Summary

Enregistrement simultané de Électrorétinographie et potentiels évoqués visuels chez des rats anesthésiés

Published: July 01, 2016
doi:

Summary

This protocol describes simultaneous measurement of electroretinogram and visual evoked potentials in anesthetized rats.

Abstract

The electroretinogram (ERG) and visual evoked potential (VEP) are commonly used to assess the integrity of the visual pathway. The ERG measures the electrical responses of the retina to light stimulation, while the VEP measures the corresponding functional integrity of the visual pathways from the retina to the primary visual cortex following the same light event. The ERG waveform can be broken down into components that reflect responses from different retinal neuronal and glial cell classes. The early components of the VEP waveform represent the integrity of the optic nerve and higher cortical centers. These recordings can be conducted in isolation or together, depending on the application. The methodology described in this paper allows simultaneous assessment of retinal and cortical visual evoked electrophysiology from both eyes and both hemispheres. This is a useful way to more comprehensively assess retinal function and the upstream effects that changes in retinal function can have on visual evoked cortical function.

Introduction

Mesure de l'électrorétinogramme (ERG) et potentiel évoqué visuel (VEP) fournir des évaluations quantitatives utiles de l'intégrité de la voie visuelle. L'ERG mesure les réponses électriques de la rétine à une stimulation lumineuse, tandis que la VEP mesure l'intégrité fonctionnelle correspondante des voies visuelles de la rétine au cortex visuel primaire suivant le même événement lumière. Ce manuscrit décrit un protocole pour l'enregistrement et l'analyse des réponses ERG et VEP dans un modèle de laboratoire couramment utilisé, le rat.

L'ERG fournit un indice de l'intégrité fonctionnelle d'un certain nombre de classes clés de cellules rétiniennes en quantifiant réponse électrique brute de la rétine à un flash de lumière. Une série coordonnée des flux ioniques initiée par l'apparition de lumière et de décalage, produire des changements détectables dans la tension qui peut être mesurée en utilisant des électrodes de surface placées à l'extérieur de l'œil. La forme d'onde résultante représente la combinaison d'un soiries de composants bien définis, différents en amplitude, le calendrier et la fréquence. Un important corpus de recherche a montré que ces composants sont relativement bien conservés dans de nombreux rétines vertébrés et que les composants peuvent être séparés les uns des autres. En choisissant judicieusement le stimulus (flash stimulus, fond, intervalle interstimulus) conditions et en choisissant des caractéristiques spécifiques de la forme d' onde composite à analyser , on peut être sûr de retourner une mesure d'un groupe spécifique de cellules rétiniennes 1,2. Ces caractéristiques sous-tendent l'utilité et donc les applications répandues de l'ERG en tant que mesure non-invasive de la fonction de la rétine. Ce manuscrit se concentre sur la méthodologie de mesure de l'ERG et l'analyse de ses caractéristiques pour renvoyer des informations sur certaines des principales classes de cellules de la rétine, à savoir photorécepteurs (la composante PIII), les cellules bipolaires (la composante PII) et les cellules ganglionnaires de la rétine (le positif seuil de réponse scotopique ou pstr).

<p class= "Jove_content"> Le VEP fournit une analyse de la réponse corticale à la lumière; la première provenant de la rétine et par la suite transmis en série par l' intermédiaire du nerf optique, l' appareil optique, le thalamus (noyau genouillé latéral, LGN) et le rayonnement optique à la zone V1 du cortex 3. Chez les rongeurs, la majorité (90 – 95%) des fibres du nerf optique de chaque decussate oculaire 4 et innervent le mésencéphale controlatéral. Contrairement à l'ERG, il est encore impossible d'attribuer les différentes composantes de la VEP à des classes de cellules spécifiques, 5 change donc partout le long de la voie visuelle pourrait affecter la forme d' onde VEP. Néanmoins, le VEP est une mesure non-invasive utile de la performance visuelle et de l'intégrité de la voie visuelle. Le VEP, lorsqu'il est utilisé en conjonction avec l'ERG, peut fournir une évaluation plus complète du système visuel (ie, la rétine / voie visuelle).

ERG et VEP enregistrements peuvent être effectuées de manière isolée ou en combinaison, en fonction de l'application. La méthodologie décrite dans ce document permet une évaluation simultanée de la rétine et du cortex électrophysiologie visuelle évoqué des deux yeux et les deux hémisphères chez les rats anesthésiés. Ceci est un moyen utile d'évaluer de façon plus complète la fonction rétinienne et les effets en amont que les changements dans la fonction rétinienne peuvent avoir sur la fonction corticale évoquée visuelle.

Protocol

Toutes les procédures expérimentales ont été réalisées selon le Code australien de pratiques pour le soin et l'utilisation des animaux à des fins scientifiques, établi par le Conseil de recherches médicales Santé nationale et en Australie. clairance éthique a été obtenue à partir de l'Université de Melbourne, Faculté des Sciences, Comité d'éthique des animaux (numéro d'homologation 0911322,1). 1. Pré-implantation chronique VEP Electrodes Remarque: Si des signaux ERG et V…

Representative Results

L'ERG a-wave (> -1.38 log cd.sm -2), b-ondes (> – 4,99 log cd.sm -2) DOD (<- 4,99 log cd.sm -2) et les VEPs (> – 0,52 log cd.sm -2) ont été enregistrés simultanément (Figure 1 et 3). A clignote très faibles, un STR positif (pstr) est observée à environ 110 ms après le flash, et ​​un STR négatif (NSTR) , à environ 220 msec (figures 1 et 2). Un ERG avec un grand b-ondes, des pics ent…

Discussion

L'ERG et VEP sont des mesures objectives de la fonction visuelle de la rétine et du cortex, respectivement. L'avantage de l'enregistrement simultané est qu'une vision plus globale de l'ensemble de la voie visuelle est assurée. Plus précisément, les informations complémentaires de leur évaluation simultanée pourrait fournir une délimitation plus claire du site de la lésion dans la voie visuelle (par exemple, pour les troubles qui se chevauchent encore ERG distincte VEP manifestation…

Declarações

The authors have nothing to disclose.

Acknowledgements

Funding for this project was provided by the National Health and Medical Research Council (NHMRC) 1046203 (BVB, AJV) and Melbourne Neuroscience Institute Fellowship (CTN).

Materials

Alligator clip generic brand HM3022 Stainless steel 26 mm clip for connecting VEP screw electrodes to cables
Bioamplifier ADInstruments ML 135 For amplifying ERG and VEP signals
Carboxymethylcellulose sodium 1.0% Allergan CAS 0009000-11-7 Viscous fluid for improving signal quality of the active ERG electrode
Carprofen 0.5% Pfizer Animal Health Group CAS 53716-49-7 Proprietary name: Rimadyl injectable (50 mg/mL). For post-surgery analgesia, diluted to 0.5% (5 mg/mL) in normal saline
Chlorhexadine 0.5% Orion Laboratories 27411, 80085 For disinfecting surgical instruments
Circulating water bath Lauda-Königshoffen MGW Lauda For maintaining body temperature of the anesthetized animal during surgery and electrophysiological recordings
Dental amalgam DeguDent GmbH 64020024 For encasing the electrode-skull assembly to make it more robust
Dental burr Storz Instruments, Bausch and Lomb #E0824A A miniature drill head of ~0.7mm diameter for making a small hole in the skull over each hemisphere to implant VEP screws
Drill Bosch Dremel 300 series An automatic drill for trepanning
Electrode lead Grass Telefactor  F-E2-30 Platinum cables for connecting silver wire electrodes to the amplifier
Faraday Cage custom-made Ensures light proof to maintain dark adaptation. Encloses the Ganzfeld setup to improve signal to noise ratio
Gauze swabs Multigate Medical Products Pty Ltd 57-100B For drying the surgical incision and exposed skull surface during surgery
Ganzfeld integrating sphere Photometric Solutions International Custom designed light stimulator: 36 mm diameter, 13 cm aperture size
Velcro VELCRO Australia Pty Ltd VELCRO Brand Reusable Wrap Hook-and-loop fastener to secure the electrodes and the animal on the recording platform
Isoflurane 99.9% Abbott Australasia Pty Ltd CAS 26675-46-7 Proprietary Name: Isoflo(TM) Inhalation anaaesthetic. Pharmaceutical-grade inhalation anesthetic mixed with oxygen gas for VEP electrode implant surgery
Ketamine  Troy Laboratories Ilium Ketamil Proprietary name: Ketamil Injection, Brand: Ilium. Pharmaceutical-grade anesthetic for electrophysiological recording
Luxeon LEDs Phillips Lighting Co. For light stimulation twenty 5 watt and one 1 watt LEDs.
Micromanipulator Harvard Apparatus BS4 50-2625 Holds the ERG active electrode during recordings
Needle electrode Grass Telefactor  F-E2-30 Subcutaneously inserted in the tail to serve as the ground electrode for both the ERG and VEP
Phenylephrine 2.5% minims  Bausch and Lomb CAS 61-76-7 Instilled with Tropicamide to achieve maximal dilation for ERG recording
Povidone iodine 10% Sanofi-Aventis CAS 25655-41-8 Proprietory name: Betadine, Antiseptic to prepare the shaved skin for surgery 10%, 500 mL
Powerlab data acquisition system ADInstruments ML 785 Controls the LEDs
Proxymetacaine 0.5% Alcon Laboratories  CAS 5875-06-9 For corneal anaesthesia during ERG recordings
Saline solution Gelflex Non-injectable, for electroplating silver wire electrodes
Scope Software ADInstruments version 3.7.6 Simultaneously triggers the stimulus via the Powerlab system and collects data
Silver (fine round wire) A&E metal 0.3 mm Used to make active and inactive ERG electrodes, and the inactive VEP electrode
Stainless streel screws  MicroFasterners 0.7 mm shaft diameter, 3 mm in length to be implanted over the primary visual cortex and serve as the active VEP electrodes
Stereotaxic frame David Kopf Model 900 A small animal stereotaxic instrument for locating the primary visual cortices according to Paxinos & Watson's 2007 rat brain atlas coordinates
Surgical blade Swann-Morton Ltd. 0206 For incising the area of skin overlaying the primary visual cortex to implant the VEP electrodes
Suture Shanghai Pudong Jinhuan Medical Products Co.,Ltd 3-0 silk braided suture non-absorbable, for skin retraction during VEP electrode implantation surgery
Tobramycine eye ointment 0.3% Alcon Laboratories  CAS 32986-56-4 Proprietary name: Tobrex. Prophylactic antibiotic ointment applied around the skin wound after surgery
Tropicamide 0.5% Alcon Laboratories  CAS 1508-75-4 Proprietary name: 0.5% Mydriacyl eye drop, Instilled to achieve mydriasis for ERG recording
Xylazine Troy Laboratories Ilium Xylazil-100 Pharmaceutical-grade anesthetic for electrophysiological recording
Pipette tip  Eppendorf Pty Ltd 0030 073.169 Eppendorf epTIPS 100 – 5000 mL, for custom-made electrodes
Microsoft Office Excel Microsoft version 2010 spreadsheet software for data analysis
Lethabarb Euthanazia Injection Virbac (Australia) Pty Ltd LETHA450 325 mg/mL pentobarbital sodium for rapid euthanazia

Referências

  1. Nguyen, C. T. O., Vingrys, A. J., Bui, B. V. Dietary omega-3 fatty acids and ganglion cell function. Invest Ophthalmol Vis Sci. 49, 3586-3594 (2008).
  2. Weymouth, A. E., Vingrys, A. J. Rodent electroretinography: methods for extraction and interpretation of rod and cone responses. Prog Retin Eye Res. 27, 1-44 (2008).
  3. Tsai, T. I., Bui, B. V., Vingrys, A. J. Effect of acute intraocular pressure challenge on rat retinal and cortical function. Invest Ophthalmol Vis Sci. 55, 1067-1077 (2014).
  4. Cowey, A., Franzini, C. The retinal origin of uncrossed optic nerve fibres in rats and their role in visual discrimination. Exp Brain Res. 35, 443-455 (1979).
  5. Weinstein, G. W., Odom, J. V., Cavender, S. Visually evoked potentials and electroretinography in neurologic evaluation. Neurol Clin. 9, 225-242 (1991).
  6. Odom, J. V., et al. Visual evoked potentials standard (2004). Doc Ophthalmol. 108, 115-123 (2004).
  7. Ridder, W. H., Nusinowitz, S., Heckenlively, J. R. Causes of cataract development in anesthetized mice. Exp Eye Res. 75, 365-370 (2002).
  8. Nixon, P. J., Bui, B. V., Armitage, J. A., Vingrys, A. J. The contribution of cone responses to rat electroretinograms. Clin Experiment Ophthalmol. 29, 193-196 (2001).
  9. Bui, B. V., et al. Using the electroretinogram to understand how intraocular pressure elevation affects the rat retina. J Ophthalmol. 2013, 262467 (2013).
  10. Nguyen, C. T., Vingrys, A. J., Bui, B. V. Dietary omega-3 fatty acids and ganglion cell function. Invest Ophthalmol Vis Sci. 49, 3586-3594 (2008).
  11. Hood, D. C., Birch, D. G. A quantitative measure of the electrical activity of human rod photoreceptors using electroretinography. Vis Neurosci. 5, 379-387 (1990).
  12. Birch, D. G., Hood, D. C., Locke, K. G., Hoffman, D. R., Tzekov, R. T. Quantitative electroretinogram measures of phototransduction in cone and rod photoreceptors – Normal aging, progression with disease, and test-retest variability. Arch Ophthalmol. 120, 1045-1051 (2002).
  13. Bui, B. V., Vingrys, A. J. Development of receptoral responses in pigmented and albino guinea-pigs (Cavia porcellus). Doc Ophthalmol. 99, 151-170 (1999).
  14. Robson, J. G., Saszik, S. M., Ahmed, J., Frishman, L. J. Rod and cone contributions to the a-wave of the electroretinogram of the macaque. J Physiol. 547, 509-530 (2003).
  15. Severns, M. L., Johnson, M. A. The care and fitting of Naka-Rushton functions to electroretinographic intensity-response data. Doc Ophthalmol. 85, 135-150 (1993).
  16. Bui, B. V., Fortune, B. Origin of electroretinogram amplitude growth during light adaptation in pigmented rats. Vis Neurosci. 23, 155-167 (2006).
  17. Bui, B. V., Fortune, B. Ganglion cell contributions to the rat full-field electroretinogram. J Physiol. 555, 153-173 (2004).
  18. Tremblay, F., Laroche, R. G., Debecker, I. The Electroretinographic Diagnosis of the Incomplete Form of Congenital Stationary Night Blindness. Vision Res. 35, 2383-2393 (1995).
  19. Bayer, A. U., Keller, O. N., Ferrari, F., Maag, K. P. Association of glaucoma with neurodegenerative diseases with apoptotic cell death: Alzheimer’s disease and Parkinson’s disease. Am J Ophthalmol. 133, 135-137 (2002).
  20. Wostyn, P., Audenaert, K., De Deyn, P. P. An abnormal high trans-lamina cribrosa pressure difference: A missing link between Alzheimer’s disease and glaucoma. Clinical Neurology and Neurosurgery. 110, 753-754 (2008).
  21. Yucel, Y. H., Zhang, Q. A., Weinreb, R. N., Kaufman, P. L., Gupta, N. Effects of retinal ganglion cell loss on magno-, parvo-, koniocellular pathways in the lateral geniculate nucleus and visual cortex in glaucoma. Prog Retin Eye Res. 22, 465-481 (2003).
  22. Gupta, N., Yucel, Y. H. What changes can we expect in the brain of glaucoma patients. Survey of Ophthalmology. 52, 122-126 (2007).
  23. Kong, Y. X., et al. Impact of aging and diet restriction on retinal function during and after acute intraocular pressure injury. Neurobiol Aging. 33, 1115-1125 (2012).
  24. Bui, B. V., Sinclair, A. J., Vingrys, A. J. Electroretinograms of albino and pigmented guinea-pigs (Cavia porcellus). Aust N Z J Ophthalmol. 26, 98-100 (1998).
  25. Jobling, A. I., Wan, R., Gentle, A., Bui, B. V., McBrien, N. A. Retinal and choroidal TGF-beta in the tree shrew model of myopia: isoform expression, activation and effects on function. Exp Eye Res. 88, 458-466 (2009).
  26. Robson, J. G., Frishman, L. J. Dissecting the dark-adapted electroretinogram. Doc Ophthalmol. 95, 187-215 (1998).
  27. Robson, J. G., Frishman, L. J. The rod-driven a-wave of the dark-adapted mammalian electroretinogram. Prog Retin Eye Res. 39, 1-22 (2014).
  28. Hudnell, H. K., Boyes, W. K. The comparability of rat and human visual-evoked potentials. Neurosci Biobehav Rev. 15, 159-164 (1991).
  29. Charng, J., et al. Conscious wireless electroretinogram and visual evoked potentials in rats. PLoS One. 8, e74172 (2013).
  30. Hetzler, B. E., Berger, L. K. Ketamine-Induced Modification of Photic Evoked-Potentials in the Superior Colliculus of Hooded Rats. Neuropharmacology. 23, 473-476 (1984).