Using a pH Meter

JoVE 과학 교육
General Chemistry
JoVE 비디오를 활용하시려면 도서관을 통한 기관 구독이 필요합니다.  전체 비디오를 보시려면 로그인하거나 무료 트라이얼을 시작하세요.
JoVE 과학 교육 General Chemistry
Using a pH Meter

332,548 Views

09:28 min

April 30, 2023

개요

Source: Laboratory of Dr. Zhongqi He – United States Department of Agriculture

Acids and bases are substances capable of donating protons (H+) and hydroxide ions (OH), respectively. They are two extremes that describe chemicals. Mixing acids and bases can cancel out or neutralize their extreme effects. A substance that is neither acidic nor basic is neutral. The values of proton concentration ([H+]) for most solutions are inconveniently small and difficult to compare so that a more practical quantity, pH, has been introduced. pH was originally defined as the decimal logarithm of the reciprocal of the molar concentration of protons Equation 1, but was updated to the decimal logarithm of the reciprocal of the hydrogen ion activity Equation 2. The former definition is now occasionally expressed as p[H]. The difference between p[H] and pH is quite small. It has been stated that pH = p[H] + 0.04. It is common practice to use the term 'pH' for both types of measurements.

The pH scale typically ranges from 0 to 14. For a 1 M solution of a strong acid, pH=0 and for a 1 M solution of a strong base, pH=14. Thus, measured pH values will lie mostly in the range 0 to 14, though values outside that range are entirely possible. Pure water is neutral with pH=7. A pH less than 7 is acidic, and a pH greater than 7 is basic. As the pH scale is logarithmic, pH is a dimensionless quantity. Each whole pH value below 7 is 10x more acidic than the next integer. For example, a pH of 4 is 10x more acidic than a pH of 5 and 100x (10 x 10) more acidic than a pH of 6. The same holds true for pH values above 7, each of which is 10x more basic (or alkaline) than the next lower whole value. For example, a pH of 10 is 10x more basic than a pH of 9.

Principles

The pH of a solution may be accurately and easily determined by electrochemical measurements with a device known as a pH meter with a pH (proton)-sensitive electrode (usually glass) and a reference electrode (usually silver chloride or calomel). Ideally, the electrode potential, E, for the proton can be written as

Equation 3

where E is a measured potential, E0 is the standard electrode potential at aH+= 1 mol/L, R is the gas constant, T is the temperature in kelvin, F is the Faraday constant.

The pH electrode uses a specially formulated, pH-sensitive glass in contact with the solution, which develops the potential (E) proportional to the pH of the solution. The reference electrode is designed to maintain a constant potential at any given temperature, and serves to complete the pH measuring circuit within the solution. It provides a known reference potential for the pH electrode. The difference in the potentials of the pH and reference electrodes provides a millivolt (mV) signal proportional to pH. In practice, a combined glass electrode has a built-in reference electrode. It is calibrated against buffer solutions of known hydrogen ion activity. Most pH sensors are designed to produce a 0-mV signal at 7.0 pH, with a (theoretically ideal) slope, or sensitivity, of -59.16 mV / pH at 25 °C. Two or more buffer solutions are used in order to accommodate the fact that the "slope" may differ slightly from ideal. Commercial standard buffer solutions usually come with information on the pH value at 25 °C and a correction factor to be applied for other temperatures.

Procedure

1. pH Calibration

  1. Turn the meter's power on by pressing the "power" button.
  2. Attach the automatic temperature compensation (ATC) probe if it is available and/or is not with the electrode.
  3. Check that the measurement mode is pH. If not, press the "MODE" button until "pH" mode appears on the LCD display.
  4. Consult the quick reference guide at the bottom of the meter or nearby for help if needed.
  5. Always use fresh, unused, unexpired pH buffers for calibration. Buffers should be at the same temperature as the testing solutions.
  6. Rinse the pH electrode with distilled water and then with the buffer being used for calibration (i.e., pH 7.00).
  7. Dip the pH electrode into a neutral pH buffer (i.e., pH 7.00). Stir the buffer with a magnetic bar (at a moderate rate for ~30 s) for best results.
  8. Press the "CAL/MEAS" (calibration [or Standardization]/easurement) button to select the 'calibration (standardization)' function. Set the buffer pH value on the meter display to 7.00.
  9. When the "reading" is stable, press the "ENTER" button to accept. The primary reading will flash briefly before the secondary display begins scrolling through the remaining available buffers.
  10. Rinse the pH electrode with distilled water and then with the buffer to be used for next calibration (i.e., pH 4.01).
  11. Dip the pH electrode into the next buffer of pH 4.01. The meter display should be locked on the buffer value.
  12. When the "reading" is stable, press "ENTER" to accept. The primary reading will flash briefly, then display the percent efficiency (slope) before the secondary display begins scrolling through the remaining the available buffers.
  13. Repeat steps 2.2-2.7 to calibrate the pH 10.01 buffer.
  14. The meter will automatically return to measurement mode upon the successful completion of the 3-point calibration.
    Notes: (1) The standard buffers of pH 4.01 and 10.01 may be replaced with other appropriate buffers per the testing samples' pH range. (2) For a single (neutral)-, or 2-point calibration, press the "CAL/MEAS" button to return to the measurements after completion of the calibration. (3) Calibration with more than 3 points may be used for more precise measurements. (4) It is recommended to perform the calibration at the beginning of each day. For very precise work the pH meter should be calibrated before each measurement. (5) Manually adjust the pH values of the buffers if the temperature differs from the standard room temperature and no ATC probe is attached.

2. pH Measurements

  1. Confirm that the meter is on the pH measurement mode.
  2. Thoroughly rinse the pH electrode between measurements with distilled water to prevent carryover contamination of the tested solutions. Gently blot the electrode on a laboratory cleaning tissue to remove the excess rinse water. Do not rub the bulb since this can cause a static charge buildup. Alternatively, rinse the electrode with the testing solution.
  3. Dip the pH electrode into a testing solution or suspension. Stir the solution with a magnetic bar (~30 s) with the same stirring rate as for calibration for best results.
  4. The pH is completed when the pH reading is stable.
  5. If needed and available, press the "HOLD" button to freeze the measured reading. Press again to resume live reading.
  6. Record the pH value (and temperature if needed) by writing down or pressing the "MEMORY" button (if applicable) to store the value into memory.
  7. Repeat steps 3.2-3.6 for multiple measurements.
  8. Thoroughly rinse and store the electrode in storage solution once all measurements are completed.
    Notes: (1) The pH probe response time in each buffer should be no longer than 60 s, but may be longer for some testing solutions/slurries. (2) The electrode probe should be cleaned using pH-electrode cleaning solution once a month, or whenever it is dirty. A 0.1 M HCl solution can be used for general cleaning. Diluted liquid detergent and household laundry bleach may be used for cleaning grease and bacterial contaminations. However, to avoid unexpected problems, the best practice is to always refer to the electrode manufacturer recommendations. (3) The pH electrode bulb should be moist at all times. Keep it in the electrode storage solution that comes with the electrode. Use pH 4 buffer solution if no storage solution is available. Use pH 7 buffer solution for a short time if neither are available.

The pH meter is an electrical device that determines the acidity or basicity of aqueous solutions, one of the most commonly monitored parameters.

To use a pH meter, the pH electrode is first calibrated with standard buffer solutions with known pH values that span the range being measured. To make a pH measurement, the electrode is immersed into the sample solution until a steady reading is reached. The electrode is then rinsed after each sample and stored in a storage solution after all the measurements have been completed.

This video will demonstrate how to calibrate a pH meter and obtain pH measurements, as well as offer a few tips on handling the fragile electrode.

When an acid or a base is placed in water, hydrolysis reactions can occur. The amount of the hydronium ions or the hydroxide ions yielded in the reaction determines the acidity or basicity of the solution. This important property is commonly evaluated by the concentration of the hydronium ion, which is often shortened as hydrogen ion or proton. For most solutions, the hydrogen ion concentration, expressed in moles per liter, is very small, so a more practical quantity, pH, has been introduced.

pH is defined as the negative logarithm of the molar concentration of the hydrogen ion. The pH scale ranges from 0 to 14. Pure water is neutral with a pH of 7; pH less than 7 is acidic, and pH greater than 7 is basic. Since the pH scale is logarithmic, a unit decrease in pH equals a ten-fold increase in acidity.

So how does the pH meter measure pH? A key component of a pH meter is a hydrogen ion-sensitive electrode. The solution inside this electrode contains a known concentration of hydrogen ions. When the electrode is immersed in a solution of unknown pH, an electric potential develops as a function of the hydrogen ion concentration in the test solution. This hydrogen ion-sensitive electrode, along with a reference electrode with which it is often combined into one body, is connected to the pH meter, so that the developed potential can be measured and converted to the pH value.

Now that you understand the theory behind a pH meter, let's look at its use in an agricultural setting.

Before pH measurements, the pH meter is calibrated. Calibration should be performed at the beginning of each day or before each measurement if extremely precise data are required.

Choose buffers that span the range of pH values of the samples. In this demonstration, the pH meter is calibrated with three buffers with pH 4, 7, and 10. Make sure the buffers are fresh, unused, and unexpired.

To begin, turn on the pH meter by pressing the power button.

Next, plug the pH probe and automatic temperature compensation, or ATC probe, into the unit. On the LCD display, make sure the measurement mode is "pH". If not, press the "MODE" button until "pH" appears.

Then, remove the pH electrode from the storage buffer. Be careful not to allow the electrode bulb to dry out at any point during the experiment. Rinse the electrode with distilled water, followed by the pH 7 standard buffer.

Next, immerse the pH electrode into the buffer. Stir the buffer with a magnetic bar for best results. To avoid breaking the fragile electrode, be sure to keep a distance between the electrode and the stirring bar.

Press the "CAL(ibration) / MEAS(urement)" button to select the calibration function. Set the buffer pH value to 7.00. When the reading is stable, press "ENTER". The primary reading flashes briefly; then the secondary display shows the remaining buffers. The electrode is now ready to be calibrated with the next standard buffer.

Rinse the pH electrode as before, first with distilled water, then the pH 4 standard buffer. Then, place the electrode in the buffer. When the reading is stable, press "ENTER". The primary reading flashes briefly, followed by a display of the percent efficiency, or slope, before the remaining buffers are shown on the secondary display.

Finally, repeat the rinse and calibration steps with the pH 10 standard buffer. The steady reading should be pH 10.01. Once "Enter" is pressed, the 3-point calibration is done, and the meter will automatically return to measurement mode.

The device is now ready to be used to test soil samples from a potato field.

Start pH measurements by thoroughly rinsing the pH electrode with distilled water. Gently blot the electrode on a laboratory cleaning tissue to remove excess water. Be careful not to rub the bulb as it can cause a static charge buildup. The rinse step should be performed between each sample to prevent contamination.

Next, dip the pH electrode into a sample with stirring. The stirring speed should be the same as during calibration. Wait for the reading to become stable, which should take less than 60 s for most samples, then record the pH value. If needed, press the "HOLD" button to freeze the reading display. Press again to resume live reading. The pH value can be stored into memory by pressing the "MI" (or memory insert) button. The stored memory location value, or StO, will be briefly displayed.

Repeat the rinse and measurement steps as previously shown for all the remaining samples. Once all the measurements are completed, thoroughly rinse the electrode before placing it in storage solution.

In this experiment, the pH of multiple soil samples from agricultural fields was measured. pH has numerous effects on crop growth, including nutrient availability, toxicity, and disease control. Different crops have pH ranges of optimal growth. By controlling the pH, disease can be minimized while increasing yield.

The soil samples were collected from five potato fields under different cropping rotation practices with or without groundwater irrigation. Groundwater irrigation consistently increased soil pH in all five fields. These data are essential for providing liming recommendation for the potato fields.

Many fields of science require pH measurements or monitoring in their research.

For example, to use biomass more efficiently and better understand plant cell wall architecture, a series of reactions was carried out to synthesize biomimics of wood, so that plant cell wall architecture can be better understood.

In the first step, kraft pulp fiber was used to generate nanofibrillated cellulose. The pH of the reaction mixture decreased as the hydroxyl groups on the cellulose were oxidized.

The pH was continuously monitored and adjusted by adding sodium hydroxide. Once all the accessible hydroxyl groups were oxidized, the pH would no longer change and the reaction was complete.

In this environmental study, water runoff was analyzed in a facility simulating urban landscapes. Runoff has the potential to carry nutrients and sediments into local streams and lakes where they may contribute to eutrophication.

A facility with multiple field plots was constructed, and runoff water was collected. The pH of the runoff samples, along with other chemical constituents, was quantified.

In life science research, pH is also of great interest as it is strictly regulated in living organisms. In this example, fluorescent pH sensors were developed.

To calibrate these sensors in vitro, a pH titration was performed in a cuvette, where the pH of the sensor solution was measured with a microelectrode, and the emission spectra at each pH were obtained. This way, the fluorescence intensity could be plotted against the pH to generate a calibration curve.

These sensors were then used to measure pH in living cells.

You've just watched JoVE's introduction to using a pH meter. You should now understand what pH is, how the meter works, and how to use one to measure pH.

Thanks for watching!

Results

Figure 1 shows the pH of agricultural soils impacted by cropping management and groundwater irrigation. These soil samples were collected from 5 potato fields under different cropping rotation practices with or without groundwater irrigation. The lowest pH is observed in Field 4 soils in both rainfed and groundwater irrigated series. Groundwater irrigation consistently increased soil pH in all 5 fields. The pH information is essential for recommendation of liming the potato fields appropriately to promote optimal growth.

Figure 1
Figure 1. Soil pH of potato fields under different cropping management practices with or without underwater irrigation.

Applications and Summary

pH is one of the most commonly measured chemical parameters of aqueous solutions. It is a critical parameter in water and wastewater treatment for municipal and industrial applications, chemical production, agriculture research, and production. It is also critical in environmental monitoring, chemical and life sciences research, biochemical and pharmaceutical research, electronics production, and many more applications. Figure 2 lists pH values of some common substances.

Pure water is neutral, with a pH of 7.00. When chemicals are mixed with water, the mixture can become either acidic or basic. Vinegar and lemon juice are acidic substances, while laundry detergents and ammonia are basic. Chemicals that are very basic or very acidic are considered "reactive." These chemicals can cause severe burns. Automobile battery acid is an acidic chemical that is reactive. Automobile batteries contain a stronger form of one the acids found in acid rain. Household drain cleaners often contain lye, a very alkaline chemical that is also reactive.

In living systems, the pH of different cellular compartments, body fluids, and organs is usually tightly regulated in a process called acid-base homeostasis. The pH of blood is usually slightly basic with a value of pH 7.365. This value is often referred to as physiological pH in biology and medicine. Plaque can create a local acidic environment that can result in tooth decay by demineralization. Enzymes and other proteins have an optimum pH range and can become inactivated or denatured outside this range.

Figure 2

Figure 2. The pH scale and the pH values of some common items.

내레이션 대본

The pH meter is an electrical device that determines the acidity or basicity of aqueous solutions, one of the most commonly monitored parameters.

To use a pH meter, the pH electrode is first calibrated with standard buffer solutions with known pH values that span the range being measured. To make a pH measurement, the electrode is immersed into the sample solution until a steady reading is reached. The electrode is then rinsed after each sample and stored in a storage solution after all the measurements have been completed.

This video will demonstrate how to calibrate a pH meter and obtain pH measurements, as well as offer a few tips on handling the fragile electrode.

When an acid or a base is placed in water, hydrolysis reactions can occur. The amount of the hydronium ions or the hydroxide ions yielded in the reaction determines the acidity or basicity of the solution. This important property is commonly evaluated by the concentration of the hydronium ion, which is often shortened as hydrogen ion or proton. For most solutions, the hydrogen ion concentration, expressed in moles per liter, is very small, so a more practical quantity, pH, has been introduced.

pH is defined as the negative logarithm of the molar concentration of the hydrogen ion. The pH scale ranges from 0 to 14. Pure water is neutral with a pH of 7; pH less than 7 is acidic, and pH greater than 7 is basic. Since the pH scale is logarithmic, a unit decrease in pH equals a ten-fold increase in acidity.

So how does the pH meter measure pH? A key component of a pH meter is a hydrogen ion-sensitive electrode. The solution inside this electrode contains a known concentration of hydrogen ions. When the electrode is immersed in a solution of unknown pH, an electric potential develops as a function of the hydrogen ion concentration in the test solution. This hydrogen ion-sensitive electrode, along with a reference electrode with which it is often combined into one body, is connected to the pH meter, so that the developed potential can be measured and converted to the pH value.

Now that you understand the theory behind a pH meter, let’s look at its use in an agricultural setting.

Before pH measurements, the pH meter is calibrated. Calibration should be performed at the beginning of each day or before each measurement if extremely precise data are required.

Choose buffers that span the range of pH values of the samples. In this demonstration, the pH meter is calibrated with three buffers with pH 4, 7, and 10. Make sure the buffers are fresh, unused, and unexpired.

To begin, turn on the pH meter by pressing the power button.

Next, plug the pH probe and automatic temperature compensation, or ATC probe, into the unit. On the LCD display, make sure the measurement mode is “pH”. If not, press the “MODE” button until “pH” appears.

Then, remove the pH electrode from the storage buffer. Be careful not to allow the electrode bulb to dry out at any point during the experiment. Rinse the electrode with distilled water, followed by the pH 7 standard buffer.

Next, immerse the pH electrode into the buffer. Stir the buffer with a magnetic bar for best results. To avoid breaking the fragile electrode, be sure to keep a distance between the electrode and the stirring bar.

Press the “CAL(ibration) / MEAS(urement)” button to select the calibration function. Set the buffer pH value to 7.00. When the reading is stable, press “ENTER”. The primary reading flashes briefly; then the secondary display shows the remaining buffers. The electrode is now ready to be calibrated with the next standard buffer.

Rinse the pH electrode as before, first with distilled water, then the pH 4 standard buffer. Then, place the electrode in the buffer. When the reading is stable, press “ENTER”. The primary reading flashes briefly, followed by a display of the percent efficiency, or slope, before the remaining buffers are shown on the secondary display.

Finally, repeat the rinse and calibration steps with the pH 10 standard buffer. The steady reading should be pH 10.01. Once “Enter” is pressed, the 3-point calibration is done, and the meter will automatically return to measurement mode.

The device is now ready to be used to test soil samples from a potato field.

Start pH measurements by thoroughly rinsing the pH electrode with distilled water. Gently blot the electrode on a laboratory cleaning tissue to remove excess water. Be careful not to rub the bulb as it can cause a static charge buildup. The rinse step should be performed between each sample to prevent contamination.

Next, dip the pH electrode into a sample with stirring. The stirring speed should be the same as during calibration. Wait for the reading to become stable, which should take less than 60 s for most samples, then record the pH value. If needed, press the “HOLD” button to freeze the reading display. Press again to resume live reading. The pH value can be stored into memory by pressing the “MI” (or memory insert) button. The stored memory location value, or StO, will be briefly displayed.

Repeat the rinse and measurement steps as previously shown for all the remaining samples. Once all the measurements are completed, thoroughly rinse the electrode before placing it in storage solution.

In this experiment, the pH of multiple soil samples from agricultural fields was measured. pH has numerous effects on crop growth, including nutrient availability, toxicity, and disease control. Different crops have pH ranges of optimal growth. By controlling the pH, disease can be minimized while increasing yield.

The soil samples were collected from five potato fields under different cropping rotation practices with or without groundwater irrigation. Groundwater irrigation consistently increased soil pH in all five fields. These data are essential for providing liming recommendation for the potato fields.

Many fields of science require pH measurements or monitoring in their research.

For example, to use biomass more efficiently and better understand plant cell wall architecture, a series of reactions was carried out to synthesize biomimics of wood, so that plant cell wall architecture can be better understood.

In the first step, kraft pulp fiber was used to generate nanofibrillated cellulose. The pH of the reaction mixture decreased as the hydroxyl groups on the cellulose were oxidized.

The pH was continuously monitored and adjusted by adding sodium hydroxide. Once all the accessible hydroxyl groups were oxidized, the pH would no longer change and the reaction was complete.

In this environmental study, water runoff was analyzed in a facility simulating urban landscapes. Runoff has the potential to carry nutrients and sediments into local streams and lakes where they may contribute to eutrophication.

A facility with multiple field plots was constructed, and runoff water was collected. The pH of the runoff samples, along with other chemical constituents, was quantified.

In life science research, pH is also of great interest as it is strictly regulated in living organisms. In this example, fluorescent pH sensors were developed.

To calibrate these sensors in vitro, a pH titration was performed in a cuvette, where the pH of the sensor solution was measured with a microelectrode, and the emission spectra at each pH were obtained. This way, the fluorescence intensity could be plotted against the pH to generate a calibration curve.

These sensors were then used to measure pH in living cells.

You’ve just watched JoVE’s introduction to using a pH meter. You should now understand what pH is, how the meter works, and how to use one to measure pH.

Thanks for watching!