Summary

Gravação simultânea de Eletrorretinografia e potenciais evocados visuais em ratos anestesiados

Published: July 01, 2016
doi:

Summary

This protocol describes simultaneous measurement of electroretinogram and visual evoked potentials in anesthetized rats.

Abstract

The electroretinogram (ERG) and visual evoked potential (VEP) are commonly used to assess the integrity of the visual pathway. The ERG measures the electrical responses of the retina to light stimulation, while the VEP measures the corresponding functional integrity of the visual pathways from the retina to the primary visual cortex following the same light event. The ERG waveform can be broken down into components that reflect responses from different retinal neuronal and glial cell classes. The early components of the VEP waveform represent the integrity of the optic nerve and higher cortical centers. These recordings can be conducted in isolation or together, depending on the application. The methodology described in this paper allows simultaneous assessment of retinal and cortical visual evoked electrophysiology from both eyes and both hemispheres. This is a useful way to more comprehensively assess retinal function and the upstream effects that changes in retinal function can have on visual evoked cortical function.

Introduction

Medição do eletrorretinograma (ERG) e visual potencial evocado (VEP) fornecem avaliações quantitativas úteis da integridade da via visual. O ERG mede as respostas eléctricas do retina para a estimulação pela luz, enquanto o VEP mede a integridade funcional correspondente das vias visuais da retina para o córtex visual primário seguindo o mesmo evento de luz. Este manuscrito descreve um protocolo para a gravação e análise das respostas do ERG e VEP em um modelo de laboratório comumente usado, o rato.

O ERG fornece um índice da integridade funcional de um número de classes principais de células da retina quantificando resposta elétrica bruta da retina para um flash de luz. Uma série coordenada de fluxos iônicos iniciado pelo aparecimento de luz e compensar, produzir mudanças detectáveis ​​na tensão que pode ser medido por meio de eletrodos de superfície colocados fora do olho. A forma de onda resultante representa a combinação de um SEries de componentes bem definidas, diferindo na amplitude, tempo e frequência. Um corpo substancial de investigação tem demonstrado que estes componentes são relativamente bem conservada entre muitos retina de vertebrados e que os componentes podem ser separados uns dos outros. Ao seleccionar criteriosamente as condições de estímulo (stimulus flash, fundo, intervalo interstimulus) e escolher características específicas da forma de onda composta para analisar se pode ter a certeza de voltar uma medida de um grupo específico de células da retina 1,2. Estas características são a base da utilidade e, portanto, as aplicações difundidas do ERG como uma medida não-invasiva da função retina. Este manuscrito incide sobre a metodologia para a medição do ERG e analisar suas características para retornar informações sobre alguns dos principais classes de células na retina, nomeadamente fotorreceptores (o componente PIII), células bipolares (o componente PII) e células ganglionares da retina (o positivo limiar de resposta escotópica ou pstr).

<p class= "Jove_content"> O VEP proporciona um ensaio da resposta cortical à luz; primeiro proveniente da retina e posteriormente comunicadas em série através do nervo óptico, trato óptico, tálamo (núcleo geniculado lateral, LGN) e radiação óptica para a área do córtex V1 3. Em roedores, a maioria (90-95%) de fibras do nervo óptico de cada olho decussate 4 e inervam o cérebro médio contralateral. Ao contrário do ERG, é como ainda não é possível atribuir diferentes componentes da VEP às classes de células específicas, 5 muda, assim, em qualquer lugar ao longo da via visual pode afetar a forma de onda VEP. No entanto, o VEP é uma medida não-invasivo útil do desempenho visual e integridade da via visual. A VEP, quando usado em conjunto com o ERG, pode proporcionar uma avaliação mais completa do sistema visual (ou seja, a retina / via visual).

registos de ERG e VEP pode ser realizada de forma isolada ou em combinação, dependendo da aplicação. A metodologia descrita neste artigo permite a avaliação simultânea de eletrofisiologia visual evocado retina e cortical de ambos os olhos e ambos os hemisférios em ratos anestesiados. Esta é uma maneira útil para avaliar de forma mais abrangente a função da retina e os efeitos a montante que as mudanças na função da retina pode ter em função cortical evocado visual.

Protocol

Todos os procedimentos experimentais foram realizados de acordo com o Código australiano de Prática para o Cuidado e Utilização de animais para fins científicos, estabelecido pela National Health and Medical Research Council na Austrália. apuramento de Ética foi obtido a partir da Universidade de Melbourne, Faculdade de Ciências, Comitê de Ética Animal (número de aprovação 0.911.322,1). 1. Pré-implantação de VEP crônica Eletrodos Nota: Se os sinais ERG e VEP simultâneos devem ser recolhid…

Representative Results

O ERG uma onda (> -1,38 log cd.sm -2), b-ondas (> – 4,99 log cd.sm -2) STRs (<- 4,99 log cd.sm -2) e os PEVs (> – 0,52 cd.sm log -2) foram gravadas simultaneamente (Figura 1 e 3). No flashes muito fraca, a STR positivo (pstr) é visto em cerca de 110 milissegundos após o flash, e uma STR negativo (NSTR) a cerca de 220 ​​ms (Figuras 1 e 2). Um ERG com um grande da onda b, picos entre 50 a 100 ms…

Discussion

O ERG e VEP são medidas objetivas da função visual da retina eo córtex, respectivamente. A vantagem de gravação simultânea é que uma visão mais abrangente de toda a via visual é oferecida. Especificamente, as informações complementares na sua avaliação simultânea poderia fornecer uma definição mais clara do local da lesão na via visual (por exemplo, para distúrbios com sobreposição ERG ainda distinta VEP manifestações 18, quando neuropatia óptica pode co-existir com atrofia ce…

Declarações

The authors have nothing to disclose.

Acknowledgements

Funding for this project was provided by the National Health and Medical Research Council (NHMRC) 1046203 (BVB, AJV) and Melbourne Neuroscience Institute Fellowship (CTN).

Materials

Alligator clip generic brand HM3022 Stainless steel 26 mm clip for connecting VEP screw electrodes to cables
Bioamplifier ADInstruments ML 135 For amplifying ERG and VEP signals
Carboxymethylcellulose sodium 1.0% Allergan CAS 0009000-11-7 Viscous fluid for improving signal quality of the active ERG electrode
Carprofen 0.5% Pfizer Animal Health Group CAS 53716-49-7 Proprietary name: Rimadyl injectable (50 mg/mL). For post-surgery analgesia, diluted to 0.5% (5 mg/mL) in normal saline
Chlorhexadine 0.5% Orion Laboratories 27411, 80085 For disinfecting surgical instruments
Circulating water bath Lauda-Königshoffen MGW Lauda For maintaining body temperature of the anesthetized animal during surgery and electrophysiological recordings
Dental amalgam DeguDent GmbH 64020024 For encasing the electrode-skull assembly to make it more robust
Dental burr Storz Instruments, Bausch and Lomb #E0824A A miniature drill head of ~0.7mm diameter for making a small hole in the skull over each hemisphere to implant VEP screws
Drill Bosch Dremel 300 series An automatic drill for trepanning
Electrode lead Grass Telefactor  F-E2-30 Platinum cables for connecting silver wire electrodes to the amplifier
Faraday Cage custom-made Ensures light proof to maintain dark adaptation. Encloses the Ganzfeld setup to improve signal to noise ratio
Gauze swabs Multigate Medical Products Pty Ltd 57-100B For drying the surgical incision and exposed skull surface during surgery
Ganzfeld integrating sphere Photometric Solutions International Custom designed light stimulator: 36 mm diameter, 13 cm aperture size
Velcro VELCRO Australia Pty Ltd VELCRO Brand Reusable Wrap Hook-and-loop fastener to secure the electrodes and the animal on the recording platform
Isoflurane 99.9% Abbott Australasia Pty Ltd CAS 26675-46-7 Proprietary Name: Isoflo(TM) Inhalation anaaesthetic. Pharmaceutical-grade inhalation anesthetic mixed with oxygen gas for VEP electrode implant surgery
Ketamine  Troy Laboratories Ilium Ketamil Proprietary name: Ketamil Injection, Brand: Ilium. Pharmaceutical-grade anesthetic for electrophysiological recording
Luxeon LEDs Phillips Lighting Co. For light stimulation twenty 5 watt and one 1 watt LEDs.
Micromanipulator Harvard Apparatus BS4 50-2625 Holds the ERG active electrode during recordings
Needle electrode Grass Telefactor  F-E2-30 Subcutaneously inserted in the tail to serve as the ground electrode for both the ERG and VEP
Phenylephrine 2.5% minims  Bausch and Lomb CAS 61-76-7 Instilled with Tropicamide to achieve maximal dilation for ERG recording
Povidone iodine 10% Sanofi-Aventis CAS 25655-41-8 Proprietory name: Betadine, Antiseptic to prepare the shaved skin for surgery 10%, 500 mL
Powerlab data acquisition system ADInstruments ML 785 Controls the LEDs
Proxymetacaine 0.5% Alcon Laboratories  CAS 5875-06-9 For corneal anaesthesia during ERG recordings
Saline solution Gelflex Non-injectable, for electroplating silver wire electrodes
Scope Software ADInstruments version 3.7.6 Simultaneously triggers the stimulus via the Powerlab system and collects data
Silver (fine round wire) A&E metal 0.3 mm Used to make active and inactive ERG electrodes, and the inactive VEP electrode
Stainless streel screws  MicroFasterners 0.7 mm shaft diameter, 3 mm in length to be implanted over the primary visual cortex and serve as the active VEP electrodes
Stereotaxic frame David Kopf Model 900 A small animal stereotaxic instrument for locating the primary visual cortices according to Paxinos & Watson's 2007 rat brain atlas coordinates
Surgical blade Swann-Morton Ltd. 0206 For incising the area of skin overlaying the primary visual cortex to implant the VEP electrodes
Suture Shanghai Pudong Jinhuan Medical Products Co.,Ltd 3-0 silk braided suture non-absorbable, for skin retraction during VEP electrode implantation surgery
Tobramycine eye ointment 0.3% Alcon Laboratories  CAS 32986-56-4 Proprietary name: Tobrex. Prophylactic antibiotic ointment applied around the skin wound after surgery
Tropicamide 0.5% Alcon Laboratories  CAS 1508-75-4 Proprietary name: 0.5% Mydriacyl eye drop, Instilled to achieve mydriasis for ERG recording
Xylazine Troy Laboratories Ilium Xylazil-100 Pharmaceutical-grade anesthetic for electrophysiological recording
Pipette tip  Eppendorf Pty Ltd 0030 073.169 Eppendorf epTIPS 100 – 5000 mL, for custom-made electrodes
Microsoft Office Excel Microsoft version 2010 spreadsheet software for data analysis
Lethabarb Euthanazia Injection Virbac (Australia) Pty Ltd LETHA450 325 mg/mL pentobarbital sodium for rapid euthanazia

Referências

  1. Nguyen, C. T. O., Vingrys, A. J., Bui, B. V. Dietary omega-3 fatty acids and ganglion cell function. Invest Ophthalmol Vis Sci. 49, 3586-3594 (2008).
  2. Weymouth, A. E., Vingrys, A. J. Rodent electroretinography: methods for extraction and interpretation of rod and cone responses. Prog Retin Eye Res. 27, 1-44 (2008).
  3. Tsai, T. I., Bui, B. V., Vingrys, A. J. Effect of acute intraocular pressure challenge on rat retinal and cortical function. Invest Ophthalmol Vis Sci. 55, 1067-1077 (2014).
  4. Cowey, A., Franzini, C. The retinal origin of uncrossed optic nerve fibres in rats and their role in visual discrimination. Exp Brain Res. 35, 443-455 (1979).
  5. Weinstein, G. W., Odom, J. V., Cavender, S. Visually evoked potentials and electroretinography in neurologic evaluation. Neurol Clin. 9, 225-242 (1991).
  6. Odom, J. V., et al. Visual evoked potentials standard (2004). Doc Ophthalmol. 108, 115-123 (2004).
  7. Ridder, W. H., Nusinowitz, S., Heckenlively, J. R. Causes of cataract development in anesthetized mice. Exp Eye Res. 75, 365-370 (2002).
  8. Nixon, P. J., Bui, B. V., Armitage, J. A., Vingrys, A. J. The contribution of cone responses to rat electroretinograms. Clin Experiment Ophthalmol. 29, 193-196 (2001).
  9. Bui, B. V., et al. Using the electroretinogram to understand how intraocular pressure elevation affects the rat retina. J Ophthalmol. 2013, 262467 (2013).
  10. Nguyen, C. T., Vingrys, A. J., Bui, B. V. Dietary omega-3 fatty acids and ganglion cell function. Invest Ophthalmol Vis Sci. 49, 3586-3594 (2008).
  11. Hood, D. C., Birch, D. G. A quantitative measure of the electrical activity of human rod photoreceptors using electroretinography. Vis Neurosci. 5, 379-387 (1990).
  12. Birch, D. G., Hood, D. C., Locke, K. G., Hoffman, D. R., Tzekov, R. T. Quantitative electroretinogram measures of phototransduction in cone and rod photoreceptors – Normal aging, progression with disease, and test-retest variability. Arch Ophthalmol. 120, 1045-1051 (2002).
  13. Bui, B. V., Vingrys, A. J. Development of receptoral responses in pigmented and albino guinea-pigs (Cavia porcellus). Doc Ophthalmol. 99, 151-170 (1999).
  14. Robson, J. G., Saszik, S. M., Ahmed, J., Frishman, L. J. Rod and cone contributions to the a-wave of the electroretinogram of the macaque. J Physiol. 547, 509-530 (2003).
  15. Severns, M. L., Johnson, M. A. The care and fitting of Naka-Rushton functions to electroretinographic intensity-response data. Doc Ophthalmol. 85, 135-150 (1993).
  16. Bui, B. V., Fortune, B. Origin of electroretinogram amplitude growth during light adaptation in pigmented rats. Vis Neurosci. 23, 155-167 (2006).
  17. Bui, B. V., Fortune, B. Ganglion cell contributions to the rat full-field electroretinogram. J Physiol. 555, 153-173 (2004).
  18. Tremblay, F., Laroche, R. G., Debecker, I. The Electroretinographic Diagnosis of the Incomplete Form of Congenital Stationary Night Blindness. Vision Res. 35, 2383-2393 (1995).
  19. Bayer, A. U., Keller, O. N., Ferrari, F., Maag, K. P. Association of glaucoma with neurodegenerative diseases with apoptotic cell death: Alzheimer’s disease and Parkinson’s disease. Am J Ophthalmol. 133, 135-137 (2002).
  20. Wostyn, P., Audenaert, K., De Deyn, P. P. An abnormal high trans-lamina cribrosa pressure difference: A missing link between Alzheimer’s disease and glaucoma. Clinical Neurology and Neurosurgery. 110, 753-754 (2008).
  21. Yucel, Y. H., Zhang, Q. A., Weinreb, R. N., Kaufman, P. L., Gupta, N. Effects of retinal ganglion cell loss on magno-, parvo-, koniocellular pathways in the lateral geniculate nucleus and visual cortex in glaucoma. Prog Retin Eye Res. 22, 465-481 (2003).
  22. Gupta, N., Yucel, Y. H. What changes can we expect in the brain of glaucoma patients. Survey of Ophthalmology. 52, 122-126 (2007).
  23. Kong, Y. X., et al. Impact of aging and diet restriction on retinal function during and after acute intraocular pressure injury. Neurobiol Aging. 33, 1115-1125 (2012).
  24. Bui, B. V., Sinclair, A. J., Vingrys, A. J. Electroretinograms of albino and pigmented guinea-pigs (Cavia porcellus). Aust N Z J Ophthalmol. 26, 98-100 (1998).
  25. Jobling, A. I., Wan, R., Gentle, A., Bui, B. V., McBrien, N. A. Retinal and choroidal TGF-beta in the tree shrew model of myopia: isoform expression, activation and effects on function. Exp Eye Res. 88, 458-466 (2009).
  26. Robson, J. G., Frishman, L. J. Dissecting the dark-adapted electroretinogram. Doc Ophthalmol. 95, 187-215 (1998).
  27. Robson, J. G., Frishman, L. J. The rod-driven a-wave of the dark-adapted mammalian electroretinogram. Prog Retin Eye Res. 39, 1-22 (2014).
  28. Hudnell, H. K., Boyes, W. K. The comparability of rat and human visual-evoked potentials. Neurosci Biobehav Rev. 15, 159-164 (1991).
  29. Charng, J., et al. Conscious wireless electroretinogram and visual evoked potentials in rats. PLoS One. 8, e74172 (2013).
  30. Hetzler, B. E., Berger, L. K. Ketamine-Induced Modification of Photic Evoked-Potentials in the Superior Colliculus of Hooded Rats. Neuropharmacology. 23, 473-476 (1984).
check_url/pt/54158?article_type=t

Play Video

Citar este artigo
Nguyen, C. T., Tsai, T. I., He, Z., Vingrys, A. J., Lee, P. Y., Bui, B. V. Simultaneous Recording of Electroretinography and Visual Evoked Potentials in Anesthetized Rats. J. Vis. Exp. (113), e54158, doi:10.3791/54158 (2016).

View Video