Scientific 05-408-129
PBS 10x concentrate Sigma-Aldrich P5493-1L
Water Sigma-Aldrich W4502-1L
Methanol Sigma-Aldrich 34860-100ML
Tw-20 Sigma-Aldrich P7949-100 ml
BupH MES buffered saline Thermo Scientific 28390
SDS Sigma-Aldrich 05030-500ML-F
NaOH solution Fisher Scientific SS256-500
Safe-lock microcentrifuge tube VWR labshop 53511-997
EDC Thermo Scientific 22980
Sulfo-NHS Thermo Scientific 24510
Human VEGF capture antibody R&D Systems MAB293
Human IP-10 capture antibody R&D Systems MAB266
Human IL-8 capture antibody R&D Systems MAB208
Human EGF capture antibody R&D Systems MAB636
Human MMP-9 capture antibody R&D Systems MAB936
Human IL-1β capture antibody R&D Systems MAB601
Mouse IgG1 isotype control antibody R&D Systems MAB002
StartingBlock (TBS) buffer Thermo Scientific 37542
HCl standard solution 1.0 N Sigma-Aldrich 318949-500 ml
0.5 ml microcentrifuge tubes Fisher Scientific 05-408-120
Protein-free (PBS) buffer Thermo Scientific 37572
Recombinant human VEGF 165 R&D Systems 293-VE
Recombinant human IP-10 R&D Systems 266-IP
Recombinant human IL-8 R&D Systems 208-IL
Recombinant human EGF R&D Systems 236-EG
Recombinant human MMP-9 R&D Systems 911-MP
Recombinant human IL-1β R&D Systems 201-LB
StartingBlock T20 (PBS) buffer Thermo Scientific 37539
Blocker BSA in PBS Thermo Scientific 37525
Biotinylated VEGF detection antibody R&D Systems BAF293
Biotinylated IP-10 detection antibody R&D Systems BAF266
Biotinylated IL-8 detection antibody R&D Systems BAF208
Biotinylated EGF detection antibody R&D Systems BAF236
Biotinylated MMP-9 detection antibody R&D Systems BAF911
Biotinylated IL-1β detection antibody R&D Systems BAF201
Streptavidin, R-phycoerythrin Invitrogen S-21388
Ethanol (200 proof) Sigma-Aldrich E7023-500ML

References

  1. Schena, M., Shalon, D., Davis, R. W., Brown, P. O. Quantitative monitoring of gene-expression patterns with a complementary-DNA microarray. Science. 270, 467-470 (1995).
  2. Schena, M. . Protein Microarray. , (2004).
  3. Tam, S. W., Wiese, R., Lee, S., Gilmore, J., Kumble, K. D. Simultaneous analysis of eight human Th1/Th2 cytokines using microarrays. J. Immunol. Methods. 261 (01), 157-165 (2002).
  4. Wang, C. C., et al. Array-based multiplexed screening and quantitation of human cytokines and chemokines. J. Proteome Res. 1, 337-343 (2002).
  5. de Jager, W., Velthuis, t. e., Prakken, H., Kuis, B. J., W, G. T., Rijkers, Simultaneous detection of 15 human cytokines in a single sample of stimulated peripheral blood mononuclear cells. Clin. Diagn. Lab. Immunol. 10, 133-139 (2003).
  6. Lee, H. J., Nedelkov, D., Corn, R. M. Surface plasmon resonance imaging measurements of antibody arrays for the multiplexed detection of low molecular weight protein biomarkers. Anal. Chem. 78, 6504-6510 (2006).
  7. Vignali, D. A. A. Multiplexed particle-based flow cytometric assays. J. Immunol. Methods. 243 (00), 243-255 (2000).
  8. Rissin, D. M., et al. Single-molecule enzyme-linked immunosorbent assay detects serum proteins at subfemtomolar concentrations. Nat. Biotechnol. 28, 595-599 (2010).
  9. Zhang, H., Nie, S., Etson, C. M., Wang, R. M., Walt, D. R. Oil-sealed femtoliter fiber-optic arrays for single molecule analysis. Lab Chip. 12, 2229-2239 (2012).
  10. Blicharz, T. M., et al. Fiber-Optic Microsphere-Based Antibody Array for the Analysis of Inflammatory Cytokines in Saliva. Anal. Chem. 81, 2106-2114 (2009).
  11. Mukhopadhyay, R. Devices to drool for. Anal. Chem. 78, 4255-4259 (2006).
  12. Wong, D. T. Salivary diagnostics powered by nanotechnologies, proteomics and genomics. J. Am. Dent. Assoc. 137, 313-321 (2006).
  13. Segal, A., Wong, D. T. Salivary diagnostics: enhancing disease detection and making medicine better. Eur. J. Dent. Educ. 12, 22-29 (2008).
  14. St John, M. A. R., et al. Interleukin 6 and interleukin 8 as potential biomarkers for oral cavity and oropharyngeal squamous cell carcinoma. Arch. Otolaryngol. Head Neck Surg. 130, 929-935 (2004).
  15. Herr, A. E., et al. Microfluidic immunoassays as rapid saliva-based clinical diagnostics. Proc. Natl. Acad. Sci. U.S.A. 104, 5268-5273 (2007).
  16. Pantano, P., Walt, D. R. Ordered nanowell arrays. Chem. Mater. 8, 2832-2835 (1996).
  17. Schena, M. . Protein Microarrays. , (2005).

Play Video

Cite This Article
Nie, S., Benito-Peña, E., Zhang, H., Wu, Y., Walt, D. R. Multiplexed Fluorescent Microarray for Human Salivary Protein Analysis Using Polymer Microspheres and Fiber-optic Bundles. J. Vis. Exp. (80), e50726, doi:10.3791/50726 (2013).

View Video