To prepare a sample for imaging, plate about 5.5 x 104 stable expression cells per well of an 8-well glass-bottomed chamber slide, in 350 microliters of phenol red-free DMEM. Use fresh paraformaldehyde with glutaraldehyde to fix the cells, and after replacing the fixative with PBS or imaging buffer, vortex 100-nanometer gold particles, and add 35 microliters per well for tracking stage drift during imaging.

Approach the microscope and power on the 405 and 561-nanometer lasers. Keeping the shutters closed at this point, ensure the 561-nanometer dichroic mirror and notch filter are in place. Open the image acquisition software.

Turn on the EMCCD camera and allow it to cool down, and set the exposure time to 100 milliseconds and the EMCCD gain to an appropriate value. Add immersion oil to the objective, and secure the sample on the microscope stage. Next, with either brightfield or the 561-nanometer laser, bring the sample into focus.

For imaging Ras and other membrane proteins, use a 60X apochromat TIRF objective with a 1.49 numerical aperture, and bring the microscope into TIRF configuration.

Adjust the excitation laser so that it is off-centered when hitting the back aperture of the TIRF objective, which will cause the laser to deflect upon reaching the coverslip buffer interface. Keep adjusting the laser incident position until the critical angle is reached and the laser is being reflected back.

Search for a cell to image with several gold particles in view. Then, set a region of interest that encloses the cell or a region of it and the gold particles.

In case sufficient activation is already occurring because of high expression levels, begin acquisition with the 405-nanometer laser off and the 561-laser on. Otherwise, turn on the 405-nanometer laser at the lowest factory power setting, and increase it gradually until there are tens of molecules per frame or so that single molecules are well-separated.

As data acquisition continues, gradually increase the 405-nanometer laser power to keep the spot density roughly constant. Continue image acquisition until high 405 power does not initiate any more activation events.