This video describes a gel-clot assay to determine the presence of bacterial endotoxins in nano-formulations using Limulus Amoebocyte Lysate derived from the blood cells of horseshoe crabs. The assay helps test for endotoxin contamination of pharmaceutical products, including nano-based formulations.
Protocol
1. Preparation of Nanoparticle Samples Prepare the study sample in LAL-grade water. If the sample pH is outside of the 6-8 range, adjust the pH by using pyrogen-free sodium hydroxide or hydrochloric acid. Using LAL-grade water, prepare several dilutions of the study sample. Make sure that the highest dilution does not exceed maximum valid dilution (MVD). 2. Preparation of Reagents Common Between LAL Formats Dilute concentrated sodium hydroxide…
Disclosures
The authors have nothing to disclose.
Materials
LAL reagent
Associates of Cape Cod
G5003
This reagent is specific to the gel clot assay
Control endotoxin standard
Associates of Cape Cod
E0005
This reagent can be used with turbidity and gel clot assays
Sodium hydroxide
Sigma
S2770
When needed, it is used to adjust sample pH to be between 6-8
Hydrochloric acid
Sigma
H9892
When needed, it is used to adjust sample pH to be between 6-8
LAL grade water
Associates of Cape Cod
WP0501
This reagent can be used with any LAL format
Glucashield buffer
Associates of Cape Cod
GB051-25
Used to prevent false-positive response from beta-glucans
Disposable endotoxin-free glass dilution tubes 12 x 75 mm
Associates of Cape Cod
TB240
These tubes can be used with all three assays
Disposable endotoxin-free glass reaction tubes 10 x 75 mm
Associates of Cape Cod
TS050
These tubes are for use with the gel-clot assay
Pyrogen-free tips with volumes 0.25 and 1 mL
RAININ
PPT25, PPT10
Tips and pipettes may adsorb endotoxin and release leachables which interfere with LAL assay. These RAININ tips are used because their optimal performance in the LAL assay was verified and confirmed
Pyrogen-free microcentrifuge tubes, 2.0 mL
Eppendorf
22600044
Other equivalent supplies can be used
Pyrogen-fee combitips, 5mL
Eppendorf
30089669
Other equivalent supplies can be used
Repeat pipettor
Eppendorf
4982000020
Other equivalent supplies can be used
Microcetrifuge
Any brand
Any brand can be used
Refrigerator, 2-8° C
Any brand
Any brand can be used
Vortex
Any brand
Any brand can be used
Freezer, -20 C
Any brand
Any brand can be used
Water bath, 37° C
Any brand
Any brand can be used, however, it is important either to switch off water circulation or use a non-circulating water bath because water flow will affect clot formation and lead to false-negative results