This protocol describes the rapid and highly sensitive quantification of Förster resonance energy transfer (FRET) sensor data using a custom-made portable FRET analyzer. The device was used to detect maltose within a critical temperature range that maximizes detection sensitivity, enabling practical and efficient assessment of sugar content.
Recent improvements in Förster resonance energy transfer (FRET) sensors have enabled their use to detect various small molecules including ions and amino acids. However, the innate weak signal intensity of FRET sensors is a major challenge that prevents their application in various fields and makes the use of expensive, high-end fluorometers necessary. Previously, we built a cost-effective, high-performance FRET analyzer that can specifically measure the ratio of two emission wavelength bands (530 and 480 nm) to achieve high detection sensitivity. More recently, it was discovered that FRET sensors with bacterial periplasmic binding proteins detect ligands with maximum sensitivity in the critical temperature range of 50 – 55 °C. This report describes a protocol for assessing sugar content in commercially-available beverage samples using our portable FRET analyzer with a temperature-specific FRET sensor. Our results showed that the additional preheating process of the FRET sensor significantly increases the FRET ratio signal, to enable more accurate measurement of sugar content. The custom-made FRET analyzer and sensor were successfully applied to quantify the sugar content in three types of commercial beverages. We anticipate that further size reduction and performance enhancement of the equipment will facilitate the use of hand-held analyzers in environments where high-end equipment is not available.
Förster resonance energy transfer (FRET) has been widely used as a biometric sensor to detect small molecules such as sugars, calcium ions, and amino acids1-4. FRET biosensors contain fluorescent proteins, cyan fluorescent proteins (CFPs), and yellow fluorescent proteins (YFPs), which are fused to both ends of periplasmic-binding proteins (PBPs). Sugars bind to PBPs located in the middle of the FRET sensor, causing structural changes to the sensor that subsequently alter the distance and transition dipole orientation of the two fluorescent proteins at either end of the PBPs. This change enables quantitative analysis of sugar content by measuring the ratio of the emission wavelengths of EYFP (530 nm) and ECFP (480 nm). Owing to the high sensitivity, specificity, real-time monitoring capacity, and fast response time of FRET biosensors, these sensors are widely used in environmental, industrial, and medical applications5. Moreover, ratiometric measurement using FRET biosensors has important practical benefits, as it can be used to measure components in complex biological samples where the sensor concentration cannot be easily controlled and background fluorescence is always present.
Despite these advantages of FRET-based sensors for quantitative visualization, small structural changes with incomplete domain motion-transfer to the fluorescent proteins produce inherently weak signal intensity. This weak signal limits the application of FRET-based sensors for in vitro or in vivo analysis6. Consequently, most FRET biosensors require the use of expensive and highly sensitive equipment. Previously, we developed an inexpensive and portable FRET analyzer with capabilities similar to those of the existing fluorescence analyzers7. In this device, inexpensive 405-nm band ultraviolet light-emitting diode (LED) was used as the light source to cause excitation of the fluorescence signal, replacing an expensive lamp or laser. The detection system of the analyzer efficiently focuses the dissipating fluorescence signal onto two photodetectors with a silicon photodiode. In a more recent study, we showed that optimization of detection temperature at 50 – 55 °C could significantly magnify the ratiometric FRET signal8. This temperature-specific signal enhancement, along with the custom-made FRET analyzer, enables the use of FRET sensors in more general diagnostic applications with rapid and high sensitivity.
In this protocol, we demonstrated the general applicability of the FRET analyzer under optimal FRET temperature conditions by quantifying the sugar content of commercially-available beverages. This protocol provides the details of the FRET device operation, as well as a brief description of sensor and sample preparation. We anticipate that this report will promote the potential application of the portable analyzer in small-scale laboratory environments and provide a foundation for further development of an inexpensive on-site diagnostic device with FRET-based biosensors.
1. Preparation of Biosensor
2. Measurement of Sugar Content using the FRET Analyzer
NOTE: The details of the FRET analyzer construction were described in our previous work7.
To perform quantitative analysis of sugar content using the FRET analyzer, it is necessary to build a fitted curve estimating the target sugar concentration from the observed FRET ratio. Let r define the ratio of the emission intensity of CFP at 480 nm and the emission intensity of YFP generated at 530 nm (Eq. 1).
The dose-response curve of the FRET biosensor (CMY-BII at 53 °C) can be generated by observing the FRET ratio, r, at different sugar concentrations. The curve can then be expressed as an S-shaped sigmoidal curve as follows:
where rmax and rmin represent the signal ratio with sugar concentrations of 0 and saturated (1,000 μM), respectively; x0 represents the sugar concentration at 50% response; and p represents the slope of the response, which is close to 1 or -1. In the present study, rmax, rmin, x0, and p are 4.256, 2.672, 71.779, and 1, respectively. The concentration range from 1 µM to 1000 µM was used in the model fitting.
Using Equations 1 and 2, the sugar content of commercially- available beverages was quantified with the FRET analyzer. Two maltose FRET sensors were examined to test the signal, r, depending on various temperatures2,8. The first FRET sensor, CMY-0, is a basic FRET-based sensor consisting of CFP, maltose-binding protein (MBP), and YFP, with no linker peptides. The second sensor, CMY-BII, has a Ser-Arg linker between MBP and the two fluorescence proteins2. As Figure 1A shows, CMY-0 is not observed at measurement temperatures below 50 °C, as there is no signal difference between 0 and 1 mM maltose concentration. The signal differences of both FRET sensors were maximized between 50 and 55 °C (Figure 1)8. In order to quantify the sugar content of the three types of commercially-available beverages, a dose-response curve of the CMY-BII sensor at 53-°C was generated (Figure 2A) and the maltose content of the three samples was identified by converting the FRET ratio into the maltose concentration.
As sample A is made of grains such as rice and barley, which are important maltose sources, the sample was expected to contain relatively high maltose content (average 11.892 g/235 ml) (Figure 2A). In contrast, sample C is a sports drink that had the lowest maltose (0.29 g/250 ml) among the three beverages. These results suggest that the FRET analyzer can be used at optimal temperatures to maximize the quantification of sugar content, eliminating the need for an expensive high-end FRET-detection device.
Figure 1. FRET Signal Difference Between 0 and 1 mM Maltose using the FRET Analyzer at Various Temperatures. (A) The CMY-0 sensor showed no signal difference at different maltose concentrations at temperatures below 50 °C. (B) The CMY-BII sensor was able to distinguish the FRET signal difference between 0 and 1 mM maltose in a wide range of temperatures. In both cases, the signal difference dramatically increased in a specific temperature range (50 – 55 °C). The error bars represent standard deviation. Please click here to view a larger version of this figure.
Figure 2. Maltose Content Quantification in Three Commercially-available Beverages. (A) A dose-response curve for CMY-BII. (B) Maltose content of three beverage samples was quantified. Note that "Total sugar" indicates the quantity of all sugar (including maltose) reported by the beverage manufacturer on the beverage label. The error bar indicates the standard deviation. Please click here to view a larger version of this figure.
This protocol allows rapid and efficient quantification of the sugar content in beverage samples, using a custom-made FRET analyzer7 at an optimal temperature for FRET sensors. The analyzer was designed with a recently-developed, inexpensive 405-nm band ultraviolet-LED as the light source and two photodetectors with a silicon photodiode. This device is more cost-effective than other comparable fluorometers. The device showed high detection sensitivity, specifically when measuring the ratio of two emission wavelength bands (530 nm and 480 nm) in an optimal temperature range for FRET sensors. Its sensitivity and intensity in detecting various sugars were superior to those of a fluorescence spectrophotometer device7.
The main goal of this protocol is to support the wide applicability of the FRET-based sensors with the custom-made FRET analyzer. While the analyzer indirectly measures sugar content via FRET sensors, it is clear that the device incorporates a number of the benefits of FRET sensors, including widely extendable genetically engineered ligand specificity, modular design, sensor concentration-independent signals, and accurate targeting of subcellular small molecules. FRET sensors are actually used to detect a wide range of small molecules, including ions9, heme10, and others. Moreover, more than 20 types of FRET constructs can be easily found and ordered through the nonprofit depository AddGene11.
Despite the wide applicability of the FRET analyzer, there are two main issues with the operation of the device. First, because operation of the device is relatively simple, sample preprocessing is the critical step that affects quality of detection, except in cases of device malfunction. In this protocol, one step (sample dilution) was sufficient to process liquid samples that were clearly transparent and contained no insoluble particles. However, other samples may require additional processing to remove insoluble materials, such as cellular or lipid components. Any autofluorescent particles that can affect the FRET signal should also be removed, as noted following Step 2.7. Secondly, quality control and connectivity interfacing with hospital information systems need to be addressed, as with all types of point-of-care testing (POCT) tools12. Since the signal quality of the FRET analyzer largely depends on the quality of the FRET sensor and on the preprocessing steps, regular quality control checks are required to ensure that measurements remain within the standard signal range for regular quality control data analysis. The FRET sensor stability and storage period, both of which are essential for further reliable applications, should be investigated during quality control checks. Creating guidelines and developing appropriate software can also address the connectivity limitation. Current versions of the FRET analyzer are equipped with RS232 connectivity for remote command line control, but wireless communication may be a feature of the next version of the analyzer, which will have an improved interface for hospital information systems.
However, FRET sensors have been engineered for substrate specificity, an approach that normally allows broader specificity2. Consequently, the FRET signal may encounter unintended interference from other ingredients, including other types of sugars in commercial beverages. Further investigations should explore how FRET sensors respond to various sugar mixtures to accurately quantify the amount of sugar. Collaboration with companies that produce the beverages will help confirm the sugar content to calibrate of the FRET analyzer.
It is anticipated that the proposed portable FRET device with various FRET sensors will be used in POCT applications. POCT is used for assessing pregnancy, blood glucose levels, biomarker proteins, infectious bacteria, and infectious viruses. POCT methods have rapid turnaround times and generally exhibit low error rates owing to the small number of processing steps. These are important advantages of POCT over the central laboratory testing approach. Hand-held portable POCT devices, such as the device described herein, have attracted increasing attention because of their potential applications in food assessment and blood sugar monitoring. In particular, glucose monitoring of blood samples in patients with diabetes requires a rapid, accurate, and cost-effective POCT method13. After the Ames research team developed the first blood glucose test strip in 1965 (using a strip that contains glucose oxidase), several technologies were proposed for blood glucose monitoring purposes12. The FRET analyzer is also available to detect glucose in blood samples with appropriate preprocessing of blood and periplasmic glucose-binding protein (MglB)14-based FRET protein.
Simple, rapid methods for food quality assessment are needed. The consumption of sugar-containing beverages is associated with a variety of diseases and syndromes, such as increased body mass index in childhood15, pediatric obesity16, and risk of stroke17. Understanding this connection necessitates accurate measurement of sugar components in beverages. Therefore, the glucose and fructose concentrations of beverages are of interest to scientists concerned with human health. This protocol demonstrates the highly sensitive performance of the FRET analyzer with optimal temperature control. The device may be used with various FRET sensors to detect various small molecules- including glucose and fructose14,15. Our portable and rechargeable device, which has a battery life of 10 – 20 hr, depending on the heating protocol, is applicable for POCT. Its simple operational protocol makes the device easy to use and eliminates the need for complicated staff training. With technical improvements, including reduction of equipment size, minimization of pretreatment steps, and identification of practical requirements for field use, this device will promote FRET-based research development in small-scale laboratory environments.
The authors have nothing to disclose.
This research was supported by grants from the Intelligent Synthetic Biology Center of Global Frontier Project (2011-0031944) and the KRIBB Research Initiative Program.
LB | BD | #244620 | |
isopropyl β-D-thiogalactoside (IPTG) | Sigma | I6758 | |
Ampicillin | Sigma | A9518 | |
Tri-HCl | Bioneer | C-9006-1 | |
PMSF | Sigma | 78830 | |
EDTA | Bioneer | C-9007 | |
DTT | Sigma | D0632 | |
NaCl | Junsei | 19015-0350 | |
phosphate-buffered saline (PBS) | Gibco | 70011-044 | 0.8% NaCl, 0.02% KCl, 0.0144% Na2HPO4, 0.024% KH2OP4, pH 7.4 |
SOC | 2% tryptone, 0.5% Yeast extract, 10 mM NaCl, 2.5 mM KCl, 10 mM MGCl2, 20 mM Glucose | ||
Resource Q | Amersham Biosciences | 17-1177-01 | 6 × 30 mm anion-exchange chromatography column |
HisTrap HP1 | Amersham Biosciences | 29-0510-21 | |
Quartz cuvette | Sigma | Z802875 | |
AKÄKTAFPLC | Amersham Biosciences | 18-1900-26 | a fast protein liquid chromatography (FPLC) |
Cary Eclipse | VarianInc | a fluorescence spectrophotometer | |
VICTOR | PerkinElmer | 2030-0050 | a multilabel plate reader |
E. coli JM109 (DE3) | Promega | Electrocompetent cells | |
A (Beverage) | Korea Yakult Co. (Korea) | Birak | Fermented drinks |
B (Beverage) | Lotte Foods (Korea) | Epro | Soft drink |
C (Beverage) | Lotte Foods (Korea) | Getoray | Sports drink |