EPA Method 1615 uses an electropositive filter to concentrate enteroviruses and noroviruses in environmental and drinking waters. This manuscript describes the procedure for collecting samples for Method 1615 analyses.
EPA Method 1615 was developed with a goal of providing a standard method for measuring enteroviruses and noroviruses in environmental and drinking waters. The standardized sampling component of the method concentrates viruses that may be present in water by passage of a minimum specified volume of water through an electropositive cartridge filter. The minimum specified volumes for surface and finished/ground water are 300 L and 1,500 L, respectively. A major method limitation is the tendency for the filters to clog before meeting the sample volume requirement. Studies using two different, but equivalent, cartridge filter options showed that filter clogging was a problem with 10% of the samples with one of the filter types compared to 6% with the other filter type. Clogging tends to increase with turbidity, but cannot be predicted based on turbidity measurements only. From a cost standpoint one of the filter options is preferable over the other, but the water quality and experience with the water system to be sampled should be taken into consideration in making filter selections.
Human enteric viruses replicate within the gastrointestinal tract and are spread through the fecal-oral route. These viruses are often found in sewage in high concentrations 1-3. They can persist in sewage effluents 4,5, and in surface 6,7, ground 8-10, and treated drinking 11 waters. When present, the concentration of virus in environmental waters in the U.S. typically is too low for direct measurement 12,13. This requires that viruses be concentrated from large volumes of water. During the Information Collection Rule (ICR) monitoring conducted by the U.S. Environmental Protection Agency (USEPA) 14, the virus concentrations of positive samples in the source water of large utilities nationwide ranged from 0.009 to 19.7 most probable number of infectious units (MPN)/L. Median and mean concentrations of positive samples were 0.03 and 0.17 MPN/L for source waters from flowing streams, 0.01 to 0.07 MPN/L for those from lakes and reservoirs, and 0.04 to 0.74 MPN/L for those using groundwaters 11 (data from the ICR Aux1 18 month access database dated 4/25/2000). Virus concentrations of positive samples from a USEPA study of national groundwaters ranged from 0.009 to 2.12 infectious units/L with median and mean concentrations of 0.13 and 0.29 infectious units/L 8. The concentration of virus in positive groundwater samples was higher than those in flowing streams. Most of the facilities using groundwater in these studies obtained water from aquifers located in karst regions. These, along with those located in limestone and crystalline (fractured bedrock) settings are likely to have higher virus concentrations than in other settings 8,15,16. USEPA virus methods specify sampling volumes of 200 L (ICR) to 300 L (Method 1615) of surface water and 1,000 L (ICR) to 1,500 L (Method 1615) of groundwater 17,18. However, even with the use of large sample volumes, most surface and ground water samples are negative for virus 8,11,19,20.
Viruses present in surface waters pose a potential health risk to consumers of drinking water. The Surface Water Treatment Rule requires all treatment plants using surface water to reduce virus concentrations by at least 4-log. Even with a 4-log reduction, infectious virus concentrations in source water as small as 0.0044 MPN/L could lead to one infection per day assuming such average exposure and treatment conditions and the dose response parameters for rotavirus 11,21. The risk from virus in untreated groundwaters could be even greater due to the lack of treatment and viral occurrence. Borchardt and colleagues estimate that up to 22% of acute gastroenteritis in adults and 63% in children less than five years could be due to virus in drinking water in communities using untreated groundwaters 19.
USEPA Method 1615 was developed to detect enterovirus and norovirus during the Unregulated Contaminant Monitoring Regulation's third monitoring cycle (UCMR3) 22 as a national follow up to the findings of Borchardt and colleagues 19,23. The USEPA method was designed primarily for measuring virus in systems using untreated groundwaters, but was written more generally to include other water matrix types. The new method is a hybrid preserving many components from the previous virus method used during the ICR 17, the addition of molecular procedures based upon the method of Borchardt et al. 19,23, and additional primer sets for norovirus 24. The purpose of this paper is to describe the sampling procedure and the steps needed to maintain the integrity of the sample during collection and shipment. An evaluation of the overall method is described in Cashdollar et al. 25. This protocol covers simple field collection of surface and ground waters where a pump and prefilter are not necessary and where adjustments are not required for pH or the presence of a disinfectant in the water to be sampled. The more complex sampling requirements are described in Fout et al. 17,18.
1. Preliminary Procedures
2. Preparation for Sample Collection at the Sampling Site
3. Field Sample Collection
4. Shipment of Field Samples
Filter clogging is a major potential problem that can be encountered with Method 1615. Clogging decreases the flow rate of water through the filter. In some cases this can be overcome by opening the globe valve to allow greater flow. In other cases the filter will become completely clogged before the required volume has passed through it. Table 1 shows the percentage of samples with reduced volumes as a function of filter type, water type, and turbidity. Clogging occurred with both groundwater and surface water samples, with the aluminum oxide nanofiber-based filter outperforming the quaternary amine-based filter for groundwater samples while the reverse was the case for surface water samples. Overall, 6% of samples collected using the quaternary amine-based filter were deficient in volume while 10% of samples collected with the aluminum oxide nanofiber-based filters failed to meet the required minimum specified volume due to clogging. In general, clogging increased with turbidity, but a number of different components contribute to turbidity and some of these do not lead to clogging. For example, 43% of all clogging events occurring during the ICR study were confined to two river systems (data not shown). During the ICR the minimum target volume for surface water was 200 L. The average volume collected during the study was 217 ± 32 L with a median volume of 208 L (data from the ICR Aux1 18 month access database dated 4/25/2000). Thus, the full volume can be collected from many waters even with high turbidities.
The ICR required samplers to use a prefilter when turbidities were over 75 NTU, but even with a prefilter, 34% of the samples collected in waters with turbidities over 75 NTU clogged. Method 1615 recommends the use of a prefilter with waters over 50 NTU; however, since the publication of Method 1615, several different prefilter options in addition to that listed in the method have been tested. None of these resulted in a significant improvement in sample volume (data not shown).
Figure 1. Standard Filter Apparatus. The standard filter apparatus consists of intake, filter housing, and discharge modules (see Supplemental Materials for a description of each module). Please click here to view a larger version of this figure.
Total Samples a | Number of Deficient Samples b | Percent Deficient Samples | Turbidity (NTU) |
Groundwater using NanoCeram Filters c | |||
113 | 1 | 1 | < 20 |
Surface Water using NanoCeram Filters c | |||
83 | 12 | 14 | < 20 d |
8 | 4 | 50 | 20 – < 50 |
6 | 5 | 83 | ≥ 50 d |
Total NanoCeram c | |||
210 | 22 | 10 | ≥ 0 |
Groundwater using 1MDS Filters c | |||
374 | 75 | 20 | < 20 |
Surface Water using 1MDS Filters c | |||
2,693 | 27 | 1 e | < 20 e |
505 | 36 | 7 f | 20 – < 50 f |
122 | 19 | 16 e | 50 – < 75 g |
175 | 60 | 34 e,f | ≥ 75 e,f,g |
Total 1MDS c | |||
3,869 | 217 | 6 | ≥ 0 |
Table 1. Filter Capacity. a Samples are from the following studies from which volume and turbidity data are available: 1) USEPA’s ICR study (data from the ICR Aux1 18 month access database dated 4/25/2000), 2) USEPA groundwater study 8, 3) USEPA Lawrence and Lowell, MA study (unpublished data), 4) USEPA Mississippi River study (unpublished data), 5) USEPA drinking water treatment plant study (unpublished data), 6) USEPA Method 1615 evaluation study 25, and 7) the first three months of the UCMR3 monitoring (unpublished data). b Samples where the filter clogged before the minimum volume of water specified for each study was collected. This was 200 L for surface waters and 1,500 L for groundwaters, but was 200 L for source waters that were groundwater under the USEPA ICR and 1,893 L for the USEPA groundwater study. c The ability to collect the full minimum specified volume is significantly different for samples collected using NanoCeram and 1MDS filters (P = 0.002) and between surface waters and groundwaters both overall (P < 0.001) and for each filter type (P < 0.001) using the Mann-Whitney Rank Sum test. d – g Groups with the same superscript value are significantly different (P < 0.05) according to the Kruskal-Wallis One Way Analysis of Variance on Ranks test and Dunn’s Pairwise Multiple Comparison Procedure. For all statistical tests the dependent variable is the volume passed through the filter as a fraction of the minimum specified volume, but with a maximum value of 1.0.
Different filter types for concentrating viruses from environmental waters have been used over the years 26. Current methods employ ultrafilters 27, electronegative filters 13,28,29, glass wool filters 23, and electropositive filters 30. Electronegative filters were widely used for many years, but a requirement for the addition of salt and the adjustment of the water pH in the field prior to or during sampling limits their usefulness 13. The most practical filter choice for field sampling is electropositive filters. These filters allow the sampling of large volumes of water at high flow rates and without any conditioning of the water. Glass wool filters are the least expensive option, but have slower flow rates than electropositive filters and are not commercially available. Ultrafilters provide the highest virus recoveries over a large range of water quality, but the equipment required for sampling is not readily field portable and the time required to collect samples is much longer 27. Methods recently have been developed that use preconditioned electronegative filters to avoid the need for adjustment in the field, but these may not be applicable for collecting large sample volumes 28,29.
EPA Method 1615 uses electropositive cartridge filters which obtain their positive charge from either aluminum oxide nanofibers or quaternary amines. The advantages of the former over the latter is that it is less expensive and efficiently collects virus from waters over a wider range of natural pH values 30,31; however, each cartridge, as well as the glass wool filters used by Borchardt and colleagues give similar recoveries of enterovirus and norovirus from water 23,31,32 (Cashdollar, unpublished data). Cartridge filters are placed into a simple sampling apparatus that is designed to simplify the collection of samples and reduce contamination during sampling.
Standard methods are invaluable when large studies are conducted using multiple analytical laboratories. EPA Method 1615 provides standard procedures and guidance to minimize the two major sample collection issues that can affect data collected during these studies—false positive results stemming from contamination during sample collection or from inadequately disinfected apparatus components, and the clogging of filter pores by components in the water being sampled.
Just as enteric viruses can be spread person-to-person due to inadequate hygiene, viruses can be introduced into samples from poorly washed hands or hands with contaminated gloves 33. It is essential that samplers understand the potential routes of contamination and use aseptic technique during sampling. Samplers should understand that gloves are primarily used to protect the sampler from exposure and not to protect the apparatus from contamination. Hands should be washed before the start of sampling and care must be taken during the donning of gloves to prevent hand to glove contamination. Samplers with gastroenteritis or respiratory symptoms must not collect samples, as they could be shedding enteroviruses or noroviruses in high levels.
Second, care must be taken to prevent carryover of virus from previous sampling events. To minimize this potential source of contamination, the sampling apparatus in Method 1615 was modified from that of the ICR by not including a pressure regulator and pressure meter between the inlet and the cartridge housing module. These components were removed because the pressures observed during sampling events were always below the maximum housing rating (e.g., 125 psi for 5-inch cartridge housings) and because they were difficult to disinfect. The latter problem was demonstrated through the use of equipment blank controls in studies subsequent to the ICR 6,20. The degree to which it affected the ICR data is unknown, but was likely small; there were only two false positive negative performance evaluation samples during the study (data not shown). To further reduce the possibility of carryover contamination, it also is recommended that the inlet module tubing be replaced after each sampling event. It is especially important to ensure adequate disinfection if the apparatus was used for a quality or performance control that was seeded with virus. Prior to performing disinfection, the concentration of free chlorine should be measured for any loss during storage. In addition to the changes in the configuration of the apparatus described above, it is essential that regular equipment blanks be run to demonstrate that disinfection is effective. Method 1615 mandates that equipment blanks be performed using apparatuses that have been disinfected after being used for virus-seeded controls, thereby simplifying the procedure by eliminating the need to pass a virus-seeded solution through the apparatus prior to disinfection. The concentration of disinfectant was increased to 0.525% hypochlorite for Method 1615, as this concentration is required both to inactivate any viable viruses on the equipment and to degrade nucleic acids. Therefore, equipment blanks must be analyzed using both cell culture and qPCR assays.
Both types of electropositive filters were subject to clogging during the studies reported in Table 1. An unknown number of samples with reduced volumes may have been due to sampling errors, such as a misreading of the totalizer or an intentional early stop of sampling to meet another deadline, especially for waters with turbidity readings less than 20 NTU. The degree of clogging is dependent both upon the filter type and water quality parameters. Prefilters provide some measure of improvement, but if used, should be processed and analyzed separately from the electropositive filter. The current recommendation for the UCMR3 is that the sample be collected without prefilters using two aluminum oxide nanofiber-based filters if at least half the volume can be collected using the first filter.
The authors have nothing to disclose.
The authors thank numerous EPA personnel whose contributions made the monitoring conducted during the ICR and UCMR3 possible, the following lead investigators of other EPA studies reported: Daniel Dahling, Alfred Dufour, Andrey Egorov, Susan Glassmeyer, Asja Korajkic, Richard Lieberman, Robert Safferman, and Tim Wade; and Shannon Griffin and Michael Ware for critically reviewing this manuscript. The authors thank the Indian Hill Water Works for the use of one of their pump houses to demonstrate sample collection. Although this work was reviewed by USEPA and approved for publication, it may not necessarily reflect official Agency policy. Mention of trade names or commercial products does not constitute endorsement or recommendation for use.
Name of the reagent/Equipment | Company | Catalogue number | Comments/Description |
1-L polypropylene bottle | Nalgene | 2104-0032 | |
Aluminum foil squares | Cole-Parmer | 06275-40 | |
Autoclave | Steris | Amsco Lab Series | |
Bubble wrap | U.S. Plastics | 50776 | |
Closable bag | Uline | S-12283 | |
Closable bag | Fisher Scientific | S31798C | |
Commercial ice packs | Cole-Parmer | 06345-20 | |
Cool safe box | Diversified Biotech | CSF-BOX | |
Gauze sponge | Fisher Scientific | 22-415-469 | |
Graduated cylinder | Cole-Parmer | 06135-90 | 4-L or larger |
Hype Wipe | Fisher Scientific | 14-412-56 | |
iButtons temperature data logger | Maxim | DS1921G | |
Insulated storage and transport chest | Fisher Scientific | 11-676-12 | |
Packing tape | U.S. Plastics | 50083 | |
Portable chlorine colorimeter II test kit | Hach | 5870062 | |
Portable pH and temperature probe | Omega | PHH-830 | |
Portable turbidity meter | Omega | TRB-2020-E | |
PTFE thread tape | Cole-Parmer | 08270-34 | Use on all threaded connections |
Pump, Centrifugal Magnetic Drive | Cole-Parmer | 72010-20 | |
Reduction nipple | Cole-Parmer | 06349-87 | |
Sodium hypochlorite (NaClO) | Use locally available household bleach | ||
Sodium thiosulfate (Na2S2O3) | Sigma Aldrich | 217247 | |
Surgical gloves | Fisher Scientific | 19-058-800 | |
Waterproof marker | Fisher Scientific | 22-290546 | |
Media | Composition | ||
0.525% sodium hypochlorite (NaClO) | Prepare a 0.525% NaClO solution by diluting household bleach 1:10 in dH2O. Store 0.525% NaClO solutions for up to 1 week at room temperature. | ||
1 M sodium thiosulfate (Na2S2O3) pentahydrate | Prepare a 1 M solution by dissolving 248.2 g of Na2S2O3 in 1 L of dH2O. Store sodium thiosulfate for up to 6 months at room temperature. |