New tools for mechanobiology research are needed to understand how mechanical stress activates biochemical pathways and elicits biological responses. Here, we showcase a new method for selective mechanical stimulation of immobilized animals with a microfluidic trap allowing high-resolution imaging of cellular responses.
One central goal of mechanobiology is to understand the reciprocal effect of mechanical stress on proteins and cells. Despite its importance, the influence of mechanical stress on cellular function is still poorly understood. In part, this knowledge gap exists because few tools enable simultaneous deformation of tissue and cells, imaging of cellular activity in live animals, and efficient restriction of motility in otherwise highly mobile model organisms, such as the nematode Caenorhabditis elegans. The small size of C. elegans makes them an excellent match to microfluidics-based research devices, and solutions for immobilization have been presented using microfluidic devices. Although these devices allow for high-resolution imaging, the animal is fully encased in polydimethylsiloxane (PDMS) and glass, limiting physical access for delivery of mechanical force or electrophysiological recordings. Recently, we created a device that integrates pneumatic actuators with a trapping design that is compatible with high-resolution fluorescence microscopy. The actuation channel is separated from the worm-trapping channel by a thin PDMS diaphragm. This diaphragm is deflected into the side of a worm by applying pressure from an external source. The device can target individual mechanosensitive neurons. The activation of these neurons is imaged at high-resolution with genetically-encoded calcium indicators. This article presents the general method using C. elegans strains expressing calcium-sensitive activity indicator (GCaMP6s) in their touch receptor neurons (TRNs). The method, however, is not limited to TRNs nor to calcium sensors as a probe, but can be expanded to other mechanically-sensitive cells or sensors.
The sense of touch provides animals with crucial information about their environment. Depending on the applied force, touch is perceived as innocuous, pleasurable, or painful. The tissue deformation during touch is detected by specialized mechanoreceptor cells embedded in the skin that express receptor proteins, most commonly ion channels. The steps linking force perception to ion channel activation during touch and pain are not fully understood. Even less is known about how the skin tissue filters mechanical deformation and whether mechanoreceptors detect changes in strain or stress1,2,3. This gap in understanding arises, in part, from a lack of suitable tools to apply precise mechanical stimulations to the surface of the skin of a living animal while observing the responses at the cellular level. Whereas atomic force microscopy has been used extensively to apply and measure forces in isolated cells4,5 and also to activate Piezo1 receptors in living cells6, similar experiments using living animals, especially C. elegans, have been notoriously challenging due to the intrinsic mobility of the subject. This challenge is traditionally circumvented by using veterinary- or surgical-grade cyanoacrylate glue to immobilize individual animals on agar pads1,7,8,9. This approach has been productive, but has limitations related to the skill required for immobilization by gluing and the soft agar surface on mechanical compliance. A microfluidics strategy is a complimentary alternative that avoids some of the complications linked to gluing.
The nematode C. elegans is a genetic model organism with a completely mapped nervous system that, due to the animal's size, is a good fit for microfluidics technology. Microfluidics-based devices offer the advantage that the otherwise extremely mobile animals can be restrained while performing high-resolution imaging and delivery of relevant neuro-modulatory stimuli. With the help of microfluidic technologies, living animals can be immobilized without harm10,11, enabling monitoring of behavioral activity over the entire lifetime12,13 and high-resolution imaging of neuronal activity14,15,16,17. Further, many mechanoreceptor neurons needed for the sense of touch and pain can be characterized on their physiological1,8, mechanical4,18,19, and molecular level20,21,22.
C. elegans senses gentle mechanical stimuli to its body wall using six TRNs, three of which innervate the animal's anterior (ALML/R and AVM) and three of which innervate the animal's posterior (PLML/R and PVM). The ion channel molecules needed for transducing an applied force into a biochemical signal have been extensively studied in its TRNs8. This article presents a microfluidic platform23 that enables researchers to apply precise mechanical forces to the skin of an immobilized C. elegans roundworm, while reading out the deformation of its internal tissues by optical imaging. In addition to presenting well-defined mechanical stimuli, calcium transients can be recorded in mechanoreceptor neurons with subcellular resolution and correlated with morphological and anatomical features. The device consists of a central trapping channel that holds a single animal and presents its skin next to six pneumatic actuation channels (Figure 1 and Figure 2). The six channels are positioned along the trapping channel to deliver mechanical stimuli to each of the worm's six TRNs. These channels are separated from the trapping chamber by thin PDMS diaphragms, which can be driven by an external air pressure source (Figure 1). We calibrated the deflection with respect to pressure and provide the measurements in this article. Each actuator can be addressed individually and used to stimulate a mechanoreceptor of choice. The pressure is delivered using a piezo-driven pressure pump but any alternative device can be used. We show that the pressure protocol can be used to activate TRNs in vivo and demonstrate operating devices suitable for delivering mechanical stimuli to adult C. elegans, loading adult animals into devices, performing calcium imaging experiments, and analyzing the results. Device fabrication consists of two main steps: 1) photolithography to make a mold from SU-8; and 2) molding PDMS to make a device. For the sake of brevity and clarity, readers are referred to previously published articles and protocols24,25 for instructions on how to produce the molds and devices.
1. Device Fabrication
2. Preparation of the Microscope
NOTE: Transgenic animals: express a calcium indicator such as GCaMP6s28 or other genetically encoded activity probe in the neuron(s) of interest (e.g., TRN); co-express an activity-independent fluorescent protein emitting at a different wavelength to correct for small lateral and out-of-focus movement artifacts that arise due to mechanical stimulation. The automated analysis software tracks and compensates for movement-induced changes in the intensity of the activity probe. The worm strains GN69223 (or AQ323629) express GCaMP6s (or GCaMP6m) and the calcium-independent tagRFP under the control of mec-7 promotor. Additionally, GN692 contains the mutation lite-1(ce314), which prevents activation of TRNs due to the blue light sensor lite-130 during excitation of the GCaMP6s fluorescence23.
3. Animal Preparation
4. Analysis
SU-8 Lithography and Chip Bonding
The lithography protocol and PDMS molding follow standard procedures. Details can be found elsewhere23,24,25,26. The PDMS should peel off the wafer without problems after curing. If the SU-8 features rip off during PDMS peeling, either the SU-8 adhesion layer or the silanization was insufficient. If plasma-activation of the glass coverslips and the PDMS chips is not sufficient, weak bonding can result in leakage out of the trapping channel or pneumatic actuators resulting in performance degradation and absence of stimulation. If fluid leakage is evident, the chip is unusable.
Actuator Performance
All of the PDMS actuators and the tubing connecting the chip to the air reservoir must be properly sealed to ensure proper pneumatic actuation of the diaphragms. Repeated insertion and removal of the PE tubing (plumbing) will damage PDMS holes and lead to a decrease in pressure. Such a loss of pressure will limit actuator deflection and the force applied to the skin of the immobilized animal. Deflection of the diaphragm can be measured using an empty chip (no animal present) by performing several actuation cycles with the desired pressure and comparing the quantitative values with theoretical predictions. The deflection of the diaphragm was not measured in presence of an animal in the trapping channel, because the membrane and worm boundary are difficult to visualize. This makes precision calibration with a trapped animal not possible. Table 1 lists the expected deflection as a function of pressure. These values were calculated using elastic plate theory and details can be found in Nekimken et al.23 This relationship is expected to vary with differences in diaphragm thickness generated during chip casting and other factors such as the integrity of the bonding between the chip and the glass and seals connecting the chip to the pressure sources. Here, values were reproducible with slight variation not exceeding 1 µm at 450 kPa of pressure. We have characterized the performance of the diaphragms for all three different stimulus profiles and refer the interested reader to Nekimken et al.23 In summary, there are several factors to consider if the values deviate from Table 1: 1) a leak in the tubing or tubing connectors, 2) the PDMS has a different elasticity, which can result from an alternative ratio of base polymer and curing agent, or 3) the PDMS has aged and continued to crosslink over time. Therefore, use of chips which have been prepared within a month of their application is recommended.
Trapping Performance
After inserting the worms into the inlet of the chip, a gentle pressure with the syringe should place one single animal inside the trapping channel and present its skin alongside the six actuators (Figure 1). Importantly, the nose of the animal does not necessarily need to protrude into the buffer reservoir. Instead, it is more important to position the animal such that the neurite is adjacent to the nearest actuator and its cell body is within the field of view of the microscope. To achieve optimal neuronal activation, the distance between the actuator and cell body should be adjusted such that the actuator lies anterior to the cell body of interest. In our experience, for neurons stimulated posterior to the cell body, no activation will take place and calcium transients will be absent. Proper immobilization can be achieved with young adult or one-day old hermaphrodites (devices optimized for these two ages are available on the same mask). Smaller animals will slip into the collection reservoir or move laterally inside the trapping channel. Animals that are too large will be squeezed into the channel, which might result in premature activation of its TRNs and subsequent failure to detect mechanoreceptor transients. Also, removal of large worms is challenging, leading to death of the animal and/or clogging of the device. With this design, the PLM neuron usually cannot be immobilized completely, so it is free to move laterally and vertically. Although we have successfully activated PLM neurons, recording calcium transients in these neurons is difficult due to vertical movement of the tail. Another trapping design has been used to record from PLM33.
Imaging of Calcium Transients and Analysis
Once in place, the animals can be stimulated using one of the six actuators. The six actuators are positioned to deliver mechanical stimuli to each of the six TRNs. A microfluidic flow controller connected to a 450-kPa in-house pressure source was used to deliver to the animal a stimulus protocol consisting of a 2-s 275 kPa step, a 2-s 0–275 kPa ramp, and a 2-s buzz (a 75 kPa, 10 Hz sine superimposed with a 275 kPa step), each separated by a 10-s waiting period. The simultaneously recorded image sequences of the calcium transients were analyzed in Fiji, using a custom-written software available on our GitHub account (see above and https://github.com/HFehlauer/Poking-Analyzer). We found that pressure ramps and pressure steps activated TRNs only if the stimuli achieved pressures above 400 kPa. By contrast, a robust activation of all TRNs stimulated using a sinusoidal, 10-Hz buzz at pressures < 275 kPa was observed (Figure 4). In general, TRNs responded better to a brief sinusoidal stimulus (called 'buzz'), in agreement with previous reports using different methodologies1,9. The stimulation with a buzz is highly efficient, leading to ~ 90% of activation in all neurons and trials tested (Figure 4). Surprisingly, there was no strong correlation observed on whether the stimulus was applied to the dorsal or the ventral side, and was independent of the distance between the cell body and the stimulus position on the neurite. Future studies will have to investigate whether the delay of the maximum response correlates with the stimulus distance.
Figure 1: Overview of the chip design. (A) Schematic layout of the photomask and the resulting chip with all connectors indicated. There are six actuation fingers, three on each side of the central trapping channel. Scale bar = 3 mm. (B) Photograph of the microfluidic device with PDMS bonded to the coverslip and connected to tubing in the worm inlet and the worm outlets. Only one of the actuation fingers is connected to the pressure source, while the other five are not occupied. Parts of the figure have been reproduced from reference23 under creative common licenses. Please click here to view a larger version of this figure.
Figure 2: Close-up of the central trapping channel and trapping procedure. (A) Schematic drawing of the trapping channel. The pillars at the worm inlet side help to orient the worm and facilitate loading of the animals. The actuation fingers can be pressurized and indent a thin diaphragm into the trapped specimen. Scale bar = 200 µm. (B) Representative video frame of a worm being trapped in the horizontal channel. The channel is designed such that the worm comes to lay across the six pneumatic actuators. Tail is to the left, head is to the right. Location of individual touch receptor neurons (TRNs) are indicated by arrows and labeled in the figure. Scale bar = 100 µm (in all panels). Parts of the figure have been reproduced from reference23 under creative common licenses. Please click here to view a larger version of this figure.
Figure 3: Overview of the provided Fiji plugin to analyze calcium traces acquired with this protocol. The software starts a graphical user interface. The first tab ("Open video") of the user interface shows the button to open a video, it displays the path and the first frame of the opened video. The second tab ("Define ROIs") includes the button to define a touch receptor neuron, a checkbox for activating the button "Define Actuator" and the button "Define Actuator". In the lower part, the first frame of the video is displayed. In the upper part of the third tab ("Analysis"), adjustable parameters for the image analysis are displayed. The middle part shows the "Start Analysis" button. In the lower part, a progress bar shows the progress of the current analysis. In the fourth tab, "Results", the result is displayed as a graph of the relative fluorescence intensity over time. In the lower part, there is a button to save the result table. In the fifth tab, "Help", a help text is displayed showing the requirements for the analysis software and instruction on how to use it. Figure adapted from reference23 under creative common licenses. Please click here to view a larger version of this figure.
Figure 4: Individual neurons respond to mechanical stimuli. (A) Representative fluorescence micrograph of GCamP6s labeled AVM touch receptor neuron before and after mechanical stimulation with a buzz. Scale bar = 10 µm. Color scale = 1,500-3,500 gray values. (B) Stimulus protocol including 2 s diaphragm excitation representing a 275 kPa step, a 275 kPa ramp, and a sine (75 kPa; 10 Hz) superimposed with a 275 kPa step (buzz). (C) Normalized GCamP6s intensity trace (mean ± SEM as shaded area) recorded from AVM stimulated with the profile shown in B of wild type animals (green, n = 14) and animals mutant in the mechanoreceptor channel subunit mec-4 (blue, n = 10), that is known to facilitate responses to mechanical stress. This indicates that the transients are induced by the applied mechanical stimuli. Parts of the figure have been reproduced from reference23 under creative common licenses. Please click here to view a larger version of this figure.
Pressure (kPa) | Mean Deflection (µm) | SD | Expected Deflection (µm) |
0 | -0.01 | 0.01 | 0 |
50 | 0.77 | 0.26 | 0.607143 |
100 | 1.57 | 0.58 | 1.21429 |
150 | 2.32 | 0.65 | 1.82143 |
200 | 3.13 | 0.74 | 2.42857 |
250 | 3.86 | 0.79 | 3.03571 |
300 | 4.61 | 0.79 | 3.64286 |
350 | 5.3 | 0.82 | 4.25 |
400 | 5.92 | 0.82 | 4.85714 |
450 | 6.43 | 0.75 | 5.46429 |
Table 1: Expected experimental deflection values for applied pressures ranging from 0–450 kPa and their theoretical predictions. Data derived for a step stimulus. Mean ± SD. These data were acquired with an empty trapping channel.
This protocol demonstrates a method for delivering precise mechanical stimulation to the skin of a roundworm trapped in a microfluidic chip. It is intended to facilitate the integration of physical stimuli for answering biological questions and aims to streamline mechanobiology research in biological labs. This method extends previous assays to assess the function of mechanosensory neurons in C. elegans. Previous quantitative and semi-quantitative techniques measured forces1,34 and behavior35, but were difficult to integrate with high-resolution imaging of neuronal activity. We thus believe that the use of this chip extends current applications and its performance is optimized for stimulation of neurons required for gentle touch. We optimized the deflection of the actuator to reach 5 µm, which is a strong enough stimulus to reach nearly maximal neuronal responses. With slight variations, we believe that this design could be used to stimulate harsh touch receptors, such as PVD nociceptors that innervate the body surface36.
Modifications and Troubleshooting
If no clean room is available for SU-8 lithography, a benchtop clean box or SU-8 soft-lithography stations equipped with UV lamps, programmable hotplates, and spin coaters can be used. A low cost alternative for production of microfluidic devices that avoids the use of clean room facilities can also be followed26.
In the design, we implemented an existing trapping technique, which immobilizes a single animal in a microfluidic channel. The trap design is adapted from a device that is used to study olfactory sensation in the worm36, but it is not the only viable option to suppress worm movement. Other strategies have been reported to hold worms in microfluidic devices: CO2, electrophoretic, and compressive immobilization11,37,38 have been used, and could be integrated into a modified design.
We delivered pressure to the actuation channels using a commercially-available, piezoelectric pressure pump which can control and deliver up to 800 kPa from an external pressure source. We connected the pump to compressed air in the research building, which is capable of delivering ~450 kPa. If no source of compressed air is available, a tank of compressed, inert gas (e.g., N2) could be used. This would have the added benefit of delivering higher pressures and generating large deformations, but would also require high-pressure fittings to connect to the pressure controller to the chip. Fabricating the chip from softer PDMS is alternative way to increase pressure-induced deformations.
In this experiment, the outlet of the pressure pump is connected to the PDMS chip through 1-mm receiving openings (e.g., punch holes) plugged with adhesive-free and reversible interconnectors consisting of a press-fit 20 G metal-tube inserted in a PE tubing. These have been shown to resist pressures up to 700 kPa39. Care must be taken during repeated insertion, however, as the PDMS around the holes tends to tear and thus limit the performance or make the chip unusable. For this reason, the metal tubing was inserted once and left plugged into the chip, without removing and inserting it repeatedly.
Critical Steps
Stimulation is delivered by a pressurized actuation channel, which indents the skin. The resulting deformation and mechanoreceptor neuron activation can be imaged with high resolution on an optical fluorescence microscope and/or confocal set-up. Several factors can hamper the detection of the GCaMP responses. First, C. elegans express a blue-light receptor30 in many neurons, including TRNs21,23, which activates cells after illumination with blue (488 nm) light. Thus, imaging C. elegans using the genetically encoded calcium sensor GCaMP6s will activate its neurons upon illumination, obscuring calcium transients activated by mechanical stimulation23. Working in the lite-1 mutant background or using red-shifted calcium activity sensors avoids this confound. Second, mechanically-induced calcium transient might not be detected if the stimulus profile does not match the sensitivity of the mechanoreceptors. We failed to detect activation for high-amplitude steps and ramps, but evoked extremely high signals after stimulation with moderate dynamics stimuli (e.g., 200 -kPa buzz). Finally, TRNs habituate to a repeated stimuli delivered with short (<60 s) interstimulus intervals. Thus, it is important to design stimulation sequences that minimize habituation except in cases where habituation is being studied.
We recommend filtering all solutions, as debris and dust in the buffers can easily clog the small trapping channel. Further, animals must be age-synchronized and size-matched to the chip's design; animals smaller than young adults will slide through the device or fail to be restrained and larger animals cannot enter the trapping channels and are in danger of bursting and/or clogging the chip. In case of clogging, application of excessive pressure to the worm or gravity flow inlet may clean the channels, but could also lead to failure of the fitting.
As mentioned in the results section, several factors can lead to a failure in the analysis program. If problems are encountered in the compiling the code, verifying that the most recent Fiji software is running and that Fiji is using the newest Java compiler is recommended. If the program is running but there are unexpected results please make sure that the recorded movie has a 50-frame buffer before the actual mechanical stimulation is performed to ensure a proper estimation of the baseline and background fluorescence. Further, the field of view of the green, activity-dependent (GCaMP6s) and the red, activity-independent (tagRFP) channel are assumed to be projected onto the upper and lower half of the camera chip, respectively. If the optical beam splitter projects the images onto the left and right half of the sensor, turn the camera 90° to obtain the desired orientation during the recording or rotate the image stacks in a pre-processing step. Furthermore, the program identifies the neuron by its fluorescence and will fail if the signal-to-noise ratio or contrast is too low. In this case, changing the excitation intensity and/or exposure time may solve the problem. The program will also stop if the neuron leaves the field of view during the recording, in which case the neuron should be positioned in the center of the field of view before starting a new acquisition.
Limitations of the Technique
The present design enables stimulation of the body wall mechanoreceptor neurons like the PVD nociceptors and the TRNs, but not mechanoreceptors that innervate the worm's head region, such as FLP and ASH. It is limited to pressures <700 kPa. For larger pressures, different tube fittings need to be employed. Since the pressure supply here did not exceed 500 kPa, the maximum pressure possible to apply was limited to 450 kPa. The calcium transients observed here mark the upper boundary of the expected signal. Another limitation for the design of our actuators is their aspect ratio. In principle, thinner diaphragms would enable larger deformations, but structures with high aspect ratio are difficult to fabricate reliably. If larger deformations are essential, wider actuators or parallel deformations from both sides might be useful, as presented by Cho et al.29
Although we can monitor diaphragm deformation, it is difficult to measure the forces applied during pneumatic stimulation of the actuation channels. The principle limitation is that the contact area of the actuator is not well-defined. This is because the actuator itself is elastic and deforming during actuation, which changes the contact radius with increasing deflection. With these caveats in mind, we can estimate that a 5 µm indentation (around 300 kPa actuation pressure in an empty chip) is predicted to apply about 3.8 µN to the worm according to the worm stiffness estimates of Petzold et al.34 Since the force dependence of the neuronal current varies in worms with different stiffnesses1,34, however, monitoring indentation as a measure of stimulus intensity instead of force is recommended.
We hope that the dissemination of this protocol will help users take advantage of microfluidics for mechanobiology research using C. elegans as a model system, and thus accelerate the advancement of science on this important subject in the field.
The authors have nothing to disclose.
We thank Sandra N. Manosalvas-Kjono, Purim Ladpli, Farah Memon, Divya Gopisetty, and Veronica Sanchez for support in device design and generation of mutant animals. This research was supported by NIH grants R01EB006745 (to BLP), R01NS092099 (to MBG), K99NS089942 (to MK), F31NS100318 (to ALN) and received funding from the European Research Council (ERC) under the European Union's Horizon 2020 research and innovation program (grant agreement No. 715243 to MK).
Chrome mask | Compugraphics (http://www.compugraphics-photomasks.com/) | 5'', designed in AutoCAD (Autodesk, Inc.) | |
Chrome mask | Mitani-Micronics (http://www.mitani-micro.co.jp/en/) | 5'', designed in AutoCAD (Autodesk, Inc.) | |
Chrome mask | Kuroda-Electric (http://www.kuroda-electric.eu/ | 5'', designed in AutoCAD (Autodesk, Inc.) | |
4'' Silicon wafer (B-test) | Stanford Nanofabrication Facility | ||
SU-8 2002 | MicroChem | ||
SU-8 2050 | MicroChem | ||
Spin-coater | Laurell Technologies | WS-400BZ-6NPP/LITE | |
Exposure timer | Optical Associates, Inc | OAI 150 | |
Illumination controller | Optical Associates, Inc | 2105C2 | |
SU-8 developer | MicroChem | ||
2-Propanol | Fisher Scientific | A426F-1GAL | |
Acetone | Fisher Scientific | A18-4 | |
Trichloromethylsilane (TCMS) | Sigma-Aldrich | 92361-500ML | Caution: TCMS is toxic and water-reactive |
Sylgard 184 Elastomer Kit | Dow Corning | PDMS prepolymer | |
Biopsy punch, 1 mm | VWR | 95039-090 | |
Oxygen Plasma Asher | Branson/IPC | ||
Small metal tubing (0.635 mm OD, 0.4318 mm ID, 12.7 mm long); gage size 23TW | New England Small Tube Corporation | NE-1300-01 | |
Nalgene syringe filter, 0.22 μm | Thermo Scientific | 725-2520 | to filter all solution, small particles would clog the chip |
Polyethylene tubing; 0.9652 mm OD, 0.5842 mm ID | Solomon Scientific | BPE-T50 | |
Syringe, 1 ml | BD Scientific | 309628 | for worm trapping and release |
Syringe, 20 ml | BD Scientific | 309661 | for gravity-based flow |
Gilson Minipuls 3, Peristaltic pump | Gilson | to suck solutions and worms out of the chip | |
Microfluidic flow controller, equipped with 0–800 kPa pressure channel | Elveflow | OB1 MK3 | pressure delivery |
Water-Resistant Clear Poly- urethane Tubing, 4 mm ID and 6 mm OD | McMaster-Carr | 5195 T52 | connection from house air to pressure pump |
Water-Resistant Clear Polyurethane Tubing, 2.6mm ID and 4mm OD | McMaster-Carr | 5195 T51 | connect pressure pump to small tubng |
Push-to-Connect Tube Fitting for Air | McMaster-Carr | 5111K468 | metric – imperial converter |
Straight Connector for 6 mm × 1/4″ Tube OD | McMaster-Carr | 5779 K258 | |
Leica DMI 4000 B microscopy system | Leica | ||
63×/1.32 NA HCX PL APO oil objective | Leica | 506081 | |
Hamamatsu Orca-Flash 4.0LT digital CMOS camera | Hamamatsu | C11440-42U | |
Lumencor Spectra X light engine | Lumencor | With cyan and green/yellow light source | |
Excitation beam splitter | Chroma | 59022bs | in the microscope |
Hamamatsu W-view Gemini Image splitting optics | Hamamatsu | A12801-01 | to split green and red emission and project them on different areas on the camera chip |
Emission beam splitter | Chroma | T570lpxr | in the image splitter |
Emission filters GCamp6s | Chroma | ET525/50m | in the image splitter |
Emission filters mCherry | Chroma | ET632/60m | in the image splitter |