Breast cancer cells exhibit different dielectric properties compared to non-tumor breast epithelial cells. It has been hypothesized that, based on this difference in dielectric properties, the two populations can be separated for immunotherapy purposes. To support this, we model a microfluidic device to sort MCF-7 and MCF-10A cells.
Dielectrophoretic devices are capable of the detection and manipulation of cancer cells in a label-free, cost-effective, robust, and accurate manner using the principle of the polarization of the cancer cells in the sample volume by applying an external electric field. This article demonstrates how a microfluidic platform can be utilized for high-throughput continuous sorting of non-metastatic breast cancer cells (MCF-7) and non-tumor breast epithelial cells (MCF-10A) using hydrodynamic dielectrophoresis (HDEP) from the cell mixture. By generating an electric field between two electrodes placed side-by-side with a micron-sized gap between them in an HDEP microfluidic chip, non-tumor breast epithelial cells (MCF-10A) can be pushed away, exhibiting negative DEP inside the main channel, while the non-metastatic breast cancer cells follow their course unaffected when suspended in cell medium due to having conductivity higher than the membrane conductivity. To demonstrate this concept, simulations were performed for different values of medium conductivity, and the sorting of cells was studied. A parametric study was carried out, and a suitable cell mixture conductivity was found to be 0.4 S/m. By keeping the medium conductivity fixed, an adequate AC frequency of 0.8 MHz was established, giving maximum sorting efficiency, by varying the electric field frequency. Using the demonstrated method, after choosing the appropriate cell mixture suspension medium conductivity and frequency of the applied AC, maximum sorting efficiency can be achieved.
A malignant tumor that develops in and around the breast tissue is a frequent cause of breast cancer in women worldwide, causing a critical health problem1. Breast tumors before metastasis can be treated through surgery if detected at an early stage, but if ignored, they can have severe implications on the patient's life by spreading to their lungs, brain, and bones. The treatments offered at later stages, such as radiation and chemical-based therapies, have severe side effects2. Recent studies have reported that an early diagnosis of breast cancer reduces the mortality rate by 60%3. Hence, it is imperative to work toward personalized early detection methods. To this end, researchers working in different fields of science and technology have used microfluidics to develop devices for the early diagnosis of breast cancer4. These methods include cell affinity micro-chromatography, magnetic-activated micro-cell sorters, size-based cancer cell capture and separation, and on-chip dielectrophoresis (DEP)5,6. These microfluidic techniques reported in the literature enable precise cell manipulation, real-time monitoring, and sorting of well-defined samples, which serve as an intermediate step in many diagnostic and therapeutic applications5. The integration of these sorting mechanisms with microfluidics offers flexible and reliable manipulation of the target cells7,8,9,10. One of the main advantages of such an integration is the ability to work with fluid samples in nano to microliter volumes and also being able to manipulate the electrical properties of the sample fluid. By adjusting the conductivity of the suspending fluid inside microfluidic devices, the biological cells can be sorted based on their sizes and differences in their dielectric properties11,12.
Among these techniques, on-chip DEP is often preferred as it is a label-free cell sorting technique that exploits the electric properties of the biological samples. DEP has been reported to manipulate bio-samples such as DNA13, RNA14, proteins15, bacteria16, blood cells17, circulating tumor cells (CTCs)18, and stem cells19. Microfluidic devices that employ DEP for sorting biological samples have been reported extensively in literature20. Reservoir-based DEP microfluidic (rDEP) devices for sorting viable and non-viable yeast cells have been reported that protect the cells from the adverse effects of electrochemical reactions21,22. Piacentini et al. reported a castellated microfluidic cell sorter that separated red blood cells from platelets with an efficiency of 97%23. On-chip DEP devices with asymmetric orifices and embedded electrodes have also been reported to sort viable and non-viable cells24. Valero and Demierre et al. modified the castellated microfluidic cell sorter by introducing two arrays of microelectrodes on both sides of the channel25,26. This helped in focusing the cells in the center of the channel. Zeynep et al. presented a DEP-based microfluidic device to separate and concentrate MCF7 breast cancer cells from leukocytes27. They reported an efficiency of extracting MCF7 cells from leukocytes between 74%-98% with a frequency of 1 MHz and an applied voltage ranging from 10-12 Vpp. Supplementary Table 1 represents a qualitative and quantitative comparison between the DEP-based microfluidic sorting devices based on their design, electrode configuration, and operating parameters (applied frequency and voltage).
More recently, researchers have tried to measure the differences in the dielectric behavior of breast epithelial cells (MCF-10A) and non-metastatic breast cancer cells (MCF-7) inside a microfluidic chip28,29. Jithin et al. also characterized the dielectric responses of different cancer cell lines using an open-ended coaxial probe technique with frequencies between 200 MHz and 13.6 GHz30. These differences in the dielectric responses of MCF-7 and MCF-10A cell lines can be exploited to separate them in runtime and can lead to the development of personalized early-stage diagnosis devices.
In this article, we simulate the controlled sorting of non-metastatic breast cancer cells (MCF-7) and non-tumor breast epithelial cells (MCF-10A) using AC dielectrophoresis. The region of change in the electric field influences the sorting inside the microfluidic chip. The proposed technique is easy to implement and allows for the integration of the sorting technique into various microfluidic chip layouts. Computational fluid dynamics (CFD) simulations were carried out to study the separation of non-metastatic breast cancer cells and non-tumor breast epithelial cells by varying the conductivity of the fluid medium in which cells were suspended. In these simulations, it is shown that, by keeping the conductivity constant and by changing the applied frequency, the separation of cancer cells and healthy cells can be controlled.
NOTE: The protocol here uses COMSOL, a multiphysics simulation software, to simulate the controlled sorting of non-metastatic breast cancer cells (MCF-7) and non-tumor breast epithelial cells (MCF-10A) using AC dielectrophoresis.
1. Chip design and parameter selection
2. Mathematical model and computational analysis
Investigating the optimal operational parameters for effective DEP-based sorting of non-metastatic breast cancer (MCF-7) and non-tumor breast epithelial (MCF-10A) cells
To achieve a successful separation of non-metastatic breast cancer (MCF-7) and non-tumor breast epithelial (MCF-10A) cells with divergent dielectric properties when undergoing dielectrophoresis, their K factors should be distinct by keeping the applied frequency fixed37,38. The quantification of the dielectric responses of non-metastatic breast cancer cells and non-tumor breast epithelial cells under an applied electric field and the calculation of the "K" factor as a function of the applied frequency for both cell lines were achieved by using Equation 4. The results shown in Figure 1 were generated by keeping all the dielectric parameters of the non-metastatic breast cancer cells and non-tumor breast epithelial cells fixed while the applied frequency of the electric field was varied for three different values of conductivity of the cell suspension media, σm.
As shown in Figure 1, in each case, the value of K is within the range of -1 to 1, in line with previous studies39,40. Nonetheless, the plot of real(K) versus the frequency changes for both non-metastatic breast cancer cells (MCF-7) and non-tumor breast epithelial cells (MCF-10A) according to the value of the medium conductivity (σm). The results shown in Figure 1 are in agreement with a recent study in which the effect of σm on Re(K) for MCF-7 cells was quantified41.
Figure 1: Clausius-Mossotti factor. The real part of the Clausius-Mossotti factor, K, plotted as a function of frequency, for MCF-7 and MCF-10A cells suspended in a medium characterized by a conductivity of (A) σm = 0.01 S/m; (B) σm = 0.4 S/m; (C) σm = 1.2 S/m. Please click here to view a larger version of this figure.
Figure 1 was plotted using the MyDEP tool for three different σm values, keeping and varying the applied AC frequency from 100 kHz to 100MHz. Initially, the σm was chosen to be 0.01 S/m, and the applied AC frequency was varied between 100 kHz and 100 MHz, as shown in Figure 1A. At an applied AC frequency of 100 kHz, the value of Re(K) for MCF-10A cells turned out to be 0.82, which means they experience positive dielectrophoresis (pDEP) and should move toward an area of high electric field strength. Similarly, the MCF-7 cells at 100 kHz also experience pDEP with a Re(K) value of 0.76. The frequency was increased in steps of 100 kHz, and the value of the CM factor for both cell types remained on the positive side throughout the applied frequency spectrum. By keeping all the other operating parameters constant, the medium conductivity was increased to 0.4 S/m to plot the Re(K), as shown in Figure 1B. MCF-10A and MCF-7 demonstrated negative dielectrophoresis (nDEP) behavior with Re(k) values of -0.46 and -0.31, respectively, at 100 kHz. As the frequency was increased to 0.8 MHz, the DEP response of MCF-10 cells changed, and they experienced pDEP with an Re(K) value of 0.014. This behavior of MCF-7 cells is caused by the Maxwell-Wagner polarization at the interface between the cell membrane and the surrounding cell suspension medium39,41. The frequency where this change in DEP response is observed is known as the cross-over frequency, as shown in Figure 1A42,22. The MCF-7 cells, in this case, experienced nDEP. The frequency was further increased up to 100 MHz, but both cell types did not change their DEP behavior and, thus, remained unaffected by the variations in the applied electric field frequency. When the conductivity was increased to 1.2 S/m, the MCF-10A and MCF-7 cells experienced nDEP at 100 kHz. The Re(k) values for MCF-10A and MCF-7 cells, in this case, were -0.49 and -0.43, respectively, as shown in Figure 1C. As the frequency was increased to 0.8 MHz, the DEP response of the cells did not change, as they kept experiencing nDEP. The negative DEP behavior of both MCF-7 and MCF-10A cell lines at high values of cells suspension medium conductivity is in agreement with previously reported studies39,43,44. The DEP behavior of the cells at frequencies higher than the first cross-over frequency is governed by the interaction between cytoplasmic conductivity and the suspending solution45,46. On the other hand, at frequencies lower than the first cross-over frequency, the dielectric response of the cells is determined by the interaction between the cell membrane conductivity and the cell suspension medium.
Based on the results shown in Figure 1, COMSOL simulations were set up. Initially, the electric field strength was quantified using this simulation software, as shown in Figure 2. It can be seen that the maxima of the magnitude of the total electric field generated by two electrodes placed side-by-side on the top wall of the sorting channel are located near the electrode edges. The arrows show the direction of the electric field.
Figure 2: Electric field strength. The electric field generated by two electrodes placed side-by-side. Please click here to view a larger version of this figure.
The simulations to sort MCF-7 and MCF-10A cells were set up by keeping the applied AC frequency fixed at 0.8 MHz (cross-over frequency) and changing the value of σm. Three values for σm were chosen in accordance with the Re(K) plots shown in Figure 1. Initially, when σm was 0.01 S/m, both cell types experienced positive DEP, moved toward the region of high electric field strength, and moved out from the top outlet, as shown in Figure 3A. The cytoplasmic conductivities σcytoplasm for both cell lines were higher than the medium conductivity σm in this particular case, thus forcing both cell lines to move closer to the electrodes at the top of the microfluidic channel47. The DEP response of the MCF-10A cells changed, and they experienced negative DEP, when the σm was increased to 0.4 S/m with the applied frequency fixed at 0.8 MHz. Figure 3B shows that MCF-10A cells move to the top outlet, while MCF-7 cells move to the bottom outlet. The reason for this separation is that MCF-7 cells are more polarized compared to MCF-10A as their cytoplasmic conductivity σcytoplasm is greater than the medium conductivity σm, as shown in the cell sorting video (Video 2).
Figure 3: Cell sorting fixed frequency. Simulation of MCF-7 and healthy cell separation over time by DEP in the microfluidic device designed. MCF-7 and healthy cell separation at three different values of conductivity of the suspended medium: (A) 0.01 S/m; (B) 0.4 S/m; (C) 1.2 S/m. In each case, the applied frequency was 0.8 MHz, the applied voltage Vpp was 1.5 V, and the flow velocity at the injection inlet was 184 µm/s and 853 µm/s at the flow focusing inlet. In the simulation, MCF-7 cells and MCF-10A cells are represented with blue and red circles, respectively. Please click here to view a larger version of this figure.
As the medium conductivity was further increased to 1.2 S/m, both MCF-7 and MCF-10A cells both became less polarizable than the medium surrounding them due to lower cytoplasmic conductivity σcytoplasm values. Consequently, they experienced nDEP and moved away from regions of high electric field, as shown in Figure 3C.
These results demonstrate that the conductivity of the medium plays an important role in separating the MCF-7 non-metastatic breast cancer cells from MCF-10A non-tumor breast epithelial cells based on DEP. Moreover, as shown in Figure 3B, to achieve an effective separation, the medium conductivity needs to be adjusted in a way that the cells experience either pDEP or nDEP, based on their respective dielectric properties.
Finally, the effect of the applied dielectrophoretic force, FDEP, on the sorting behavior of both the cell lines was investigated by keeping the medium conductivity constant at 0.4 S/m. FDEP is a function of the frequency of the applied electric field48,49, and as the frequency of the applied electric field is changed, the cells change their DEP behavior. The simulations were started by setting the frequency at 100 kHz, and it was observed that both MCF-10A and MCF-7 cell lines experienced nDEP and moved away from the region of high electric field toward the bottom outlet, as shown in Figure 4A. As the frequency was increased, the DEP behavior of both cell lines remained unchanged until 0.8 MHz, when MCF-10A changed their DEP behavior and crossed-over to the pDEP region. This is the point with the maximum separation between the DEP response cells under investigation and maximum sorting efficiency, as shown in Figure 4B. When the frequency was increased to 100 MHz, it was observed that both cell lines experienced pDEP and moved toward the top outlet, as shown in Figure 4C. At higher frequencies above 0.8 MHz, the cells started to immobilize at the channel walls. The immobilization of cells at the channel walls can lead to cell loss during the sorting process, which, in turn, has an effect on the overall efficiency of the device. The effect of these forces can also cause serious loss in cell viability if exposed for a longer period of time. Yang et al. quantified the effect of DEP forces on a Listeria monocytogenes cell line by exposing them to an AC electric field of 5 MHz and a peak to peak voltage of 20 VPP50. Their results indicated a viable cell loss of 56%-89% when kept under the effect of DEP force for 4 h. Similarly, DEP forces have also been reported to have an effect on the movement of cells when suspended in a polarizable medium and have been used to immobilize cells. Ettehad et al. reported a microfluidic device with interdigitated electrodes (IDEs) that used an AC frequency of 1 MHz and 20 VPP to immobilize yeast cells51. They showed that immobilization of their yeast cells was dependent on the aspect ratio of the spacing between their IDEs and applied voltage. The increase in the aspect ratio of IDE spacing resulted in a sharp decrease in cell immobilization, and in order to immobilize cells in devices with greater spacing between IDEs, higher VPP was required. Cell immobilization is a desired application when cells are required to be trapped for analysis or growth. The previous results clearly showed that applied AC frequency and voltage have an effect on cell immobilization. In applications where high-throughput sorting or screening is the desired outcome, cell immobilization results in cell loss and reduces the output efficiency of the device.
In order to quantify the effect of applied frequency and voltage on cell immobilization, a set of simulations were run from the kilohertz to megahertz frequency range at a fixed applied voltage of 1.5 VPP. The results are shown in Supplementary Figure S4. The results revealed that, at frequencies in the kHz range, the immobilization of the cells at the channel walls was far less compared to frequencies in the MHz range. As DEP force is directly proportional to the applied AC frequency, we can deduce that, at high DEP force, the immobilization of cells is more pronounced. For this microfluidic device, there will be a cell loss during the sorting of MCF7 and MCF-10A cells as it is required to operate at frequencies greater than 0.8 MHz. The effect of the random distribution of cells at the inlet was further investigated by choosing a random distribution boundary condition. More cell-channel wall interactions were observed in this case, as shown in Supplementary Figure 5.
Figure 4: Cell sorting with fixed medium conductivity. Effect of the frequency of the applied AC field on the separation of non-metastatic breast cancer cells (MCF-7) and non-tumor breast epithelial cells (MCF-10A) in the microfluidic device simulated. (A) f= 100 KHz; (B) f= 0.8 MHz; (C) f= 100 MHz. The fluid medium conductivity was fixed at σm = 0.4 S/m. Please click here to view a larger version of this figure.
Dielectric properties for simulation | εcytoplasm | σcytoplasm (S/m) |
εmembrane | σmembrane (S/m) |
MCF-7 | 50 | 0.8 | 11 | 6× 10-6 |
MCF-10A | 100 | 0.1 | 11 | 6 |
f0 | 800 [kHz] | 1.2 x 103 Hz | Frequency of the electric field |
sigma_f | 0.4 [S/m] | 0.8 S/m | Fluid medium conductivity |
epsilon_f | 80 | 80 | Fluid relative permittivity |
rho_f | 1000 [kg/m3] | 1000 kg/m³ | Fluid density |
mu_f | 1 x 10-3 [Pa·s] | 0.001 Pa·s | Fluid dynamic viscosity |
rho_p | 1050 [kg/m3] | 1050 kg/m³ | Particle density |
dp1 | 17 [µm] | 1.7 x 10-5 m | Particle diameter |
dp2 | 16 [um] | 1.6 x 10-5 m | Particle diameter |
sigma_p1 | 0.8 [S/m] | 0.6 S/m | Particle conductivity |
sigma_p2 | 0.1 [S/m] | 1.1 S/m | Particle conductivity |
epsilon_p1 | 50 | 55 | Healthy relative permittivity |
epsilon_p2 | 100 | 65 | Cancer relative permittivity |
sigma_s1 | 6 x 10-6 [S/m] | 6 x 10-6 S/m | Shell electrical conductivity |
sigma_s2 | 6 [S/m] | 6 S/m | Shell electrical conductivity |
epsilon_s1 | 11 | 11 | Shell relative permittivity |
epsilon_s2 | 11 | 11 | Shell relative permittivity |
th_s1 | 7 [nm] | 7 x 10-9 m | Shell thickness |
th_s2 | 7 [nm] | 7 x 10-9 m | Shell thickness |
Table 1: Operating parameters. Dielectric properties of MCF-7 and MCF-10A
Video 1: A video showing the protocol steps. Please click here to download this Video.
Video 2: Cell sorting video. Please click here to download this Video.
Supplementary File 1: The governing equations and theoretical background. Please click here to download this File.
Supplementary Figure 1: Device design and parameters. Microfluidic device design highlighting different parts of the device. Please click here to download this File.
Supplementary Figure 2: Gap between electrodes. The gap between two patch electrodes. Please click here to download this File.
Supplementary Figure 3: Mesh independence study. A mesh independence study depicting the effect of different mesh sizes on the sorting of MCF-7 and MCF-10A cells. (A) The different mesh sizes for the microfluidic device, depicting the number of elements for each mesh. The number of elements that constitutes the mesh increase from coarser to finer mesh. (B) The sorting of MCF7 and MCF-10A cells on different mesh sizes by keeping all the other operating parameters constant. The fine and finer mesh sizes produce the best results for sorting. Please click here to download this File.
Supplementary Figure 4: Cell immobilization and random distribution test. Simulations performed for frequencies between 10 KHz and 6 MHz to validate the effect of DEP force on cell immobilization. (A) At f = 10 kHz, no sorting and cell immobilization are observed. (B) At f = 200 kHz, no sorting and cell immobilization are observed. (C) At f = 0.8 MHz, sorting and cell immobilization at the outlet walls are observed. Please click here to download this File.
Supplementary Figure 5: Random distribution. Randomly distributed particles at the inlet of the chip. Please click here to download this File.
Supplementary Table 1: Comparison of different DEP-based microfluidic sorting devices. Please click here to download this File.
Microfluidic devices have been reported previously for cell culture, trapping, and sorting47,52,53. The fabrication of these devices in the cleanroom is an expensive process, and it is imperative to quantify the output and efficiency of a proposed microfluidic device through CFD simulations. This study presents the design and simulations of an AC-dielectrophoretic microfluidic device for the continuous separation of non-metastatic breast cancer cells (MCF-7) and non-tumor breast epithelial cells (MCF-10A) based on their dielectric properties23.
The device is operated by applying an AC electric field via a set of two microelectrodes embedded into a single microfluidic sorting channel to separate MCF-7 and MCF-10A cells based on their dielectric properties. The separation efficacy of the device was computationally simulated for different values of medium conductivity and for a range of applied AC frequencies. The optimal applied AC frequency and media conductivity values were found to be 0.8 MHz and 0.4 S/m, respectively. A low voltage of 1.5 Vp-p was used throughout the simulations. The applied AC frequency range and the applied voltage are comparable to previously reported literature23,47. At frequencies above 1 MHz, the cell immobilization effect was observed, which should be taken into consideration for future device designs and fabrication. We list this cell immobilization as a limitation of our method in the context of cell sorting applications. We believe cell immobilization at higher frequencies can be used for cell differentiation as previously reported in literature54, giving this proposed design a new direction. This application would be of great interest to researchers in synthetic biology.
The critical steps for the correct implementation of this protocol include the choice of suitable physics nodes and sub-nodes (steps 1.5-1.9). These steps form the basis of the entire simulation protocol and help to choose the parameter values for each cell type, applied force, and applied voltage. Another critical step is to choose the correct fluid medium conductivity and applied frequency. This can be achieved by running a troubleshooting step of the parametric sweep. The parametric sweep of these two parameters can assist in deciding the optimal values for any future simulations. Lastly, a mesh independence study is also critical in the context of choosing the right mesh size for any future simulations. It is highly recommended that a mesh independence study is performed as a troubleshooting step before finalizing any future simulations.
This study provides the first simulation-based example of inline separation of non-metastatic breast cancer cells (MCF-7) and non-tumor breast epithelial cells (MCF-10A) based on their dielectric properties. We believe this design can be further implemented for viable and non-viable cell sorting.