概要

Developing a Salivary Antibody Multiplex Immunoassay to Measure Human Exposure to Environmental Pathogens

Published: September 12, 2016
doi:

概要

In the current climate of scarce resources, new technologies are emerging that allow researchers to conduct studies cheaper, faster and with more precision. Here we describe the development of a bead-based salivary antibody multiplex immunoassay to measure human exposure to multiple environmental pathogens simultaneously.

Abstract

The etiology and impacts of human exposure to environmental pathogens are of major concern worldwide and, thus, the ability to assess exposure and infections using cost effective, high-throughput approaches would be indispensable. This manuscript describes the development and analysis of a bead-based multiplex immunoassay capable of measuring the presence of antibodies in human saliva to multiple pathogens simultaneously. Saliva is particularly attractive in this application because it is noninvasive, cheaper and easier to collect than serum. Antigens from environmental pathogens were coupled to carboxylated microspheres (beads) and used to measure antibodies in very small volumes of human saliva samples using a bead-based, solution-phase assay. Beads were coupled with antigens from Campylobacter jejuni, Helicobacter pylori, Toxoplasma gondii, noroviruses (G I.1 and G II.4) and hepatitis A virus. To ensure that the antigens were sufficiently coupled to the beads, coupling was confirmed using species-specific, animal-derived primary capture antibodies, followed by incubation with biotinylated anti-species secondary detection antibodies and streptavidin-R-phycoerythrin reporter (SAPE). As a control to measure non-specific binding, one bead set was treated identically to the others except it was not coupled to any antigen. The antigen-coupled and control beads were then incubated with prospectively-collected human saliva samples, measured on a high throughput analyzer based on the principles of flow cytometry, and the presence of antibodies to each antigen was measured in Median Fluorescence Intensity units (MFI). This multiplex immunoassay has a number of advantages, including more data with less sample; reduced costs and labor; and the ability to customize the assay to many targets of interest. Results indicate that the salivary multiplex immunoassay may be capable of identifying previous exposures and infections, which can be especially useful in surveillance studies involving large human populations.

Introduction

Eighty-eight percent of diarrhea-related illness worldwide is associated with human exposure to contaminated water, unsafe food, and poor sanitation/hygiene, causing approximately 1.5 million deaths, the majority of whom are children1. This is a major cause of concern for public health officials and policy makers. In an effort to investigate exposures and illnesses associated with waterborne and other environmental pathogens, we developed a multiplex immunoassay to measure antibodies in human samples2-4. This method can be applied to epidemiological studies to determine human exposure to these pathogens and to better define immunoprevalence and incident infections.

Saliva holds considerable promise as an alternative to serum for human biomarker research. Among the advantages of using saliva are the non-invasiveness and ease of sample collection, low cost, and samples can easily be collected from children5-7. Serum and saliva samples have been studied extensively for antibodies against H. pylori2,3,8, Plasmodium falciparum9, Entamoeba histolytica10, Cryptosporidium parvum3,11, Streptococcus pneumonia12, hepatitis viruses A and C13-14, noroviruses2-4,15, T. gondii2-4, dengue virus16, human immunodeficiency virus (HIV)17, and Escherichia coli O157:H718.

A multiplex immunoassay allows for the analysis of multiple analytes simultaneously within a single sample volume and within a single cycle or run. Multiplexed antigens from C. jejuni, T. gondii, H. pylori, hepatitis A virus, and two noroviruses were used to measure human salivary IgG2-4 and IgA3,4 and plasma IgG2,3 antibody responses to these pathogens using a bead-based multiplexing immunoassay. When used in conjunction with epidemiological studies of exposure to microbes in water, soil and food, the type of assay described in this study may provide valuable information to enhance the understanding of infections caused by environmental pathogens. Moreover, salivary antibody data obtained from such studies can be used to improve risk assessment models19-22.

Protocol

Approval was obtained from the Institutional Review Board (IRB # 08-1844, University of North Carolina, Chapel Hill, NC, USA) for the collection of stimulated crevicular saliva samples from beachgoers at Boquerón Beach, Puerto Rico, as part of the United States Environmental Protection Agency (USEPA) National Epidemiological and Environmental Assessment of Recreational (NEEAR) Water Study23 to assess swimming associated exposures and illnesses. Study subjects provided informed consent and were instructed on the use of the saliva collection device by trained USEPA contractors. The saliva samples were shipped on ice and, upon receipt, they were centrifuged and stored at -80 °C as described4.

1. Bead Activation

  1. Resuspend bead set stocks by vortexing and sonicating for 20 sec and transfer approximately 5.0 x 106 of the stock beads (400 µl) to microcentrifuge tubes.
    Note: The beads are supplied at a concentration of 12.5 x 106 beads/ml.
  2. Pellet the stock beads by centrifuging at 10,000 x g for 2 min.
  3. Remove the supernatant and resuspend the pelleted beads in 100 µl distilled water (dH2O) by vortex and sonication for 20 sec. Repeat step 1.2.
  4. Remove the supernatant and resuspend the washed beads in 80 µl of 100 mM sodium phosphate monobasic, pH 6.2 by vortex and sonication for 20 sec.
  5. Immediately before use, make a 50 mg/ml N-hydroxysulfosuccinimide (Sulfo-NHS) solution by adding 200 µl dH2O to the 10 mg Sulfo-NHS aliquot. Mix by vortex.
  6. Add 10 µl of the 50 mg/ml Sulfo-NHS to the beads. Mix by vortex.
  7. Immediately before use, make a 50 mg/ml 1-ethyl-[3dimethylaminopropyl] carbodiimide hydrochloride (EDC) solution by adding 200 µl distilled water (dH2O) to the 10 mg EDC aliquot. Mix by vortex.
  8. Add 10 µl of 50 mg/ml EDC solution to the beads. Mix by vortex. Incubate the beads for 20 min at room temperature, in the dark, with mixing by vortex at 10 min intervals. Pellet the activated beads by microcentrifugation at 10,000 x g for 2 min.
  9. Remove the supernatant and resuspend the beads in 250 µl of 50 mM 2-[N-Morpholino]ethanesulfonic acid (MES), pH 5.0 by vortex and sonication for 20 sec.
  10. Pellet the activated beads by microcentrifugation at 10,000 x g for 2 min.
  11. Repeat steps 1.9 and 1.10.
    Note: This provides a total of two washes with 50 mM MES, pH 5.0.
  12. Resuspend the beads in 100 µl of 50 mM MES, pH 5.0 by vortex and sonication for 20 sec.

2. Bead Coupling

  1. Couple the antigens to the bead sets using the concentrations shown in Table 1.
  2. Add each antigen to the activated beads and bring the total volume to 500 µl in 50 mM MES, pH 5.0. Mix the antigens and beads by vortex.
  3. Incubate the antigens and beads for 2 hr with mixing by rotation (~15 rpm) at room temperature in the dark. Pellet the coupled beads by microcentrifugation at 10,000 x g for 2 min.
  4. Remove the supernatant and resuspend the beads in 500 µl of phosphate buffered saline (PBS)-bovine serum albumin (BSA)-polyoxyethylenesorbitan monolaurate (Tween-20)-sodium azide (PBS-TBN) pH 7.4 by vortex and sonication. Pellet the beads by microcentrifugation at 10,000 x g for 2 min and remove the supernatant.
    Caution: Sodium azide is an acutely toxic chemical. It is fatal if swallowed or gets in contact with skin. Do not breathe dust/fume/gas/mist/vapors or spray. Wear appropriate personal protective equipment (PPE's) when handling and dispose of in accordance with appropriate laws.
  5. Resuspend the beads in 1 ml of PBS-TBN by vortex and sonication for 20 sec.
  6. Pellet the beads by microcentrifugation at 10,000 x g for 2 min.
  7. Repeat steps 2.5 and 2.6.
    Note: This provides a total of two washes with PBS-TBN.
  8. Resuspend the coupled and washed beads in 1 ml of PBS, 1% BSA, 0.05% Azide, pH 7.4. Store the coupled beads in a 2-8 °C refrigerator in the dark.

3. Bead Count

  1. Prepare a 1:10 dilution of the coupled beads in water or PBS buffer.
  2. Load 10 µl of the bead dilution onto a hemocytometer at the sample introduction point.
  3. Count the beads seen in one of the 4 x 4 corner grids. Calculate the total number of coupled beads using the following formula: Count (1 corner of 4 x 4 grid) x (1 x 104) x (dilution factor) x resuspension volume in ml.

4. Confirmation of Antigen Coupling

  1. Resuspend the stock mixture of beads coupled to antigens of interest by vortex and sonication for 20 sec.
  2. Prepare a working bead mixture by diluting the coupled bead stocks to a final concentration of 100 beads /µl of each unique bead set in PBS-1% BSA buffer (PBS-1% BSA, pH 7.4).
  3. Prepare at least 7 two-fold serial dilutions of anti-species IgG primary antibody according to manufacturer's recommendations in 96-well round bottom plate with PBS-1% BSA buffer.
  4. Pre-wet a separate 8-well column (8 rows) of a 96-well filter bottom plate for each antigen coupling confirmation test with 100 µl of wash buffer and remove supernatant by vacuum. Add 50 µl of working bead mixture (antigen-coupled beads) to the pre-wet wells.
  5. Add 50 µl of antibody dilutions to rows 1-7 of each column of the 96-well filter plate and 50 µl of PBS-1% BSA buffer to row 8 in lieu of diluted antibody to serve as background wells. Mix with a multi-channel pipettor by pipetting up and down 5 times. Perform the same procedure for every antigen-coupled bead set to be confirmed.
  6. Cover and allow to incubate in the dark at room temperature for 1 hr on a microplate shaker at 500 rpm. Remove supernatant by vacuum.
  7. Wash wells with 100 µl of wash buffer and remove supernatant by vacuum. Repeat 1x. Resuspend the beads in 50 µl of PBS-1% BSA with a multi-channel pipettor.
  8. Dilute biotinylated anti-species specific IgG secondary detection antibody to 16 µg/ml in PBS-1% BSA.
  9. Add 50 µl of diluted secondary antibody to each well by pipetting up and down 5 times.
  10. Cover filter plate and allow to incubate in the dark at room temperature for 30 min on a plate shaker. Remove supernatant by vacuum. Wash wells with 100 µl of wash buffer and remove supernatant by vacuum. Repeat wash 1x.
  11. Resuspend the beads in 50 µl of PBS-1% BSA with a multi-channel pipettor.
  12. Dilute streptavidin-R-phycoerythrin reporter (SAPE) to 24 µg/ml in PBS-1% BSA.
  13. Add 50 µl of reporter to each well and mix by pipetting up and down 5 times.
  14. Cover filter plate and allow to incubate in the dark at room temperature for 30 min on a plate shaker. Remove supernatant by vacuum. Wash wells with 100 µl of wash buffer and remove supernatant by vacuum. Repeat wash 1x.
  15. Resuspend beads in 100 µl of PBS-1% BSA and analyze 50 µl using the analyzer29.
    Note: Results of the bead-based multiplex immunoassay are measured in Median Fluorescence Intensity (MFI). Always refer to the latest version of the software manual, if available to avoid errors.

5. Salivary Multiplex Immunoassay

  1. Remove saliva from the -80 °C freezer and allow to thaw at room temperature.
  2. Resuspend antigen coupled bead stocks by vortex and sonication for 20 sec.
  3. Prepare a working bead mixture by diluting the coupled bead stocks to a final concentration of 100 beads/µl of each unique bead set in PBS-1% BSA buffer.
  4. Prepare a 1:4 dilution of saliva with PBS-1% BSA buffer in a 96 well, deep well plate.
  5. Pre-wet filter plate with 100 µl of wash buffer and remove supernatant by vacuum.
  6. Add 50 µl of a working bead mixture and an equal volume of diluted saliva to 95 wells of the 96 well filter plates for a 1:8 final dilution. Mix reactions with a multi-channel pipettor. To the one control well, add 50 µl antigen-coupled beads plus 50 µl of PBS-1% BSA buffer (as a replacement for saliva).
  7. Cover and allow to incubate in the dark at room temperature for 1 hr on a microplate shaker at 500 rpm. Remove supernatant by vacuum. Wash wells with 100 µl of wash buffer and remove supernatant by vacuum. Repeat wash 1x.
  8. Resuspend beads in 50 µl of PBS-1% BSA with a multi-channel pipettor.
  9. Dilute biotinylated goat anti-human IgG secondary detection antibody to 16 µg/ml in PBS-1% BSA.
  10. Add 50 µl diluted secondary antibody to each well and mix contents with a multi-channel pipettor.
  11. Cover filter plate and allow to incubate in the dark at room temperature for 30 min on a plate shaker. Remove supernatant by vacuum. Wash wells with 100 µl of wash buffer and remove supernatant by vacuum. Repeat wash 1x.
  12. Resuspend beads in 50 µl of PBS-1% BSA with a multi-channel pipettor.
  13. Dilute streptavidin-R-phycoerythrin reporter (SAPE) to 24 µg/ml in PBS-1% BSA.
  14. Add 50 µl reporter to each well and mix with a multi-channel pipettor.
  15. Cover filter plate and allow to incubate in the dark at room temperature for 30 min on a plate shaker. Remove supernatant by vacuum. Wash wells with 100 µl of wash buffer and remove supernatant by vacuum. Repeat wash 1x.
  16. Resuspend beads in 100 µl of PBS-1% BSA and analyze 50 µl using the analyzer29.
    Note: Results of the multiplex immunoassay are measured in Median Fluorescence Intensity (MFI) units. Always refer to the latest version of the software manual, if available to avoid errors.

Representative Results

One unique bead set was used as a control to measure non-specific binding and sample to sample variability. These beads were treated identically to the antigen coupled beads with the exception that they were not incubated with any antigen in the coupling step. MFI values >500 obtained from the control beads incubated with all saliva samples were removed from further analyses due to suspected contamination from serum and the remaining responses were log distributed. The saliva can be contaminated with serum if the gums are rubbed too vigorously with the sponge attached to the collection device or if the study participant has periodontal disease. The log transformed MFI data were used to calculate an immunopositive cut-off point of 505 MFI based on the mean plus 3 standard deviations of the log MFI values. Additionally, antigen-coupled beads were added to specific wells that were treated as test wells in every manner except diluted saliva was replaced with PBS-1% BSA to evaluate background fluorescence and cross-reactivity. The MFI values obtained in these background wells were subtracted from the MFI values from each antigen-coupled bead set for each saliva sample.

Bead coupling was confirmed using animal derived species-specific primary detection antibodies specific for each antigen. To ensure that the antigen coupled beads were capable of approaching the dynamic range of the assay, we defined proper coupling as an MFI ≥18,000, observation of dose response and cross-reactivity between primary antibodies and non-targets of <10% as described2. Representative coupling confirmations for C. jejuni and T. gondii of the antigens in the assay are shown in Figure 1.

Based on the cut-off point established (505 MFI), the immunoprevalence rates for the samples (n = 2,078) ranged from about 2% (n = 41) for T. gondii to nearly 50% (n = 1,009) for norovirus GI.1 (Figure 2). The data indicate that the immunoprevalence rate was highest for the noroviruses followed by hepatitis A virus and H. pylori.

While 32% (n = 672) of the samples were immunonegative for all of the pathogens in the assay, 68% (n = 1,406) were immunopositive to one or more pathogens. Figure 3 shows the breakdown of immunopositivity to one or more of the pathogens.

Figure 1
Figure 1: Coupling confirmation analyses for two of the organisms in the multiplex immunoassay. Coupling confirmation was performed for all of the antigens and the graphs for C. jejuni and T. gondii in this figure are representative of the coupling confirmations for the antigens in the multiplex immunoassay. Please click here to view a larger version of this figure.

Figure 2
Figure 2: Immunoprevalence to specific pathogens. Immunoprevalence to specific pathogens in the multiplex immunoassay ranged from about 2% for T. gondii to nearly 50% for norovirus GI.1. Antibodies against the norovirus antigens were more commonly observed followed by HAV (hepatitis A virus) and H. pylori. T. gondii and C. jejuni antibodies appeared less frequently in the saliva samples analyzed. Please click here to view a larger version of this figure.

Figure 3
Figure 3: Breakdown of immunopositivity to multiple pathogens simultaneously. The multiplex immunoassay allows for the analysis of immunopositivity to multiple pathogens simultaneously in a 50 µl sample volume. Almost a third of the saliva samples assayed was immunonegative for antibodies against all of the antigens in the multiplex. Approximately 31% of the samples was immunopositive for one antigen while a quarter was positive to two antigens. 9.4% was immunopositive to three antigens. 3% was positive to four or more antigens simultaneously. Please click here to view a larger version of this figure.

Antigen Bead Set Coupling buffer Ag Concentration (µg) # of beads in 1 corner # of beads/µl Vol. for 100 B/uL for 4 ml (µl)
C. jejuni 8 MES 5.0 50 64 6,400 94
H. pylori 33 MES 5.0 25 71 7,100 85
Hepatitis A 42 MES 5.0 100 91 9,100 66
T. gondii 30 MES 5.0 25 85 8,500 71
Norovirus GII.4 55 MES 5.0 5 72 7,200 83
Norovirus GI.1 67 MES 5.0 5 63 6,300 95
Uncoupled 80 MES 5.0 60 6,000 100
Total volume of beads (µl) 594
Total volume of buffer needed (µl) 3,406

Table 1: Working bead mix for the multiplex immunoassay. After coupling and confirmation were completed, a master mix consisting of each bead set was prepared as shown. The master mix was distributed among all of the wells in the 96 well plate. All of the wells were incubated with saliva with the exception of one background control well which contained only beads and PBS-1% BSA buffer.

Discussion

These results indicate that the multiplex immunoassay method is useful for discriminating between saliva samples that are immunopositive or immunonegative. To determine immunopositivity, a single cut-off point was developed by calculating the mean plus three standard deviations of the log transformed MFI responses of the control uncoupled beads tested with all of the saliva samples. The cut-off point afforded the ability to assess exposure and immunoprevalence to either a single or multiple pathogens. This discriminative power may be useful in population studies to investigate health risks associated with exposure to environmental and other pathogens.

There are several other methods that can be used to determine a cut-off point depending on the distribution of the MFIs, uncoupled controls and what question the study is designed to answer. Some examples for determining cut-offs include finite mixed modeling24-26, mean plus 3 standard deviations of responses to controls, mean plus 2 standard deviations of responses to controls, and three times the mean of responses to controls27,28. For simple screenings, less stringent criteria may be used but for epidemiological studies, it may be more appropriate to employ a more stringent definition to reduce the probability of reporting false positives. Our initial application is to look at immunoconversion and immunoprevalence between swimmers and non-swimmers. In this study, the uncoupled control MFIs were not normally distributed and the results were log transformed.

Advantages to performing a bead-based multiplex immunoassay include the use of very low sample volumes, reduction in time and labor and the ability to measure responses for up to 500 analytes simultaneously while requiring a minimal amount of reagents. By contrast, traditional methods of assessing antibody levels indicative of exposure and/or infection (e.g., ELISA testing) can take several hours to complete, are labor intensive, and can only be used to measure one analyte at a time. These traditional methods also require larger volumes of patient/participant samples. For example, most ELISA's require at least 100 µl of sample while a multiplex immunoassay may require 50 µl or less.

Limitations of a multiplex immunoassay include the need for rigorous optimization to determine the optimal concentrations of primary capture and secondary detection antibodies, antigens and reporter. Cross-reactivity is also a major concern in immunoassays as antibodies against a particular antigen may cross-react with antigens from other organisms and thereby lead to false positive readings. Optimization, cross-reactivity and non-specific binding were addressed previously2-4. The success of any such assay is largely dependent on the ability to obtain highly immunogenic antigens as well as specific antibodies to these antigens for use in the bead coupling and coupling confirmation steps. Another limitation is the limited availability of characterized (diagnostically positive and negative) samples to validate the assays.

There are a number of critical steps within the protocol. These include: ensuring the antigens that will be coupled to the beads do not contain foreign proteins, azide, glycine, Tris or any primary amines. If any of these agents exist in the antigen preparation, they should be removed by dialysis or chromatography. It is recommended that one should start with 5 µg of antigen per 5 million beads and titrate to find the optimal concentration. Choose the optimal detection antibody concentration that gives the best sensitivity and dynamic range. Bear in mind that signals tend to decrease as antigen and detection antibody concentrations increase (hook effect). For example, if the signal decreases as antigen concentration increases then detection antibody may be limiting. Conversely, if signals decrease as detection antibody increases then reporter (SAPE) may be limiting. Check the multiplex for cross- reactivity by combining the optimal coupling for each protein (5,000 of each bead set per well). Test the multiplexed bead sets by combining with the optimal detection antibody amount. Optimize the reporter and make a standard curve of the antigen by testing each antigen individually.

Troubleshooting a multiplex immunoassay to improve sensitivity and to increase median fluorescent signal are critically important. To improve sensitivity, decrease either the amount of antigen coupled to the beads or the detection antibody concentration. The use of a higher affinity antibody for capture and/or detection may also improve sensitivity. Increasing the amount of antigen coupled to the beads may increase the median fluorescent signal. One study demonstrated that pre-incubating serum samples with polyvinylalcohol plus polyvinylpyrrolidone (PVX) buffer is able to suppress non-specific binding in serological assays using the bead-based multiplex assay30 such as the one that we are describing here. However, another study by our collaborators found no significant effect between PBS-BSA and PVX buffers. They concluded further that because of the higher viscosity of the PVX buffer, it created problems with bead acquisition in the analyzer and formed foam in the microplate wells3.

In conclusion, we have presented a method that is capable of measuring the presence of human salivary IgG antibodies to multiple pathogens simultaneously in a very small sample volume. In the future, this assay will be used to detect the presence of salivary antibodies associated with swimming related exposures or infections and to help inform risk assessment models. Additionally, the assay can be used to measure antibodies associated with airborne and foodborne exposures, determine incident infections or immunoconversions and provide critical immunological information to epidemiologists conducting questionnaire-type population studies.

開示

The authors have nothing to disclose.

Acknowledgements

Clarissa Curioso was supported through an appointment to the Research Participation Program at the U.S. Environmental Protection Agency administered by the Oak Ridge Institute for Science and Education through an interagency agreement between the U.S. Department of Energy and U.S. EPA.

Materials

Equipment and Software
Microcentrifuge Thermo Electron Corporation 75002446 Used to centrifuge samples
Vortex Mixer VWR G560 Used to mix samples
Sonicator (mini) Fisher Scientific 15-337-22 Used to separate beads
Pipettors P10, P20, P100, P1000, 8 ch. Capp Various
Hemacytometer (Bright Line) Housser Scientific  3200 Used to count coupled beads
Multiscreen Vacuum Manifold Millipore MSVMHTS00 Used in washing steps to remove supernatant
MicroShaker VWR 12620-926 Used to agitate beads during incubations
Tube rack (1.5mL and 0.5mL) (assorted) VWR 30128-346
Weighing Scale Mettler or other Used to measure wash reagents for making buffers
Dynabead Sample Mixer Invitrogen 947-01 Used during coupling incubation step
MatLab (R2014b) The MathWorks, Inc. Used to analyze antibody response data
Microsoft Excel 2014 Microsoft Corporation Used to analyze antibody response data
Luminex Analyzer with xPonent 3.1 software Luminex Corporation LX200-XPON3.1 Instrument and software used to run assay
Antigens
GI.1 Norwalk Virus : p-particle Xi Jiang (CCHMC)* NA  *Cincinnati Childrens' Hospital. Final conc. 5 µg.
GII.4 Norovirus VA387 : p-particle Xi Jiang (CCHMC)* NA  *Cincinnati Childrens' Hospital. Final conc. 5 µg.
Hepatitis A Virus : grade II concentrate from cell culture Meridian Life Sciences 8505 Antigen coupled at 100 µg
Helicobacter pylori : lysate Meridian Life Sciences R14101 Antigen coupled at 25 µg
Toxoplasma gondii : recombinant p30 (SAG1) Meridian Life Sciences R18426 Antigen coupled at 25 µg
Campylobacter jejuni : heat killed whole cells KPL 50-92-93 Antigen coupled at 50 µg
Primary Antibodies
Guinea pig anti-Norovirus (CCHMC)* NA Used for coupling confirmation
Mouse anti-Hepatitis A IgG Meridian Life Sciences C65885M Used for coupling confirmation
Mouse anti-Hepatitis A IgG Meridian Life Sciences C65885M Used for coupling confirmation
BacTraceAffinity Purified Antibody to Helicobacter pylori KPL 01-93-94 Used for coupling confirmation
Goat pAb to Toxoplasma gondii Abcam Ab23507 Used for coupling confirmation
BacTrace Goat anti-Campylobacter species KPL 01-92-93 Used for coupling confirmation
Secondary  Antibodies
Biotin-SP-Conjugated AffiniPure Donkey anti-Goat IgG (H+L) Jackson 705-065-149 Used for coupling confirmation
Biotinylated Rabbit anti-Goat IgG (H+L) KPL 16-13-06 Used for coupling confirmation
Biotinylated Goat anti-Mouse IgG (H+L) KPL 16-18-06 Used for coupling confirmation
Affinity Purified Antibody Biotin Labeled Goat anti-Rabbit IgG(H+L) KPL 176-1506 Used for coupling confirmation
Affinity Purified Antibody Biotin Labeled Goat anti-Human IgG(ᵞ)  KPL 16-10-02 Used for Salivary Immunoassay
Consumables
1.5 mL copolymer microcentrifuge tubes USA Scientific 1415-2500 Used as low binding microcentrifuge tubes
10 µL pipette tip refills BioVentures 5030050C
200 µL pipette tip refills BioVentures 5030080C
1000 µL pipette tip refills BioVentures 5130140C
Aluminum foil Various Vendors Used keep beads in the dark during incubations
Deep Well plates VWR 40002-009 Used for diluting saliva samples
Multiscreen Filter Plates Millipore MABVN1250 Used to run assays
Oracol saliva collection system Malvern Medical Developments Limited Used for saliva collection
Reagents
Carboxylated microspheres (beads) Luminex Corporation Dependent on bead set Antigens are coupled to the microspheres
EDC (1-ethyl-3-[3dimethylaminopropyl] carbodiimide hydrochloride) Pierce 77149 or 22980 Used in bead activation
Sulfo-NHS (N-hydroxysulfosuccinimide) Pierce 24510 Used in bead activation
Steptavidin-R-phycoerythrin (1mg/mL) Molecular Probes S-866 Used as reporter
MES (2-[N-Morpholino]ethanesulfonic acid) Sigma M-2933 Used for coupling
Tween-20 (Polyoxyethylenesorbitan monolaurate) Sigma P-9416 Used in wash buffer to remove non-specific binding
Protein Buffers
PBS-TBN Blocking/ Storage Buffer (PBS, 0.1% BSA, 0.02% Tween-20, 0.05% Azide, pH 7.4)** Filter Sterilize and store at 4°C
PBS, pH 7.4 Sigma P-3813 138 mM NaCl, 2.7 mM KCl
BSA Sigma A-7888 0.1% (w/v)
Tween-20 Sigma P-9416 0.2% (v/v)
Sodium Azide (0.05% azide)** Sigma S-8032 **Caution: Sodium azide is acutely toxic. Avoid contact with skin and eyes. Wear appropriate PPE's. Dispose of according to applicable laws.
MES/ Coupling Buffer (0.05 M MES, pH 5.0)
MES (2-[N-Morpholino]ethanesulfonic acid) Sigma S-3139
5 N NaOH Fisher SS256-500
Assay Buffer (PBS, 1% BSA, pH 7.4)  Filter Sterilize and store at 4°C
PBS, 1% BSA, pH 7.4 Sigma P-3688 138 mM NaCl, 2.7 mM KCl, 1% BSA
Activation Buffer (0.1 M NaH2PO4, pH 6.2) Filter Sterilize and store at 4°C
NaH2PO4 (Sodium phosphate, monobasic anhydrous) Sigma S-3139 0.1M NaH2PO4
5 N NaOH Fisher SS256-500
Wash Buffer (PBS, 0.05% Tween-20, pH 7.4) Filter Sterilize and store at 4°C
PBS, 0.05% Tween-20, pH 7.4 Sigma P-3563 138 mM NaCl, 2.7 mM KCl, 0.05% TWEEN

参考文献

  1. Prüss-Üstün, A., Gore, F., Bartram, J. . Safer water, better health: costs, benefits and sustainability of interventions to protect and promote health. , (2008).
  2. Augustine, S., et al. Statistical approaches to developing a multiplex immunoassay for determining human exposure to environmental pathogens. J Immunol Methods. 425, 1-9 (2015).
  3. Griffin, S., et al. Application of salivary antibody immunoassays for the detection of incident infections with Norwalk virus in a group of volunteers. J Immunol Methods. 424, 53-63 (2015).
  4. Griffin, S., Chen, I., Fout, G., Wade, T., Egorov, A. Development of a multiplex microsphere immunoassay for the quantitation of salivary antibody responses to selected waterborne pathogens. J Immunol Methods. 364 (1-2), 83-93 (2011).
  5. Ferguson, D. Current diagnostic uses of saliva. J Dent Res. 66, 420-424 (1987).
  6. Mandel, I. The diagnostic uses of saliva. J Oral Pathol Med. 19, 119-125 (1990).
  7. Malamud, D. Saliva as a Diagnostic Fluid. Brit Med J. 305, 207-208 (1992).
  8. Ballam, L., et al. Western blotting is useful in the salivary diagnosis of Helicobacter pylori infection. J Clin Pathol. 53, 314-317 (2000).
  9. Estevez, P., Satoguina, J., Nwakanma, D., West, S., Conway, D., Drakeley, C. Human saliva as a source of anti-malarial antibodies to examine population exposure to Plasmodium falciparum. Malar J. 10, 104 (2011).
  10. Abd-Alla, M., Jackson, T., Reddy, S., Ravdin, J. Diagnosis of invasive amebiasis by enzyme-linked immunosorbent assay of saliva to detect amebic lectin antigen and anti-lectin immunoglobulin G antibodies. J Clin Microbiol. 38 (6), 2344-2347 (2000).
  11. Toyoguchi, A., et al. Antibody reactivity to Cryptosporidium parvum in saliva of calves after experimental infection. J Vet Med Sci. 62 (11), 1231-1234 (2000).
  12. Choo, S., Zhang, Q., Seymour, L., Akhtar, S., Finn, A. Primary and booster salivary antibody responses to a 7-valent pneumococcal conjugate vaccine in infants. J Infect Dis. 182 (4), 1260-1263 (2000).
  13. Tourinho, R., et al. Importance of the cutoff ratio for detecting antibodies against hepatitis A virus in oral fluids by enzyme immunoassay. J Virol Methods. 173 (2), 169-174 (2011).
  14. Moorthy, M., Daniel, H., Kurian, G., Abraham, P. An evaluation of saliva as an alternative to plasma for the detection of hepatitis C virus antibodies. Indian J Med Microbiol. 26 (4), 327-332 (2008).
  15. Moe, C., Sair, A., Lindesmith, L., Estes, M., Jaykus, L. Diagnosis of norwalk virus infection by indirect enzyme immunoassay detection of salivary antibodies to recombinant norwalk virus antigen. Clin Diagn Lab Immunol. 11 (6), 1028-1034 (2004).
  16. Yap, G., Sil, B., Ng, L. Use of saliva for early dengue diagnosis. PLoS Negl Trop Dis. 5 (5), e1046 (2011).
  17. Schramm, W., Angulo, G., Torres, P., Burgess-Cassler, A. A simple saliva-based test for detecting antibodies to human immunodeficiency virus. Clin Diagn Lab Immunol. 6 (4), 577-580 (1999).
  18. Chart, H., Perry, N., Willshaw, G., Cheasty, T. Analysis of saliva for antibodies to the LPS of Escherichia coli O157 in patients with serum antibodies to E. coli O157 LPS. J Med Microbiol. 52 (Pt 7), 569-572 (2003).
  19. Ashbolt, N., Bruno, M. Application and refinement of the WHO risk framework for recreational waters in Sydney, Australia. J Water Health. 1 (3), 125-131 (2003).
  20. Westrell, T., Schonning, C., Stenstrom, T., Ashbolt, N. QMRA (quantitative microbial risk assessment) and HACCP (hazard analysis and critical control points) for management of pathogens in wastewater and sewage sludge treatment and reuse. Water Sci Technol. 50 (2), 23-30 (2004).
  21. Ashbolt, N. Microbial contamination of drinking water and disease outcomes in developing regions. Toxicology. 198 (1-3), 229-238 (2004).
  22. Eisenberg, J., Soller, J., Scott, J., Eisenberg, D., Colford, J. A dynamic model to assess microbial health risks associated with beneficial uses of biosolids. Risk Anal. 24, 221-236 (2004).
  23. Wade, T. J., et al. . Report on 2009 National Epidemiologic and Environmental Assessment of Recreational Water Epidemiology Studies, US EPA. US EPA Report Number: EPA/600/R-10/168: US EPA2011. , (2011).
  24. Fujii, Y., et al. Serological surveillance development for tropical infectious diseases using simultaneous microsphere-based multiplex assays and finite mixture models. PLoS Negl Trop Dis. 8, e3040 (2014).
  25. Rota, M., Massari, M., Gabutti, G., Guido, M., De Donno, A., Ciofi degli Atti, M. Measles serological survey in the Italian population: interpretation of results using mixture model. Vaccine. 26 (34), 4403-4409 (2008).
  26. Vyse, A., Gay, N., Hesketh, L., Pebody, R., Morgan-Capner, P., Miller, P. Interpreting serological surveys using mixture models: the seroepidemiology of measles, mumps and rubella in England and Wales at the beginning of the 21st century. Epidemiol Infect. 134 (6), 1303-1312 (2006).
  27. Baughman, A., et al. Establishment of diagnostic cutoff points for levels of serum antibodies to pertussis toxin, filamentous hemagglutinin, and fimbriae in adolescents and adults in the United States. Clin Diagn Lab Immunol. 11 (6), 1045-1053 (2004).
  28. Clotilde, L., Bernard, C., Hartman, G., Lau, D., Carter, J. Microbead-based immunoassay for simultaneous detection of Shiga toxins and isolation of Escherichia coli O157 in foods. J Food Prot. 74 (3), 373-379 (2011).
  29. . . Luminex User Software Manual (RUO) xPONENT 3.1 Rev. 2. 3, 6-102 (2014).
  30. Waterboer, T., Sehr, P., Pawlita, M. Suppression of non-specific binding in serological Luminex assays. J Immunol Methods. 309 (1-2), 200-204 (2006).

Play Video

記事を引用
Augustine, S. A. J., Eason, T. N., Simmons, K. J., Curioso, C. L., Griffin, S. M., Ramudit, M. K. D., Plunkett, T. R. Developing a Salivary Antibody Multiplex Immunoassay to Measure Human Exposure to Environmental Pathogens. J. Vis. Exp. (115), e54415, doi:10.3791/54415 (2016).

View Video