Precise determination of the evolved gases' flow rate is key to study the details of reactions. We provide a novel quantitative analysis method of equivalent characteristic spectrum analysis for thermogravimetry-mass spectrum analysis by establishing the calibration system of the characteristic spectrum and relative sensitivity, for obtaining the flow rate.
During energy conversion, material production, and metallurgy processes, reactions often have the features of unsteadiness, multistep, and multi-intermediates. Thermogravimetry-mass spectrum (TG-MS) is seen as a powerful tool to study reaction features. However, reaction details and reaction mechanics have not been effectively obtained directly from the ion current of TG-MS. Here, we provide a method of an equivalent characteristic spectrum analysis (ECSA) for analyzing the mass spectrum and giving the mass flow rate of reaction gases as precise as possible. The ECSA can effectively separate overlapping ion peaks and then eliminate the mass discrimination and temperature-dependent effect. Two example experiments are presented: (1) the decomposition of CaCO3 with evolved gas of CO2 and the decomposition of hydromagnesite with evolved gas of CO2 and H2O, to evaluate the ECSA on single-component system measurement and (2) the thermal pyrolysis of Zhundong coal with evolved gases of inorganic gases CO, H2, and CO2, and organic gases C2H4, C2H6, C3H8, C6H14, etc., to evaluate the ECSA on multi-component system measurement. Based on the successful calibration of the characteristic spectrum and relative sensitivity of specific gas and the ECSA on mass spectrum, we demonstrate that the ECSA accurately gives the mass flow rates of each evolved gas, including organic or inorganic gases, for not only single but multi-component reactions, which cannot be implemented by the traditional measurements.
Understanding in depth the real features of a reaction process is one critical issue for the development of advanced materials and the establishment of a new energy conversion system or metallurgy production process1. Almost all reactions are carried out under unsteady conditions, and because their parameters, including the concentration and flow rate of reactants and products, always change with the temperature or pressure, it is difficult to clearly characterize the reaction feature by only one parameter, for instance through the Arrhenius Equation. In fact, the concentration implies only the relationship between the component and the mixture. Real reaction behavior might not be affected, even though the concentration of a component in one complicated reaction is adjusted to a great extent since the other components might have a stronger influence on it. On the contrary, the flow rate of each component, as an absolute quantity, can give persuasive information to understand the characteristics of the reactions, especially very complicated ones.
At present, the TG-MS coupling system equipped with the electron ionization (EI) technique has been used as a prevalent tool for analyzing the features of reactions with evolved gases2,3,4. However, first, it should be noted that the ion current (IC) obtained from an MS system makes it difficult to directly reflect the flow rate or concentration of the evolved gas. The massive IC overlap, fragment, severe mass discrimination, and diffusion effect of gases in the furnace of a thermogravimeter can greatly hamper the quantitative analysis for TG-MS5. Second, EI is the most common and readily available strong ionization technique. An MS system equipped with EI easily results in fragments and does not often directly reflect some organic gases with a larger molecular weight. Therefore, MS systems with different soft ionization techniques (e.g., photoionization [PI]) are required simultaneously to be hyphenated to a thermobalance and applied to evolved gas analysis6. Third, the intensity of the IC at some mass-to-charge ratios (m/z) cannot be used to determine the dynamic characteristic of any reaction gas, because it is often affected by the other ICs for a complex reaction with multicomponent evolved gases. For example, the drop in the IC curve of a specific gas does not necessarily indicate a decrease in its flow rate or concentration; instead, maybe it is affected by the other gases in the complex system. Thus, it is important to take into account all gases' ICs, certainly with a carrier gas and inert gas.
In fact, quantitative analysis based on mass spectrum greatly depends on the determination of the calibration factor and relative sensitivity of the TG-MS system. Maciejewski and Baiker7 investigated in a thermal analyzer-mass spectrometer (TA-MS) system, in which the TA is connected by a heated capillary to a quadrupole MS, the effect of experimental parameters, including the concentration of gases species, temperature, flow rate, and properties of the carrier gas, on the sensitivity of the mass spectrometric analysis. The evolved gases were calibrated by the decomposition of the solids via a well-known, stoichiometric reaction and injecting a certain amount of gas into the carrier gas stream with a constant rate. The experimental results show that there is a negative linear correlation of the MS signal intensity of evolved gas to that of the carrier gas flow rate, and the evolved gas MS intensity is not influenced by the temperature and the amount of analyzed gas. Further, based on the calibration method, Maciejewski et al.8 invented the pulse thermal analysis (PTA) method, which provides an opportunity to determine the flow rate by simultaneously monitoring the changes of the mass, enthalpy, and gas composition resulted from the reaction course. However, it is still hard to give persuasive information about the complicated reaction (e.g., coal combustion/gasification) by using the traditional TG-MS analysis or PTA methods.
In order to overcome the difficulties and disadvantages of the traditional measuring and analysis method for the TG-MS system, we developed the quantitative analysis method of ECSA9. The fundamental principle of ECSA is based on the TG-MS coupling mechanism. The ECSA can take into account all gases' ICs, including the reaction gases', carrier gases', and inert gases'. After building the calibration factor and relative sensitivity of some gas, the real mass or molar flow rate of each component can be determined by the calculation of the IC matrix (i.e., the mass spectrum of TG-MS). Compared with the other methods, ECSA for the TG-MS system can effectively separate the overlapping spectrum and eliminate the mass discrimination and the temperature-dependent effect of TG. The data produced by ECSA have proved to be reliable via a comparison between the mass flow rate of evolved gas and mass loss data by differential thermogravimetry (DTG). In this study, we used an advanced TG-DTA-EI/PI-MS instrument10 to carry out the experiments (Figure 1). This instrument consists of a cylindrical quadrupole MS and a horizontal thermogravimetry-differential thermal analyzer (TG-DTA) equipped with both EI and PI mode, and with a skimmer interface. ECSA for the TG-MS system determines the physics parameters of all evolved gases by utilizing the actual TG-MS coupling mechanism (i.e., an equal relative pressure) to implement the quantitative analysis. The overall analysis process includes a calibration, the test itself, and data analysis (Figure 2). We present two example experiments: (1) the decomposition of CaCO3 with only evolved gas of CO2 and the decomposition of hydromagnesite with evolved gas of CO2 and H2O, to evaluate the ECSA on a single-component system measurement and (2) the thermal pyrolysis of brown coal with evolved gases of inorganic gases CO, H2, and CO2, and organic gases CH4, C2H4, C2H6, C3H8, C6H14, etc., to evaluate the ECSA on a multi-component system measurement. ECSA based on the TG-MS system is a comprehensive solution method for quantitatively determining the amount of evolved gas in thermal reactions.
1. Calibration of ECSA for the TG-MS System
2. Testing Process of ECSA for the TG-MS System
3. Qualitative and Quantitative Analysis
The thermal decomposition of CaCO3 is a relatively simple reaction, which was used to demonstrate the applicability of the ECSA method. After calibrating the characteristic peak and relative sensitivity of CO2 to carrier gas He, the actual mass flow rate of CO2 evolved by the thermal decomposition of CaCO3 was calculated by the ECSA method and was compared with the actual mass loss (Figure 3). It is shown that there is a good agreement between the mass flow rate of CO2 calculated by ECSA and the mass loss data by DTG during the entire measurement process. The relative error of mass flow rate of evolved gas to that of DTG is significantly lower, as shown by the blue and yellow lines in Figure 4. Also, the thermal decomposition process of hydromagnesite was analyzed by ECSA and the calibration data of CO2 and H2O (Figure 4). The carrier gas flow rate was chosen as 100 mL/min and the heating rate was set at 5, 10, 15, and 20 K/min. The calculated results were also in good agreement with the experimental TG/DTG data.
To further demonstrate the qualitative analysis of organic gases and the ability of ECSA to quantitatively determine the flow rate of a complicated reaction system, pyrolysis of Zhundong coal was carried out10. Combining both the PI and EI measuring modes, 16 types of volatile gases, including H2, CH4, H2O, CO, CO2, C2H4 (ethene), C3H6 (propene), C4H8 (butylene), C5H10 (pentene), C6H10 (hexadiene), C7H8 (toluene), C6H6O (phenol), C8H10 (ethylbenzene), C7H8O (anisole), C9H12 (propyl benzene), and C10H14 (butylbenzene), were clearly identified (Figure 5). After a detailed determination of mass spectrum and sensitivity of each gas to the carrier gas, the mass flow rate of each gas can be calculated. Straightforwardly, the ion current from the mass spectrum can be used to compare based on the same operating parameters (Figure 6).
Figure 1: Schematic diagram of the TG-DTA-EI/PI-MS system equipped with the EI and PI devices and skimmer-type interface. This TG-DTA-EI/PI-MS system mainly consists of a cylindrical quadrupole MS and a horizontal thermogravimetry-differential thermal analyzer (TG-DTA) equipped both the EI and PI devices. The MS and TG-DTA are connected by the skimmer interface. This figure has been modified from Li et al.10. Please click here to view a larger version of this figure.
Figure 2: Process diagram of ECSA for testing the thermal reactions. The overall analysis process can be divided into three parts that are calibration, test, and data analysis. The calibration part first provides the information of the characteristic spectrum and relative sensitivity of each gas in the reaction; this information is used for the subsequent calculation of the physics parameters, such as the flow rate, following the test. Please click here to view a larger version of this figure.
Figure 3: Comparison of the mass flow rate of evolved gas with the mass loss of DTG for the thermal decomposition of CaCO3. A comparison in mass loss between ECSA calculation results and measurement results from DTG was used to validate the reliability of the ECSA method. It is shown that there is a good agreement between the calculation by the ECSA and the measurements by DTG, and the relative error of mass flow rate of CO2 to that of DTG is significantly lower. Please click here to view a larger version of this figure.
Figure 4: Thermal decomposition process of hydromagnesite. These panels show (a) a 3-D mass spectrum graph plotted against temperature and m/z, (b) mass flow rates of CO2 calculated by ECSA at a heating rate of 5, 10, 15, and 20 K/min, (c) mass flow rates of H2O calculated by ECSA at a heating rate of 5, 10, 15, and 20 K/min, and (d) a comparison between ECSA-based flow rates and experimental TG/DTG data. Here, the carrier gas flow rate was chosen as 100 mL/min. Please click here to view a larger version of this figure.
Figure 5: 3-D mass spectrum graph of raw coal with the temperature and m/z in modes of EI and PI. (a) The EI mode was mainly used to identify inorganic gases such as CO2 and H2O, while (b) the PI mode was mainly done to identify organic gases such as C6H6 and C7H8. A joint use of EI and PI provides a comprehensive information for the pyrolysis of coal. This figure has been modified from Li et al.10. Please click here to view a larger version of this figure.
Figure 6: Mass ion curves of CH4 and C6H6O evolved from the raw coal sample and the pretreated Zhundong coal samples. One kind of inorganic gas, (a) CH4, and one organic gas, (b) C6H6O, were chosen to be represented in mass ion curves, for interpreting the function of ECSA on a quantitative analysis of the pyrolysis characteristics of different pretreated coals. Here the pretreated method includes the H2O-washed coal and HCl-washed coal. This figure has been modified from Li et al.10. Please click here to view a larger version of this figure.
This protocol could be easily modified to accommodate other measurements for studying evolved gases and pyrolysis reactions by a TG-MS system. As we know, the evolved volatile from the pyrolysis of biomass, coal, or other solid/liquid fuel does not always include only the inorganic gases (e.g., CO, H2, and CO2) but also the organic ones (e.g., C2H4, C6H5OH, and C7H8). Moreover, massive fragments would result from the organic gases, and secondary reactions would occur during pyrolysis11. Even though several conventional measurement methods, like the normal TG-MS coupling system, Fourier transform infrared (FTIR) spectrometry12, high-performance liquid chromatography (HPLC)13, and UV-vis absorption and fluorescence spectroscopies14, have been employed for the characterization of the volatile gases and tar, there are still some issues to be solved, including the minimization of the secondary reactions, the mitigation of recondensation of evolved volatile gases during measurement, and the reduction of excessive fragments. ECSA developed based on a TG-DTA-EI/PI-MS system can be used to investigate accurately the pyrolysis characteristics for real-time, in situ measurement. Because of the fundamentality, applicability, and generality of the ECSA, the quantitative analysis for massive volatile gases from pyrolysis can be easily implemented9.
It should be considered that ECSA based on the TG-MS system is a powerful tool to analyze the thermal reaction processes with evolved gases for not only the simple system but also for the complicated one. A key step to implement the ECSA method is to successfully build the calibration factor and the relative sensitivity of the required gases. It also should be noted that the testing conditions of MS must be the same (or very similar) as those for calibration. Specifically, the reference gas for calibrating the relative sensitivity must be the same as the reference gas for the testing process, and it must never react with the evolved gases. In this study, helium is chosen as the reference gas in order to analyze CO2 and H2O in the measurement. Further, we believe that the ECSA can be used to characterize the elementary reactions if the calibration factor and relative sensitivity of the reactants or products in the elementary reactions are successfully built. On the other hand, since the ECSA separates the mass spectrum of all evolved gases into the spectra of different components, the matrix constructed by the ion current of various evolved gases must be solved before the quantitative results are obtained. The matrix can be expected to be large, should there be a large quantity of evolved gas species. Therefore, matrix solution is also key to the implementation of ECSA.
Finally, ECSA has much more advantages than the traditional TG-MS analysis methods. The key one is that the ECSA can provide the exact quantitative information (i.e., the flow rate, the concentration, and the partial pressure) for all gases. Another advantage is that, since ECSA treats with the IC of mass spectrum from the point of coupling characteristics (i.e., the equal relative pressure between TG and MS), it radically eliminates the mass discrimination of MS and the temperature-dependent effect of TG. And further, the issue of time delay during the measurement of reactions with evolved gases (especially solid particle reactions) can also be effectively resolved by varying the flow rate of the carrier gas and the temperature of TG. However, due to the MS, the ECSA cannot be used to determine reactions without evolved gases, and there is still some difficulty in dealing with the elementary reactions. Since all reactions accompany a change of heat, we are developing a new method to correlate the heat change into the ECSA to provide quantitative information for reactions without evolved gases but with heat change.
The authors have nothing to disclose.
The authors gratefully acknowledge the financial support from the National Natural Science Foundation of China (Grant No. 51506199).
CaCO3 and Ca(OH)2 | Sinopharm Chemical Reagent | ||
hydromagnesite | Bangko Coarea in Tibet | ||
Zhundong coal | the coal field in the Mori Kazak Autonomous County, Junggar basin, Xinjiang province of China | ||
ThermoMass Photo/H | Rigaku Corporation | ||
The STA449F3 synchronous thermal analyzer and QMS403C quadrupole MS analyzer | NETZSCH |