Studying the Excitability of Fluorescent Neurons using a Whole-Cell Patch Clamp

Studying the Excitability of Fluorescent Neurons using a Whole-Cell Patch Clamp

DEŞİFRE METNİ

Transfer a slice to the recording chamber using a Pasteur pipette or a small brush. Hold down the slice with a harp, and perfuse it with ACSF at a rate of 2 milliliters per minute. To patch a GFP-positive neuron, locate the area of interest through the microscope at 10x. Then, find a GFP-positive cell using the 60x objective.

Next, fill the recording electrode with intracellular solution. Subsequently, place a glass pipette in the pipette holder. Afterward, place the pipette tip in the bath and focus on the tip. Once the pipette is in the bath, apply positive pressure through the back pressure control system.

Approach the cell of interest under visual guidance while maintaining back pressure in the pipette. Upon the appearance of a small dimple on the cell surface, release the pressure. At this point, a tight seal with a resistance larger than 1 gigaohm may be formed. Otherwise, apply a light negative pressure to facilitate it.

While the seal is being formed, bring the holding voltage-clamp to negative 60 millivolts. Once the gigaohm seal is formed, apply a pulse of suction to rupture the cell membrane and break into the whole-cell mode. Once it is in whole-cell mode, switch from voltage-clamp to current-clamp mode and start recording.

İlgili Videolar