This paper details the fabrication protocol of microfluidic chips developed for on-chip protein crystallization with the dialysis method and in situ X-ray diffraction experiments. The microfabrication process makes it possible to integrate a semipermeable regenerated cellulose dialysis membrane with any molecular weight cut-off, between two layers of the chip.
This protocol describes the manufacturing of reproducible and inexpensive microfluidic devices covering the whole pipeline for crystallizing proteins on-chip with the dialysis method and allowing in situ single-crystal or serial crystallography experiments at room temperature. The protocol details the fabrication process of the microchips, the manipulation of the on-chip crystallization experiments and the treatment of the in situ collected X-ray diffraction data for the structural elucidation of the protein sample. The main feature of this microfabrication procedure lies on the integration of a commercially available, semipermeable regenerated cellulose dialysis membrane in between two layers of the chip. The molecular weight cut-off of the embedded membrane varies depending on the molecular weight of the macromolecule and the precipitants. The device exploits the advantages of microfluidic technology, such as the use of minute volumes of samples (<1 µL) and fine tuning over transport phenomena. The chip coupled them with the dialysis method, providing precise and reversible control over the crystallization process and can be used for investigating phase diagrams of proteins at the microliter scale. The device is patterned using a photocurable thiolene-based resin with soft imprint lithography on an optically transparent polymeric substrate. Moreover, the background scattering of the materials composing the microchips and generating background noise was evaluated rendering the chip compatible for in situ X-ray diffraction experiments. Once protein crystals are grown on-chip up to an adequate size and population uniformity, the microchips can be directly mounted in front of the X-ray beam with the aid of a 3D printed holder. This approach addresses the challenges rising from the use of cryoprotectants and manual harvesting in conventional protein crystallography experiments through an easy and inexpensive manner. Complete X-ray diffraction data sets from multiple, isomorphous lysozyme crystals grown on-chip were collected at room temperature for structure determination.
Elucidating the three-dimensional (3D) structure of biological macromolecules is an unceasing pursuit in structural biology where X-ray crystallography remains the principal investigation technique. Applied for unraveling the structural details of complex macromolecules, such as proteins, it aims at facilitating the understanding of their mechanisms of actions and their involvement in various biological functions. Powerful X-ray sources at synchrotrons and X-ray free-electron lasers (XFELs) provide all the tools required for a deeper insight into the proteins' structure at near atomic resolution. Despite the advantages that come along with the use of X-rays for structural studies, there are intrinsic limitations to X-ray radiation and the crystallization process itself. Radiation damage provoked by high X-ray flux and long exposure times of the protein crystal in front of the X-ray beam are restrictive parameters that crystallographers have to surpass using cryogenic cooling1. However, finding the optimal cryocooling conditions can be laborious since conformational changes from the native protein structure or artifacts can be concealed2,3. Moreover, recent studies indicate that performing diffraction experiments at room temperature leads to lower specific radiation damage4. Another bottleneck in structural biology is the acquisition of well-diffracting crystals with a sufficient size5. Small crystals are easier to produce, especially in the case of membrane proteins, but are more susceptible to radiation damage even under cryocooling conditions because a high radiation dose must be directed in a smaller volume compared to the case of larger protein crystals6. The novel approach of serial crystallography7,8 at synchrotrons and XFELs can circumvent the restrains of radiation damage and at the same time exploit smaller crystals (200 nm to 2 µm)7 by merging data sets from multiple, isomorphous and randomly oriented protein crystals and profiting from the associated technological advances such as femtosecond pulses, shorter exposure times and micro-focused X-ray beams5,7,9,10.
Microfluidic technology is valuable to X-ray crystallography, exhibiting manifold advantages for the crystallization of biological macromolecules and their structural investigation. Conducting crystallization experiments in microfluidic devices requires small volumes of protein sample, therefore constraining the production cost of these high-valued bio macromolecules and facilitating high-throughput screening and optimization of numerous crystallization conditions. Moreover, the inherent large surface area-to-volume ratio at the microfluidic scale and diffusion-limited transport phenomena enable fine control over flows and temperature or concentration gradients11,12,13,14, rendering microfluidic devices suitable for growing uniformly sized crystals and exploring phase diagrams15,16,17,18,19. Moreover, microfluidic tools display a distinctive potential to address another hurdle in protein crystallography, which is the sample delivery, and the necessity to handle and harvest protein crystals prior to their use for X-ray diffraction experiments. The method of on-chip and in situ X-ray crystallography eliminates the crystal manipulation and the potential deterioration of crystal quality prior to data collection. A wide range of microfluidic chips compatible for in situ X-ray protein crystallography have been designed, developed, and tested by many research groups confronting the related restrictions arising from the nature of the microfabrication materials and their interactions with X-rays14,19,20,21,22,23. The fabrication materials must be optically transparent, biologically inert and demonstrate high transparency to X-ray radiation and an optimal signal-to-noise ratio during data collection.
Most of the crystallization methods applied in conventional protein crystallography24,25 have also been implemented at the microfluidic scale11,14 for on chip crystallization and in situ X-ray diffraction analysis. Simple, hybrid, or multi-layered microfluidic apparatus incorporating vapor diffusion26, evaporation27, free interface diffusion (FID)28, microbatch26, or even seeding29 have been used to crystallize soluble and membrane proteins. High throughput screening and optimization of crystallization conditions can be achieved30,31 in well-based32, droplet-based33, or valve-actuated34 devices. In situ X-ray diffraction experiments of challenging protein targets at room temperature have been conducted in microchips fabricated from various materials such as PDMS (polydimethylsiloxane), COC (cyclic olefin copolymer), PMMA (poly(methyl methacrylate))21,22,26,28,29, graphene films23, Kapton35, epoxy glue6, or NOA (Norland Optical Adhesive)19 and the materials' transparency to X-ray radiation and their contribution to background noise have been evaluated. Moreover, microchips have been designed to couple the in situ and the serial data collection strategies in a single tool for X-ray protein crystallography experiments at synchrotron sources23,35,36 and XFELs7.
Room temperature in situ data collection has also been implemented in various delivery methods and devices. For example, Nogly et al.54 used a lipidic cubic phase (LCP) injector in order to study the structure of the light-driven photon pump bacteriorhodopsin (bR) by serial femtosecond crystallography (SFX) using an XFEL source. The crystal structure of bR was solved to 2.3 Å resolution, demonstrating the compatibility of an LCP injector with time-resolved serial femtosecond crystallography (TR-SFX). Baxter et al.55 designed a high-density multi-crystal grid, fabricated by a 100 or 200 µm thick polycarbonate plastic with laser-cut holes of various sizes. An additional 5 µm thick polycarbonate film can be fixed to one side of the grid when using the device for sitting- or hanging-drop crystallization experiments. This high-density grid can be used in multiple ways as crystals can be loaded directly onto the ports of the device or crystals can be grown on the device by vapor diffusion or the LCP method. Moreover, the grid can be adjusted in a standard magnetic base and used for in situ X-ray data collection at cryogenic or room temperature conditions. More recently, Feiler et al.56 developed a sample holder for macromolecular in situ X-ray crystallography at cryogenic and ambient temperature with minimal background noise contribution. Specifically, the holder comprises of a plastic support, a transparent COC foil and a microporous structured polyimide foil. It was designed to replace the commonly used cover slides for setting up crystallization drops, while allowing in-place manipulation such as ligand soaking, complex formation, and cryogenic protection without opening the crystallization drop or manually handling the crystals. Moreover, the sample holder can be removed from the crystallization plate and placed onto a magnetic base for in situ data collection at standard goniometer-based beamlines. For ambient temperature data collection, the COC foil is removed prior to the experiment and only the 21 µm-thick polyimide foil contributes to background scattering, which in this case is minimal. These examples compose only a small fraction of the ongoing research and the multitude of versatile microchips developed for X-ray protein crystallography.
However, the dialysis protein crystallization method has not been widely incorporated within microfluidics. Dialysis is a diffusion-based method aiming for the equilibration of precipitant concentration through a semi-permeable membrane in order to approach the nominal concentration for protein crystallization and enables precise and reversible control over the crystallization conditions24. The Molecular Weight Cut-Off (MWCO) of the semi-permeable dialysis membrane can be chosen depending on the molecular weight of the macromolecule and the precipitants to allow the diffusion of small precipitant molecules while retaining the macromolecule of interest. Due to the reversibility of the dialysis process, it can be used in combination with temperature control to decouple and optimize nucleation and crystal growth independently37 for investigating phase diagrams by altering the precipitant concentration while using the same protein sample. The integration of membranes in microfluidics is reviewed by de Jong et al.38 and the case studies in biology implanting dialysis into microchips can be principally listed in sample preparation, concentration or filtration applications39,40,41,42 or cell-related studies43,44. Pervaporation through PDMS was used by Shim et al.37 to study the nucleation and growth of xylanase in various conditions. Water permeated through the 15 µm thick PDMS membrane into the protein reservoir of the microfluidic device, subsequently altering the protein and precipitant concentration.
The protocol developed by Junius et al.19,45 for the fabrication of a microfluidic chip compatible for both on-chip protein crystallization via microdialysis and in situ X-ray diffraction experiments at room temperature is presented. The protocol for the device fabrication is directly inspired by the pioneering work accomplished by Studer and coworkers12,46 for micro-patterned stickers of photo-curable thiolene-based resin NOA 81 embedding commercially available membranes, using soft imprint lithography. An innovative modification of the method resulted in microchips enabling the use of microdialysis to accurately monitor and control the experimental parameters for the on-chip growth of protein crystals and simultaneously exploit the advantages of microfluidics, such as reduced consumption of protein samples per experiment (<1 µL). In a previous work, the principles of dialysis applied to a macro-scale system (typical volume >20 µL) for screening and optimizing crystallization conditions by mapping temperature-precipitant concentration phase diagrams were demonstrated47. In this work, a protocol is described for producing dialysis microchips incorporating regenerated cellulose (RC) dialysis membranes of different MWCO in order to perform crystallization assays on-chip and in situ X-ray diffraction data collection. The materials comprising the microchips have been evaluated for their transparency to X-rays19 and the devices can be set directly in front of the X-ray beam for room temperature in situ diffraction experiments, excluding the manual handling and minimizing the degradation of fragile protein crystals. In a case study, hen egg-white lysozyme crystals were grown on-chip via microdialysis generating a uniformly sized population. The microchip was then mounted in front of the X-ray beam with a 3D-printed support19 and complete in situ diffraction data sets were collected at room temperature from multiple, isomorphous crystals, demonstrating the high potential and relevance of the chips for synchrotron serial crystallography studies of challenging macromolecular targets.
1. Mask design and master fabrication
2. PDMS molds fabrication
NOTE: The following steps of the protocol can be performed in any laboratory as long as a laminar flow hood is used, yellow light in the room is used when working with the NOA 81 resin (steps 3.6-3.11) and a source of UV light is available for polymerizing the NOA 81 resin (steps 3.7 and 3.11).
3. Dialysis chip fabrication
4. Fluidic connectors
NOTE: The design of the microfluidic chip consists of a linear fluidic channel for the crystallization solution and a central reservoir for the protein sample (protein reservoir), both shown from a top view of two microchips in Figure 2A. A RC dialysis membrane is embedded between these two microstructures (Figure 2D) and the crystallization process evolves while precipitants from the crystallization solution diffuse across the membrane due to a concentration gradient between the two compartments of the chip that are separated by the membrane. The microfluidic channel is imprinted on the bottom PDMS mold (Figure 1F). Once the fabrication protocol for the chips is completed, the linear channel is located on the bottom layer of the NOA 81 sticker in contact with the PMMA substrate, as shown in Figure 1K. An inlet and an outlet access point for the crystallization solution are located at each end of the linear channel and look like holes (total height 90 µm) as can be seen in Figure 2A. For the handling of the crystallization solution, connectors must be added on the access points.
5. Protein encapsulation
NOTE: The pattern of the chip dedicated to being used as the protein reservoir remains so far open to the atmosphere. The following protocol is proposed to carefully confine the protein sample within the microfluidic chip.
6. On-chip protein crystallization
7. In situ and on-chip X-ray diffraction
The microfluidic chips developed by Junius et al.19,45 are compatible for both on-chip protein crystallization with the microdialysis method and in situ X-ray diffraction data collection at room temperature. Pictures of the microchips, their detailed design, the fluidic connectors, and the RC dialysis membrane are illustrated in Figure 2. The crystallization experiments are set up by manually pipetting the protein sample directly into the protein reservoir and introducing the crystallization solution into the linear fluidic channel with an automated pressure-driven system or syringe pump or manually with the aid of a syringe. The protein reservoir and the fluidic channel can be distinguished in Figure 2A. Designs for fabricating chips with 0.1 µL or 0.3 µL maximum volume of the protein reservoir are shown in Figure 2A on the left and right, respectively. Chips with a 0.2 µL or 0.7 µL maximum capacity for the protein sample are shown elsewhere19. The highlight of the protocol for the device fabrication can be narrowed down on the use of the photocurable thiolene-based resin NOA 81 embedding commercially available RC dialysis membranes of various MWCOs. During the fabrication of the microfluidic devices, the linear fluidic channel is imprinted on the bottom PDMS mold (Figure 1F), while the upper PDMS mold consists only of the patterned pillars for the protein reservoir and the inlet and outlet ports (Figure 1F). Once NOA 81 is crosslinked and the PDMS molds are removed from the assembly (Figure 1K), the fluidic channel is located at the bottom layer of the microchip and the protein channel and inlet/outlet ports are located on both layers. Figure 1L illustrates a side view schematic of the dialysis chip where all the layers of the device and their respective thickness are indicated. The height of the patterns imprinted on the bottom layer of the chips (fluidic channel) is approximately 45 µm, while the total height of the inlet and outlet ports is approximately 90 µm. The protein reservoir (45 µm height) is also illustrated in Figure 2D and 2E. The alignment of the two layers was investigated under an optical microscope and the piece of the RC dialysis membrane incorporated within the microchip can be clearly distinguished in Figure 2D. In the same figure, air has been entrapped within the fluidic channel during the injection of the crystallization solution, as can be seen in the upper-left part of the protein reservoir. Figure 2E is a close-up photograph of the protein reservoir after the manual deposition of the protein droplet with a pipette and before the encapsulation of the droplet with a piece of PMMA and Kapton tape, as described in steps 5.2 and 5.3 of the protocol. The microfluidic chip ready to be used for crystallization experiments, after the encapsulation of the protein sample and the gluing of the fluidic connectors, is depicted in Figure 2C. The air-tight assembly ensures that leakages cannot occur. The fluidic connectors for the inlet and outlet ports of the microfluidic channel can be either the commercially available ones as described in step 4.1 of the protocol and shown in Figure 2C, or disposable laboratory pipette tips can be used for the same purpose (Figure 2B, protocol step 4.4).
For the fabrication of the microfluidic chips, optically transparent and biologically inert materials were chosen, demonstrating high compatibility for in situ X-ray diffraction experiments at room temperature. The interactions of X-rays, absorption and scattering, with the materials composing the microfluidic device and the surrounding atmosphere (air) generate a signal known as background noise. This noise sums up to the diffraction signal of the protein crystals recorded by the detector, decaying the signal-to-noise ratio and should be maintained as low as possible during X-ray diffraction data collection. We have evaluated the background noise generated by the materials comprising the protein reservoir, which is in the direct path of the X-ray beam. The protein reservoir consists of the RC dialysis membrane, the Kapton tape and two PMMA pieces, one used as a substrate for the microchip and one used for the encapsulation of the protein sample. The thickness of the PMMA is 2 x 175 µm, of the Kapton tape 20 µm, and the RC dialysis membrane is approximately 40 µm thick (Figure 1L). The total thickness of these layers is about 410 µm and the NOA 81 layer is not in the direct X-ray path. Apart from the thickness of the fabrication materials, their density is also crucial for measuring the background scattering noise, as X-ray scattering increases with the elemental atomic number. For this reason, helium flux (a feature provided at BM30A-FIP at ESRF) was used instead of air during the data collection for material characterization and for protein diffraction experiments. Figure 3C illustrates the background noise generated by the Kapton tape, the RC dialysis membrane, the PMMA sheet, and their assembly in helium atmosphere. Each material was exposed for 20 s to X-rays of 0.98 Å wavelength and the sample-detector distance was 200 mm. The experiments were performed at the BM30A-FIP beamline at ESRF, as explained in step 7 of the protocol. Diffuse rings attributed to the interactions of the X-ray beam with the materials can be distinguished for the Kapton tape at a resolution lower than 4 Å, the PMMA sheet between 4-8 Å, and the dialysis membrane between 4-5 Å resolution. The background noise generated by the dialysis chip is mainly observed at a resolution lower than 6 Å that does not affect the treatment of high-resolution diffraction data of the large lysozyme crystals. The background scattering intensity as a function of the resolution for the microchip and the separate materials are shown elsewhere19. In the measurement presented in Figure 3C, the dialysis chip was empty of any solution (protein or precipitant solution) and the contribution of the presence of solution to the background noise has not been measured. The chips were mounted in front of the X-ray beam with a 3D-printed support (Figure 3B) designed for in situ diffraction experiments19. However, the same support with dimensions equal to the dimensions of a 96-well/SBS standard crystallization plate, can be used for performing 1 to 3 crystallization experiments concurrently, as it can hold up to 3 chips simultaneously (Figure 3A).
Experiments were conducted to evaluate the efficiency of the microfluidic devices for the on-chip crystallization of model soluble proteins with the microdialysis method. The fluidic channel was filled as described in step 4 of the protocol, while steps 5 and 6 described how to encapsulate the protein sample within the dedicated protein reservoir and how to set up the crystallization experiments. Figure 4 shows lysozyme crystals grown at 293 K under 1.5 M sodium chloride (NaCl) with 0.1 M sodium acetate (CH3COONa) pH 4.0 (A) and under 1 M NaCl, 0.1 M CH3COONa pH 4.5 with 30% polyethylene glycol (PEG) 400 (B). The lyophilized lysozyme powder was dissolved in water to a final concentration of ~30 mg mL-1 or in 20 mM CH3COONa pH 4.2 buffer to a final concentration of ~20 mg mL-1 for the experiments illustrated in Figures 4A and 4B, respectively. The volume of the protein sample in both experiments was about 0.3 µL and the MWCO of the RC dialysis membrane embedded within the microchips lies in the range of 6-8 kDa. The lysozyme crystals shown in Figure 3A grew within 1 h and the crystals in Figure 3B grew within 30 min from the onset of the experiment. The crystallization experiments were carried out under static conditions. However, it has been shown19 that conducting the experiments under flowing conditions provides the possibility to dynamically exchange the crystallization conditions and study phase diagrams, verifying the reversibility of the microdialysis method.
In situ X-ray diffraction data from the lysozyme crystals shown in Figure 4A were collected to demonstrate the suitability of the dialysis chips for such experiments. The data collection was carried out at BM30A-FIP beamline (ESRF) at room temperature, as described in step 7.2.1 of the protocol. The microchips were mounted at the beamline with the aid of the 3D-printed support (Figure 3B) and complete X-ray diffraction data sets were collected from two single lysozyme crystals grown on-chip under the conditions given in the second line of Table 1. The observed reflections of the data sets were processed, indexed, and integrated using XDS48 and the molecular replacement and refinement were accomplished using Phaser49 and Phenix52, respectively. The crystallographic statistics for the complete data set of each lysozyme crystal and for the merging of the two data sets are provided in Table 2. For the molecular replacement, the PDB entry 193L was used.
Electron density maps from a single lysozyme crystal and the merged data set of the two crystals have been obtained at 1.95 Å and 1.85 Å, respectively, and are illustrated in Figures 5A and 5B. Both electron density maps show detailed structural information that can be obtained by in situ X-ray diffraction experiments conducted directly on the dialysis microchip at room temperature from a single crystal or from multiple crystals, rendering the chips compatible for in situ X-ray crystallography studies.
Figure 1: Schematic illustration of the dialysis chip fabrication. (A) SU-8 resin is deposited on two silicon wafers and spin coated. (B) A SU-8 master is acquired after irradiating the photoresist with UV light through a photomask and developing the unexposed parts. (C) PDMS is dispensed upon the SU-8 masters and after being cured at 338 K for 1 h, (D) the 2 PDMS molds produced by replica molding and imprinting the micro-patterns are peeled of the masters and (E) cut to the appropriate size. (F) The PDMS molds are supported on a glass slide incorporating the RC dialysis membrane in between the two central pillars. (G) The 2 PDMS molds are then aligned and desiccated for ~30 min in a vacuum chamber. (H) The NOA 81 resin is poured in between the two molds and (I) fills the space by capillarity. (J) After the first exposure to UV light, the bottom PDMS mold is removed and the assembly is deposited on a PMMA sheet. (K) The second UV exposure follows in order to fully polymerize the NOA 81 resin and the dialysis chip is ready-to-use after removing the remaining upper PDMS mold. (L) Side view schematic of the dialysis chip where all the layers of the device and their respective thickness are indicated. Please click here to view a larger version of this figure.
Figure 2: Dialysis chips embedding an RC dialysis membrane for on-chip protein crystallization and in situ X-ray diffraction experiments. (A) NOA 81 microchips on 175 µm thick PMMA substrate with a protein reservoir of 0.1 µL (left) and 0.3 µL (right) nominal volume. (B) Microchips with pipette tips as fluidic connectors glued on the inlet and outlet ports of the fluidic channel. (C) Picture of a microchip during a crystallization experiment. The protein sample is encapsulated with a piece of 175 µm thick PMMA sheet and Kapton tape. Peek Nanoport connectors are used for the inlet and outlet ports of the fluidic channel. (D) Top view of the protein reservoir during the circulation of the crystallization solution within the fluidic channel. Air is trapped in the upper part of the reservoir right under the RC dialysis membrane, which can be clearly detected. (E) Top view of the dialysis reservoir through an optical microscope during the deposition of the protein sample. The protein droplet is deposited right above the embedded RC dialysis membrane. Please click here to view a larger version of this figure.
Figure 3: The 3D-printed support (A) for the microchips used during crystallization experiments and (B) mounted in front of the X-ray beam at BM30A-FIP beamline at the ESRF for in situ X-ray diffraction experiments. (C) Background noise generated by the interaction of X-rays with Kapton, RC dialysis membrane, PMMA, and the dialysis chip (from left to right). Please click here to view a larger version of this figure.
Figure 4: On-chip crystallization of lysozyme with the microdialysis method. (A) Lysozyme (~30 mg mL-1) crystals grown on-chip under 1.5 M NaCl and 0.1 M CH3COONa pH 4.0 and (B) lysozyme (~20 mg mL-1) crystals grown under crystallization conditions containing 1 M NaCl, 0.1 M CH3COONa pH 4.5, and 30% PEG 400. Both experiments were conducted at 293 K. Please click here to view a larger version of this figure.
Figure 5: Electron density maps of the refined lysozyme structure from (A) a single crystal and (B) the merged data set of two crystals grown on-chip via microdialysis. The maps were obtained at 1.95 Å and 1.84 Å, respectively, contoured at 1σ. Please click here to view a larger version of this figure.
Protein | Protein concentration (mg mL-1) |
Protein bufffer | Initial concentration of precipitant solution |
MWCO of RC dialysis membrane (kDa) |
Temperature (K) |
||
Lysozyme | ~ 30 | Water | 1.5 M NaCl 0.1 M CH3COONa pH 4.0 |
6 – 8 | 293 | ||
Lysozyme | ~ 20 | 20 mM CH3COONa pH 4.2 | 1 M NaCl 0.1 M CH3COONa pH 4.5 30% PEG 400 |
6 – 8 | 293 |
Table 1: Composition of the protein buffer and the precipitant solution for on-chip crystallization of lysozyme with the microdialysis method. The lysozyme crystals grown on-chip with the conditions provided in the second line were used for in situ X-ray diffraction data collection.
Protein | Lysozyme | Lysozyme | Lysozyme |
Number of crystals | 1 | 1 | 2 |
Number of diffraction frames | 40 | 30 | 70 |
Oscillation (°) per exposure | 1 | 1 | |
Exposure time (s) | 30 | 30 | |
Temperature (K) | 293 | 293 | 293 |
Space group | P43212 | P43212 | P43212 |
Unit cell parameters | 78.86 78.86 37.87 90.0 90.0 90.0 |
79.17 79.17 37.95 90.0 90.0 90.0 |
78.47 78.47 37.65 90.0 90.0 90.0 |
Resolution range (Å) | 27.31 – 1.95 (2.02 – 1.95) |
27.39 – 1.96 (2.03 – 1.96) |
27.17 – 1.85 (1.91 – 1.85) |
Mosaicity (°) | 0.319 | 0.121 | |
Total reflections (observed) | 25127 (3552) | 19991 (3001) | |
Unique reflections (observed) | 8641 (1357) | 8295 (1321) | 10404 (975) |
Redudancy | 2.90 (2.61) | 2.41 (2.27) | |
Completeness (%) | 95.0 (94.8) | 91.9 (93.3) | 98.23 (93.15) |
Mean I/σ | 6.83 (1.16) | 7.09 (1.66) | 3.7 |
CC(1/2) | 99.1 (42.4) | 97.9 (37.0) | 97.0 |
R-merge | 0.184 | ||
R-meas | 0.139 | 0.221 | 0.219 |
R-pim | 0.116 | ||
Reflections used in refinement | 8645 (787) | 8451 (857) | 10391 (965) |
Reflections used for R-free | 864 (78) | 846 (85) | 1039 (96) |
R-work | 0.1988 (0.2968) | 0.1853 (0.2872) | 0.1839 (0.3102) |
R-free | 0.2430 (0.3437) | 0.2297 (0.3622) | 0.2207 (0.3703) |
Number of non-hydrogen atoms | 1069 | 1071 | 1096 |
macromolecules | 1012 | 1012 | 1012 |
water | 55 | 57 | 82 |
ligand | 2 | 2 | 2 |
Protein residues | 131 | 131 | 131 |
Rms (bonds, Å) | 0.008 | 0.009 | 0.005 |
Rms (angles, °) | 1.17 | 1.26 | 1.05 |
Ramachandran favored (%) | 98.43 | 97.64 | 99.21 |
Ramachandran allowed (%) | 1.57 | 2.36 | 0.79 |
Ramachandran outliers (%) | 0.00 | 0.00 | 0.00 |
Avegare B-factor | 34.26 | 28.54 | 24.34 |
protein | 33.94 | 28.14 | 23.62 |
water | 40.23 | 35.57 | 33.16 |
ligands | 33.23 | 29.63 | 24.77 |
Table 2: Data collection parameters, crystallographic and refinement statistics of lysozyme crystals grown on-chip via the microdialysis method. Values provided in parentheses correspond to the highest resolution shell. The fourth column corresponds to values obtained after merging the data sets of the second and third column.
A microfluidic device has been developed for on-chip protein crystallization with the microdialysis method and in situ X-ray diffraction experiments at room temperature. NOA 81 chips integrating RC dialysis membranes of any MWCO in order to use microdialysis for on-chip protein crystallization can be fabricated. Fabrication materials with a relatively high X-ray transparency were used, rendering the chips compatible for in situ protein crystallography. The fabrication materials that compose the compartment for protein crystallization of the device (PMMA, Kapton, RC dialysis membrane) were evaluated to generate low background noise. Specifically, the background noise generated by the dialysis chip is mainly observed at low resolution (> 6 Å) and does not affect the treatment of high-resolution diffraction data of the large lysozyme crystals required for protein structure determination. The automation of the data collection is amplified using a 3D printed support that can be mounted directly in macromolecular crystallography beamlines and carry up to three microchips simultaneously. This way, manual harvesting and manipulation of the fragile protein crystals is avoided. Moreover, the data collection takes place at room temperature, avoiding the need for cryoprotection which can be related to conformational changes from the native protein structure2,3.
The use of microdialysis as a method to grow crystals on-chip allows to accurately monitor and control the crystallization process. As discussed in the introduction, most of the conventional protein crystallization methods have been implemented using microfluidic devices11,14. However, the advantages of dialysis for protein crystallization had not yet been fully exploited at the microscale. On-chip microdialysis provides the possibility to study phase diagrams and perform screening and optimization of crystallization conditions with the same protein sample19. For the prototypes presented in this work, the protein consumption per chip is limited down to 0.1 or 0.3 µL. Based on the experimental work so far, the most critical steps of the protocol do not derive from the chips' fabrication procedure but from the crystallization process. The fabrication protocol includes many steps but it is straightforward and enables the fabrication of numerous devices (20 to 30 chips) in a single day in the clean room, with relatively inexpensive materials. However, the on-chip crystallization of proteins can be a delicate procedure due to the intrinsic stochastic nature of nucleation and crystal growth, especially in the microscale. A case study has been described, where well-established conditions were used for the crystallization of lysozyme that yielded robust, well-defined crystals suitable for in situ X-ray diffraction data collection. Nevertheless, difficulties may arise by the use of more challenging protein targets, such as membrane proteins, where the crystallization medium is much more complicated, phase diagrams are not known and well-working crystallization conditions are not yet well established. The dialysis chip offers the possibility to surpass these difficulties and study phase diagrams on-chip, without disposing the valuable and frequently costly protein sample, by merely exchanging the crystallization solution within the microfluidic channel.
The versatility of the microfluidic devices stems from exploiting microdialysis for on-chip protein crystallization in order to reversibly control crystallization conditions and map concentration and temperature varied phase diagrams using low protein volume. Moreover, the device is compatible with in situ X-ray diffraction experiments and the prototyping of the devices is inexpensive and rapid. Numerous, isomorphous crystals of soluble and membrane proteins (in preparation) can be grown on-chip and it is expected that all these features can be utilized for serial X-ray crystallography studies of challenging protein targets at synchrotron and XFEL facilities. Finally, performing on-chip and in situ time-resolved studies is a future possibility that could be of significant interest to the crystallographic community. Therefore, by growing crystals on the dialysis chip and introducing the reagents into the microfluidic channel, either manually (using a syringe) or automatically (with a pressure-control fluid system or a syringe-pump), future efforts will focus on proving that the microfluidic chips can be successfully used to trigger time-resolved experiments on synchrotron beamlines.
The authors have nothing to disclose.
MBS acknowledges the support of the MI / CNRS under the contract Instrumentation at the limits 2014-2015. NJ acknowledges CEA's International Doctoral Research Program (Irtelis) for the PhD Fellowship. MBS and SJ acknowledge funding from the European Union's Horizon 2020 Research and Innovation Programme under Marie Skłodowska-Curie grant agreement number 722687. MBS, SJ, and NJ thank LIPhy (UGA) for the clean room establishment for microfabrication experiments. IBS acknowledges integration into the Interdisciplinary Research Institute of Grenoble (IRIG, CEA).
3 in wafer | Silicon Materials Inc. | Silicon wafer | |
Centrifuge | Eppendorf | Minispin | Bench-top centrifuge |
CleWin 3.0 | WieWeb software | Designing software | |
Epoxy glue | Devcon | 5 minutes epoxy glue | |
Fluidic connectors | Cluzeau Info Lab | N-333 | NanoPort kit for 1/16" OD tubing |
Hen egg-white lysozyme | Roche | 10 837 059 001 | Lyophilized protein powder |
High-vacuum silicone grease | Sigma-Aldrich | Z273554 | Dow Corning high-vacuum silicone grease |
HMDS | Sigma-Aldrich | 440191 | Silane, chemical |
Hot plate | Sawatec | HP-200-Z-HMDS BM | Hot plate |
Isopropyl alcohol | Sigma-Aldrich | Solvent | |
Kapton tape | DuPont | Polyimide tape | |
Mask aligner | SUSS MicroTec | MJB4 | Mask aligner, UV source |
Membrane filter | Millipore | GSWP04700 | 0.22 μm pore size filter |
Microscope glass slide | Fisher Scientific | 12164682 | 3 x 1 in glass slides |
NOA81 | Norland Products Inc. | NOA81 | Photocurable resin |
Oven | Memmert | Oven | |
Parafilm | Sigma-Aldrich | P6543 | Parafilm M roll size 20 in. × 50 ft |
PDMS | Dow Corning | Sylgard 184 | Silicone |
PEG 400 | Hampton Research | HR2-603 | Chemical |
Petri dish | Sigma-Aldrich | P5731 | 100 x 15 mm |
PGMEA | Sigma-Aldrich | 484431 | Developer |
Plasma equipment | Diener Electronic | ZEPTO | Plasma treatment |
PMMA | Goodfellow | 137-745-63 | PMMA sheets 150×150 mm, 0.175 mm thickness |
Pressure driven system | Elveflow | OB1 MK3+ | Pressure/vacuum controller |
PTFE tubing | Elveflow/Darwin microfluidics | LVF-KTU-15 | PTFE tubing roll 1/16" OD X 1/32" ID |
RC dialysis membrane | Spectra/Por | Various MWCOs | |
Scalpel | Swann-Morton | Carbon steel surgical blades | |
Sodium acetate | Sigma-Aldrich | S2889 | Chemical |
Sodium chloride | Sigma-Aldrich | 746398 | Chemical |
Solidworks | Dassault Systemes | 3D-CAD designing software | |
Spin coater | SPS | Spin150 | Wafer spinner |
SU-8 3000 series | MicroChem Corp. | SU-8 3050 | Photoresist |
Syringe | BD | 309628 | 1 mL Luer-Lok syringe |
UV crosslinker | Uvitec | CL-508 | UV crosslinker |