Özet

Yüksek frekanslı Ultrason Kullanımı Fare Embriyolar Kontrast Görüntüleme

Published: March 04, 2015
doi:

Özet

Burada, biz yaşam içine ultrason mikrokabarcık kontrast maddeler, izole geç gebelik sahne kemirgen embriyolar enjekte etmek için bir protokol mevcut. Bu yöntem perfüzyon parametrelerinin ve kontrastlı yüksek frekanslı ultrason görüntüleme kullanılarak embriyo içinde vasküler moleküler belirteçlerin çalışma sağlar.

Abstract

Ultrasound contrast-enhanced imaging can convey essential quantitative information regarding tissue vascularity and perfusion and, in targeted applications, facilitate the detection and measure of vascular biomarkers at the molecular level. Within the mouse embryo, this noninvasive technique may be used to uncover basic mechanisms underlying vascular development in the early mouse circulatory system and in genetic models of cardiovascular disease. The mouse embryo also presents as an excellent model for studying the adhesion of microbubbles to angiogenic targets (including vascular endothelial growth factor receptor 2 (VEGFR2) or αvβ3) and for assessing the quantitative nature of molecular ultrasound. We therefore developed a method to introduce ultrasound contrast agents into the vasculature of living, isolated embryos. This allows freedom in terms of injection control and positioning, reproducibility of the imaging plane without obstruction and motion, and simplified image analysis and quantification. Late gestational stage (embryonic day (E)16.6 and E17.5) murine embryos were isolated from the uterus, gently exteriorized from the yolk sac and microbubble contrast agents were injected into veins accessible on the chorionic surface of the placental disc. Nonlinear contrast ultrasound imaging was then employed to collect a number of basic perfusion parameters (peak enhancement, wash-in rate and time to peak) and quantify targeted microbubble binding in an endoglin mouse model. We show the successful circulation of microbubbles within living embryos and the utility of this approach in characterizing embryonic vasculature and microbubble behavior.

Introduction

Kontrastlı ultrason görüntüleme görselleştirmek ve damar ortamı karakterize etmek kabarcık kontrast maddelerin kullanımını kolaylaştırır. Bu ajanlar mikrosirkülasyon, damarlanma ve kardiyovasküler fonksiyonu noninvaziv değerlendirmesini sağlar. Buna ek olarak, baloncuk yüzeyinin modifikasyonu mümkün vasküler olay moleküler ultrason görüntüleme yapmak anjiyojenez, ateroskleroz ve inflamasyon 1,2 klinik öncesi uygulamalarda gösterildiği gibi, endotelyal biyolojik bağlanma hedeflenen mikro-sonuçlanabilir. Kontrastlı ultrason nedenle de sağlıklı ve hastalıklı damar durumları 3-5 etkileyen karmaşık ve çeşitli ortamlar belirlemek için kullanılabilir.

Yıl geçmiş sayıda, kabarcık görüntüleme programı ilgi yönlü fare embriyo modeli genişletmiştir. Memeli bir gelişme modeli olarak, embriyonik damar içine kabarcıkların giriş fizyolojik artırırGelişmekte olan dolaşım sistemi (örneğin, kan akımı, kardiyak output) ve transgenik durumlarda ve kalp hastalığı 6,7 hedeflenen mutant fare modellerinde, çalışma kardiyovasküler fonksiyonu nasıl değiştirdiğini genetik faktörler anlayışlar verebilir. Aslında, nicel ve nitel 2D embriyonik beyin damar analizleri zaten 8 elde edilmiştir. Bundan başka, fare embriyosu canlı dokuda vasküler belirteçlerle hedeflenen mikro-kabarcıklar bağlanmasının incelenmesi için mükemmel bir model olarak sunulur. Bartelle ve diğ. 9, örneğin, Biotag-BirA transgenik embriyo bağlanma ve damar anatomisinin incelemek hedeflenen değerlendirir embriyo kalp ventriküllere, avidin mikro-kabarcıkları getirmiştir. kliniğe Bu teknik çeviri önemli bir kriter – heterozigot ve homozigot fare modelleri üretimi de moleküler ultrason kantitatif tabiatının belirlemek amacıyla yapılan tümör modeli araştırmaları için bir vekil olarak kullanılabilir edilebilir.

<p clgöt = "jove_content"> mikro-en sık laparotomi 8-10 aracılığıyla maruz tek embriyo içi kardiyak enjeksiyonlar aracılığıyla embriyonik dolaşıma tanıtıldı. utero enjeksiyonları Ancak, bir takım zorluklarla karşı karşıya. Bu enjeksiyon rehberlik, anne ve batın embriyo hareketi karşı ihtiyacını, anne hemodinamik canlılığının bakım ve 11 kanama nedeniyle anestezi ve komplikasyon uzun vadeli etkileri ele batın embriyolar içerir. Bu nedenle, soruşturmanın amacı izole yaşayan geç evre embriyoların 12 içine mikrokabarcıkları enjekte için bir teknik geliştirmek oldu. Bu seçenek, enjeksiyon kontrol ve konumlandırma, tıkanma olmadan görüntüleme uçağın tekrarlanabilirlik ve basitleştirilmiş görüntü analizi ve ölçümü açısından daha fazla özgürlük sunuyor.

Bu çalışmada, biz fo yaşayan kemirgen embriyolara kabarcıkların enjeksiyon için yeni bir prosedür anahatr kabarcık kinetik davranışları okuyan ve amaçları endojen endotel yüzey belirteçleri bağlayıcı Kabarcığa hedeflenen Derslerinizi. Lineer olmayan bir kontrast özel ultrason görüntüleme tepe geliştirilmesinde uygulanmıştır (PE), yıkama-oranı ve süresi izole E17.5 embriyolar (TTP) tepe dahil olmak üzere temel perfüzyon bir dizi parametre ölçülmesi için kullanılır. Bu, aynı zamanda, aktif bir anjiyojenez 13 bölgelerinde damar endotel hücrelerinde yüksek ifade için endoglin klinik olarak anlamlı bir hedef işlev transjenik fare modeli, bir embriyonik endoglin kaybına moleküler ultrason kantitatif tabiatının değerlendirmek için embriyo modelinin geçerliliğini göstermek . endoglin hedefli (MB E) fare izotip IgG2 kontrol (MB C) ve hedefsiz (MB U) mikro-kabarcıkların yapışma heterozigot endoglin (İngilizce +/-) ve homozigoz endoglin (İngilizce + / +) sentezleyen embriyolar değerlendirilir. Hedeflenen bindinin analizing moleküler ultrason endoglin genotipleri arasında ayırt ve ölçülebilir moleküler ultrason seviyelerine reseptör yoğunlukları ilgili yeteneğine sahip olduğunu ortaya koymaktadır.

Protocol

NOT: Bu çalışmada gerçekleştirilen deneysel prosedürler Hayvan Bakım Sunnybrook Araştırma Enstitüsü Komitesi (Toronto, Ontario, Kanada) tarafından onaylanmıştır. Hayvanların insancıl için prosedürler her zaman dikkat edilmelidir. Bu araştırmacı bir ultrason görüntüleme sisteminin temel operasyonda eğitimli olduğu varsayılır. Bu protokol, iki kişi ile iyi çalışır. 1. Hayvan Modelleri Mate CD-1 erkek ve dişi Mus musculus perfüzyon çalı?…

Representative Results

Eski in utero fare embriyoları ultrason kontrast ajanlarının enjekte enjeksiyonu ve ilgili ultrason görüntüleme boyunca rahim ve canlılığının korunması ile ilgili oturma, geç gebelik aşamasında embriyoların başarılı bir şekilde izole bağlıdır. Embriyo batın ve Şekil 1 'de gösterildiği gibi, konumlandıktan sonra, embriyonik damar içine kontrast maddenin dikkatli bir enjeksiyon mümkündür. Bir E17.5 fare embriyo tipik bir B-kipi ultrason görüntüsü Ş…

Discussion

Ultrason kontrast ajanlar geç evre gebelik fare embriyoları ve doğrusal olmayan kontrast görüntüleri enjekte edildi perfüzyon parametreleri ölçmek için edinilen ve bağlayıcı Kabarcığa hedef alındı. Embriyonik damar sistemi içerisinde mikro-kabarcıkların başarılı görüntüleme bir dizi faktöre, ilk olarak embriyo canlılığı bağımlıydı. Tüm ekipman ve cihaz enjeksiyonun başlangıcında rahim embriyoların izolasyonu için gerekli olan zamanı en aza indirmek için önceden hazırlanmışt…

Açıklamalar

The authors have nothing to disclose.

Acknowledgements

This work was supported by the Terry Fox Program of the National Cancer Institute of Canada.

Materials

Reagents Company Catalog Number Comments/Description
Antibodies (biotinylated, eBioscience) Antibody choice depends on the experiment
      rat isotype IgG2 control eBioscience 13-4321-85 This antibody/microbubble combination is often required as experimental control 
      biotin anti-mouse CD309 eBioscience 13-5821-85
Biotinylated rat MJ 7/18 antibody to mouse endoglin In house hybridoma Outside antibodies may also be appropriate: we  have used eBioscience (13-1051-85 ) in the past
Distilled water
Embryo media
     500 mL Dulbecco’s Modified Eagle’s Medium with high glucose Sigma D5796
     50 mL Fetal Bovine Serum ATCC 30-2020 lot # 7592456
     Hepes  Gibco 15630 5mL, 1M
     Penicillin-Streptomycin  Gibco 15140-122 5 mL, 10,000 units Pen., 10,000 ug Strep
Ethanol, 70%
Ice
Paraformaldehyde Sigma 76240 4%
Phosphate Buffered Saline [1x]  Sigma D8537 1x, w/o calcium chloride & magnesium chloride
Pregnant mouse, CD-1 Charles River Laboratories Inc. 
0.9% sodium chloride (saline) Hospira 0409-7984-11
Ultrasound contrast agent, target ready and untargeted MicroMarker; VisualSonics Inc.
Ultrasound gel (Aquasonic 100, colourless) CSP Medical 133-1009
Equipment
Cell culture plates (4) :  100×20 mm Fisher Scientific 08-772-22
Cell culture plates (12) : 60×15 mm Sigma D8054
Centrifuge Sorvall Legend RT centrifuge 
Conical tubes, 50 mL BD Falcon VWR 21008-938
Diluent Beckman Coulter Isoton II Diluent, 8448011
Dissection scissors (Wagner) Fine Science Tools Wagner 14068-12
Forceps (2), Dumont SS (0.10×0.06 mm) Fine Science Tools 11200-33
Forceps, splinter VWR 25601-134
Glass beaker, 2 L (Griffin Beaker) VWR 89000-216
Glass capillaries, 1×90 mm GD-1 with filament Narishige GD-1
Glass needle puller Narishige PN-30
Gloves Ansell 4002
Gross anatomy probe Fine Science Tools 10088-15
Hot plate VWR 89090-994
Ice bucket Cole Parmer RK 06274-01
Imaging Platform VisualSonics Inc. Integrated Rail System
Light source, fiber-optic Fisher Scientific 12-562-36 Ideally has adjustable arms
Luers (12), polypropylene barbed female ¼-28 UNF thread Cole Parmer 45500-30
Micro-ultrasound system, high-frequency VisualSonics Inc. Vevo2100
Needles, 21 gauge  (1”) VWR 305165
Particle size analyzer Beckman Coulter Multisizer 3 Coulter Counter
Perforated spoon (Moria) Fine Science Tools MC 17 10373-17
Pins (6), black anodized minutien 0.15 mm Fine Science Tools 26002-15
Pipettors [2-20 uL, 20-200uL, 100-1000uL] Eppendorf Research Plus  adjustable 3120000038;       3120000054;       3120000062
Pipettor tips [2-200uL, 50-1000uL] Eppendorf epT.I.P.S.                   22491334;             022491351
Scissors
Sylgard 184 Silicone Elastomer Kit Dow Corning
Tubing, Tygon laboratory 1/32×3/32” VWR 63010-007
Wooden applicator stick (swab, cotton head) VWR CA89031-270
Surgical microscope 5-8x magnification Fisher Scientific Steromaster
Syringes, 1 mL Normject Fisher 14-817-25
Syringes (10), 30 mL VWR CA64000-041
Syringe infusion pump  Bio-lynx  NE-1000
Thermometer, -20-110oC VWR 89095-598
Timer VWR 33501-418
Tubes, Eppendorf VWR 20170-577
Tube racks (3) VWR 82024-462
Ultrasound transducer, 20 MHz VisualSonics Inc. MS250
Vannas-Tubingen, angled up Fine Science Tools 15005-08

Referanslar

  1. Voigt, J. U. Ultrasound molecular imaging. Methods. 48 (2), 92-97 (2009).
  2. Klibanov, A. Preparation of targeted microbubbles: Ultrasound contrast agents for molecular imaging. Medical Biological Engineering Computing. 47 (8), 875-882 (2009).
  3. Cosgrove, D., Lassau, N. Imaging of perfusion using ultrasound. European Journal Of Nuclear Medicine And Molecular Imaging. 37 (S1), 65 (2010).
  4. Williams, R., et al. Dynamic microbubble contrast-enhanced US to measure tumor response to targeted therapy: A proposed clinical protocol with results from renal cell carcinoma patients receiving antiangiogenic therapy. Radiology. 260 (2), 581 (2011).
  5. Burns, P. N., Wilson, S. R. Focal liver masses: Enhancement patterns on contrast-enhanced Images – Concordance of US scans with CT scans and MR images. Radiology. 242 (1), 162 (2006).
  6. Phoon, C. K. L., Aristizabal, O., Turnbull, D. H. 40 MHz doppler characterization of umbilical and dorsal aortic Blood flow in the early mouse embryo. Ultrasound. In Medicine And Biology. 26 (8), 1275-1283 (2000).
  7. Phoon, C. K. L., Aristizabal, O., Turnbull, D. H. Spatial velocity profile in mouse embryonic aorta and doppler-derived volumetric flow: A preliminary model. Am J Physiol Heart Circ Physiol. 283, H908-H916 (2002).
  8. Aristizábal, O., Williamson, R., Turnbull, D. H. . 12A-4 in vivo 3D contrast-enhanced imaging of the embryonic mouse vasculature. Paper presented at Ultrasonics Symposium. , (2007).
  9. Bartelle, B. B., et al. Novel genetic approach for in vivo vascular imaging in mice. Circ.Res. 110 (7), 938-947 (2012).
  10. Endoh, M., et al. Fetal gene transfer by intrauterine injection with microbubble-enhanced ultrasound. Molecular Therapy. 5 (5), 501-508 (2002).
  11. Yamada, M., Hatta, T., Otani, H. Mouse exo utero development system: Protocol and troubleshooting. Congenital Anomalies. 48 (4), 183-187 (2008).
  12. Denbeigh, J. M., Nixon, B. A., Hudson, J. M., Purin, M. C., Foster, F. S. VEGFR2-targeted molecular imaging in the mouse embryo: An alternative to the tumor model. Ultrasound in medicine and biology. 40 (2), 389-399 (2014).
  13. Paauwe, M., Dijke, t. e. n., P, L. J. A. C., Hawinkels, Endoglin for tumor imaging and targeted cancer therapy. Expert Opinion On Therapeutic Targets. 17 (4), 421-435 (2013).
  14. Bourdeau, A., Faughnan, M. E., Letarte, M. Endoglin-deficient mice, a unique model to study hereditary hemorrhagic telangiectasia. Trends Cardiovasc. Med. 10 (7), 279-285 (2000).
  15. Whiteley, K. J., Adamson, S. L., Pfarrer, C. D. Vascular corrosion casting of the uteroplacental and fetoplacental vasculature in mice. Placenta And Trophoblast: Methods And Protocols. 121 (121), 371-392 (2006).
  16. Kulandavelu, S., et al. Embryonic and neonatal phenotyping of genetically engineered mice. ILAR Journal. 47 (2), 103-10 (2006).
  17. Kalaskar, V. K., Lauderdale, J. D. Mouse embryonic development in a serum-free whole embryo culture system. Journal of Visualized Experiments. 85, (2014).
  18. Willmann, J. K., et al. Targeted contrast-enhanced ultrasound imaging of tumor angiogenesis with contrast microbubbles conjugated to integrin-binding knottin peptides. The Journal of Nuclear Medicine. 51 (3), 433-440 (2010).
  19. Deshpande, N., Ren, Y., Foygel, K., Rosenberg, J., Willmann, J. K. Tumor angiogenic marker expression levels during tumor growth: Longitudinal assessment with molecularly targeted microbubbles and US imaging. Radiology. 258 (3), 804-811 (2011).
  20. Lyshchik, A., et al. Molecular imaging of vascular endothelial growth factor receptor 2 expression using targeted contrast-enhanced high-frequency ultrasonography. Journal Of Ultrasound In Medicine. 26 (11), 1575-1586 (2007).
  21. Jerkic, M., et al. Endoglin regulates nitric oxide-dependent vasodilatation. The FASEB Journal. 18 (3), 609-611 (2004).
  22. Denbeigh, J. M., Nixon, B. A., Lee, J. J. Y., et al. . Contrast-Enhanced Molecular Ultrasound Differentiates Endoglin Genotypes in Mouse Embryos. , (2014).
  23. Adamson, S. L., Lu, Y., Whiteley, K. J., et al. Interactions between trophoblast cells and the maternal and fetal circulation in the mouse placenta. Dev Biol. 250, 358-35 (2002).
  24. Needles, A., et al. Nonlinear contrast imaging with an array-based micro-ultrasound system. Ultrasound. Medicine Biology. 36 (12), 2097 (2010).
  25. Watson, E. D., Cross, J. C. Development of structures and transport functions in the mouse placenta. Physiology. 20 (3), 180-193 (2005).
  26. Shalaby, F., Rossant, J., Yamaguchi, T. P., et al. Failure of blood-island formation and vasculogenesis in Flk-1-deficient mice. Nature. 376, 62-66 (1995).
  27. Kwee, L., Baldwin, H. S., Shen, H. M., et al. Defective development of the embryonic and extraembryonic circulatory systems in vascular cell adhesion molecule (VCAM-1) deficient mice. Development. 121, (1995).
  28. Mercurio, A. M. Lessons from the α2 integrin knockout mouse. The American journal of pathology. , 161-163 (2002).
  29. Hodivala-Dilke, K. αvβ3 integrin and angiogenesis: a moody integrin in a changing environment. Curr Opin Cell Biol. 20, 514-519 (2008).
  30. Pysz, M. A., Gambhir, S. S., Willmann, J. K. Molecular imaging: current status and emerging strategies. Clinical radiology. 65, 500-516 (2010).
  31. Cybulsky, M. I., Iiyama, K., Li, H., et al. A major role for VCAM-1, but not ICAM-1, in early atherosclerosis. J Clin Invest. 107, 1255-1262 (2001).
  32. Xu, H., Gonzalo, J. A., St Pierre, ., Y, , et al. Leukocytosis and resistance to septic shock in intercellular adhesion molecule 1-deficient mice. J Exp Med. 180, 95-109 (1994).
  33. Gerwin, N., Gonzalo, J. A., Lloyd, C., et al. Prolonged eosinophil accumulation in allergic lung interstitium of ICAM-2-deficient mice results in extended hyperresponsiveness. Immunity. 10, 9-19 (1999).
  34. Johnson, R. C., Mayadas, T. N., Frenette, P. S., et al. Blood cell dynamics in P-selectin-deficient mice. Blood. 86, 1106-1114 (1995).
  35. Corrigan, N., Brazil, D., McAuliffe, F. High-frequency ultrasound assessment of the murine heart from embryo through to juvenile. Reproductive Sciences. 17 (2), 147-14 (2010).
  36. Turnbull, D. H., Bloomfield, T. S., Baldwin, H. S., Foster, F. S., Joyner, A. L. Ultrasound backscatter microscope analysis of early mouse embryonic brain development. Proc Natl Acad Sci U S A. 92, 2239-2243 (1995).
  37. Greco, A., Mancini, M. L., Gargiulo, S., et al. Ultrasound Biomicroscopy in Small Animal Research: Applications in Molecular and Preclinical Imaging. Journal of Biomedicine and Biotechnology. 2012, (2012).
  38. Pysz, M. A., Guracar, I., Foygel, K., Tian, L., Willmann, J. K. Quantitative assessment of tumor angiogenesis using real-time motion-compensated contrast-enhanced ultrasound imaging. Angiogenesis. 15, 433-442 (2012).
  39. Larina, I. V., et al. Live imaging of blood flow in mammalian embryos using doppler swept-source optical coherence tomography. J.Biomed.Opt. 13 (6), 060506-06 (2008).
  40. Garcia, M. D., Udan, R. S., Hadjantonakis, A. K., Dickinson, M. E. . Live imaging of mouse embryos. 4 (4), 104-10 (2011).
  41. Teichert, A., et al. Endothelial nitric oxide synthase gene expression during murine embryogenesis. Commencement of expression in the embryo occurs with the establishment of a unidirectional circulatory system. Circulation Research. 103 (1), 24-33 (2008).
  42. Walls, J. R., Coultas, L., Rossant, J., Henkelman, R. M. Three-dimensional analysis of vascular development in the mouse embryo. PLoS One. 3 (8), (2008).