Laser-induced breakdown spectroscopy performed on thin organ and tumor tissue successfully detected natural elements and artificially injected gadolinium (Gd), issued from Gd-based nanoparticles. Images of chemical elements reached a resolution of 100 μm and quantitative sub-mM sensitivity. The compatibility of the setup with standard optical microscopy emphasizes its potential to provide multiple images of a same biological tissue.
Emission spectroscopy of laser-induced plasma was applied to elemental analysis of biological samples. Laser-induced breakdown spectroscopy (LIBS) performed on thin sections of rodent tissues: kidneys and tumor, allows the detection of inorganic elements such as (i) Na, Ca, Cu, Mg, P, and Fe, naturally present in the body and (ii) Si and Gd, detected after the injection of gadolinium-based nanoparticles. The animals were euthanized 1 to 24 hr after intravenous injection of particles. A two-dimensional scan of the sample, performed using a motorized micrometric 3D-stage, allowed the infrared laser beam exploring the surface with a lateral resolution less than 100 μm. Quantitative chemical images of Gd element inside the organ were obtained with sub-mM sensitivity. LIBS offers a simple and robust method to study the distribution of inorganic materials without any specific labeling. Moreover, the compatibility of the setup with standard optical microscopy emphasizes its potential to provide multiple images of the same biological tissue with different types of response: elemental, molecular, or cellular.
The wide development of nanoparticles for biological applications urged the parallel improvement of analytical techniques for their quantification and imaging in biological samples. Usually the detection and the mapping of the nanoparticles in organs are made by fluorescence or confocal microscopy. Unfortunately these methods require the labeling of the nanoparticles by a near infrared dye that can modify the biodistribution of the nanoparticles, especially for very small nanoparticles due to its hydrophobic properties. The detection of labeled nanoparticles, and especially the very small nanoparticles (size < 10 nm), might thus interfere with their biodistribution at the whole body scale but also at the tissue and cell levels. The development of new devices able to detect nanoparticles without any labeling offers new possibilities for the study of their behavior and kinetics. Moreover, the role of trace elements such as iron and copper in brain illnesses and neurodegenerative diseases such as Alzheimer1, Menkes2,3, or Wilson4 suggest the interest to study and localize these elements in tissues.
Various techniques have been used to provide elemental mapping or microanalysis of different materials. In their review paper published in 2006, R. Lobinski et al. provided an overview of available standard techniques for elemental microanalysis in biological environment, one of the most challenging environments for analytical sciences5. The electron microprobe, which consists of energy dispersive X-ray microanalysis in a transmission electron microscope, can be applied to numerous studies if the element concentration is sufficient (>100–1,000 μg/g). To reach lower detection limits, the following techniques have been used:
The above-mentioned techniques provide micrometric resolution as shown in the Table 1 extracted from Lobinski et al.
3D reconstruction of serial 2D investigations could also be proposed for the reconstruction of deeper tissues11. However, all devices and systems require both qualified professionals, moderate to highly expensive equipment and long-lasting experiments (typically more than 4 hr for a 100 µm x 100 µm for µ-SXRF and 10 mm x 10 mm for LA-ICP-MS)12. Altogether, these requirements make elemental microanalysis very restricting and incompatible with conventional optical imaging systems, fluorescence microscopy or nonlinear microscopy. Another point that we can mention here is that the quantitative measurement capability is still quite limited and depends on the availability of matrix-matched laboratory standards. The further generalization of the use of elemental microanalysis in industry processes, geology, biology and other domains of applications will generate significant conceptual and technological breakthroughs.
The purpose of the present manuscript is to propose solutions for quantitative elemental mapping (or elemental microanalysis) in biological tissues with a tabletop instrumentation fully compatible with conventional optical microscopy. Our approach is based on the laser-induced breakdown spectroscopy (LIBS technology). In LIBS, a laser pulse is focused on the sample of interest to create the breakdown and spark of the material. The atomic radiation emitted in the plasma is subsequently analyzed by a spectrometer and the elemental concentrations can be retrieved with calibration measurements performed beforehand13,14. The advantages of LIBS include sensitivity (µg/g for almost all the elements), compactness, very basic sample preparation, absence of contact with the sample, instantaneous response and precisely localized (micro) surface analysis. However, the application of tissue chemical imaging remains challenging since the laser ablation of tissue must be finely controlled to perform maps with high spatial resolution together with sensitivity in the µg/g range15,16.
With such solution, the adjunction of tracers or labeling agents is not needed, which allows detecting inorganic elements directly in their native environment in biological tissues. The LIBS instrument developed in our laboratory offers a current resolution inferior to 100 µm with an estimated sensitivity for Gd below 35 µg/g, equivalent to 0.1 mM 16, which allows the mapping of large samples (>1 cm2) within 30 min. In addition, homemade software facilitates the acquisition and exploitation of the data. This instrument is used to detect, map, and quantify the tissue distribution of gadolinium (Gd)-based nanoparticles17–18 in kidneys and tumor samples from small animals, 1 to 24 hr after intravenous injection of the particles (size <5 nm). Inorganic elements, which are intrinsically contained in a biological tissue, such as Fe, Ca, Na, and P, have also been detected and imaged.
1. Biological Sample Preparation
All the experiments described in this study were approved by the Animal Care and Use Committee of the CECCAPP (Lyon, France) (authorization #LYONSUD_2012_004), and the experiments were carried out under the supervision of authorized individuals (L. Sancey, DDPP authorization #38 05 32).
2. Sample Preparation for Calibration
3. LIBS Experiment
With the same experimental parameters, measure the calibration samples (for preparation details, see section 2). Perform a map or record 25 spectra (obtained from measurement sites in the center part of the drop) in each of the calibration drops.
4. LIBS Spectrum Analysis: Constructing of Chemical Images
As shown in Figure 1, the beam of a Nd:YAG laser in the fundamental wavelength of 1,064 nm was focused vertically down on the tissue slice by a quartz lens of 50 mm focal distance. The pulse energy was 4 mJ and the repetition rate 10 Hz. In order to avoid the generation of plasma in air, the laser beam was focused around 100 µm under the surface of the sample. No air plasma was observed in this condition. During the experiments, the sample was moved by a step motor in order to generate one plasma in only one position on the sample (single shot). A microscopic observation of the craters generated on the substrate showed a crater diameter lower than 50 µm. The generated plasma was imaged by a couple of lens onto the entrance of an optical fiber connected at the other end to a Czerny-Tuner spectrograph equipped with a grating of 1,200 lines per mm (blazed at 300 nm). An ICCD camera was mounted on the output focal plane of the spectrograph to record the spectrum. During the measurements, the height of the sample surface (with regard to the laser focus position) is monitored using trigonometric method with an obliquely incident beam from a low power cw laser diode on the sample surface and a CCD monitor camera. The ICCD was triggered by the laser Q-switch and the 400 nsec and 2 µsec were set respectively for the delay and gate parameters. During a mapping, the sample was shifted by an offset of 100 µm after each laser shot. One spectrum was recorded for each laser shot. Homemade software developed in the LabVIEW environment controlled the entire equipment and allowed to perform automated sequence to scan the area of interest of the tissue sample with specific lateral resolution.
Example of single shot spectra, recorded on different regions of a kidney tissue after IV injection of Gd-based nanoparticles is shown in Figure 2. The nanoparticles were synthesized as described in Mignot et al17. Briefly, the nanoparticles are composed of a polysiloxane matrix, holding DOTAGA-Gd3+ chelates on their surface. They are developed for (i) multi-modal imaging as they can be used in MRI, nuclear imaging and fluorescence imaging, separately or simultaneously, and (ii) therapeutic function as radiosensitizer (i.e. increasing locally the efficiency of the radiotherapy)18. The spectral range used (286 to 320 nm) allows detecting different elements such as Gd, Ca, Fe, Si, and Al. The blue spectrum was recorded in the central region of the kidney (the medulla), the red spectrum corresponded to the kidney membrane (capsule) and the green spectrum to the peripheral area (cortex). It is also noteworthy that the Gd, Si, Al, Ca, and Fe intensities are widely variable at different regions, suggesting a large heterogeneity of these elements concentrations within the tissue. Basically Gd and Si were containing in the Gd-based nanoparticles, Fe was specific for the blood vessels, Ca from the biological sample itself.
The tissue slice samples were analyzed spot by spot, moving the sample in the X and Y position. Lateral resolution was close to 100 µm. One single spectrum was recorded for each position (single shot measurement). Line intensities were extracted from the spectra using a background subtraction. The chemical images were then constructed from these intensities. An example is shown in Figure 3 for Gd, Si and Fe. Lines used to build these chemical images are shown in Table 2. The analysis was correlated to the colored photography of the biological sample before the experiment to improve the localization and mapping of the different elements. In the kidney, the Gd and Si signals were co-localized, but this co-localization did not fit with the distribution of the Fe, indicating a specific distribution.
Similar results were obtained with tumor samples. An example of measurement on SQ20B tumor tissue is shown in Figure 4. In this case, the chemical map of Gd, shown with a false color scale, has been superposed to the natural light picture. For this measurement, quantitative analysis has been performed to retrieve the local gadolinium concentration (see the procedure in the protocol section). For this experiment, the nanoparticles were administered directly into the tumor, which was removed 1 hr after injection and sliced for the analysis. As seen in Figure 4, the particles diffused in the tissue from the injection point in the tumor center to the periphery. 1 hr after the injection, approximately half of the tumor volume contained some particles. Such a mapping can help to provide information about the therapeutic protocol. For an optimal efficacy of the treatment, the nanoparticles should diffuse within the entire tumor; after 1 hr, only half of the tumor contained some particles, suggesting that a longer diffusion time would be required for an optimal diffusion and thus effective treatment.
Table 1. Main spatially resolved analytical techniques for chemical elemental imaging. (Lobinski et al.5). Please click here to view a larger version of this table.
Figure 1. Schematic presentation of the used LIBS experimental setup. (Motto-Ros et al.15).
Figure 2. LIBS single-shot spectra obtained in three different regions of the kidney tissue. These regions are the medulla (central part of the kidney), the capsule in red and the cortex in green. Spectra have been shifted vertically for clarity.
Figure 3. LIBS elemental mapping of Gd, Si, and Fe in a slice of mouse kidney prepared according to the conditions specified in the text. The scale intensity of chemical images is expressed in an arbitrary unit. A natural light photography is also shown in the upper left of the figure.
Table 2. Spectral lines used for the detection of Gd, Si, and Fe in a mouse kidney (atomic lines are labeled I and ionic lines are labeled II).
Figure 4. LIBS quantitative mapping of Gd inside of a slice of tumor tissue. The false color scale runs from 0.1 mM (green) to 20 mM (violet).
Applied to biological sample, this technique allows the chemical imaging, i.e. the mapping and quantification, of Gd and Si from injected Gd-based nanoparticles in different organs. From the main critical settings, the control of laser properties (wavelength, pulse energy, focusing, and stability) is critical for a precise and fine tissue ablation (i.e. mapping resolution) as well as for sensitivity. Working at high energy provides a better sensitivity but unfortunately generates degraded spatial resolution. Besides, the type of spectrometer used must be checked carefully. Basically, a broadband investigation (using for example an echelle spectrometer) will allow a large choice of elemental species but with low sensitivity, while using a Czerny-Turner spectrometer (equipped with an ICCD or a PMT) will allow detecting fewer elements but with much more sensitivity. All the chosen settings of the probed wavelength range should be adjusted to the purpose of the investigation.
Concerning the biological sample itself, its thickness and hardness might also interfere with the quality of the spectrum. A very thin sample would be totally sprayed and the sample support would be attacked also by the laser shot, whereas a too thick sample might suffer from its in-homogeneity of its constitution that will be reverberated on the in-homogeneity of the amount of sprayed matter (presence of large vessels etc.). From the elemental study, the choice of the support might be anticipated; for example, if the investigated element is Si, the sample should be filed on pure plastic but glass should be avoid as it contains large amounts of Si. Similarly, the sample might be prepared and fixed in specific conditions to prevent any component of the fixative to contaminate the level of the studied element.
This technique might be very useful for the analysis of pre-clinical samples. LIBS could allow the detection of abnormal metallic elements such as Fe and Cu, in particular into brain samples. A particular limitation with Fe should be the high content of Fe in hemoglobin; the sample prepared carefully and the assessment of a calibration assay is mandatory. In the field of biotechnologies applied to biology or medicine, LIBS could allow the detection of any compound or particle that contains a specific element such as Au, Gd, Cu, etc.
The bench-top instrumentation is fully compatible with the standard optical microscopy, which shows its large potential use in Biology and Medicine as a tool for complementary observation of trace metallic elements. This type of elemental imaging, combining high lateral resolution (<50 µm) with low detection limits (in the range of mg/kg), generally involves high level of required equipment, such as synchrotron radiation microanalysis (SXRF) or laser ablation inductively coupled mass spectrometry (LA-ICP-MS), making it very restrictive and incompatible with commercially available microscopic imaging systems. The LIBS instrumentation is not very expensive, depending of the resolution and sensitivity required (price between 100 and 200 k€). Currently a 1 cm2 sample could be analyzed within 30-40 min; this acquisition time might not change after the addition of other microscopic techniques such as fluorescence detection and colored imaging of the sample. The current 100 µm spatial resolution already shows the efficacy for mapping of biological tissues. However, this resolution could be greatly improved by using an optical microscope instead of a simple lens to focus the ablation laser pulse. Micrometer resolution can theoretically be reached, even though the sensitivity will drop, as less material will be ablated. Increasing the resolution to 10 µm would allow an accurate localization of nanoparticles to be observed at the cell level.
To conclude, LIBS is a new technology applicable to the detection and quantification of elements in tissue samples. This approach is sensitive (µg/g), fast (currently 10 measurements/sec), affordable, easy to implement and to use, and fully compatible with conventional optical microscopy.
The authors have nothing to disclose.
The authors gratefully acknowledge financial support by the Labex-Imust.
Laser nanosecond Nd:YAG | Quantel | Brillant | 5ns pulse witdh, wavelength 1064 nm |
Spectrometer | Andor Technology | Shamrock 303 | with 1200 l/mm blazed at 300 nm grating |
Detector ICCD | Andor Technology | Istar | 2 ns temporal resolution |
LIBS Unit | ILM | Homemade Instrumentation | |
Gd-based nanoparticles | Nano-H | particles | |
HEPES | Sigma-Aldrich | H4034 | for particle's dilution |
CaCl2 | Sigma-Aldrich | 21108 | for particle's dilution |
NaCl | Sigma-Aldrich | S5886 | for particle's dilution |
mice | Charles River | depending of animal breeding | |
isoflurane | Coveto / Virbac | for anaesthesia – Isofluranum | |
isopentane | Sigma-Aldrich | 59060 | to froze the sample slowly |
liquide nitrogen | Air Liquide | to cool down the isopentane | |
cryostat | Leica | CM-3050S | to slide the samples |
petri dishes | Dutscher | 353004 | to stick the sample |