This article describes a setup and method for the in situ visualization of oil samples under a variety of temperature and pressure conditions that aim to emulate refining and upgrading processes. It is primarily used for studying isotropic and anisotropic media involved in the fouling behavior of petroleum feeds.
To help address production issues in refineries caused by the fouling of process units and lines, we have developed a setup as well as a method to visualize the behavior of petroleum samples under process conditions. The experimental setup relies on a custom-built micro-reactor fitted with a sapphire window at the bottom, which is placed over the objective of an inverted microscope equipped with a cross-polarizer module. Using reflection microscopy enables the visualization of opaque samples, such as petroleum vacuum residues, or asphaltenes. The combination of the sapphire window from the micro-reactor with the cross-polarizer module of the microscope on the light path allows high-contrast imaging of isotropic and anisotropic media. While observations are carried out, the micro-reactor can be heated to the temperature range of cracking reactions (up to 450 °C), can be subjected to H2 pressure relevant to hydroconversion reactions (up to 16 MPa), and can stir the sample by magnetic coupling.
Observations are typically carried out by taking snapshots of the sample under cross-polarized light at regular time intervals. Image analyses may not only provide information on the temperature, pressure, and reactive conditions yielding phase separation, but may also give an estimate of the evolution of the chemical (absorption/reflection spectra) and physical (refractive index) properties of the sample before the onset of phase separation.
The study of the phase behavior of oil samples in a wide range of temperatures, pressures, and reactive conditions can yield very useful information for the operator of a refinery that processes a variety of feeds. In particular, the fouling of process units and lines by an uncontrolled formation of coke or sediments can severely affect production (loss of throughput) and energy efficiency (increase in heat transfer resistance)1,2,3. Possible plugging caused by the accumulation of fouling material may require a shutdown for clean-up purposes, which would have a highly negative economic impact4. Conducting an assessment of the fouling propensities of feeds can be highly valuable for the optimization of process conditions5 and the blending of refinery streams.
We have developed an in situ analyzer of petroleum stability in our laboratory to allow the visualization of oil samples subject to refinery process conditions. This apparatus relies on a specifically designed reactor made of stainless steel fittings and equipped with a sealed sapphire window at the bottom. The main principle of the device is the illumination of the sample inside the reactor at the desired range of temperature and pressure and the imaging of the resulting cross-polarized reflection. While previous published work relative to this setup focused on thermal cracking processes to emulate visbreaking conditions6,7,8,9 (which do not require high pressure), the reactor design was overhauled to investigate the behavior of samples under hydroconversion (catalytic cracking under high H2 pressure) and aquathermal10 (thermal cracking under high-pressure steam) conditions. Thus, the device was revised in order to operate in the 20-450 °C temperature range and the 0.1-16 MPa pressure range, with the ability to sustain both 450 °C and 16 MPa for reaction times of up to 6 h.
The first level of analysis on the visual information of the samples under a particular range of temperature, pressure, and reactive conditions is to determine whether the sample is single-phase or multiphase. This system is unique in that it allows for the visualization of opaque isotropic material and is not limited to the visualization of anisotropic material described in other work11. While the main indicator of the fouling propensity of samples is the tendency to drop sediments out of the bulk liquid; gas-liquid, liquid-liquid, liquid-solid, and more complex phase behaviors can be observed. However, valuable information can also be extracted from the visual evolution of a liquid as it remains homogeneous (single-phase). In particular, the brightness of the images is related to the refractive index and the extinction coefficient of the sample, while the color of the sample is a subset of its spectral information in the visible light range (380-700 nm), which can be used as a descriptor of its chemistry9.
Caution: Use all appropriate safety practices when performing an experiment under high temperature and pressure conditions, including the use of engineering controls (H2 flow limiter, pressure regulators, and rupture disc assembly) and personal protective equipment (safety glasses, temperature-resistant gloves, lab coat, full-length pants, and closed-toe shoes). Consult all relevant material safety data sheets (MSDS) before use. Carry out micro-reactor loading and clean-up in a fume hood, as these steps involve the use of harmful volatile organic solvents (toluene and dichloromethane).
NOTE: Setup description (see supplemental file).
1. Micro-reactor Loading
2. Micro-reactor Installation
3. Regular Procedure for the Visualization of Cracking Reactions
4. Shutdown and Clean-up
5. Image Analysis9
Eq. 1
Eq. 2
Eq. 3
Eq. 4
Eq. 5
Eq. 6
The visual evolution of Athabasca Vacuum Residue is representative of the behavior of asphaltenic heavy crude oil samples and asphaltenic vacuum residue samples under thermal cracking conditions. However, using different samples and/or different temperature, pressure, or reaction conditions can give rise to a wide variety of phase behaviors. Micrographs corresponding to the thermal cracking experiment on an Athabasca Vacuum Residue sample at final set-point conditions of 435 °C and Patm (N2) are given in Figure 3, while Figure 4 shows the evolution of temperature during the experiment.
At room temperature, this sample is a pasty solid; thus, the sapphire window is mostly not wetted by the sample and is in contact with gas (in this case, N2). An air/sapphire interface yields a much brighter reflection than an oil/sapphire interface, so the appropriate illumination and exposure settings to image a liquid sample will always yield white regions if the sapphire surface is in contact with gas. At a higher temperature (> 150 °C), the sample becomes fluid enough to flow and wet the window surface. Small mineral solids inside the sample, which can be identified by small bright elements (Figure 3 A), can serve as an indicator of the stirring efficiency. As the sample is heated to higher temperatures, the images brighten correspondingly, with no color change as long as no significant reaction is taking place. Thermal cracking reactions in asphaltenic vacuum residues cause color and brightness changes that correspond to the chemical transformation of the sample. At extended reaction times, the formation of domains of anisotropic carbonaceous phase (mesophase) can be detected as stationary heterogeneities on the window (Figure 3 D).
An image analysis of the series of micrographs is shown in Figures 5 and 6, which show the evolution of brightness intensity and color with reaction time, respectively. At very early reaction times, the increase in image brightness follows the evolution of the temperature inside the reactor. As the temperature inside the reactor approaches the 435 °C set-point, thermal cracking reactions in the Athabasca Vacuum Residue become prevalent. Thermal cracking reactions in Athabasca Vacuum Residue induce a brightness change in the sample that follows a decreasing exponential trend. In the same period, the color of the sample remains stable in the first part of the reaction before undergoing a shift towards a blue color. The formation of mesophase has the effect of increasing the overall brightness intensity and enhancing the blue color shift9.
Figure 1: Photographs of the micro-reactor, held upside-down by a clamp. Pre-loading arrangement, with the bottom face opened (A). The loaded and sealed micro-reactor (B). Please click here to view a larger version of this figure.
Figure 2: Examples of preferable fields-of-view, as outlined by red rectangles, with respect to the inner surface of the sapphire window. Please click here to view a larger version of this figure.
Figure 3: Micrographs taken during a thermal cracking experiment on Athabasca Vacuum Residue with a condition set-point of 435 °C and Patm (N2) after 0 min (A), 25 min (B), 50 min (C), and 80 min (D). Scale bar = 100 µm. Please click here to view a larger version of this figure.
Figure 4: The temperature inside the reactor during a thermal cracking experiment on Athabasca Vacuum Residue with a set-point of 435 °C and Patm (N2). Please click here to view a larger version of this figure.
Figure 5: Evolution of the brightness intensity (I) of the micrographs taken during a thermal cracking experiment on Athabasca Vacuum Residue under 435 °C and Patm (N2), normalized by the brightness of the micrograph taken at 350 °C. Please click here to view a larger version of this figure.
Figure 6: Evolution of the hue and saturation (H and S in polar coordinates) of the micrographs taken during a thermal cracking experiment on Athabasca Vacuum Residue under 435 °C and Patm (N2). Please click here to view a larger version of this figure.
Figure 7: Evolution of the brightness intensity (I) of the micrographs taken during a thermal cracking experiment on Cold Lake bitumen under 435 °C and Patm (N2), normalized by the brightness of the micrograph taken at 350 °C. The data points outlined in red correspond to pictures taken with an overheated objective. Please click here to view a larger version of this figure.
Figure 8: The main incident rays (blue arrows) and reflected rays (red arrows) involved in the illumination of a sample through a window. Please click here to view a larger version of this figure.
Figure 12: Evolution of the brightness intensity (I) of the micrographs taken during a hydroconversion experiment, normalized by the brightness of the micrograph taken at 350 °C. The hydroconversion experiment was carried out on a heavy vacuum gasoil sample under 420 °C and 15 MPa (H2), with 12.3 wt.% Ni/Mo catalyst. Please click here to view a larger version of this figure.
Critical Steps within the Protocol
The first critical step in the protocol is ensuring the integrity of the metal-to-sapphire seal, especially if the experiment is to be carried out under pressure. Thus, the parallelism, the smoothness, and the cleanliness of the sealing surfaces should be carefully inspected, and the leak tests should be thorough. Since the modulus of rupture of sapphire is a decreasing function of temperature14, thicker sapphire windows should be used for work at high pressure and high temperature. As a guideline, 8 mm-thick sapphire windows are used in our experiments aiming to emulate hydroconversion conditions (400-450 °C and 16 MPa H2).
The second critical step relates to obtaining high-quality images, which require the bright illumination of the sample; a clean train of optics; adapted microscope settings (wide iris aperture and long working-distance objectives); and the proper alignment between the microscope objective, the reactor window, and the supporting platform.
For the quantitative analysis of image information, it is critical not to overheat the microscope objective while performing the observations. The method described in step 3.6.1 of the Protocol prevents such overheating. If the operator omits to remove the objective from underneath the reactor between two snapshots taken one min apart, the second picture will appear noticeably brighter as a result. To illustrate this issue, the data points outlined in red on Figure 7 correspond to pictures where the operator had left the objective underneath the reactor in the previous min.
Lastly, it is important to clean the reactor thoroughly between experiments in order to avoid cross-contamination.
Modifications and Troubleshooting
Poor data quality generally results from a poorly controlled operating variable (temperature, pressure, or stirring), or a problem in the train of optics. Possible problems in the train of optics include: poor illumination; a small iris aperture; misaligned cross-polarizers; dirty mirrors, filters beam splitters, or objectives; a misaligned reactor or supporting platform above the objective; a dirty or scratched sapphire window; a maladjusted field of view; overheated objectives; and out-of-focus objectives.
Limitations of the Technique
For present configuration of the experimental setup, the main limitation of this technique is the lack of ability to reproduce the same level of image brightness across different experiments. In addition to the cleanliness and the alignment of the train of optics, image brightness was found to be very sensitive to the positioning and tilt of the reactor over the objective, which currently are not tightly controlled parameters. However, normalizing the image brightness of a series of micrographs in a given experiment by the image brightness of a micrograph taken at a reference temperature within the same series provides a satisfying workaround, as it yields reproducible data.
Significance of the Technique with Respect to Existing/Alternative Methods
The combination of cross-polarizers in the train of optics of an inverted microscope with a reactor window made of sapphire allows the observation of high-contrast images of the sample in situ. When shining light on an opaque sample through a window, two main reflections are involved, as shown in Figure 8: the reflection of the light on the outside surface of the window in contact with air, and the reflection of the light on the inside surface of the window in contact with the sample. The intensity of the reflection at each interface is given by the following equation15:
Eq. 7
where indices 1 and 2 refer to the media located before and beyond the interface, respectively; n describe refractive indices; and κ is the extinction coefficient. In air/sapphire and sapphire/oil reflections, the contribution of the extinction coefficient to the reflection can be neglected. Considering the refractive index of sapphire in the C-axis direction (extraordinary ray) as 1.765 (average in the 380-700 nm range)16, the intensity of the first reflection at the air/sapphire interface is about 7.7% of the incident light. Since most oil samples have refractive indices ranging from 1.45 to 1.617, the intensity of the second reflection at the sapphire/oil interface can be considered as less than 0.9% of the incident light. On first approximation, the air/sapphire reflection is at least more than 9 times brighter than the sapphire/oil reflection. Hence, when observations are made under bright-field settings (using unpolarized light), visuals of the sample are outshined by the air/sapphire reflection. To illustrate this issue, micrographs taken under bright-field settings during the thermal cracking experiment on an Athabasca Vacuum Residue sample at final set-point conditions of 435 °C and Patm (N2) are presented in Figure 9 (the microscope lamp voltage was reduced to 10 V and the camera exposure was reduced to 25 ms to avoid blow-outs).
Figure 9: Micrographs taken during a thermal cracking experiment on Athabasca Vacuum Residue with a condition set-point of 435 °C and Patm (N2) after 0 min (A), 25 min (B), 50 min (C), and 80 min (D), taken using bright-field microscope settings instead of cross-polarizers. Scale bar = 100 µm. Please click here to view a larger version of this figure.
As can be seen by comparing Figure 9 with Figure 3, the presented method for observing the sample using cross-polarized light and a sapphire window has the advantage of yielding high-contrast images that are able to describe isotropic media.
As the light is reflected at the air/sapphire interface, its polarization plane does not change. Thus, the cross-polarizer setting cancels this reflection before it hits the CCD camera. As light travels through sapphire, however, its polarization plane rotates because of the sapphire birefringence. This phenomenon ultimately allows sample imaging, even if the oil sample itself is isotropic and the polarization plane of the light does not change upon the sapphire/oil reflection. If the cross-polarizer setting is used in combination with an optically isotropic window (such as fused silica or yttrium-aluminum-garnet, YAG), then only an anisotropic medium (changing the polarization plane of the light at the window/sample interface) and depolarized fluorescence can be viewed. Figure 10 presents micrographs taken during a thermal cracking experiment on an Athabasca Vacuum Residue sample at final set-point conditions of 435 °C and Patm (N2) using the cross-polarizer setting and a 4 mm-thick YAG window.
Figure 10: Micrographs taken during a thermal cracking experiment on Athabasca Vacuum Residue with a condition set-point of 435 °C and Patm (N2) after 0 min (A), 25 min (B), 50 min (C), and 80 min (D), taken using a YAG window instead of a sapphire window. Scale bar = 100 µm. Please click here to view a larger version of this figure.
In comparison to the presented technique, the top-down, hot-stage configuration used in other works11,18 has the disadvantage of featuring a gap of gas between the inner surface of the reactor window and the liquid sample. In such a configuration, using a sapphire window would produce images dominated by the brightness of the sapphire/gas reflection, very similar to the use of bright field with an inverted microscope. Thus, the operators of the top-down hot stage used a reactor window made of YAG, which only allows for the observation of anisotropic material, as explained previously.
The optical properties of a sample may evolve as it undergoes a change in temperature, pressure, or reaction time. The formation of a multiphase system can be characterized by the formation of heterogeneity on the window surface. Figure 11 shows examples of gas-liquid-anisotropic solid, liquid-isotropic solid, liquid-anisotropic semi-solid, and liquid-liquid crystal multiphase systems.
Figure 11: Examples of varied phase behaviors observed during thermal cracking (A, B, and C) and coal liquefaction (D) experiments. Gas-liquid-anisotropic solid (A), liquid-isotropic solid (B), liquid-anisotropic semi-solid (C), and liquid-liquid crystal (D) multiphase systems. Scale bar = 100 µm. Please click here to view a larger version of this figure.
For homogeneous, single-phases systems, changes in the brightness and color of the sample can be related to physical and chemical properties. Following Equation 7, changes in sample brightness are attributed to changes in the refractive indices. In particular, the greater the difference in refractive indices between the sample and the sapphire, the brighter the reflection. For instance, as a heavy oil sample is heated to temperatures below 300 °C, the refractive index of the oil decreases while the refractive index of the sapphire slightly increases, yielding brighter images. During isothermal cracking reactions of vacuum residue samples, images undergo an exponential decrease in brightness; this is attributed to an increase in refractive index due to an increase in aromaticity and density. Conversely, hydroconversion reactions at constant temperature produce a gradual increase in sample brightness, which corresponds to a decrease in refractive index following a decrease in the density of the sample.
Color changes follow the evolution of the spectral properties of the sample, which correspond to its chemistry. Most notably, vacuum residue samples have exhibited a red-to-blue color shift when subjected to thermal cracking reactions for an extended amount of time prior to the formation of sediments. Given enough thermal cracking reaction time, such samples undergo an increase in aromaticity and begin to form oligomers. The formation of more conjugated species leads to a change in spectral properties, where the predominant light absorption of the sample shifts from shorter wavelengths to longer wavelengths. Since reflection spectra are the counterpart of absorption spectra, the corresponding spectral shift in the reflected light goes from longer wavelengths to shorter wavelengths, matching the color change from red to blue9.
Future Applications or Directions after Mastering This Technique
While our studies involving the use of this setup have been primarily related to phase separation phenomena during visbreaking and hydroconversion of heavy petroleum samples in downstream operation, the technique could be applied to the investigation of other phase separation mechanisms occurring in oil processing units and lines (wax crystallization, demulsification, etc.). More generally, this technique could be applied to any system where tracking the optical properties of a sample in situ is of great importance.
Our current research efforts are focused on establishing more relationships between the spectral properties and the physical properties (solubility in particular) of petroleum samples. At present, the spectral information contained in images is limited, since it is expressed in three color channels (RGB). Therefore, the most promising development of this technique lies in the implementation of hyperspectral characterization.
The authors have nothing to disclose.
The authors acknowledge Daniel Palys for supplying Figure 12 and for his assistance in managing laboratory supplies.
Sapphire window, C-plane, 3mm thick – 20 mm diam., Scratch/Dig: 80/50 | Guild Optical Associates | ||
C-seal | American Seal & Engineering | 31005 | |
Type-K thermocouple | Omega | KMQXL-062U-9 | |
Ferrule (1/16") | Swagelok | SS-103-1 | Inserted for creating a clearance gap between the magnet and the window surface |
Coil Heater | OEM Heaters | K002441 | |
Temperature controller | Omron | E5CK | |
Inverted microscope | Zeiss | Axio Observer.D1m | Require cross-polarizer module |
Toluene, 99.9% HPLC Grade | Fisher | Catalog # T290-4 | Harmful, to be handled in fume hood |
Methylene chloride, 99.9% HPLC Grade | Fisher | Catalog # D143-4 | Harmful, to be handled in fume hood |
Acetone, 99.7 Certified ACS Grade | Fisher | Catalog # A18P-4 |