Özet

Measuring the Complete-arch Distortion of an Optical Dental Impression

Published: May 30, 2019
doi:

Özet

Here, we present a protocol to measure the degree of distortion at each part of the compete-arch digital impression acquired from an intraoral scanner with 3D-printed metal phantom with standard geometries.

Abstract

Digital workflows have actively been used to produce dental restorations or oral appliances since dentists started to make digital impressions by acquiring 3D images with an intraoral scanner. Because of the nature of scanning the oral cavity in the patient's mouth, the intraoral scanner is a handheld device with a small optical window, stitching together small data to complete the entire image. During the complete-arch impression procedure, a deformation of the impression body can occur and affect the fit of the restoration or appliance. In order to measure these distortions, a master specimen was designed and produced with a metal 3D printer. Designed reference geometries allow setting independent coordinate systems for each impression and measure x, y, and z displacements of the cylinder top circle center where the distortion of the impression can be evaluated. In order to evaluate the reliability of this method, the coordinate values of the cylinder are calculated and compared between the original computer-aided design (CAD) data and the reference data acquired with the industrial scanner. The coordinate differences between the two groups were mostly less than 50 µm, but the deviations were high due to the tolerance of 3D printing in the z coordinates of the obliquely designed cylinder on the molar. However, since the printed model sets a new standard, it does not affect the results of the test evaluation. The reproducibility of the reference scanner is 11.0 ± 1.8 µm. This test method can be used to identify and improve upon the intrinsic problems of an intraoral scanner or to establish a scanning strategy by measuring the degree of distortion at each part of the complete-arch digital impression.

Introduction

In the traditional dental treatment process, a fixed restoration or a removable denture is made on a model made of gypsum and impregnated with a silicone or irreversible hydrocolloid material. Because an indirectly made prosthesis is delivered in the oral cavity, a lot of research has been done to overcome the errors caused by a series of such manufacturing processes1,2. Recently, a digital method is used to fabricate a prosthesis through the CAD process by manipulating models in the virtual space after acquiring 3D images instead of making impressions3. In the early days, such an optical impression method was used in a limited range such as a dental caries treatment of one or a small number of teeth. However, as the base technology of the 3D scanner was developed, a digital impression for the complete arch is now used for the fabrication of large-scale fixed restorations, removable restorations such as a partial or full denture, orthodontic appliances, and implant surgical guides4,5,6,7. The accuracy of the digital impression is satisfactory in a short region such as the unilateral arch. However, since the intraoral scanner is a handheld device that completes the entire dentition by stitching together the image obtained through a narrow optical window, the distortion of the model can be seen after completing the U-shaped dental arch. Thus, an appliance of a large range made on this model might not fit well in the patient's mouth and require a lot of adjustment.

Various studies have been reported on the accuracy of the virtual impression body obtained with an intraoral scanner, and there are various research models and measurement methods. Depending on the research subject, it can be divided into clinical research8,9,10,11,12 for actual patients and in vitro studies13,14,15,16 conducted in models separately produced for research. Clinical studies have the advantage of being able to evaluate the conditions of an actual clinical setting, but it is difficult to control the variables and increase the number of clinical cases indefinitely. The number of clinical studies is not large because there is a limit to being able to evaluate the desired variables. On the other hand, many in vitro studies that evaluate the basic performance of the intraoral scanner by controlling variables have been reported17. The research model also includes a partial or complete arch of natural teeth18,19,20,21,22 and a fully edentulous jaw with all teeth lost23, or the case where the dental implant is installed and spaced apart at a certain interval24,25,26,27, or a form in which the majority of the teeth remain and only a part of a tooth is missing16,28. However, studies on the distortion of the virtual impression body made by a handheld intraoral scanner have been limited to the qualitative evaluation of deviations through a color map created by superimposing it with reference data and expressed as one numerical value per data. It is difficult to accurately measure the 3D distortion of the complete arch because most studies only examine the localized portion of the dental arch with a nondirectional distance deviation.

In this study, the distortion of the dental arch during optical impression with an intraoral scanner is investigated by using a standard model with a coordinate system. The aim of this study is to provide information on a method for evaluating the accuracy performance of the intraoral scanners which exhibit various characteristics by the difference in optical hardware and processing software.

Protocol

1. Master specimen preparation

  1. Model preparation
    1. Remove the artificial teeth (left and right canines, second premolar, and the second molar) on the mandibular complete-arch model with only 1/5 of the cervical portion left.
  2. CAD design
    1. Acquire the data of the master specimen with a reference scanner.
    2. Design the cylinders (with a top diameter of 2 mm and a cylinder height of 7 mm) on top of the trimmed six teeth with the reverse engineering software.
    3. Add three reference spheres (3.5 mm in diameter) posterior to the left second molar for the purpose of defining the reference 3D coordinate system from the reverse engineering software.
    4. Locate one sphere on the distal side of the distal and buccal side of the cylinder on the left second molar so that the coordinates of all the cylinders have positive values.
    5. Design the left second molar cylinder so that it is inclined 30° medially and the right second molar cylinder so that it is tilted 30° distally. Set the other cylinders at right angles from the model.
  3. Metal 3D printing
    1. Manufacture a phantom model with CoCr alloy by a metal 3D printer to serve as a patient’s dentition (Figure 1).

2. Reference data acquisition and software analysis

  1. Scan the phantom with the testing intraoral scanner.
    1. Obtain the reference image by scanning the metal phantom model with the industrial-level model scanner.
  2. Establish a coordinate system by extracting points from reference spheres.
    1. Load the reference image to the reverse engineering analysis software to calculate the reference coordinates of each cylinder position.
    2. Extract the sphere by selecting the Ref. geometry | Create | Sphere | Pick boundary points command and picking the four points on the surface of the reference sphere that are furthest apart from each other (Supplemental Figure 1 and Supplemental Figure 2).
    3. Calculate the center of three reference spheres.
    4. Use the Ref. geometry | Create | Plane | Pick points command to connect the centers of three spheres and create a plane (Supplemental Figure 3).
    5. Set the formed plane as XY plane.
    6. Select the Ref. geometry | Create | Plane | Offset plane command to create a tangent plane above the xy plane (Supplemental Figure 4).
    7. Create points where the tangent plane and two lingual spheres meet by choosing the Ref. geometry | Create | Point | Project on ref. plane command (Supplemental Figure 5).
    8. Generate a plane between the points created and the center of the two lingual spheres by using the Ref. geometry | Create | Plane | Pick points command (Supplemental Figure 6).
    9. Measure the distance from this plane to the center of the buccal sphere with the Inspection | Dimension | Linear command (Supplemental Figure 7).
    10. Create a parallel plane that passes through the midpoint of the buccal sphere with the Geometry | Create | Plane | Offset Plane command (Supplemental Figure 8).
    11. Set the formed plane as YZ plane (Supplemental Figure 9).
  3. Set the x, y, and z axes.
    1. Set the center of the buccal sphere as the ‘origin’ of the coordinate system.
    2. Set a line parallel to the line connecting the center points of the remaining two spheres while traveling in the forward and backward direction of the model through the origin as the Y-axis.
    3. Set the line on the xy plane that passes the origin and is perpendicular to the y axis as the X-axis.
    4. Use the Ref. geometry | Create | Coordinate | Pick origin & X, Y direction command to create a new coordinate system with the buccal sphere center as the origin (Supplemental Figure 10).
    5. Set the line perpendicular to the xy plane and passing through the origin as the Z-axis (Supplemental Figure 11).
  4. Transfer this detail from the scan coordinate system to the newly established coordinate system.
    1. Use the Ref. geometry | Bind to shell command to fix the geometries created during this process on top of the scan data (Supplemental Figure 12).
    2. Execute the Ref. geometry | Transform | Coordinate | Align coordinate command to transit from the basic coordinate system to the newly created coordinate system (Supplemental Figure 13).
    3. In this way, assign a coordinate system to the metal master specimen with reference to the three reference spheres (Supplemental Figure 14).
  5. Extract the measurement points from the cylinders at the main area.
    1. Extract the x, y, and z coordinates for the upper circle centers of six cylinders to be analyzed for the distortion of the specified regions by the reverse engineering process.
    2. For this, use the Ref. geometry | Create | Cylinder | Pick boundary points command and specify at least 10 points on the top border of the cylinder and designate the same amount of points on the ellipse that meets the tooth at the bottom of the cylinder (Supplemental Figure 15, Supplemental Figure 16, and Supplemental Figure 17).
    3. Obtain the extracted coordinates of the cylinder top center. Evaluate the 3D deformation at each position by comparing it with the coordinate values of the same cylinder of the digital impression acquired by the intraoral scanner to be evaluated.

Representative Results

The coordinates of each cylinder calculated from the originally designed CAD data and the reference scan image of the 3D-printed metal master specimen scanned by the industrial-level model scanner are shown in Table 1. The difference between the two showed a value of lower than 50 µm, but the z coordinate value of the right second molar cylinder from the 3D-printed master specimen was low. Although the metal phantom was produced from a high-end industrial 3D printer, a small difference in the height of one cylinder was found. While the design was done with CAD software, the metal phantom was used as a reference which was scanned with the various testing intraoral scanners, and the difference was negligible. If another evaluator fabricates a new phantom from the same shared data and executes the same process, the phantom should be scanned again with an industrial-level reference scanner to obtain reference coordinates and then proceed with the subsequent process. Table 2 shows the coordinates of the master specimen which was scanned five times with an industrial scanner. Evaluating from the standard deviation, the average deviation was 45 µm, showing a large deviation in the y coordinate value of the right second molar cylinder. It could be concluded that the precision of the reference scanner was good enough for extracting the reference coordinates of point zero and six cylinders.

The evaluation of the reproducibility of the reference scanner was conducted through overlap comparison among five datasets of the metal master specimen scanned with the reference scanner. A total of 10 pairs were aligned and evaluated. The deviation of each pair resulted in reproducibility of 0.011 ± 0.002 mm (Table 3). The reproducibility of the reference scanner was calculated differently, and it was concluded that the results from both methods were reliable and the latter could be omitted.

Figure 1
Figure 1: Design and manufacturing process of a phantom model for the distortion evaluation. (A) Originally designed CAD data. (B) 3D-printed master specimen made of CoCr alloy. Please click here to view a larger version of this figure.

Figure 1
Supplemental Figure 1: Extract sphere picking points. Please click here to view a larger version of this figure.

Figure 2
Supplemental Figure 2: Picking points on the surface of the reference sphere. Please click here to view a larger version of this figure.

Figure 3
Supplemental Figure 3: Creation of the XY plane by picking the center of three spheres. Please click here to view a larger version of this figure.

Figure 4
Supplemental Figure 4: Creation of the offset plane, half a diameter of the sphere above the XY plane. Please click here to view a larger version of this figure.

Figure 5
Supplemental Figure 5: Creation of the points where the offset plane and two lingual spheres meet. Please click here to view a larger version of this figure.

Figure 6
Supplemental Figure 6: Creation of the plane that passes both centers of the lingual spheres by picking four points. Please click here to view a larger version of this figure.

Figure 7
Supplemental Figure 7: Measurement of the distance from this plane to the center of the buccal sphere. Please click here to view a larger version of this figure.

Figure 8
Supplemental Figure 8: Creation of the parallel plane that passes through the center of the buccal sphere. Please click here to view a larger version of this figure.

Figure 9
Supplemental Figure 9: Setting of the formed plane as YZ plane. Please click here to view a larger version of this figure.

Figure 10
Supplemental Figure 10: Creation of a new coordinate system with the center of the buccal sphere as origin. Please click here to view a larger version of this figure.

Figure 11
Supplemental Figure 11: Setting of the line perpendicular to the XY plane and passing through the buccal sphere center as the Z-axis. Please click here to view a larger version of this figure.

Figure 12
Supplemental Figure 12: Fixing of the created geometries to the scan data. Please click here to view a larger version of this figure.

Figure 13
Supplemental Figure 13: Transfer of the basic coordinate system to the newly created coordinate system. Please click here to view a larger version of this figure.

Figure 14
Supplemental Figure 14: Checking whether the origin and coordinate system are correctly moved to the one extracted from the scan data. Please click here to view a larger version of this figure.

Figure 15
Supplemental Figure 15: Using the Pick boundary points command to extract the cylinder. Please click here to view a larger version of this figure.

Figure 16
Supplemental Figure 16: Picking sufficient points on the top circle and bottom ellipse around the cylinder. Please click here to view a larger version of this figure.

Figure 17
Supplemental Figure 17: Checking whether the extracted cylinder was reverse engineered correctly. Please click here to view a larger version of this figure.

CAD 3D printed Difference
data metal master specimen
37i x 7.897 7.875 0.022
y 6.418 6.373 0.045
z 7.312 7.265 0.047
35i x 8.481 8.427 0.054
y 26.045 25.99 0.055
z 7.846 7.846 0
33i x 11.889 11.85 0.04
y 40.16 40.106 0.054
z 8.346 8.409 -0.063
43i x 37.246 37.196 0.051
y 45.738 45.686 0.052
z 9.445 9.5 -0.055
45i x 47.21 47.178 0.032
y 35.115 35.081 0.034
z 8.707 8.708 -0.001
47i x 56.397 56.386 0.011
y 13.038 13.041 -0.002
z 7.558 7.451 0.107

Table 1: Differences in cylinders’ coordinates between CAD data and the 3D-printed metal master specimen. Unit: mm.

Ref. 1 Ref. 2 Ref. 3 Ref. 4 Ref. 5 Mean ± SD
37i x 7.856 7.874 7.871 7.89 7.885 7.875 ± 0.013
y 6.406 6.375 6.358 6.356 6.368 6.373 ± 0.020
z 7.259 7.274 7.269 7.265 7.258 7.265 ± 0.007
35i x 8.435 8.379 8.393 8.471 8.46 8.427 ± 0.040
y 26.032 25.98 25.996 25.962 25.979 25.990 ± 0.026
z 7.838 7.883 7.837 7.858 7.816 7.846 ± 0.025
33i x 11.839 11.779 11.794 11.925 11.91 11.850 ± 0.066
y 40.129 40.085 40.112 40.097 40.106 40.106 ± 0.017
z 8.372 8.485 8.391 8.414 8.381 8.409 ± 0.046
43i x 37.177 37.115 37.155 37.269 37.262 37.196 ± 0.068
y 45.711 45.723 45.725 45.622 45.65 45.686 ± 0.047
z 9.437 9.568 9.541 9.498 9.457 9.500 ± 0.055
45i x 47.15 47.123 47.142 47.246 47.23 47.178 ± 0.056
y 35.109 35.148 35.135 34.988 35.025 35.081 ± 0.071
z 8.609 8.785 8.728 8.738 8.681 8.708 ± 0.067
47i x 56.369 56.373 56.371 56.409 56.407 56.386 ± 0.020
y 13.085 13.122 13.114 12.923 12.959 13.041 ± 0.093
z 7.349 7.445 7.457 7.527 7.478 7.451 ± 0.065

Table 2: Cylinders’ coordinates of reference datasets acquired from the 3D-printed metal master specimen. Unit: m.

Precision 1 2 3 4 5 6 7 8 9 10 Mean ± SD
Reference scanner 8.3 12.4 9.5 13.2 11.7 8 12.1 10.7 12.1 11.8 11.0 ± 1.8

Table 3: Precision of the dataset acquired from the reference scanner. Unit: µm.

Discussion

Among the studies evaluating the accuracy of the intraoral scanner by evaluating the resultant digital impression body, the most common method is to superimpose the digital impression data on the reference image and calculate the shell-to-shell deviation12,13,14,15,20,23. However, this method is limited to calculating the deviation value from the paired data or evaluating the distribution qualitatively through the color map. In a study which measured the deviation of the local site by selecting points to be analyzed on the color map, the deviation in the x, y, and z direction was not considered29. In addition, these methods have limitations in that they should be analyzed after overlapping with reference data. Alignment can vary from one data point to another, and the results vary depending on the sorting criteria. In clinical trials involving patients, it is difficult to apply these methods because it is not possible to scan the complete dental arch through the mouth with an industrial scanner located outside the oral cavity.

In this study, a master specimen made of metal, which is less affected by temperature and humidity, was proposed. The coordinate system for the specific 3D-printed metal specimen was set and the six cylinders’ position coordinates were calculated in advance. In this way, regardless of the intraoral scanner, an individual coordinate system was formed from each digital impression through the reference spheres of the scan data so that the analysis could be performed with the scanned data only, without the reference image superimposition. The reference image obtained with the high-precision industrial reference scanner was used only to acquire the coordinate values of the six cylinders when the metal master specimen was first produced. The comparative evaluation between the reference and intraoral scan data was done just by simple arithmetic calculation through coordinate values. In addition, since the deviations in the x, y, and z directions of the cylinder coordinates were expressed as positive and negative values, 3D positional changes were shown for each region. Therefore, the method used in this study is suitable for evaluating the data distortion of the handheld device, the oral scanner. Since the deviation of the cylinder coordinates in the x, y, and z direction was displayed with positive and negative values, a 3D position change of each location becomes obvious. Therefore, the method used in this study is suitable for evaluating the distortion of digital impression data acquired with the handheld intraoral scanner.

Most of the coordinate values of each cylinder calculated from the original CAD data and the reference image of the metal master specimen showed values of less than 50 µm. This is related to the performance peculiar to metal 3D printers. Since the master specimen after 3D printing is used as a new reference rather than using standard CAD data, the limitations of these 3D printers do not need to be considered. The change in the master specimen was large at the z coordinate of the right second molar. It was because the cylinder was tilted distally and the length of the cylinder exposed above the tooth was short, which was disadvantageous for the reverse engineering process. Also, the cylinder’s upper circle of this tooth was inclined to the xy plane of the 3D printer when metal printing was performed in this study. It seems that the characteristics of the 3D printer, in which xy accuracy and z accuracy are expressed separately, were also reflected. In future research, designing and using all cylinders without tilting can be a good alternative.

If there is a cost problem in fabricating a master specimen with a metal 3D printer, it can be made of gypsum or resin. As the new coordinate system was set and the coordinates of the six cylinders were calculated after specimen fabrication, the dimensional change that could be caused by the expansion and contraction of the material during the manufacturing process does not affect the final result. However, when using such a specimen for a long period of time, there may be a slight volume change due to moisture and temperature, and there is a possibility that it will be deformed due to breakage or abrasion. Therefore, a calibration procedure is required to periodically calculate the cylinder coordinate value with a reference scanner. In addition, instead of using an industrial reference scanner, the coordinate measuring machine (CMM) may be used to measure the reference coordinates of the master specimen. In this case, it is recommended to carry out a superimposing investigation with reference data for the purpose of evaluating the complicated tooth surface in addition to the deviation inspection through the coordinates of cylinders.

The limitations of this method are that the time required for reverse engineering analysis becomes longer when the number of digital impressions to be evaluated increases. However, recently introduced 3D image analysis software enables inspection automation through a macro function. Since the global shape of the master specimen is the same, it is possible to shorten the analysis time by automating the coordinate system setting and the cylinder coordinate calculation of the acquired digital impression.

By measuring the degree of distortion in each part of the complete-arch digital impression as a numerical value, it can be used to find and improve the inherent problems of the intraoral scanner to be evaluated for its performance. Since the intraoral scanner is a complicated optical device consisting of a projection lamp, a lens, a lens barrel, a camera, etc., hardware consideration factors are large. Also, a software algorithm that enables stitching together the acquired 3D data in real-time at more than 30 frames per second is also important19. It is possible to evaluate and improve the performance of the intraoral scanner by understanding the relationship between the recurrence pattern of the metal master specimen and the consideration factors of intraoral scanners. The scanning strategy determined by the direction and sequence of acquiring images is also an important element for acquiring digital impressions30. This method can be used to establish a strategy that minimizes deformation.

Açıklamalar

The authors have nothing to disclose.

Acknowledgements

This study was supported by a grant of the Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare (grant number: HI18C0435).

Materials

EOS CobaltChrome SP2 Electro Oprical Systems H051601 Powder type metal alloy for 3D printing
Geomagic Verify 3D Systems 2015.2.0 3D inspection software
Prosthetic Restoration Jaw Model Nissin Dental Products Inc. Mandibular complete-arch model
Rapidform Inus technology RF90600-10004-010000 Reverse engineering software
stereoSCAN R8 AICON 3D Systems GmbH Industrial-level model scanner

Referanslar

  1. McLean, J. W., von Fraunhofer, J. A. The estimation of cement film thickness by an in vivo technique. British Dental Journal. 131 (3), 107-111 (1971).
  2. Park, J. M., Hong, Y. S., Park, E. J., Heo, S. J., Oh, N. Clinical evaluations of cast gold alloy, machinable zirconia, and semiprecious alloy crowns: A multicenter study. Journal of Prosthetic Dentistry. 115 (6), 684-691 (2016).
  3. Keul, C., et al. Fit of 4-unit FDPs made of zirconia and CoCr-alloy after chairside and labside digitalization–a laboratory study. Dental Materials. 30 (4), 400-407 (2014).
  4. Ritter, L., et al. Accuracy of chairside-milled CAD/CAM drill guides for dental implants. International Journal of Computerized Dentistry. 17 (2), 115-124 (2014).
  5. Grunheid, T., McCarthy, S. D., Larson, B. E. Clinical use of a direct chairside oral scanner: an assessment of accuracy, time, and patient acceptance. American Journal of Orthodontics and Dentofacial Orthopedics. 146 (5), 673-682 (2014).
  6. Penarrocha-Oltra, D., Agustin-Panadero, R., Bagan, L., Gimenez, B., Penarrocha, M. Impression of multiple implants using photogrammetry: description of technique and case presentation. Medicina Oral, Patolodia Oral y Cirugia Bucal. 19 (4), e366-e371 (2014).
  7. Kattadiyil, M. T., Mursic, Z., AlRumaih, H., Goodacre, C. J. Intraoral scanning of hard and soft tissues for partial removable dental prosthesis fabrication. Journal of Prosthetic Dentistry. 112 (3), 444-448 (2014).
  8. Kim, J., et al. Comparison of experience curves between two 3-dimensional intraoral scanners. Journal of Prosthetic Dentistry. 116 (2), 221-230 (2016).
  9. Lim, J. H., Park, J. M., Kim, M., Heo, S. J., Myung, J. Y. Comparison of digital intraoral scanner reproducibility and image trueness considering repetitive experience. Journal of Prosthetic Dentistry. 119 (2), 225-232 (2018).
  10. Muhlemann, S., Greter, E. A., Park, J. M., Hammerle, C. H. F., Thoma, D. S. Precision of digital implant models compared to conventional implant models for posterior single implant crowns: A within-subject comparison. Clinical Oral Implants Research. 29 (9), 931-936 (2018).
  11. Park, J. M., Hammerle, C. H. F., Benic, G. I. Digital technique for in vivo assessment of internal and marginal fit of fixed dental prostheses. Journal of Prosthetic Dentistry. 118 (4), 452-454 (2017).
  12. Ender, A., Zimmermann, M., Attin, T., Mehl, A. In vivo precision of conventional and digital methods for obtaining quadrant dental impressions. Clinical Oral Investigations. 20 (7), 1495-1504 (2016).
  13. Kim, R. J., Park, J. M., Shim, J. S. Accuracy of 9 intraoral scanners for complete-arch image acquisition: A qualitative and quantitative evaluation. Journal of Prosthetic Dentistry. 120 (6), 895-903 (2018).
  14. Ender, A., Mehl, A. Accuracy in dental medicine, a new way to measure trueness and precision. Journal of Visualized Experiments. (86), e51374 (2014).
  15. Ender, A., Mehl, A. In-vitro evaluation of the accuracy of conventional and digital methods of obtaining full-arch dental impressions. Quintessence International. 46 (1), 9-17 (2015).
  16. Ajioka, H., Kihara, H., Odaira, C., Kobayashi, T., Kondo, H. Examination of the Position Accuracy of Implant Abutments Reproduced by Intra-Oral Optical Impression. PLOS ONE. 11 (10), e0164048 (2016).
  17. Patzelt, S. B., Lamprinos, C., Stampf, S., Att, W. The time efficiency of intraoral scanners: an in vitro comparative study. Journal of Americal Dental Association. 145 (6), 542-551 (2014).
  18. Gan, N., Xiong, Y., Jiao, T. Accuracy of Intraoral Digital Impressions for Whole Upper Jaws, Including Full Dentitions and Palatal Soft Tissues. PLOS ONE. 11 (7), e0158800 (2016).
  19. Rehmann, P., Sichwardt, V., Wostmann, B. Intraoral Scanning Systems: Need for Maintenance. International Journal of Prosthodontics. 30 (1), 27-29 (2017).
  20. Patzelt, S. B., Emmanouilidi, A., Stampf, S., Strub, J. R., Att, W. Accuracy of full-arch scans using intraoral scanners. Clinical Oral Investigations. 18 (6), 1687-1694 (2014).
  21. Muallah, J., et al. Accuracy of full-arch scans using intraoral and extraoral scanners: an in vitro study using a new method of evaluation. International Journal of Computerized Dentistry. 20 (2), 151-164 (2017).
  22. Treesh, J. C., et al. Complete-arch accuracy of intraoral scanners. Journal of Prosthetic Dentistry. 120 (3), 382-388 (2018).
  23. Patzelt, S. B., Vonau, S., Stampf, S., Att, W. Assessing the feasibility and accuracy of digitizing edentulous jaws. Journal of Americal Dental Association. 144 (8), 914-920 (2013).
  24. Andriessen, F. S., Rijkens, D. R., van der Meer, W. J., Wismeijer, D. W. Applicability and accuracy of an intraoral scanner for scanning multiple implants in edentulous mandibles: a pilot study. Journal of Prosthetic Dentistry. 111 (3), 186-194 (2014).
  25. Gimenez, B., Ozcan, M., Martinez-Rus, F., Pradies, G. Accuracy of a digital impression system based on parallel confocal laser technology for implants with consideration of operator experience and implant angulation and depth. International Journal of Oral and Maxillofacial Implants. 29 (4), 853-862 (2014).
  26. Gimenez, B., Ozcan, M., Martinez-Rus, F., Pradies, G. Accuracy of a digital impression system based on active wavefront sampling technology for implants considering operator experience, implant angulation, and depth. Clinical Implant Dentistry and Related Research. 17 Suppl 1, e54-e64 (2015).
  27. Papaspyridakos, P., et al. Digital versus conventional implant impressions for edentulous patients: accuracy outcomes. Clinical Oral Implants Research. 27 (4), 465-472 (2016).
  28. Flugge, T. V., Att, W., Metzger, M. C., Nelson, K. Precision of Dental Implant Digitization Using Intraoral Scanners. International Journal of Prosthodontics. 29 (3), 277-283 (2016).
  29. Kim, S. Y., et al. Accuracy of dies captured by an intraoral digital impression system using parallel confocal imaging. International Journal of Prosthodontics. 26 (2), 161-163 (2013).
  30. Ender, A., Mehl, A. Influence of scanning strategies on the accuracy of digital intraoral scanning systems. International Journal of Computerized Dentistry. 16 (1), 11-21 (2013).

Play Video

Bu Makaleden Alıntı Yapın
Park, J., Shim, J. Measuring the Complete-arch Distortion of an Optical Dental Impression. J. Vis. Exp. (147), e59261, doi:10.3791/59261 (2019).

View Video