JoVE 비디오를 활용하시려면 도서관을 통한 기관 구독이 필요합니다.  전체 비디오를 보시려면 로그인하거나 무료 트라이얼을 시작하세요.
Inducing Axonal Varicosities with a Micromechanical Stimulus on Neurons

Inducing Axonal Varicosities with a Micromechanical Stimulus on Neurons

내레이션 대본

In this procedure, insert the pipette into the screw top of the micromanipulator to hold it into place. Angle the pipette at a 45-degree angle with the surface to the cell culture dish. Turn on the fluorescence filter. Open the aperture, and ensure a fluorescent spot can be seen coming through the objective.

Using the micromanipulator, move the pipette into a position that lines up the tip with the fluorescent spot, and lower the pipette into a position just above the height of the cell culture dish. Once it is in position, turn on the transmitted light and turn off the fluorescence. Using tweezers, transfer one coverslip with the cells facing toward the pipette from the cell culture plate to the cell culture dish containing 2 milliliters of Hank's buffer at room temperature.

Using the fine focus knob and 20-fold objective, focus on a plane about four full turns above the cells. Subsequently, use the micromanipulator to position the pipette tip so that it is focused and in the center left-hand side of the plane, as seen through the eyepieces. Once the pipette is in position, focus the microscope on a plane that contains the region of interest.

In the imaging of 7DIV, GFP-transfected mouse hippocampal neurons showed an axon pre- and post-puffing.

Related Videos

Read Article