Procedures are described to perform simultaneous recordings of membrane potential or current and changes of intracellular calcium concentration. Suprachiasmatic nucleus neurons are filled with the calcium indicator bis-fura-2 using a patch clamp electrode in the whole cell patch clamp configuration.
Simultaneous electrophysiological and fluorescent imaging recording methods were used to study the role of changes of membrane potential or current in regulating the intracellular calcium concentration. Changing environmental conditions, such as the light-dark cycle, can modify neuronal and neural network activity and the expression of a family of circadian clock genes within the suprachiasmatic nucleus (SCN), the location of the master circadian clock in the mammalian brain. Excitatory synaptic transmission leads to an increase in the postsynaptic Ca2+ concentration that is believed to activate the signaling pathways that shifts the rhythmic expression of circadian clock genes. Hypothalamic slices containing the SCN were patch clamped using microelectrodes filled with an internal solution containing the calcium indicator bis-fura-2. After a seal was formed between the microelectrode and the SCN neuronal membrane, the membrane was ruptured using gentle suction and the calcium probe diffused into the neuron filling both the soma and dendrites. Quantitative ratiometric measurements of the intracellular calcium concentration were recorded simultaneously with membrane potential or current. Using these methods it is possible to study the role of changes of the intracellular calcium concentration produced by synaptic activity and action potential firing of individual neurons. In this presentation we demonstrate the methods to simultaneously record electrophysiological activity along with intracellular calcium from individual SCN neurons maintained in brain slices.
Changes in gene expression are known to occur in neurons as a consequence of synaptic signaling. Signaling by the excitatory neurotransmitter glutamate can depolarize the neuronal membrane potential eventually leading to gene transcription and translation1,2. Activation of ionotropic receptors by glutamate allows extracellular calcium ions to enter the cell, which is thought to play a critical role as a second messenger in activating gene transcription. Evaluating the relationship between membrane electrical activity, such as action potential firing frequency, and changes of intracellular calcium concentration requires the combination of two methods – whole cell patch clamping and quantitative imaging of fluorescent calcium probes3-5, allowing the relationship to be studied in individual neurons. The single cell recording technique allows the recording of the activity of individual neurons in identifiable portions of the brain. The whole cell recording technique allows the membrane voltage or current to be controlled allowing for experimental manipulation of specific ion channel currents. Using micropipettes filled with fluorescent calcium probes also ensures that the neuron is well filled with calcium probe. This technique has a clear advantage when working with brain slice preparations from adult slice preparations, since these neurons are particularly difficult to load using the more common cell permeant probes and reduces potential background fluorescence issues6-8.
Light is the principal way mammals adjust their circadian clock, which is located in the hypothalamic suprachiasmatic nucleus (SCN). Light information transduced in the retina is transmitted 9-11 via the retinohypothalamic tract (RHT) where glutamate is released in the SCN12,13. Glutamate opens NMDA and AMPA ionotropic receptors located on SCN neurons producing an influx of calcium and sodium, and initiating an intracellular signaling cascade that ultimately leads to altering the expression of a family of clock genes14-17 and shifts in phase of the circadian clock18-20. However, calcium can enter neurons either directly through ionotropic glutamate receptors or through membrane depolarization and activation of voltage-dependent calcium channels (VDCC)21. We therefore developed an experimental protocol to investigate the relationship between the intracellular calcium concentration and membrane electrical activity in SCN neurons, such as occurs with action potential firing and from synaptic input22.
1. Preparation of Hypothalamic Brain Slices
2. Patch Clamp Recording
3. Measurement of Intracellular Ca2+
Using hypothalamic brain slices, we simultaneously recorded changes in intracellular calcium in whole cell mode under both voltage and current clamp conditions. The microelectrode shown in Figure 1A is lowered into position under high magnification (40X or 63X UV objectives) with a small amount of applied positive pressure. After touching the neuron, a gigaohm seal is formed with gentle suction. In voltage-clamp mode after setting the cell membrane potential to -60 mV, additional suction breaks the membrane creating the whole cell configuration. The neuron will rapidly fill with probe when viewed on the monitor. Dendritic processes in a similar focal plane as the soma can be visualized (Figure 1A inset). Alternatively, the focus could be adjusted to measure calcium changes in distal dendrites. Recording in voltage clamp mode at -60 mV we simulated action potential firing using brief (2 msec) steps to +20 mV at 10 Hz, which is in the physiological range for the daytime action potential firing rate of many SCN neurons (Figures 1B and 1C). Different regions of interest (ROI) of the neuron including the proximal dendrite were quantified and the background at each wavelength was subtracted to determine the ratio of emitted fluorescence from excitation at 340 nm and 380 nm of light (Figure 1B, inset). The bis-fura-2 filled microelectrode can be seen but is mainly out of the focal plane. Pseudocolor applied to the image with software can sometimes facilitate visualization. Differences in the rate of intracellular calcium influx can be observed between different regions of the neuron and the magnitude of the response increased with depolarizing current steps. Combining location data from images taken with 63X and 4X objectives (Figure 1A) we can assess the anatomical location within the SCN from each neuron recorded.
Similarly, recording in current-clamp mode the intracellular calcium response to changes in membrane potential can be measured. A small change of current near the action potential firing threshold can induce action potential firing and a corresponding elevation of calcium (Figure 2). Recorded in the soma, the calcium concentration increase lags behind the changes in membrane voltage but decays after cessation of action potential firing to a new plateau. Using these methods we have previously shown that synaptically-evoked excitatory postsynaptic potentials (EPSPs) do not significantly alter somatic intracellular calcium without subsequently induced action potential firing22, suggesting that postsynaptic action potential firing is required for calcium to affect subsequent transcriptional and translation events which may represent a way these neurons differentiate synaptic input signaling at different times of the day and night.
Figure 1. Calcium imaging of a voltage clamped SCN neuron. A: Bright field image of a coronal slice of hypothalamus containing the SCN and optic chiasm (OC). A glass microelectrode filled with a bis-fura-2 containing internal solution is sealed onto a single neuron. Inset: Recorded neuron with bis-fura-2 fluorescence (380 nm excitation) images in grayscale and pseudocolor. B: Calcium responses to a series of 50 voltage steps recorded from a voltage-clamped SCN neuron. Voltage-gated sodium currents were inhibited with TTX (0.5 µM). Inset: Screen capture image of somatic and dendritic regions of interest corresponding to the calcium responses shown in the graph. C: Calcium responses to a variable number of voltage steps. The right recording is the same data shown in B. Click here to view larger image.
Figure 2. Simultaneous recording of membrane voltage and intracellular calcium in an SCN neuron. The current-clamp experiment was performed in the presence of picrotoxin (50 µM). A small depolarization of the membrane potential triggers several bouts of action potential firing with a corresponding rise and subsequent fall of intracellular calcium. Click here to view larger image.
The methods described above provide a powerful tool to simultaneously record the link between neuron membrane electrical activity and the intracellular calcium concentration. The method has a number of strengths in that it combines two very well characterized methods – whole cell patch clamp recording and measurement of intracellular calcium using fluorescent dyes. Our approach is similar to those described by other investigators6,23.
A number of items must be taken into consideration when using the method. First is the use of the whole cell patch configuration, which leads to the dialyzation of the cellular contents with the intracellular microelectrode solution. The dialyzation can lead to the rundown of voltage-gated ion channels and the loss of second messenger mediated signals24-26. Second, cell injury or damage to the gigaohm seal can result in increasing levels of calcium that may be associated with decreasing cellular input resistance24. Third, the intercellular solution containing the probe should be filtered just before filling the microelectrode to avoid plugging the microelectrode tip. By minimizing these factors, intracellular calcium and membrane electrical activity may be simultaneously recorded in patch clamped neurons for more than 30 min. We have had some success using the beta-Escin perforated patch method27 and bis-fura-2 to slow dialyzation and increase recording time. Fourth, calcium and a chelator EGTA or BAPTA are common constituents of intercellular solutions used to stabilize the intracellular calcium concentration. However, we have not included these under our recording conditions and dialyzing the neuron may alter calcium homeostasis and lower the basal calcium levels. We also did not add magnesium to the internal solution since it can also bind to fura-2. However, since fura-2 binds Mg2+ with a much lower affinity than calcium (Kd 5.6 mM)28 this may be optional. Fifth, the affinity of bis-fura-2 for calcium can be altered by a number of intracellular factors within a cell including pH and cell viscosity4,29. Finally, bis-fura-2 can buffer the intracellular calcium concentration and this should be considered when deciding on the probe concentration in the internal solution5. Thus the constituents of the internal solution may be modified based on neuronal type and conditions, experimental questions being addressed, and the equipment available for the study.
It is important to note that quantitative calcium imaging using bis-fura-2 typically uses the ultraviolet light wavelengths at 340 nm and 380 nm. Protective equipment must be used to avoid injury to eyes (UV eye protection) and skin (lab coats, etc.). In addition, UV producing bulbs may produce ozone requiring adequate ventilation especially important since such equipment is often located in small rooms where ambient light can be controlled.
The affinity of the calcium probe for calcium is an important consideration. If a high affinity probe is chosen then large calcium currents may saturate the probe leading to an under estimation of the magnitude of the calcium signal5. Whereas a low affinity probe may be more useful in regions where high calcium concentrations occur such as in distal dendritic processes and spines. Similarly, low affinity probes may miss changes at low calcium concentrations. In general, somatic changes of calcium can be measured using a quantitative ratiometric calcium probe such as bis-fura-2 or fura-2.
Data may be presented as a simple ratio quantified from a region of interest from a pair of images following background (BG) subtraction at each wavelength. While not as quantitative, ratiometric probes may also be excited at a single wavelength as one would use other single excitation wavelength probes for experiments requiring very fast recording time frames or if a monochronometer or other wavelength changing (e.g. filter wheel) device is not available.
We chose to use an in vitro calibration to obtain the parameters needed to estimate the intracellular calcium concentration along with a published Kd. However, more detailed techniques are available for in situ calibrations and to measure the Kd of the probe under specific experimental conditions. We used the excitation wavelengths of 340 nm and 380 nm for bis-fura-2 and estimated the intracellular Ca2+ concentration using the formula: Est[Ca2+]i = (R-Rmin)/(Rmax-R) x (Ff380/Fb380) x Kd, where F is fluorescence intensity at 380 nm of the probe in it's free (f) and bound (b) states, R is the emission ratio following excitation at 340 nm divided by the emission following excitation at 380 nm (i.e. F340/F380), Rmin is the minimum (Ff340/Ff380) and Rmax the maximum (Fb340/Fb380) ratio obtained in the free and bound states respectively, and Kd is the Ca2+ disassociation constant of the indicator4. Note that both Kd and (Ff380/Fb380) are scalar terms while Rmin and Rmax can affect the magnitude of the change in calcium. Care must be taken when estimating the concentration values for calcium. There are a number of unknowns that may be altered by the intracellular chemical environment that may affect the values in the equation.
Table 1. Slicing solution.
Chemicals | Final Conc (mM) | FW (g/M) | Final (g/L) |
NaCl | 120 | 58.44 | 7.013 |
KCl | 2.5 | 74.55 | 0.186 |
MgCl2 | 5 | 203.31 | 1.017 |
CaCl2•2H2O | 0.5 | 147.02 | 0.074 |
NaH2PO4•H2O | 1.2 | 137.99 | 0.166 |
NaHCO3 | 26 | 84.01 | 2.184 |
D-Glucose | 10 | 180.16 | 1.802 |
Table 2. Recording solution.
Chemicals | Final Conc (mM) | FW (g/M) | Final (g/L) |
NaCl | 120 | 58.44 | 7.013 |
KCl | 2.5 | 74.55 | 0.186 |
MgCl2•6H2O | 1.2 | 203.31 | 0.244 |
CaCl2•2H2O | 2.4 | 147.02 | 0.353 |
NaH2PO4•H2O | 1.2 | 137.99 | 0.166 |
NaHCO3 | 26 | 84.01 | 2.184 |
D-Glucose | 10 | 180.16 | 1.802 |
HEPES (optional) | 10 | 238.3 | 2.383 |
Table 3. Intracellular (microelectrode) solution.
Chemicals | Final Conc (mM) | FW (g/M) | Final (mg/25 ml) |
KCl | 5 | 74.55 | 9.32 |
Potassium D-gluconate | 140 | 234.2 | 819.70 |
HEPES | 10 | 238.3 | 59.58 |
Adenosine 5'-triphosphate dipotassium salt | 4 | 583.4 | 58.34 |
Guanosine 5'-triphosphate tris salt | 0.4 | 523.2 | 5.23 |
Table of Specific Reagents and Equipment:
Name of Reagent | Company | Catalog Number | Comments |
NaCl | Fisher Scientific Co. | S271-3 | |
KCl | Fisher Scientific Co. | P217-500 | |
NaH2PO4•H2O | Sigma Chemical Co. | S-9638 | |
MgCl2•6H2O | Fisher Scientific Co. | M33-500 | |
CaCl2•2H2O | Fisher Scientific Co. | C79-500 | |
D-glucose | Fisher Scientific Co. | D16-500 | |
NaHCO3 | Fisher Scientific Co. | S233-500 | |
Sucrose | Fisher Scientific Co. | S5-500 | |
Potassium D-gluconate | Sigma-Aldrich | G4500 | |
HEPES | Sigma-Aldrich | H4034 | |
Adenosine 5′-triphosphate dipotassium salt dihydrate | Sigma-Aldrich | A8937 | |
Guanosine 5′-triphosphate tris salt | Sigma-Aldrich | G9002 | |
Agarose (ultrapure) | Life Technologies | 15510-027 | Gel pored into sterile Petri dish 6 mm thick layer |
KOH | Sigma-Aldrich | P-6310 | |
NaOH | Sigma-Aldrich | S-5881 | |
bis-fura-2, hexapotassium salt | Invitrogen | B6810 | cell impermeant |
Tetrodotoxin | Alomone Labs | T-550 | |
Picrotoxin | Sigma-Aldrich | P-1675 | |
Name of Equipment | Company | Catalog Number | Comments |
Microtome | Leica | VT1000S | Tissue slicing |
Cyanoacrylic glue (Roti-Coll1) | Carl Roth GmbH+Co | Art-Nr. 0258.1 | |
Microelectrode puller | Narshige International USA | PP-83 | |
Microelectrode Capillary Tubes | World Precision Instruments | 1B150F-4 | |
Microfil 34g | World Precision Instruments | MF34G-5 | |
Syringe filter | Corning | #431212 | |
Microscope | Leica | DM LFS | With 4X, 40 UV and 63X UV objectives, and epifluorescence |
CCD Camera | Hamamatsu | ORCA ER | 12 bit CCD |
Fura-2 Filter Cube | Chroma | 71500A | Set with UG11 filter |
Polychrome IV | Till Photonics GmBH | Monochronometer | |
Ultraviolet blocking safety glasses | Ultra-Violet Products | ||
EPC-9 amplifier | HEKA Eletronik | ||
Metafluor | Molecular Devices | Imaging Software | |
Patchmaster | HEKA Eletronik | Data Acquisition Software | |
Igor version 6 | Wavemetrics | Electrophysiology & Ca2+ Data Analysis | |
VAPRO 5520 | Westcor | Vapor pressure osmometer |
The authors have nothing to disclose.
The work was funded by a grant from the National Institute of General Medical Sciences (GM096972).
NaCl | Fisher Scientific Co. | S271-3 | |
KCl | Fisher Scientific Co. | P217-500 | |
NaH2PO4•H2O | Sigma Chemical Co. | S-9638 | |
MgCl2•6H2O | Fisher Scientific Co. | M33-500 | |
CaCl2•2H2O | Fisher Scientific Co. | C79-500 | |
D-glucose | Fisher Scientific Co. | D16-500 | |
NaHCO3 | Fisher Scientific Co. | S233-500 | |
Sucrose | Fisher Scientific Co. | S5-500 | |
Potassium D-gluconate | Sigma-Aldrich | G4500 | |
HEPES | Sigma-Aldrich | H4034 | |
Adenosine 5′-triphosphate dipotassium salt dihydrate | Sigma-Aldrich | A8937 | |
Guanosine 5′-triphosphate tris salt | Sigma-Aldrich | G9002 | |
Agarose (ultrapure) | Life Technologies | 15510-027 | Gel pored into sterile Petri dish 6 mm thick layer |
KOH | Sigma-Aldrich | P-6310 | |
NaOH | Sigma-Aldrich | S-5881 | |
Bis-fura-2, hexapotassium salt | Invitrogen | B6810 | Cell impermeant |
Name of Equipment | Company | Catalog Number | Comments |
Microtome | Leica | VT1000S | Tissue slicing |
Cyanoacrylic glue (Roti-Coll1) | Carl Roth GmbH+Co | Art-Nr. 0258.1 | |
Microelectrode puller | Narshige International USA | PP-83 | |
Microelectrode Capillary Tubes | World Precision Instruments | 1B150F-4 | |
Microfil 34 g | World Precision Instruments | MF34G-5 | |
Syringe filter | Corning | #431212 | |
Microscope | Leica | DM LFS | With 4X, 40X UV and 63X UV objectives, and epifluorescence |
CCD Camera | Hamamatsu | ORCA ER | 12 bit CCD |
Fura-2 Filter Cube | Chroma | 71500A | Set with UG11 filter |
Polychrome IV | Till Photonics GmBH | Monochronometer | |
Ultraviolet blocking safety glasses | Ultra-Violet Products | ||
EPC-9 amplifier | HEKA Eletronik | ||
Metafluor | Molecular Devices | Imaging Software | |
Patchmaster | HEKA Eletronik | Data Acquisition Software | |
Igor version 6 | Wavemetrics | Electrophysiology & Ca2+ Data Analysis | |
VAPRO 5520 | Westcor | Vapor pressure osmometer |