This protocol describes simultaneous measurement of electroretinogram and visual evoked potentials in anesthetized rats.
The electroretinogram (ERG) and visual evoked potential (VEP) are commonly used to assess the integrity of the visual pathway. The ERG measures the electrical responses of the retina to light stimulation, while the VEP measures the corresponding functional integrity of the visual pathways from the retina to the primary visual cortex following the same light event. The ERG waveform can be broken down into components that reflect responses from different retinal neuronal and glial cell classes. The early components of the VEP waveform represent the integrity of the optic nerve and higher cortical centers. These recordings can be conducted in isolation or together, depending on the application. The methodology described in this paper allows simultaneous assessment of retinal and cortical visual evoked electrophysiology from both eyes and both hemispheres. This is a useful way to more comprehensively assess retinal function and the upstream effects that changes in retinal function can have on visual evoked cortical function.
Meting van de electroretinogram (ERG) en visuele evoked potential (VEP) bruikbare kwantitatieve beoordeling van de integriteit van de visuele route. De ERG meet de elektrische respons van het netvlies lichtstimulatiebron, terwijl de VEP meet de overeenkomstige functionele integriteit van de visuele trajecten van de retina naar de primaire visuele cortex na hetzelfde licht event. Dit manuscript beschrijft een protocol voor de registratie en analyse van de ERG en VEP reacties in een algemeen gebruikt laboratorium model, de rat.
De ERG biedt een index van de functionele integriteit van een aantal belangrijke retinale cel klassen door het kwantificeren van de bruto elektrische reactie van de retina om een lichtflits. Een gecoördineerde reeks ionische fluxen gestart door het aanzetten van licht en offset, detecteerbare veranderingen in de spanning die kan worden gemeten met behulp van oppervlakte-elektroden buiten het oog geplaatst. De resulterende golfvorm vertegenwoordigt de combinatie van een SERies van goed gedefinieerde componenten, die verschillen in amplitude, timing en frequentie. Een aanzienlijke hoeveelheid onderzoek is gebleken dat deze onderdelen relatief goed geconserveerd zijn over vele gewervelde retina en de componenten kunnen worden gescheiden van elkaar. Door oordeelkundig selecteren van de stimulus (flash stimulus, achtergrond, interstimulus interval) voorwaarden en de keuze van de specifieke kenmerken van de samengestelde golfvorm te analyseren kan men overtuigd van het terugsturen van een maatregel van een specifieke groep van retinale cellen 1,2 zijn. Deze kenmerken ten grondslag aan de bruikbaarheid en daarmee de wijdverspreide toepassing van de ERG als een niet-invasief meten van retina functie. Dit handschrift is gericht op de methode voor het meten van de ERG en de analyse van de mogelijkheden ervan om informatie terug te keren over een aantal van de grote mobiele klassen in het netvlies, namelijk fotoreceptoren (de PIII component), bipolaire cellen (de PII component) en retinale ganglion cellen (de positieve scotopisch drempel reactie of pstr).
<p class= "Jove_content"> The VEP biedt een analyse van de corticale reactie op licht; eerst afkomstig van het netvlies en daarna serieel gecommuniceerd via de oogzenuw, optische darmkanaal, thalamus (laterale geniculate nucleus, LGN) en optische straling tot gebied V1 van de cortex 3. Bij knaagdieren, de meerderheid (90-95%) van de optische zenuwvezels van elk oog kruisvormige 4 en innerveren de contralaterale midden hersenen. In tegenstelling tot de ERG, is het nog niet mogelijk verschillende componenten van de VEP toeschrijven aan specifieke klassen cellen, 5 verandert dus overal langs de gezichtsbaan kunnen hebben voor de VEP golfvorm. Toch is de VEP een nuttige niet-invasief meten van visuele prestatie en gezichtsbaan integriteit. De VEP, bij gebruik in combinatie met de ERG, kan een volledige beoordeling van de visuele systeem (bijv retina / gezichtsbaan) verschaffen.ERG en VEP opnamen kan worden uitgevoerd of in combinatie, afhankelijk van de toekation. De in dit document beschreven methode maakt gelijktijdige evaluatie van het netvlies en corticale visuele opgeroepen elektrofysiologie van beide ogen en beide halfronden in verdoofde ratten. Dit is een handige manier om de retinale functie en de upstream-effecten die veranderingen in de functie van de retina op de visuele opgeroepen corticale functie kan hebben meer uitvoerig te evalueren.
De ERG en VEP objectieve metingen van de visuele functie van het netvlies en de cortex, respectievelijk. Het voordeel van gelijktijdige opname is dat een meer uitgebreid overzicht van het gehele visuele route wordt geboden. Daarbij zouden de aanvullende informatie van de samenlopende beoordeling een duidelijker afbakening van de plaats van letsel te voorzien in de visuele banen (bijvoorbeeld bij aandoeningen met overlappende Nog ERG afzonderlijke VEP manifestaties 18, wanneer optische neuropathie kan…
The authors have nothing to disclose.
Funding for this project was provided by the National Health and Medical Research Council (NHMRC) 1046203 (BVB, AJV) and Melbourne Neuroscience Institute Fellowship (CTN).
Alligator clip | generic brand | HM3022 | Stainless steel 26 mm clip for connecting VEP screw electrodes to cables |
Bioamplifier | ADInstruments | ML 135 | For amplifying ERG and VEP signals |
Carboxymethylcellulose sodium 1.0% | Allergan | CAS 0009000-11-7 | Viscous fluid for improving signal quality of the active ERG electrode |
Carprofen 0.5% | Pfizer Animal Health Group | CAS 53716-49-7 | Proprietary name: Rimadyl injectable (50 mg/mL). For post-surgery analgesia, diluted to 0.5% (5 mg/mL) in normal saline |
Chlorhexadine 0.5% | Orion Laboratories | 27411, 80085 | For disinfecting surgical instruments |
Circulating water bath | Lauda-Königshoffen | MGW Lauda | For maintaining body temperature of the anesthetized animal during surgery and electrophysiological recordings |
Dental amalgam | DeguDent GmbH | 64020024 | For encasing the electrode-skull assembly to make it more robust |
Dental burr | Storz Instruments, Bausch and Lomb | #E0824A | A miniature drill head of ~0.7mm diameter for making a small hole in the skull over each hemisphere to implant VEP screws |
Drill | Bosch | Dremel 300 series | An automatic drill for trepanning |
Electrode lead | Grass Telefactor | F-E2-30 | Platinum cables for connecting silver wire electrodes to the amplifier |
Faraday Cage | custom-made | Ensures light proof to maintain dark adaptation. Encloses the Ganzfeld setup to improve signal to noise ratio | |
Gauze swabs | Multigate Medical Products Pty Ltd | 57-100B | For drying the surgical incision and exposed skull surface during surgery |
Ganzfeld integrating sphere | Photometric Solutions International | Custom designed light stimulator: 36 mm diameter, 13 cm aperture size | |
Velcro | VELCRO Australia Pty Ltd | VELCRO Brand Reusable Wrap | Hook-and-loop fastener to secure the electrodes and the animal on the recording platform |
Isoflurane 99.9% | Abbott Australasia Pty Ltd | CAS 26675-46-7 | Proprietary Name: Isoflo(TM) Inhalation anaaesthetic. Pharmaceutical-grade inhalation anesthetic mixed with oxygen gas for VEP electrode implant surgery |
Ketamine | Troy Laboratories | Ilium Ketamil | Proprietary name: Ketamil Injection, Brand: Ilium. Pharmaceutical-grade anesthetic for electrophysiological recording |
Luxeon LEDs | Phillips Lighting Co. | For light stimulation twenty 5 watt and one 1 watt LEDs. | |
Micromanipulator | Harvard Apparatus | BS4 50-2625 | Holds the ERG active electrode during recordings |
Needle electrode | Grass Telefactor | F-E2-30 | Subcutaneously inserted in the tail to serve as the ground electrode for both the ERG and VEP |
Phenylephrine 2.5% minims | Bausch and Lomb | CAS 61-76-7 | Instilled with Tropicamide to achieve maximal dilation for ERG recording |
Povidone iodine 10% | Sanofi-Aventis | CAS 25655-41-8 | Proprietory name: Betadine, Antiseptic to prepare the shaved skin for surgery 10%, 500 mL |
Powerlab data acquisition system | ADInstruments | ML 785 | Controls the LEDs |
Proxymetacaine 0.5% | Alcon Laboratories | CAS 5875-06-9 | For corneal anaesthesia during ERG recordings |
Saline solution | Gelflex | Non-injectable, for electroplating silver wire electrodes | |
Scope Software | ADInstruments | version 3.7.6 | Simultaneously triggers the stimulus via the Powerlab system and collects data |
Silver (fine round wire) | A&E metal | 0.3 mm | Used to make active and inactive ERG electrodes, and the inactive VEP electrode |
Stainless streel screws | MicroFasterners | 0.7 mm shaft diameter, 3 mm in length to be implanted over the primary visual cortex and serve as the active VEP electrodes | |
Stereotaxic frame | David Kopf | Model 900 | A small animal stereotaxic instrument for locating the primary visual cortices according to Paxinos & Watson's 2007 rat brain atlas coordinates |
Surgical blade | Swann-Morton Ltd. | 0206 | For incising the area of skin overlaying the primary visual cortex to implant the VEP electrodes |
Suture | Shanghai Pudong Jinhuan Medical Products Co.,Ltd | 3-0 silk braided suture non-absorbable, for skin retraction during VEP electrode implantation surgery | |
Tobramycine eye ointment 0.3% | Alcon Laboratories | CAS 32986-56-4 | Proprietary name: Tobrex. Prophylactic antibiotic ointment applied around the skin wound after surgery |
Tropicamide 0.5% | Alcon Laboratories | CAS 1508-75-4 | Proprietary name: 0.5% Mydriacyl eye drop, Instilled to achieve mydriasis for ERG recording |
Xylazine | Troy Laboratories | Ilium Xylazil-100 | Pharmaceutical-grade anesthetic for electrophysiological recording |
Pipette tip | Eppendorf Pty Ltd | 0030 073.169 | Eppendorf epTIPS 100 – 5000 mL, for custom-made electrodes |
Microsoft Office Excel | Microsoft | version 2010 | spreadsheet software for data analysis |
Lethabarb Euthanazia Injection | Virbac (Australia) Pty Ltd | LETHA450 | 325 mg/mL pentobarbital sodium for rapid euthanazia |