The usefulness of an analytical method to determine inorganic arsenic in a wide range of food matrices is demonstrated. The method consists of selective extraction of inorganic arsenic into chloroform with a final determination by hydride generation-atomic absorption spectrometry.
The European Food Safety Authority (EFSA) underlined in its Scientific Opinion on Arsenic in Food that in order to support a sound exposure assessment to inorganic arsenic through diet, information about distribution of arsenic species in various food types must be generated. A method, previously validated in a collaborative trial, has been applied to determine inorganic arsenic in a wide variety of food matrices, covering grains, mushrooms and food of marine origin (31 samples in total). The method is based on detection by flow injection-hydride generation-atomic absorption spectrometry of the iAs selectively extracted into chloroform after digestion of the proteins with concentrated HCl. The method is characterized by a limit of quantification of 10 µg/kg dry weight, which allowed quantification of inorganic arsenic in a large amount of food matrices. Information is provided about performance scores given to results obtained with this method and which were reported by different laboratories in several proficiency tests. The percentage of satisfactory results obtained with the discussed method is higher than that of the results obtained with other analytical approaches.
Since January 2016 maximum levels for inorganic arsenic (iAs) in several rice commodities have been included in Commission Regulation (EC) 1881/2006 setting maximum levels for certain contaminants in foodstuffs1 with 0.10 µg/L for rice destined for the production of food for infants and young children, 0.20 µg/L for non-parboiled milled rice (polished or white rice), 0.25 µg/L for parboiled rice and husked rice and 0.30 µg/L for rice waffles, rice wafers, rice crackers and rice cakes. This update of the European legislation for contaminants in food followed the Scientific Opinion on Arsenic in Food of the European Food Safety Authority (EFSA)2 in which it is estimated that the exposure through diet to iAs for average and high consumers in Europe is such that can pose a risk to some consumers, keeping in mind that chronic exposure to iAs causes cancer of the lung, skin and bladder, and skin lesions. In the scientific report of EFSA on Dietary exposure to inorganic arsenic in the European population3, published in 2014, it is concluded that the main contributors to iAs in the diet for consumers of all ages are processed products made of cereals other than rice and that also rice, milk, dairy products and drinking water contribute significantly to iAs intake, with milk and dairy products being the main contributors for toddlers and infants.
In 2010 the European Union Reference Laboratory for Heavy Metals in Feed and food, EURL-HM, ran a proficiency test, IMEP-107, for the determination of iAs in rice, demonstrating that it was possible to determine iAs in rice with sufficient accuracy, regardless of the analytical method used4.
Several analytical methods have been validated for the determination of iAs in foodstuffs. China was the first country to introduce in its legislation a maximum level for iAs in rice. To make the implementation of legislation possible, a standard method was published in 2003 for the determination of what in the standard is called "abio-arsenic"5. The European Committee for Standardisation (CEN), published in 2008 a standardized method, EN 15517:2008, for the determination of iAs in seaweed6. The two methods are based on the use of optimized conditions to generate arsine only from iAs. In that way separation of iAs from other arsenic species that can also generate arsenic hydride is not needed. The final determination is done by atomic fluorescence5 or by hydride generation atomic absorption spectrometry, HG-AAS6. However, it is difficult to set the exact conditions to generate arsenic hydride without suffering from interference of other arsenic compounds and all the iAs mass fractions in algae reported in IMEP-112 (PT organized by the EURL-HM) obtained with those two methods, were scored as unsatisfactory7. Organoarsenic species, such as monomethylarsonic acid (MMA), dimethylarsinic acid (DMA), and arsenosugars present in algae samples, can generate volatile hydrides too and could interfere in the determination of iAs leading to a positive bias in the results8.
Recently, CEN published a new standard method, EN 16802:2016, for the determination of iAs in foodstuffs of marine and plant origin using HPLC-ICP-MS9. Not all laboratories are equipped with that type of instrumentation and non-expensive, straight-forward methods are needed, in particular in countries with less developed laboratory infrastructures.
In 2012 CEN standardized a method for the determination of iAs in animal feeding stuffs by HG-AAS after microwave extraction and off-line separation of iAs by solid phase extraction (SPE), EN 16278:201210. This method which proved to be fit for analyzing iAs in feed could lack the sensitivity required to determine iAs in food of non-marine origin, which according to EFSA seems to be the main dietary contributors in Europe3. However, the same group that developed and validated EN 16278:2012 tested and successfully applied and validated the method to determine iAs in seafood and rice in a collaborative trial11,12.
An alternative method for the determination of iAs in food matrices after selective extraction of iAs into chloroform and further quantification by HG-AAS, was recently validated by the Joint Research Centre (JRC) in a collaborative trial13. The selectivity of the method is better than that of direct HG-AAS and is easy to implement not requiring the use of sophisticated instrumentation such as HPLC-ICP-MS. In this manuscript, the feasibility of using this method to determine iAs in a wide range of food matrices: vegetables, grains, mushrooms and food of marine origin, has been evaluated. Furthermore, the performance of laboratories that used the method in proficiency tests organized by the EURL-HM and the JRC covering several matrices is described.
NOTE: All the material used needs to be decontaminated with 10% (m/v) HNO3 and rinsed at least twice with deionized water.
1. Hydrolysis
2. Extraction
3. Clean-up of the Chloroform Phase
4. Back-extraction
5. Sample Mineralization
Note: This step allows elimination of interferences and pre-concentration in samples in which the iAs mass fraction is close to or below the quantification limit, and it is frequently omitted by laboratories which use this protocol with ICP-MS for final determination instead of HG-AAS .
6. Calibration
NOTE: For quantification purposes use an external calibration curve of As(III) in the range 0.5 – 10 µg/L. Use a 1000 mg/L As(V) commercially available certified standard solution to construct the calibration curve applying subsequent dilutions.
7. Determination
8. Quantification
The method was applied to determine the iAs mass fraction in several food commodities purchased from various Spanish markets. The results obtained with this method for a series of different matrices are classified in Table 4 following the categories used by EFSA3 in a report in which dietary exposure to inorganic arsenic in the European population is evaluated on the basis of data reported by Official Control Laboratories (OCL). The results in Table 4 represent the mean of three replicates ± the reproducibility standard deviation (SR) for the different food categories, calculated during the collaborative trial in which the present method was validated13. The results shown in Table 4 are in good agreement with other previously published in similar matrices11,12,14.
Of particular relevance are the results obtained for iAs in different types of rice because maximum limits are included for them in the European legislation for contaminants in food1. The highest values being obtained for brown rice and the lowest for white rice, in agreement with the findings of the OCLs3. The highest levels were found for the sea weed Hizikia fusiforme, whose consumption has been discouraged by several authorities as indicated in the report by EFSA.
The performance of laboratories that participated in PTs organized by the EURL-HM and the JRC and that used this method for the determination of iAs, has been compared to the performance of laboratories using other methods. Most of the other methods are based on HPLC-ICP-MS (around 50% of the evaluated results) and on HG-AAS without previous separation of iAs from other arsenic species (25% of the total), Figure 1. Other approaches used (around 15% of the evaluated results), were based on electrothermal atomization (ETAAS), fluorescence detection and ICP coupled to atomic emission spectroscopy (ICP-AES), with and without hydride generation, and are evaluated together under the name "Other methods" because the individual numbers would be too few to be of any statistical significance.
Some of the laboratories that have used the evaluated method introduced some variations to the original protocol and used ICP-MS instead of FI-HG-AAS. Frequently those laboratories did not apply the dry ashing step (Step 5 in the protocol) and just introduced the 1 M HCl phase into the ICP-MS. The PTs evaluated covered various matrices: rice15,16, wheat, spinach, algae17 and chocolate18.
The performance of laboratories was expressed as z-score:
Where:
xlab is the measurement result reported by a participant in a PT
Xref is the assigned value (used to benchmark laboratories). In all PTs dealt within this paper the assigned value was established by a group of expert laboratories in the field of iAs analysis using different analytical methods.
σ is the standard deviation for proficiency assessment, fixed by the PT provider taking into consideration the state of the art in a certain area of analysis. In the PTs considered in this paper σ was 15% of the assigned value for rice and wheat, 22% in algae and 25% for spinach and chocolate.
The interpretation of the z score is done according to ISO 17043:201019:
|score| ≤ 2 satisfactory (S) performance
2 < |score| < 3 questionable (Q) performance
|score| ≥ 3 unsatisfactory (U) performance
Seventy-five per cent of the results obtained with the method described above, got a satisfactory z-score. The determination of the iAs mass fraction in algae turned out to be challenging as expected, taking into consideration the complex distribution of arsenic species in matrices of marine origin. Two out of the three values reported in IMEP-112 for iAs in algae, using this method, got an unsatisfactory z-score. The same difficulty was observed among the results obtained with other methods. Excluding the results reported for iAs in algae, 85% of the results obtained with the evaluated method were satisfactory.
Figure 1: Comparison of Performances (expressed as z-scores) of Laboratories Taking Part in PTs (IMEP-107, IMEP-112, EURL-HM-20 and IRMM-PT-43) with the Method Described in this Paper and with Other Commonly Applied Methods. S: satisfactory, Q: questionable and U: unsatisfactory. Please click here to view a larger version of this figure.
Expected iAs mass fraction lower than 0.010 mg/kg |
Expected iAs mass fraction higher than what is covered by the calibration curve |
|
6 mol L-1 HCL volume used to re-dissolve the ashes (mL) | 2 | 10 |
Pre-reducing agent volume (mL) | 2 | 10 |
Final volume (mL) | 10 | 50 |
Table 1: Modifications of the Protocol when Analyzing Samples in which Very Low or Very High iAs Concentrations are Expected.
Concentration in the calibration curve (µg/L) |
Aliquot (mL) |
0.5 | 1 |
1 | 2 (QC1) |
2.5 | 5 |
5 | 10 (QC2) |
7.5 | 15 |
10 | 20 |
All As(III) calibration standard solutions shall be prepared freshly before each calibration. |
Table 2: Aliquots to be taken from the 25 µg/L As(V) standard solution to construct the As(III) calibration curve in a 50 mL final volume.
Flow injection Hydride generation |
· Loop sample: 0.5 mL (To be adapted when the reconstitution volume of the final pre-reducing solution is different from 25 mL). |
· Reducing agent: 0.2 % (w/v) NaBH4 in 0.05 % (w/v) NaOH; 5 mL/min flow rate. | |
· HCl solution 10 % (v/v), 10 mL/min flow rate. | |
· Carrier gas: Argon, 100 mL/min flow rate. | |
Atomic absorption spectrometer |
· Wavelength: 193.7 nm |
· Spectral band-pass: 0.7 nm | |
· Electrodeless discharge lamp system 2 | |
· Lamp current setting: 400 mA | |
· Cell temperature: 900 °C |
Table 3: Instrumental Conditions used for iAs Quantification by HG-AAS.
Food | i-As (µg/kg fresh weight) | |
Grain and grain-based products | ||
Rice | White | 113 ± 18 |
73 ± 12 | ||
56 ± 9 | ||
Brown | 197 ± 32 | |
125 ± 20 | ||
275 ± 44 | ||
Parboiled | 134 ± 21 | |
159 ± 25 | ||
Wafers | 162 ± 26 | |
127 ± 20 | ||
Vegetable and vegetable products | ||
Dehydrated mushroom | Boletus edulis | 174 ± 10 |
Galocybe gambosa | 74 ± 4 | |
Marasmius oreades | 104 ± 6 | |
Cantharellus lutescens | 16 ± 1 | |
Lentinula edodes | 96 ± 6 | |
Sea weed | Hizikia fusiforme | 97000 ± 14550 |
44943 ± 6742 | ||
Fucus vesiculosus | 288 ± 43 | |
433 ± 65 | ||
Fish and other seafood | ||
Fish meat | Flathead grey mullet | 53 ± 12 |
21 ± 5 | ||
European eel | 72 ± 16 | |
42 ± 9 | ||
Crayfish | 33 ± 7 | |
20 ± 4 | ||
Tuna | 11 ± 2 | |
5 ± 1 | ||
Molluscs | Clam | 243 ± 54 |
133 ± 29 | ||
Mussel | 32 ± 32 | |
139 ± 31 |
Table 4: Results Obtained for a Range of Different Matrices Applying the Described Method.
A critical step in the described protocol is the clean-up of the chloroform phase (Step 3.2) because any acid phase residues remaining in the chloroform phase will lead to overestimated iAs results since all other arsenic species in the sample are present in the acid phase. This is of particular relevance when analyzing marine samples due to the presence of a plethora of organic species, which could account for most of the arsenic mass fraction present in the sample. The use of a hydrophobic PTFE (3.3) membrane is of paramount importance. If an emulsion is formed during the extraction of iAs into chloroform, the speed of centrifugation (3.1) can be increased. Other traditional approaches to eliminate emulsions can also be applied. Another critical step is the mineralization (Step 5.3). The rate of increase in temperature must be strictly implemented to avoid any projections which would reduce the iAs recovery leading to an uncontrolled negative bias and could be dangerous for the analyst.
As mentioned above some laboratories have used the evaluated method using ICP-MS instead of FI-HG-AAS. In such a case the dry ashing step (Step 5 in the protocol) is not needed and the 1 M HCl phase can be introduced into the ICP-MS. In the case of HG-AAS, due to its higher detection limit, a pre-concentration step which also eliminates possible interferences, is needed.
The percentage of satisfactory results obtained with the method described in this paper, both with and without the results reported for algae, is comparable to that of HPLC-ICP-MS and higher than that of HG-AAS. The latter technique (HG-AAS) is widely available but prone to interferences from organic arsenic species, especially in food commodities with a complex arsenic species distribution pattern. The lowest percentage of satisfactory results characterizes those obtained with "Other methods" but it must be kept in mind that it covers several analytical approaches, each one of them represented by a small amount of results, Figure 1. The method presented in this paper is an alternative to the more sophisticated/expensive HPLC-ICP-MS, still being characterized by a similar performance even in complex matrices. Frequently the use of hyphenated techniques, such as HPLC-ICP-MS, requires highly qualified operators and expensive infra-structures. The method presented in this paper can be implemented by any analyst trained in basic analytical chemistry.
There are some main drawbacks associated to the method. It is time-consuming since several steps must be followed to separate iAs from other arsenic species and to pre-concentrate iAs down to even sub-ppm levels. It implies the use of chloroform. There is a tendency to avoid the use of chlorinated compounds in the laboratories, due to the negative health effects that they could have. Nevertheless, if good laboratory practices are kept and samples are handled in fume hoods, those negative effects could be avoided. MMA will interfere in the determination of iAs. This must be kept in mind when analyzing samples in which MMA could be present, such as algae, fish and other seafood. However, MMA is normally present in small amounts which would be covered by the uncertainty associated to the results obtained for iAs.
The authors have nothing to disclose.
The authors thank Dr. F. Cordeiro of the JRC for the useful discussions about statistical treatment of data. Expert laboratories in the analysis of iAs in biological matrices that provided results to be used as assigned value in PTs and laboratories that took part in the studied PTs are acknowledged.
Deionised water | Any available | 18.2 MΩ cm | |
Concentrated hydrochloric acid (HCl). | Any available | Not less than 37 % m/v, c(HCl) = 12 mol/L, with a density of approx. ρ (HCl) 1.15 g/L |
|
Concentrated nitric acid (HNO3) | Any available | Not less that 65 % m/v, c(HNO3) = 14 mol/L, with a densitiy of approx. ρ 1.38 g/L |
|
Chloroform | Any available | Harmful by inhalation and if swallowed. Irritating to skin. Wear suitable protective clothing and gloves. |
|
Hydrogen bromide (HBr) | Any available | Not less than 48 % m/v | |
Hydrazine sulphate (N2H6SO4) | Any available | Harmful if swallowed. Causes burns. May cause cancer. |
|
Magnesium nitrate hexahydrate [Mg(NO3)6H2O] | Any available | ||
Magnesium oxide (MgO) | Any available | ||
Potassium iodide (KI) | Any available | ||
Ascorbic acid (C6H8O6) | Any available | ||
Sodium hydroxide (NaOH) | Any available | ||
Sodium borohydride (NaBH4) | Any available | ||
Arsenic (V) standard solution | Any available | 1000 mg/L Use certified standard solutions commercially available |
|
Centrifuge | Any available | ||
Mechanical shaker | Any available | ||
Sand bath | Any available | ||
Muffle furnace | Any available | ||
Polypropylene centrifuge (PC) tubes | Any available | 50 mL with screw cap | |
Syringe filters with hydrophobic PTFE membrane | Any available | 25 mm diameter | |
Pyrex glass beaker | Any available | Tall form 250 mL, capable of withstanding 500 °C |
|
Watch glasses | Any available | ||
Volumetric flasks | Any available | 10, 25, 100 or 200, Class A. |
|
Plastic funnels | Any available | ||
Whatman n° 1 paper or equivalent | Any available | ||
Atomic absorption spectrometer equipped with a flow injection system (FI-AAS) | Any available |