Cellular Respiration

Lab Manual
생물학
JoVE 비디오를 활용하시려면 도서관을 통한 기관 구독이 필요합니다.  전체 비디오를 보시려면 로그인하거나 무료 트라이얼을 시작하세요.
Lab Manual 생물학
Cellular Respiration

17,296 Views

05:44 min

January 29, 2019

Procedure

  1. Quantifying Respiration using Microrespirometers
    • NOTE: In this experiment, you will measure the rate of cellular respiration for germinating seeds by measuring the rate of exchange for oxygen. As oxygen is consumed to provide energy, germinating seeds release carbon dioxide. This carbon dioxide is absorbed by potassium carbonate and thus the overall gaseous pressure of the respirometer will be reduced. HYPOTHESES: The experimental hypothesis is that germinating seeds will show a greater rate of respiration than control glass beads. Additionally, that at higher temperatures, the rate of cellular respiration in the seeds will increase. The null hypotheses are that both glass beads and germinating seeds will display a similar rate of cellular respiration and that temperature has no effect on respiration.
    • To set up the control, first remove the plunger from a prepared microrespirometer and label it with a “CR” or “CH” respectively, depending on whether it is the control room temperature or heated treatment.
    • Add the glass microbeads to each of the microrespirometers up to the 0.5 mL mark, and push the plunger in the syringe to the 1 mL mark.
    • Place 3 – 4 washers onto the flared end of each microrespirometer to weigh them down while in the water bath. NOTE: The microrespirometer will sit capillary tube end up.
    • Set the control devices to the side.
    • To set up the experimental seated respirometers, remove the plunger from a prepared microrespirometer and label it with “ER” or “EH” respectively, depending on whether it is the experimental room temperature or heated treatment.
    • Measure 0.5 mL of germinating seeds into the tuberculin syringe of the microrespirometer.
    • Then, push the syringe to the 1 mL mark.
    • Place 3 – 4 washers onto the flared end of each microrespirometer to weigh them down while in the water bath.
    • To use the microrespirometers, carefully press the “CR” and “ER” microrespirometers into the room temperature water bath. The entire syringe should be submerged, but the capillary tubes should be entirely exposed. Remove water from the bath until this position is achieved. Be sure the top of the capillary tube is open.
    • Perform the same procedure (step 9) for the “CH” and “EH” microrespirometers to set up the higher temperature experimental group.
    • Wait 5 minutes before moving on to allow the temperature in the microrespirometers to equalize with the water bath.
    • Add a single drop of manometer fluid to the top of each of the four capillary tubes which seals the microrespirometer chambers. NOTE: If the chambers are working properly, the manometer fluid will be sucked into the capillary chamber.
    • Use the plunger set in the bath to suck the manometer fluid into the capillary. Stop when the fluid is about halfway down the capillary tube. IMPORTANT: The microrespirometers should not be disturbed once they are added to the bath. Do not bump the table or insert or remove anything to or from the water bath once the chambers have been sealed and equalized.
    • Mark the bottom edge of the manometer fluid that is closer to the chamber. This represents time point zero.
    • Set a timer to sound in 5 minutes.
    • After 5 minutes, create a new mark where the manometer fluid is on each microrespirometer. Repeat every 5 minutes until 25 minutes has passed or until the manometer fluid has traveled the entire length of the capillary tube.
    • Then, remove the respirometers from the water baths.
    • Measure the distance between each of the marks and record them in Table 1 next to the correct time point. Click Here to download Table 1
    • During clean up, manometers should either be deconstructed to be reused or disposed of in the trash.
  2. Results
    • To analyze the data you collected, first plot the data points with the x-axis as time and the y-axis as the distance the manometer moved, representing oxygen consumption.
    • Calculate the slope of each line using this equation.
    • Record these values for each microrespirometer in Table 2. NOTE: This value represents the amount of cellular respiration that took place over the course of the experiment. Click Here to download Table 2
    • Plot the respiration rates as a bar graph.
    • Examine both of the graphs. Consider how your results fit with the hypotheses. Note any observed differences in respiration rate between the control beads and the germinating seeds. Determine if increasing the temperature had any effect on the respiration rate.