Microscale thermophoresis obtains binding constants quickly at low material cost. Either labeled or label free microscale thermophoresis is commercially available; however, label free thermophoresis is not capable of the diversity of interaction measurements that can be performed using fluorescent labels. We provide a protocol for labeled thermophoresis measurements.
The ability to determine the binding affinity of lipids to proteins is an essential part of understanding protein-lipid interactions in membrane trafficking, signal transduction and cytoskeletal remodeling. Classic tools for measuring such interactions include surface plasmon resonance (SPR) and isothermal titration calorimetry (ITC). While powerful tools, these approaches have setbacks. ITC requires large amounts of purified protein as well as lipids, which can be costly and difficult to produce. Furthermore, ITC as well as SPR are very time consuming, which could add significantly to the cost of performing these experiments. One way to bypass these restrictions is to use the relatively new technique of microscale thermophoresis (MST). MST is fast and cost effective using small amounts of sample to obtain a saturation curve for a given binding event. There currently are two types of MST systems available. One type of MST requires labeling with a fluorophore in the blue or red spectrum. The second system relies on the intrinsic fluorescence of aromatic amino acids in the UV range. Both systems detect the movement of molecules in response to localized induction of heat from an infrared laser. Each approach has its advantages and disadvantages. Label-free MST can use untagged native proteins; however, many analytes, including pharmaceuticals, fluoresce in the UV range, which can interfere with determination of accurate KD values. In comparison, labeled MST allows for a greater diversity of measurable pairwise interactions utilizing fluorescently labeled probes attached to ligands with measurable absorbances in the visible range as opposed to UV, limiting the potential for interfering signals from analytes.
Microscale thermophoresis is a relatively new technique in determining disassociation constants (KD) as well as inhibition constants (IC50) between biochemically relevant ligands. The leading commercial retailer for MST (e.g., NanoTemper) offers two popular MST technologies: 1) Label free MST requiring a fluorescent tag, and 2) labeled thermophoresis using the inherent fluorescence of proteins dependent on the number of aromatic residues present in each protein1. A disadvantage of label-free thermophoresis is that in most cases, it does not allow for the measurement of protein-protein interactions. However, it may be possible to engineer proteins without aromatic amino acids such as tryptophan for use in label free thermophoresis2.
MST measures the movement of particles in response to the induction of microscopic temperature fields initiated by an infrared laser in currently available technologies1. MST can be used to measure protein-protein interactions, protein-lipid interactions, protein-small molecule, competition experiments, and even interactions between small-molecules so long as one can produce enough signal separation. Additionally, MST allows for the measurement of membrane-protein based interactions embedded in either liposomes or nanodiscs. Labeled thermophoresis takes advantage of the use of fluorescently labeled tags allowing for chemically controllable separation of signal between ligand and analyte. KD values can be obtained using thermophoresis for interactions involving protein binding at low nanomolar concentrations, which in most cases is a much lower concentration of protein than what is required for isothermal calorimetry (ITC)3. Additionally, MST does not have strict buffering requirements as required for surface plasmon resonance (SPR)4 and labeled thermophoresis can even be used to measure binding constants of proteins of interest from non-fully purified protein solutions5 with genetically inserted fluorescent tags6. A disadvantage of MST is that kinetic parameters cannot be obtained readily for MST as in SPR2.
Thermophoresis measurements depend on the local temperature difference of a solution. This heat can be generated from an infrared laser. The MST device has a fluorescence detector coupled to an infrared (IR) beam and can pick up changes in fluorescence from local concentration changes of the fluorescent molecules at the point where the IR laser is targeted. The MST device utilizes an IR targeted laser coupled directly to a fluorescence detector focused at the same point in which the heat is generated in the solution. This allows for robust detection of changes in temperature corresponding to the depletion of molecules at the point of heat generated by the IR laser. Measured fluorescence generally decreases closer to the IR laser in response to temperature increases. The differences measured as a result can be due to multiple factors including charge, size, or solvation entropy. These differences are measured as changes in fluorescence in response to induction of heat or movement of molecules from hot to colder parts of the capillary.
When loading a capillary with a given solution, it is important to leave air at either end of the capillary and not load the capillary completely full. The commercial capillary holds about 10 μL of solution. One can achieve accurate measurements with 5 μL of solution so long as the solution is manipulated to the center of the capillary, there are no air bubbles (potentially degas prior to loading capillary), and one is careful loading the rack to not jostle the solution from the center of the capillary, where the infrared laser is targeted. If the laser does not come in contact fully with solution, the result will most likely be one of three unusable outputs for that concentration: 1) no or low fluorescence detection, 2) higher fluorescence detection (potentially with jagged peak), or 3) fluorescence detection within other values from given titration, but with a jagged and unrounded peak.
For labeled thermophoresis it is optimal to have a fluorescence signal above 200 and below 2000 fluorescence units7. The MST device uses a range of LED intensities from 0 to 100, which can be selected to achieve a signal above 200 or below 2000. Alternatively, one can use different concentrations of the labeled ligand to modify the fluorescence signal to an optimal level. It is important to run a cap scan with a given MST measurement as a reference when analyzing data, as a poor cap scan can often result in a point that may later be determined to be an outlier. Each run should take approximately 30 min if measuring a single MST power with a cap scan. The commercial devices allow changes in MST power. In older software versions this could be set from 0 to 100; and in later versions one can select low, medium, or high MST. To achieve robust traces, a researcher may need to try each of these and decide which MST setting results in the most robust data for a given interaction.
1. Preparation of materials
2. Preparation of the MST device
3. Preparation of samples for labeled MST
4. MST of samples
5. Analysis of MST data
NOTE: The analysis software provided by Nanotemper is proprietary and is performed using M.O. Affinity Analysis. There are different ways to measure binding affinities based on either fluorescence or thermophoresis. Newer versions of this software are preset to automatically evaluate data using thermophoresis and are preset to use Thermophoresis with Tjump taking advantage of both measurements. Alternatively, one can select either Tjump alone or Thermophoresis alone. Additionally, the analysis software allows estimated affinity measurements using initial fluorescence. These settings can be accessed in expert mode only.
This is a sample output using the affinity analysis. The labeled MST was used to determine the binding constant of the Vam7-His8 to the soluble dioctanoyl (DiC8) PA of one of its natural substrates9. Figure 1 presents the thermophoretic traces from one trial of a 1:1 titration of DiC8 PA starting at 500 μM against 50 nM of Vam7-His8. Initial fluorescence (time before infrared laser turned on), Tjump (time initially after infrared laser turned on), and thermophoresis (once particles reach equilibrium with temperature) are shown. One can calculate a KD from any one of these measurements alone or can use a combination of Tjump and thermophoresis considering two of these measurements.
In Figure 2, a saturation curve is shown for the thermophoresis with Tjump output from the analysis software. As shown in Figure 2, saturation curve can be plotted from these results; however, it may be difficult to calculate a saturation curve as Fnorm does not start from zero. In order to get around this issue, the data can be manually normalized, or the output can be set to the fraction-bound determined by the analysis software. Generally, MST results are determined using a log-scale as shown in Figure 3. The analysis software automatically accounts for account protein concentration in order to determine binding affinity by selecting the KD model and inputting the protein concentration. The Hill model in the analysis software can also be used, which does not consider protein concentration, but can potentially give a measure of cooperativity for a given interaction. Exporting either output from the analysis software and plotting in third party software, one can obtain a KD measurement as shown in Figure 3. It should be noted that the critical micelle concentration (CMC) of dioctanoyl glycerophospholipids is >3 mM, indicating that these experiments are operating far below the CMC10.
Figure 1: MST traces of Vam7–His8 to DiC8 PA. Traces are shown for PA analyte to NTA-Atto 647N labeled Vam7-His8 ligand measured in the red channel. Initial fluorescence was measured at room temperature for 5 s (A) prior to turning on infrared laser. Tjump (B) is measured in the initial seconds following induction of heat from the built-in infrared laser and measures fluorescence differences present when IR laser is first turned on prior to thermophoretic movement of particles being measured. Thermophoresis (C) is measured after molecules equilibrate from the movement induced by heat from the infrared laser due to thermophoretic motion differences due to factors such as size, charge, solvation entropy, or conformational change of particle being measured with respect to titrated analyte. Please click here to view a larger version of this figure.
Figure 2: Saturation binding determined from exported MST results using third party software. Normalized fluorescence results exported from the analysis software. The data were exported and plotted using one-site specific binding model. Please click here to view a larger version of this figure.
Figure 3: Binding affinity of Vam7-His8 to DiC8 PA using a sigmoidal model. Analysis of exported data from Figure 1 exported from the analysis software using the normalization Fnorm [%]. Taking the log of the concentrations of DiC8 PA plotted against Fnorm using Graphpad Prism v.7 using Sigmoidal, 4PL, X is log(concentration) model allows a fit that results in a KD of 89 ± 18.0 μM for Vam7-His8 binding affinity to DiC8 PA. Please click here to view a larger version of this figure.
Supplemental File. Please click here to download this file.
The determination of Vam7-His8binding to DiC8-PA provided a robust fitted KD for the given interaction, which is slightly lower affinity than the measured KD of Vam7-His8 to PA liposomes (unpublished). This difference is most likely due to the lack of a membrane, which generally results in lower affinity for membrane specific lipid binding interactions and therefore demonstrates the role for the liposome membrane scaffold to this interaction11.
In order to further determine the strength of MST data, a researcher needs to examine both the shape of the traces from Figure 1 and the separation of the traces from the chosen analysis method. Looking at the traces from Figure 1, most concentrations resulted in a clear trace as determined by the leveling off of the thermophoresis portion of the curve. In some instances traces corresponding to reactions containing higher concentrations of analyte towards the end of the fluorescent measurement could be due to aggregation in the sample adherence to the sample capillaries, which can be remedied by additions of either or both Pluronic or Tween to the buffer12. However, detergents such as these may not be suitable for protein-lipid interactions13. To test for non-specific binding for this type of protein-lipid interaction one can add in BSA, increase salt concentration, or use glycerol in the buffer and test whether this affects the estimated KD14,15.
Separation is determined by the difference in Fnorm between the highest and lowest concentration measurement for the given interaction. As shown in Figure 2, the separation for this interaction was approximately 75 fluorescence units. Generally, speaking, a separation of at least 5 fluorescence units should be achieved to confidently rely on MST data of unknown chemical affinity measurements. If a robust KD fit is achieved at slightly lower separation, one might be able to still consider such data if it corresponded to affinity measurements used via another technique such as SPR or ITC.
Because some of the thermophoretic traces showed some stickiness in the sample, the output chosen for Figure 3 was a combination of thermophoresis with Tjump. This setting is the preselected setting in most later versions of the analysis software. Figure 3 indicates a robust fit to a sigmoidal binding affinity model as there was clear saturation with a 12 point 1:1 titration. The device allow for 16 points to be taken in a given run, and many interactions require all of these points to achieve a strong binding affinity measurement. In order to achieve greater confidence, this trial should be repeated with new capillaries two or more times requiring a total of 36 capillary tubes for this experiment. Additionally, one might try another technique such as SPR to corroborate this data in order to ensure reliable reporting of affinity constants as we previously have done8,16.
The authors have nothing to disclose.
This research was supported by a grant from National Science Foundation (MCB 1818310) to RAF. This work was supported in part by the Chemical Biology Core Facility/Protein Crystallography Unit at the H. Lee Moffitt Cancer Center (NIH/NCI: P30-CA076292).
Cy5 Maleimide Mono-Reactive Dye | GE Healthcare | PA23031 | For protein labeleing |
Graphpad Prsim | Graphpad software | ||
Monolith NT.115 Capillaries (1000 count) | Nanotemper | MO-K022 | Capillaries for MST |
Monolith NT.115 machine | Nanotemper | University equipment | |
NTA-Atto 647 N | Sigma | 2175 | label for His tags |
Phosphatidylinositol 3-phosphate diC8 (PI(3)P diC8) | Echelon | P-3008 | Lipid for binding experiments |