This study introduces a method for the simultaneous recording of local field potentials in the brain, electrocardiograms, electromyograms, and breathing signals of a freely moving rat. This technique, which reduces experimental costs and simplifies data analysis, will contribute to the understanding of the interactions between the brain and peripheral organs.
Monitoring the physiological dynamics of the brain and peripheral tissues is necessary for addressing a number of questions about how the brain controls body functions and internal organ rhythms when animals are exposed to emotional challenges and changes in their living environments. In general experiments, signals from different organs, such as the brain and the heart, are recorded by independent recording systems that require multiple recording devices and different procedures for processing the data files. This study describes a new method that can simultaneously monitor electrical biosignals, including tens of local field potentials in multiple brain regions, electrocardiograms that represent the cardiac rhythm, electromyograms that represent awake/sleep-related muscle contraction, and breathing signals, in a freely moving rat. The recording configuration of this method is based on a conventional micro-drive array for cortical local field potential recordings in which tens of electrodes are accommodated, and the signals obtained from these electrodes are integrated into a single electrical board mounted on the animal’s head. Here, this recording system was improved so that signals from the peripheral organs are also transferred to an electrical interface board. In a single surgery, electrodes are first separately implanted into the appropriate body parts and the target brain areas. The open ends of all of these electrodes are then soldered to individual channels of the electrical board above the animal’s head so that all of the signals can be integrated into the single electrical board. Connecting this board to a recording device allows for the collection of all of the signals into a single device, which reduces experimental costs and simplifies data processing, because all data can be handled in the same data file. This technique will aid the understanding of the neurophysiological correlates of the associations between central and peripheral organs.
The central nervous system controls body states in response to various environmental changes, and this control is typically represented as changes in heart rate, breathing rate, and muscle contractions. However, few studies have tested how such peripheral physiological factors are associated with cortical activity. To address this issue, a large-scale recording method for monitoring electrical biosignals from both central and peripheral tissues is necessary. In the cerebral cortex, local field potential (LFP) signals are extracellularly recorded by electrodes that are inserted into the cortical tissues1,2,3. To simultaneously record multiple LFP signals from the cortical regions of small mammals, such as rats and mice, a number of studies have developed various types of custom-made electrode assemblies that are termed micro-drives. A conventional micro-drive is composed of metal screws attached to the middle parts of the electrodes (which are typically tetrodes), a core body that accommodates the screws and electrodes, and an electrical interface board (EIB) that accommodates metal holes to connect the open ends of the electrodes (Figure 1, Figure 2, and Figure 3). This electrode assembly enables the operator to control the depth of many electrodes inserted into the brain over the course of days to weeks, and allows the conducting of long-term chronic recordings of neuronal activity as the animal is challenged with various behavioral tasks. In the peripheral organs, heartbeat signals are recorded as electrocardiograms (ECGs) by a pair of electrodes that are implanted on or around the heart area4,5,6, and skeletal muscle signals are recorded as electromyograms (EMGs) with electrodes that are inserted into the muscle tissue7,8,9. The relationship between electrical signals of the olfactory bulb and breathing (BR) rhythm has been studied with single unit recordings10,11. In conventional recording systems, these signals from different tissues have been captured by independent recording devices, which means that an additional experimental system is required to precisely synchronize these multiple devices for simultaneous recordings of brain-body signals. This system was developed to overcome this issue. In this system, all electrical signals recorded from the peripheral organs, including ECGs, EMGs, and electrical signals from the olfactory bulb that reflect the breathing rhythm, are integrated into a single micro-drive array1,2,3, here termed an integrative micro-drive array. This system requires only one multi-channel recording device, and is applicable to any conventional micro-drive array. The advantages of this technique are that it does not require any special devices or trigger signals to match the recording time of multiple devices, and it allows for more convenient data processing, since all of the signals are recorded as similar data types. This technique will aid the understanding of the neurophysiological correlates of the associations between central and peripheral organs. This paper describes the procedures associated with the technique and presents representative datasets obtained from a rat.
All of the procedures involving animal subjects were performed according to the NIH guidelines for the care and use of animals.
1. Preparation of the Integrative Micro-Drive Array
2. Implantation of the ECG/EMG Electrodes
NOTE: Perform all surgical steps with aseptic technique using sterilized gloves and autoclaved instruments. For all steps involving the creation of an incision, sterilize the skin with 70% ethanol before, and cover the incision with surgical drapes.
3. Implantation of the Integrative Micro-Drive Array and the BR Electrodes
4. In Vivo Recordings
NOTE: All signals are amplified, sampled at 2 kHz, and band-pass filtered (0.1 – 500 Hz) except for unit activities (sampled at 30 kHz and band-pass filtered (500 – 6 kHz)).
This method can simultaneously capture bioelectrical signals from multiple organs that represent the neuronal activity of the brain, heart rate, breathing rhythm, and skeletal muscle contractions (Figure 1). Figure 4 provides representative recording data from a freely moving rat that was freely foraging in a rectangular box (25 × 40 cm2). The example dataset includes typical behavioral transitions between moving and resting states. A power spectrum was computed from a hippocampal LFP trace by wavelet analysis. The BR signal recorded from the surface area of the olfactory bulb was used to roughly estimate the relative changes in breathing frequencies, such as those that occur during exploratory sniffing behavior.
Figure 1: Illustration of the recording system for monitoring multiple brain-body signals from a freely moving rat. All bioelectrical signals (LFP, ECG, EEG, BR signals) from a freely moving rat are collected into the integrative-microdrive array mounted on the head. Please click here to view a larger version of this figure.
Figure 2: An integrative micro-drive array. All LFP, ECG, EEG, and BR signals are transmitted to the holes on the EIB as indicated by the arrows. The dotted area is magnified in the right panel and displays some of the tetrodes protruding from the micro-drive array that are inserted into the brain tissue. Please click here to view a larger version of this figure.
Figure 3: A top view of the EIB. The EIB includes 24 cortical LFP (LFP) channels that are connected to the tetrodes, 2 ECG channels, 2 EMG channels, 2 BR channels, and 2 ground (Gr) channels. All channels except the LFP channels were connected to insulated wires. Please click here to view a larger version of this figure.
Figure 4: An example of the simultaneous multi-channel recording of bioelectrical signals.
(From top to bottom) LFP signals in the somatosensory cortex (scale bar: 250 µV). LFP signals in the hippocampal CA1 region (scale bar: 500 µV). A color-coded power spectrum of the hippocampal LFP trace. An ECG signal (band-pass filtered at 20-200 Hz, scale bar: 500 µV). An EMG signal (band-pass filtered at 100-500 Hz, scale bar: 100 µV). A BR signal (scale bar: 500 µV). A color-coded power spectrum of the BR signal that indicates sniffing behavior as defined by a transient increase in breathing rate. Please click here to view a larger version of this figure.
For understanding how the brain modulates peripheral activity levels, and vice versa, large-scale recording methods to simultaneously capture electrical biosignals from multiple body areas are necessary. This study described a surgical procedure, and a recording system for monitoring cerebral local field potentials, heart rates, the magnitude of muscle construction and respiratory rates, which have been improved on a recording system that is used for extracellular recordings in brain tissue. This system collects electrical signals from both the brain and the peripheral organs into a single EIB on an integrative micro-drive array. The preparation of the integrative micro-drive array should be started at least several hours before surgery, as it takes some time. Here, the drive array includes tetrodes for recording brain local field potentials, but the other types of metal electrodes, such as platinum and tungsten electrodes, can be attached to the EIB if the ends of these electrodes are properly soldered. When following the surgical procedures in the protocol, experienced experimenters have been able to complete all the procedures within 2-3 h.
A critical step within this protocol is the positioning of the electrodes on the tissue, especially for the ECG and EMG electrodes. Several training repetitions might be required to obtain stable recordings. To date, all recordings have been stable at least one month after surgery. Experimenters should note that, if the signal-to-noise ratios of the recorded data become low, this problem is frequently due to loose fixations of the ground/reference electrodes, or insufficient soldering of the ends of wires between the EIBs or the other ends of wires. Compared with a conventional electrophysiological recording using some independent devices, the advantages of this method are that (1) it is technically simple to conduct, if trained several times, (2) it does not require a communication system to synchronize multiple devices, (3) it reduces the overall experimental costs, as only one recording device is required, and (4) all recorded data files can be handled by the same processing method and program, which increases the efficiency of data analysis. Furthermore, the methodological concept is applicable to many combinations of multiple brain regions and peripheral organs, including the respiratory system, circulatory system, and autonomic nervous system. It is also applicable to any commercially available electrical recording device. Simultaneous monitoring of systemic physiological activity patterns using this method will be helpful in elucidating neuronal activity patterns in various physiological states against emotional challenges, external sensory modulation, and pathological diseases, leading to an increased understanding of the biological mechanisms underlying the brain-body association.
The authors have nothing to disclose.
This work was supported by Kaken-hi (17H05939; 17H05551), the Nakatomi Foundation, and the Suzuken Memorial Foundation.
FEP Hookup Wire Stranded Stainless Steel | Cooner Wire Company, Chatsworth, CA | AS 633 | Bioflex wire |
EIB-36-PTB | Neuralynx, Inc., Bozeman, MT | EIB-36-PTB | EIB |
Cereplex M | Blackrock Microsystems, Salt Lake City, UT | Digital headstage | |
Cereplex Direct | Blackrock Microsystems, Salt Lake City, UT | Data acquisition system | |
UEW polyurethane magnet wire | Oyaide.com, Tokyo, Japan | UEW 0.14mm 20m | Enamel wire |
SD-102 | Narishige, Tokyo, Japan | SD-102 | High-speed drill |
Minimo ONE SERIES ver.2 | Minitor Co.,Ltd, Tokyo, Japan | C2012 | High-peed drill Power Supply |
Provinice 250 mL | Shofu Inc., Kyoto, Japan | 213620136 | Dental cement |
Small Animal Anesthetizer | Biomachinery, Chiba, Japan | TK-7 | Anesthetizer |
Buprenorphine hydrochloride | Sigma-Aldrich, St. Louis, MO | B7536-1ML | Analgesic |
Isoflurane | DS Pharma Animal Health, Osaka, Japan | Isoflu 250mL | |
Vaseline, White | Wako Pure Chemical Industries, Ltd., Osaka, Japan | 224-00165 | Vet ointment |
Sodium alginate | Nacalai tesque, Kyoto, Japan | 31131-85 | |
Calcium Chloride Dihydrate | Wako Pure Chemical Industries, Ltd., Osaka, Japan | 031-00435 | |
Stainless steel screw M1.0×4.0 | MonotaRO, Hyogo, Japan | 42617504 | Stainless steel screw for BR electrodes |
Stainless steel screw M1.4×3.0 | MonotaRO, Hyogo, Japan | 42617687 | Stainless steel screw for g/r electrodes and anchors |