This protocol describes how resin embedded brain tissue can be prepared and imaged in the three dimensions in the focussed ion beam, scanning electron microscope.
This protocol describes how biological samples, like brain tissue, can be imaged in three dimensions using the focussed ion beam/scanning electron microscope (FIB/SEM). The samples are fixed with aldehydes, heavy metal stained using osmium tetroxide and uranyl acetate. They are then dehydrated with alcohol and infiltrated with resin, which is then hardened. Using a light microscope and ultramicrotome with glass knives, a small block containing the region interest close to the surface is made. The block is then placed inside the FIB/SEM, and the ion beam used to roughly mill a vertical face along one side of the block, close to this region. Using backscattered electrons to image the underlying structures, a smaller face is then milled with a finer ion beam and the surface scrutinised more closely to determine the exact area of the face to be imaged and milled. The parameters of the microscope are then set so that the face is repeatedly milled and imaged so that serial images are collected through a volume of the block. The image stack will typically contain isotropic voxels with dimenions as small a 4 nm in each direction. This image quality in any imaging plane enables the user to analyse cell ultrastructure at any viewing angle within the image stack.
1. Sample fixation and resin embedding
2. Preparing the sample for the FIB/SEM
3. Imaging in the FIB/SEM
4. Secrets to success
Step 1) Ensure the samples are not thicker than 150 microns. This will enable adequate penetration of the different stains and resin into the material. This can be achieved by sectioning the samples with a vibratome immediately after the fixation. This thickness is appropriate for brain tissue. Other tissues need to be tested to ensure adequate fixation and staining.
Step 3) The samples must be freely distributed throughout the solution, and not clumped together in one part of the vial while this secondary fixative is introduced. This will ensure maximum penetration of the sample and reduce the sample becoming distorted during this fixation process. This is achieved by gently swirling the samples within the vial immediately after the solution is added.
Step 15) Ensure that the region of interest is located immediately below (< 30 microns) the surface of the block. This will reduce the amount of milling by the ion beam during the image acquisition phase.
Step 22) The scanning time of the ion beam needs to be adequte to ensure that the entire imaging face is completely milled. Inadequate milling will be shown as white streaks, or ‘curtains’ appearing vertically down the imaging face.
Step 26) The final milled face must be larger than the face that is imaged. This is because the milled resin is ejected from the block and redeposits on the block surface in the immediate vicinity. This redeposition can encroach onto the imaging field of view, and is avoided by milling a much large area than is imaged. For an imaging window that is 20 microns wide, the milled block face is greater than 30 microns wide.
Step 29) It is critical that the electron beam parameters are such that the image is generated by electrons that emerge from only the first few nanometres of the block’s surface. This a acheved by minimising the voltage to below 2 kV and ensuring that no electrons penetrate too far. This is also helped by using a grid voltage on the detector so that only the electron with the highest energies contribute to the final image. Typically, a grid tension of 1.3 kV is used for an imaging voltage of 1.5 kV.
5. Representative Results:
Figure 1. Preparation of resin block. a) Resin embedded coronal section (80 microns thick) through an adult rat brain viewed in a steromicroscope. The image clearly shows the different brain regions that can be cut from the section (b) using a scalpel blade. Here, a 1 mm square from the cortex is removed and (c) stuck to a blank resin block. d) The resin block is then trimmed with a glass knife mounted in an ultramicrotome, and once the block has been reduced to only leave the region of interest (e), a step is cut perpendicular to the front surface of the embedded material. f) This entire trimmed region is then cut from the rest of the resin and mounted onto a aluminium stub ready for metal coating and imaging in the scanning electron microscope.
Figure 2. Preparing the block face for FIB/SEM imaging. a) and b) show illustrations of the block and the edge that is milled to create the imaging face (indicated with the black double-headed arrow). The position of the ion beam (white) is shown parallel to the imaging face, and the electron beam (grey) is shown hitting the face at 54° to the ion beam. c) shows a view of the block taken with the electron beam and imaged with the secondary electrons. Along this side of the block a darker region (indicated with dotted white box) shows a portion of the face that has been roughly milled (shown with double-headed black arrow) to reveal the underlying sample. d) When this region is imaged using only the backscattered electrons the embedded tissue can be seen. Here, the full width of the embedded tissue section is indicated with a double headed white arrow. Scale bar in c) is 100 microns.
Figure 3. Imaging the a volume of brain tissue with isotropic voxels. a) Reversed contrast backscattered image of the block face showing the ultrastructure within a region of the rat brain. All membranes are visible as well as large macromolecular structures. b) A total of 1600 images were collected at 5 nm spacing resulting in an image stack with isotropic voxels. c) This stack can be imaged in any plane with the same resolution. This image shows single synaptic connection that is imaged in the x y plane and the y z plane.
The technique of FIB/SEM works optimally when the region of interest is not too large. The upper limit for an imaged volume depends on the ion beam and the area that it can consistently mill repeatedly over many hours. So far this has been done with areas of around 60 x 60 microns, however, imaging this entire area is not possible with a suitable image resolution, mainly due to the time factor in collecting such a large image. As an example, a 4 k x 4 k image with a pixel dwell time to 10 microseconds would take 2 minutes and 40 seconds to acquire, and for a field of view of 60 microns this would only give a pixel size of 15 nm. For imaging the ultrastructure of cells and tissue, a smaller pixel size is more suitable; usually between 4 and 10 nm / pixel. For an area of 60 x 60 microns the image would be acquired in 10 hours, even if the microscope had this capability. For these reasons, the microscope is an ideal tool for image volumes in the order of 10 x 10 x 10 microns, or significant regions of whole cells. This can easily be done within a 48 hour period on samples that are well-contrasted with this procedure.
So far the protocol has been used mainly with brain tissue, as well as various types of cultured mammalian cells adhering to coverslips. However, the fixation and staining procedure can be used with many other types of biological material, including plant material to produce equally well-contrasted image stacks.
Therefore the main advantage of this technique is its capacity to image volumes with isotropic voxels. Image stacks of this type allow for analyses of the 3d volume using images taken at any plane in the stack. The benefit of this is to allow the observer to image features in the image series from any direction and providing a greater image detail (Figure 3).
The authors have nothing to disclose.
Name of Reagent | Company | Catalogue number |
---|---|---|
Cyanoacrylate glue | ||
Dissecting microscope: Leica MZ8 | Leica Microsystems, Germany | |
Durcupan resin | Sigma-Aldrich Chemie GmbH, Switzerland | |
FIB/SEM microscope (Zeiss, NVision 40) | ||
Glass histology slides | Menzel-Gläser, Germany | AA00008032E |
Glass knife maker: Leica EM KMR2 | Leica Microsystems, Germany | |
Glass scintillation vials (20ml) | EMS, USA | 72634 |
Glutaraldehyde | EMS, USA | 16222 |
Jeweler’s saw | EMS, USA | 72010 |
Mould separating agent | Glorex, Switzerland | 62407445 |
Osmium tetroxide | EMS, USA | 19110 |
Paraformaldehyde | EMS, USA | RT 19208 |
Phosphate salts for phosphate buffer | Sigma-Aldrich Chemie GmbH, Switzerland | 71642 and 71496 |
Sodium cacodylate | Sigma-Aldrich Chemie GmbH, Switzerland | 20840 |
Sputter Coater with gold target | Cressington, USA | |
Tris base (TBS) | Sigma-Aldrich Chemie GmbH, Switzerland | T1378 |
Ultramicrotome: Leica UCT | Leica Microsystems, Germany | |
Uranyl acetate | EMS, USA | RT 22451 |