A method for implanting electrodes into the subthalamic nucleus (STN) of rats is described. Better localization of the STN was achieved by using a microrecording system. Furthermore, a stimulation set-up is presented that is characterized by long-lasting connections between the head of the animal and the stimulator.
Deep brain stimulation (DBS) is a widely used and effective therapy for several neurologic disorders, such as idiopathic Parkinson’s disease, dystonia or tremor. DBS is based on the delivery of electrical stimuli to specific deep anatomic structures of the central nervous system. However, the mechanisms underlying the effect of DBS remain enigmatic. This has led to an interest in investigating the impact of DBS in animal models, especially in rats. As DBS is a long-term therapy, research should be focused on molecular-genetic changes of neural circuits that occur several weeks after DBS. Long-term DBS in rats is challenging because the rats move around in their cage, which causes problems in keeping in place the wire leading from the head of the animal to the stimulator. Furthermore, target structures for stimulation in the rat brain are small and therefore electrodes cannot easily be placed at the required position. Thus, a set-up for long-lasting stimulation of rats using platinum/iridium electrodes with an impedance of about 1 MΩ was developed for this study. An electrode with these specifications allows for not only adequate stimulation but also recording of deep brain structures to identify the target area for DBS. In our set-up, an electrode with a plug for the wire was embedded in dental cement with four anchoring screws secured onto the skull. The wire from the plug to the stimulator was protected by a stainless-steel spring. A swivel was connected to the circuit to prevent the wire from becoming tangled. Overall, this stimulation set-up offers a high degree of free mobility for the rat and enables the head plug, as well as the wire connection between the plug and the stimulator, to retain long-lasting strength.
Deep brain stimulation (DBS) is a treatment based on the delivery of electrical impulses via implanted electrodes to specific cerebral structures, such as the internal globus pallidus1, the subthalamic nucleus (STN)2–4 or the ventral intermediate thalamus5. In the last two decades, this treatment has been established as a powerful therapeutic tool for Parkinson's disease1–4, dystonia6 and tremor7, and is also used to modulate chronic pain7, psychiatric disorders (i.e., obsessive–compulsive disorder8, major depression9) or intractable epilepsy10,11. Furthermore, DBS might, in the future, become a treatment option for refractory arterial hypertension12 or orthostatic hypotension13.
The physiological mechanisms underlying the effects of DBS remain poorly understood. Studies in anesthetized rodents have provided insight into neural responses to high-frequency stimulation that mimic clinically applied DBS14. However, these studies not only lack behavioral corroboration of the DBS effect but also result in considerable variability depending on the stimulation parameters applied14.
To investigate more concisely the behavioral effects and underlying mechanisms of DBS in conscious rodents, a stimulation set-up is needed that fulfills specific requirements. DBS is mostly used as a long-term therapy (e.g., Parkinson's disease, chronic pain). Thus, the stimulation set-up in rodents should be designed so that the unit consists of an electrode with a plug, as well as a wire from the plug to an external stimulator; and this unit should be lightweight but unbreakable when fixed onto the skull. Furthermore, freedom of movement is indispensable for rats during stimulation over a prolonged period. The target structures of DBS are small; for example, the STN in rats has a length of 1.2 mm and a volume of 0.8 mm3,15. Therefore, electrodes must be designed such that the nucleus is not lesioned during insertion and targeting needs to be precise. As most DBS studies conducted in rodents have used landmark based stereotactic insertion of the electrode to the target structure, the error rate can be relatively high, even when using the coordinates according to Paxinos and Watson16. This results in a larger number of animals needed to reach a statistically meaningful result.
In the present study an electrode implantation technique is introduced, that targets the STN with high accuracy by using a microrecording system while advancing the electrode. In addition, a stimulation system is presented that does not only allow a high degree of mobility for the stimulated animal but also guarantees continuous stimulation via secure fixation of the stimulation wire (which is protected by a stainless-steel spring) onto the head of the rat.
Animal experiments were approved by the University of Würzburg and the legal state authorities (Lower Franconia, approval number: 54-2531.01-102/13) and performed according to the recommendations for research in experimental stroke studies17 and the current Animal Research: Reporting of In Vivo Experiments Guidelines (http://www.nc3rs.org.uk/arrive-guidelines).
1. Anesthesia
2. Surgery
3. Stimulation
Implanting an electrode into the STN of a rat using a recording system – as presented here – is an effective and accurate procedure for DBS that takes approximately 1 hr per animal. This model is a fairly minor procedure: out of 10 rats subjected to surgery, all survived the intervention. Twenty-four hr after intervention, the state of each rat was monitored and no animal achieved more than 1 of 3 points according to the severity code. During the period of continuous stimulation (14 days, 24 hr a day), no wire detached, broke or was bitten through. None of the 10 rats lost the cap of dental cement nor did they get hurt by the equipment during the phase of stimulation. The impedance measured in these 10 animals before stimulation was 353 ± 101 kΩ . Rats were stimulated at a frequency of 130 Hz and a pulse width of 60 µsec. The mean stimulus intensity was 60 µA, which was set at 20% below the intensity threshold of orofacial or contralateral forepaw dyskinesia, thereby preventing problems with feeding or locomotion during the period of stimulation.
Fourteen days after intervention and continuous stimulation, all 10 rats were euthanized by decapitation after deep anesthesia and brains were rapidly harvested. In a rat brain matrix, a 2 mm thick brain block encompassing the STN was cut and immediately frozen at –80 °C. These brain blocks were cut in coronal sections (8 µm thick). Each section was stained with hematoxylin & eosin to visualize the position where the tip of the electrode was located, as well as to detect sings of inflammation or scar tissue due to the electrode. The success rate for localizing the electrode in the STN was 8 of 10 animals. In these 8 rats, the tip of the implanted electrode was situated in the STN, as shown histologically. Figure 11 illustrates the electrode location in the STN. A small lesion developed after continuous stimulation was found in all rats. This lesion was surrounded by a small number of inflammatory cells (Figure 11).
Figure 1. Fixation of the head in the stereotaxic instrument. The rat is fixed by the ear bars of the stereotaxic frame, as well as by the gas anesthesia mask. Please click here to view a larger version of this figure.
Figure 2. Exposing the skull. After a midline incision, the skin and periosteum are rolled to the wound edges and kept away from the surgical area using four clamps. Please click here to view a larger version of this figure.
Figure 3. Fixation of the electrode in a probe holder. Using forceps, the pin of the electrode is inserted in the plug and fixed with a probe holder. The plug is connected with the recording apparatus via a wire. Please click here to view a larger version of this figure.
Figure 4. Insertion of the electrode in the brain. After determining the exact AP and ML coordinates of the subthalamic nucleus, the tip of the electrode is advanced to the level of the punctured dura and the dorsoventral vernier scale reading is taken. Please click here to view a larger version of this figure.
Figure 5. Protection of electric interference. A Faraday cage (or, alternatively, aluminum foil) is put over the rat in the stereotaxic instrument and the instrument, as well as the animal, is grounded. Please click here to view a larger version of this figure.
Figure 6. Recording of brain activity. The subthalamic nucleus (STN) shows an irregular firing pattern and a high firing rate (mean frequency: 40.9 ± 12.9 Hz)18. Before entering the STN, the electrode passes a relatively silent region, which is consistent with the zona incerta; the vertical size of this area measures about 0.5–1 mm. Thereafter, the number of spikes increases, indicating that the insertion into the STN is complete. Please click here to view a larger version of this figure.
Figure 7. Fixation of the electrode. When the subthalamic nucleus is identified by means of recording, the electrode is fixed by applying dental cement around the electrode shank and the screws. This allows unplugging of the connector from the electrode pin without shifting the position of the electrode. Please click here to view a larger version of this figure.
Figure 8. Attaching the plug to the electrode pin. The plug for the connector of the stimulator is attached to the electrode pin. The ground wire, which is soldered onto the plug, is fixed with a screw onto the skull. Please click here to view a larger version of this figure.
Figure 9. Fixation of the plug. (A) Frontal and (B) Lateral views. Dental cement is applied around the plug and a cap is formed; sharp edges should be avoided. Please click here to view a larger version of this figure.
Figure 10. Connection of the rat to the stimulator. A swivel was joined into the circuit to prevent the wire from becoming tangled. A stainless-steel spring protects the wire if the rat starts to bite the wire. Please click here to view a larger version of this figure.
Figure 11. Brain section through the subthalamic nucleus (STN) (hematoxylin & eosin staining). (A) Overview, magnification 2.5. A solid line surrounds the STN. A small lesion is visible where the electrode tip was located during a 14 day period of stimulation. It is of note that there is no penetration canal of the electrode visible (shank diameter: 125 μm), indicating that the electrode is conserving the tissue. (B) Image detail from picture A (box), magnification 100. A small number of inflammatory cells was detectable around the lesion due to the reaction of the brain tissue to the electrode tip. Arrow: indicating an example of an inflammatory cell. Please click here to view a larger version of this figure.
This study presents a step-by-step set of instructions for implanting a monopolar chronic electrode into the STN of rats. Although tungsten electrodes with low impedance are often used for DBS18,19, a monopolar electrode made of platinum/iridium (Pt/Ir) was employed that had an impedance of about 1 MΩ. Pt/Ir electrodes are also used in patients with Parkinson’s disease because of their favorable properties: they demonstrate minimal erosion20 and do not produce relevant tissue damage21 if no high-charge densities are used. Since the aim of this study was a long-term stimulation set-up and, in order to achieve a translational approach, electrodes with the aforementioned specifications were applied in the present experiment. Histologic examination of brain slices showing the localization of the electrode tip corroborated the attenuated foreign-body response of Pt/Ir22 in this experiment.
In the present study, electrodes of Pt/Ir with an impedance of 1 MΩ were used. Electrodes with lower, or even higher impedance, are only suitable for either recording cerebral activity or for stimulating cerebral structures, but not both. In contrast, an electrode impedance of 1 MΩ , as used in our study, is suitable for both, recording the activity of deep brain areas and stimulating cerebral structures such as the STN. The major advantage of recording is the identification of the STN location in a short amount of time. Recording allows reliable localization of the STN, as our results have shown: histologic control yielded a high success rate of targeting the STN (8 of 10 animals). The electrode tip was implanted into the upper cell layers of the dorsal-lateral portion of the STN (DV: 7.7 mm) which is known to receive motor inputs mainly from the motor cortex23.
Using a wiring system for DBS may be limited by potential complications such as breaking of wires or a low degree of freedom of movement of animals. However, in our set-up, the wires were connected to swivels, which allowed the animals to move freely. Wireless in vivo stimulating systems (often fixed at the head or implanted into the trunk of the animal) are also limited by the requirement for batteries. As the batteries need to be small, the voltage is therefore low. When using 1 MΩ electrodes, a high voltage is required to achieve the desired stimulus intensity and, in turn, results in larger batteries or frequent replacement of batteries. However, an advantage of the stimulus system used in our study is the large voltage compliance range of the stimulator and the option of constant current stimulation. In this mode the stimulator adjusts the voltage to changes in tissue impedance in order to provide a constant current output at the electrode. An impedance change is expected over the long-term course of DBS with the formation of a stable tissue-electrode interface, e.g., glial encapsulation of the electrode tip22.
In summary, the presented method of electrode implantation is technically simple to perform, reliable, and robust, allowing accurate and safe stimulation of the STN in rats without restricting the freedom of movement or even injuring the animal during the long-term course. With small modifications (e.g., using a plug with additional electrical outputs), this protocol is also applicable by implanting microelectrodes in both STNs or other cerebral structures, long-term recordings or both, stimulation of deep brain structures and recording activity of another cerebral region.
The authors have nothing to disclose.
We wish to thank Mr Wabbel for preparing the wires and Mr Tietsch for constructing the plugs and cages according to our plans. This work was supported by the Deutsche Forschungsgemeinschaft (Sonderforschungsbereich 688). Felix Fluri holds a fellowship of the Interdisziplinäre Zentrum für Klinische Forschung (IZKF), University Clinics Würzburg, Germany.
Pt/Ir electrode | FHC Inc. | UE | Custom-made: Specification: UEPSEGSECN1M |
Plugs | GT Labortechnik (Arnstein/Germany) | Custom-made | |
Pin header | DISTRELEC | 143-95-324 | single-row, 90° 1×3 datamate, Type M80-8420342 |
Socket | DISTRELEC | 143-95-621 | single-row,straight 2 mm pole no.1×3 datamate, Type M80-8400342 |
Stainless steel spring | Plastics ONE | SS0102 | Part-#: .120 X .156 Spring ID (mm): 3.0 Spring OD (mm): 4.0 |
Dental cement/Paladur | Heraeus Kulzer | 64707938 | Liquid, 500 ml |
Dental cement/Paladur | Heraeus Kulzer | 64707954 | Powder, rose, 500g |
Head screw | Hummer & Reiss | V2ADIN84 M1.6×3 | |
Jodosept PVP | Vetoquinol | 435678/E04 | |
Mepivacain 1% | AstraZeneca | PZN03338515 | |
Epinephrine | Sanofi-Aventis | PZN00176118 | |
Tramadolhydrochloride | Rotexmedica | 38449.00.00 |