Here, we present a protocol to perform portable cellular aerosol exposures and measure cellular response. The method uses cells, grown at the air-liquid interface, mimicking in vivo physiology. Cellular response to copper nanoparticle aerosols was observed as oxidative stress through reactive oxygen species generation and cytotoxicity as lactate dehydrogenase release.
This protocol introduces a new in vitro exposure system, capable of being worn, including its characterization and performance. Air-liquid interface (ALI) in vitro exposure systems are often large and bulky, making transport to the field and operation at the source of emission or within the breathing zone difficult. Through miniaturization of these systems, the lab can be brought to the field, expediting processing time and providing a more appropriate exposure method that does not alter the aerosol prior to contacting the cells. The Portable In vitro Exposure Cassette (PIVEC) adapts a 37 mm filter cassette to allow for in vitro toxicity testing outside of a traditional laboratory setting. The PIVEC was characterized using three sizes of copper nanoparticles to determine deposition efficiency based on gravimetric and particle number concentration analysis. Initial cytotoxicity experiments were performed with exposed lung cells to determine the ability of the system to deposit particles while maintaining cell viability. The PIVEC provides a similar or increased deposition efficiency when comparing to available perpendicular flow in vitro exposure devices. Despite the lower sample throughput, the small size gives some advantages to the current in vitro ALI exposure systems. These include the ability to be worn for personal monitoring, mobility from the laboratory to the source of emission, and the option to set-up multiple systems for spatial resolution while maintaining a lower user cost. The PIVEC is a system capable of collecting aerosols in the field and within the breathing zone onto an air-interfaced, in vitro model.
Personal sampling using in vitro techniques could provide comprehensive information regarding the biological effects of aerosols in the workplace.1 Exposures to contaminants in the air include exposures to the chemical itself, to the collected air samples, under submerged conditions where the gas is introduced to the cell suspension, intermittent exposures using a device such as a rocker, or direct exposures at the air-liquid interface (ALI).2 Many of these techniques are performed with cells grown in suspension or the collection of samples prior to the exposure, each of which can affect the toxicological study due to potential changes in the aerosol.3 To avoid these changes, the laboratory can be brought to the field using several in vitro ALI culture exposure systems that are used in literature,4,5,6,7,8,9,10,11,12,13 however, few are commercially available.8,9,12 These systems are often bulky, especially when including instruments to regulate the temperature and humidity of the cellular environment and the flow rate of the sample aerosol. By using the PIVEC, aerosol exposures can be performed outside of a traditional lab setting or within the breathing zone while mimicking inhalation conditions.
The determination of aerosol deposition in vitro is important to the investigation of health effects due to inhalation. The breathing zone, the area within 30 cm from the mouth and nose,14 is crucial for understanding the exposure to nanoparticles and for linking to the biological effects in the lungs.2 Often, the deposition on cells is defined as a deposition efficiency, the particles deposited onto and taken up by the cells divided by the particles administered to the system6,15 or on a mass basis of the same amounts.4,16 The current methods for measuring aerosols in the breathing zone are filter based, capturing particles over a given sampling period and using the filters to conduct further testing.17 Personal monitoring requires a small system that comes with the tradeoff of fewer samples.
There are many approaches to determine the health effects from exposure to an aerosol. The ALI model allows for the aerosol to be administered directly to the cells through the air as in a real exposure scenario, yet it is more cost-effective and less time intensive than in vivo studies while mimicking the air-liquid barriers such as the eyes, skin, and lungs. Lung cells grown at the ALI have the ability to generate a polarized barrier layer,18,19 which produces physiological traits that resemble the in vivo lung epithelium, including mucus and surfactant production in specific bronchial or alveolar cell lines, cilia beating,19 tight junctions,19,20 and cell polarization.18 Changes such as these can affect the cellular response measured in toxicity studies.21 In addition, ALI in vitro model results are often more sensitive than cells exposed via suspension models22 and are able to model acute in vivo inhalation toxicity.23,24 Therefore, an ALI exposure system that is able to perform measurements within the breathing zone is a natural next step.
By exposing the cells to aerosol directly at the source of emission, investigation of the effects of all gases, semi-volatile compounds, and particles involved in the mixture occurs. When the mixture is collected on a filter, the gases and volatile compounds are not captured and the whole mixture cannot be investigated. In addition, reconstitution of particles into a powder or a liquid suspension can lead to the aggregation or particle-fluid interactions, such as dissolution, in liquid suspension.25,26 When aerosol particles are added to the liquid, there is a higher potential for agglomeration,25,27 formation of a protein corona,28 or interaction with compounds in the liquid, which can affect deposition and influence the biological response.29,30
Exposure at the ALI is based on three main aerosol profiles, cloud settling, parallel flow, and perpendicular flow. Cloud settling, used by the Air-Liquid Interface Cell Exposure (ALICE),4 is a batch system where particles deposit through gravitational and diffusional settling as the aerosol is treated as one unit. Parallel flow, used by the Electrostatic Aerosol in vitro Exposure System (EAVES)5 and Multiculture Exposure Chamber (MEC) II,6 allows for deposition through the addition of Brownian motion through the flow profile. Perpendicular flow, used by a microsprayer,7 Nano Aerosol Chamber for In-Vitro Toxicity (NACIVT),11 and commercial ALI systems8,9,10,12, adds the impaction of particles within the deposition region. Many of these exposure systems are large and bulky, requiring excess systems for aerosol pre-conditioning, pumps for flow, or even heating chambers for incubation of cells. This large size decreases the portability of the system. Instead of sampling directly at the source of emission, these systems often have samples brought to the lab or model aerosols generated for analysis. The complexity of the emitted aerosol can be lost in translation from the field to the lab. The PIVEC is smaller than current systems, with an external surface area of approximately 460 cm2 and weighing only 60 grams, with thermal and humidity control incorporated into the system allowing for a highly portable device. The decreased size and weight allow the system to be worn or taken to the source of exposure, permitting direct sampling.
The large size of current exposure systems also decreases the ability to perform sampling to investigate spatial gradients in concentrations. This resolution is key when determining toxicological effects of many potential environmental and occupational hazards such as vehicular exhaust particulate matter or workplace activities where aerosolization occurs. Immediately post-emission, there becomes a spatial variance in particle concentration. This grows with time as the particles disperse throughout the atmosphere and these effects can change based on the ambient conditions, such as temperature, pressure, wind, and sun. Particles can begin to age and oxidize as well once emitted31,32 and dispersal rates are affected by the topography; higher concentrations will be found in canyons and tunnels, where dispersion effects are slowed, and lower concentrations can be found where there is a large area for dispersion.33 These changes in dispersion rates can have significant effects on human health and can be seen when comparing the number of asthmatic adults living in urban versus in rural settings.34 While many exposure systems provide multiple samples at once, multiple systems are necessary with an abundance of large equipment to perform spatial resolution.
By bringing the lab to the field, the time of analysis can be decreased by using the whole cell as a sensor. Following known biological mechanisms and endpoints can aid in the determination of the aerosol composition and size. Due to slow clearance methods, including mucociliary clearance, phagocytosis, and translocation, these particles are often interacting with cells for approximately days to weeks3 generating oxidative stress, inflammation, and even cell death. These biological endpoints can be the starting points for adverse outcome pathways for cardiovascular disease or chronic obstructive pulmonary disease. In addition, Wiemenn et al. performed an array of in vitro assays to compare with literature values for short term in vivo inhalation toxicity.35 In vivo response was predicted with two of four positive results from testing cytotoxicity via lactate dehydrogenase release, oxidative stress from glutathione reduction and hydrogen peroxide formation and release, and inflammation potential from the tumor necrosis factor alpha gene. Out of ten nanosized metal oxides tested, six tested as active (titanium oxide, zinc oxide, and four different cerium oxide) using exposures in vitro with confirmation in vivo.
In order to study the effects of aerosols in an occupational setting, our lab developed the PIVEC for exposures in the field. Additionally, the PIVEC can be worn for personal sampling to monitor and investigate inhalation exposure like the 37 mm filter cassette36 or multiple systems can be used to achieve spatial resolution within a given area. In this protocol, the characterization and use of the PIVEC is discussed. After exposure, the biological effects are observed through cytotoxicity assays.
Operators must wear personal protective equipment (e.g. lab coat, gloves, goggles) when performing steps 1, 2, 3, 5, and 6.
1. Preparation of Materials
2. Generation of Dry Aerosol
NOTE: Operators should perform aerosol generation in a fume hood.
3. Deposition Efficiency Measurement using PIVEC
NOTE: Operators should perform aerosol exposures in a fume hood.
4. Calculation of Deposited Dose and Deposition Efficiency
NOTE: Knowledge of the deposition is important for aerosol administration and interpretation of cellular response.
5. Aerosol Exposure of Cells
NOTE: For the cell culture at the air-liquid interface the reader is referred to Blank et al.38 Operators should perform cell culture insert loading (steps 5.1.2-5.1.4) within a biosafety cabinet. Operators should perform aerosol exposures in a fume hood.
6 Statistical Methods
Occupational in vitro toxicology involves maintaining cellular viability while performing aerosol exposure. The PIVEC system is shown in Figure 2, including the temperature and humidity control and the worn PIVEC. The temperature was maintained using a battery-powered resistive heater and the aerosol humidified using increased natural humidification through a porous, wetted tube. In a controlled aerosol setting inside a laboratory, the PIVEC can be set-up for exposure shown in Figure 1. Characterization of the system allows for the determination of the deposited dose onto the cells which can then be correlated to the cellular response.
The deposition efficiency is dependent on the size of particles added to the system since deposition forces include impaction, sedimentation, and diffusion. In addition, the deposition is dependent on flow rate, exposure time, and the characteristics of the exposure system. With this information, the deposition can be predicted. The deposition efficiency of the PIVEC has been determined through two methods, gravimetric analysis, and particle number counting. Equations 1 and 2 were used to determine the deposition efficiencies in Table 1 using the filter based, scanning mobility particle sizer (SMPS), and optical particle sizer (OPS) measurements, Figure 3. Increased deposition is observed overall for the 24 well design than the 6 well design and slightly decreases for 100 nm in comparison to 40 nm and 800 nm copper particles. The deposition in the 24 wells is very uniform over the insert, however, deposition in the 6 well design is lacking uniformity as most of the particles deposit near the center of the insert. Post-exposure analysis can be expedited by determining the dose relationship between the 37 mm filter and the cell culture insert, decreasing the necessity to determine the dose after cellular exposure. Comparison of the deposition within the 37 mm filter cassette and the PIVEC shows a strong correlation for all sizes and wells with a Pearson correlation of above 0.7, however only for 800 nm is the correlation significant with a p<0.05, see Figure 4. Compared to similar systems throughout literature,11,12 the deposition efficiency of the PIVEC over the range of particle sizes tested is comparable or increased over reported values, observed in Figure 5. The deposition efficiency within the PIVEC can be improved through minimizing losses to the system using electrostatic dissipative or conductive plastic or similar material to design the PIVEC.
If filters are not well dried in a humidity-controlled environment before exposure, excess water increases the initial mass and can produce non-physical, negative deposited doses. Variable flow rates can also promote inconsistent deposited doses within the system. To avoid these issues, allow at least one day prior to and after exposure for filters to dry in a humidity-controlled environment.
Cellular responses will be affected by the deposited dose and may be affected by exposure duration if the cells are not properly maintained. A549 cells, an alveolar epithelial carcinoma cell line, were exposed for 10 min to varying sizes of copper nanoparticles at a flow rate of 0.5 LPM. Alternative perpendicular flow exposure systems use between 0.005 LPM and 1.5 LPM for a sustained exposure period whereas this method uses a moderate flow rate during a rapid exposure. Using the mass-based deposition efficiency measured and administered dose, 1.58 ± 0.04 mg/cm2 of 40 nm copper, 0.30 ± 0.00 mg/cm2 of 100 nm copper particles, and 0.32 ± 0.01 mg/cm2 of 800 nm copper particles were deposited onto the cells. Cytotoxicity and oxidative stress were observed within the first twenty-four hours post-exposure. Cytotoxicity was measured using the release of lactate dehydrogenase (LDH) from damaged cells immediately, 4 hours, and 24 hours post-exposure. There was no significant toxicity from copper nanoparticles below 1.62 mg/cm2 within 4 hours of exposure, Figure 6b. Twenty-four hours post-exposure there was a statistically significant decrease in cell viability of less than 20% viability for cells exposed to copper nanoparticles. The intracellular oxidative stress was investigated using the DCFH-DA assay through the oxidation of DCFH by reactive oxygen species within the first two hours post-exposure, Figure 6a. Significant levels of oxidative stress were measured at thirty minutes post-exposure for cells exposed to copper nanoparticles of all sizes. The level of oxidative stress continued to increase at similar rates for all sizes tested within the two hours observed.
Success of the exposures can be determined through the repeatability of cellular response. Two acceptance criteria proposed for cytotoxicity assays include: 1) the % total LDH leakage for the positive control should be greater than 50%, and 2) the positive and sample replicate coefficient of variations should be within 50%.39 If these criteria are not met, this suggest differences between experiments, such as altered deposition amounts, or poor humidity or temperature control. In addition, observing cytotoxicity measurements can lead to understanding ofthe experimental controls and aid in determining the error. When the flow rate is too high, cells may die from high amounts of shear stress. By lowering the flow rate, the stress upon the cells from the flow can be decreased.
Figure 1. Schematic of Dry Dispersal System and Experimental Set-up. Please click here to view a larger version of this figure.
Figure 2. PIVEC Design. A) Full Design Pictured with 30 cm Tall Box for Comparison. B) PIVEC with Temperature and Humidity Control Pictured with 30 cm Tall Box for Comparison. C) Worn PIVEC. D) PIVEC Top Piece. E) Cell Culture Adapter for 24 well (left) and 6 well (right). F) Bottom Piece. Please click here to view a larger version of this figure.
Number Based Deposition Efficiency (%) | Mass Based Deposition Efficiency (%) | ||||||
6 well | 40 nm | 17.83 | ± | 32.13 | 5.85 | ± | 0.85 |
100 nm | 0.47 | ± | 4.06 | 5.11 | ± | 0.94 | |
800 nm | 3.70 | ± | 35.00 | 6.39 | ± | 1.01 | |
24 well | 40 nm | 1.43 | ± | 2.43 | 12.61 | ± | 1.34 |
100 nm | 1.37 | ± | 19.45 | 2.95 | ± | 0.75 | |
800 nm | 6.98 | ± | 3.93 | 15.95 | ± | 0.53 |
Table 1. PIVEC Deposition Efficiency.
Figure 3. Particle number concentration of copper nanoparticle aerosols. A) SMPS measurement. B) OPS measurement. Please click here to view a larger version of this figure.
Figure 4. Relationship between deposition on cell insert and SKC 37 mm filter. Error ± 0.002 mg. A) 40 nm copper particles. B) 100 nm copper particles. C) 800 nm copper particles. Please click here to view a larger version of this figure.
Figure 5. Deposition Efficiency of PIVEC Compared to Perpendicular Flow Exposure Systems in Literature. Literature Values from Perpendicular Flow Exposure Systems are Plotted in Solid Markers and PIVEC Values are in Empty Markers. Please click here to view a larger version of this figure.
Figure 6. Cellular Response to Copper Nanoparticles Post-Exposure (PE). For all measurements, n=3 and p<0.05. A) Oxidative Stress determined using the DCFH-DA Assay. B) Cytotoxicity determined using the LDH Assay. Please click here to view a larger version of this figure.
Filter cassettes provide a simple, inexpensive method of collecting aerosols in the breathing zone; however, aerosol samples extracted from filters do not represent the entire aerosol (i.e. gases, volatiles, and particulates) and consequently limit the assessment of related biological effects. Using the initial design of the 37 mm filter cassette, the PIVEC is designed to maintain portability and mimic the in vivo deposition of particles from inhalation. The PIVEC is significantly smaller than current ALI exposure systems, approximately the size of a tissue box with temperature and humidity control included. The size is similar to personal cascade impactors while offering data on cellular response to the aerosol in addition. While the PIVEC contains space for one sample in comparison to other current exposure systems, the small size permits multiple systems to be used at once, therefore, increasing the sample size and allowing for spatial distribution to be monitored.
There are several critical steps in the protocol. To determine the deposition efficiency, it is critical to properly condition the filters, as described in Step 1. Additionally, it is important to properly weigh the copper nanoparticles prior to exposure and the filter post exposure to determine the mass-based deposition efficiency. The number deposition efficiency must be determined using the difference in deposition between the initial aerosol concentration and final concentration determined using the particle counters. If the deposition efficiency is not properly determined, it is difficult to determine the biological response differences between aerosol types. Modifications to the deposition include altering the deposition. Deposition can be increased by changing the flow rate and exposure duration. This may also influence the cell viability with the potential of drying out the cells. The conditioning of aerosols is often performed to compensate for physiological attributes such as body temperature and humidification in the airways. Increased humidity, over 50%, mimics inhaled air and decreases cell death due to vehicle exposure.43 When temperature and humidity are not well controlled, the cellular response can be influenced. By decreasing the flow rate, additional particles of all sizes will deposit, increasing the deposition. Exposure duration is proportional to deposition, allowing more particles to deposit over an extended experimental period. Conditioning of the aerosol is important when increasing the exposure duration so that the cells do not dry out which can affect biological responses.
The PIVEC uses perpendicular flow to deposit particles via impaction, sedimentation, and diffusion onto the cells which have the potential to dry out cells through the flow and added stress on the cells. The deposition efficiency is dependent on the size of the particle due to the forces of deposition. Many of the current exposure systems with perpendicular flow have a deposition efficiency of below 10% and much closer to 1%. The PIVEC has a deposition efficiency determined by gravimetric analysis of 3% to 16% for a 24 well cell culture insert. The particle number-based deposition efficiency is approximately 1% to 7% for the same insert. When using a 6 well cell culture inserts, these numbers decrease, with the exception of the smallest particle size, due to the streamlining of the aerosol as more particles are confined in the cell culture insert without space for the flow to develop. The 6 well cell culture inserts permit the aerosol streams to bypass deposition. Other perpendicular flow systems have been characterized on a mass basis using 60 nm fluorescein particles40, 80 nm and 180 nm copper particles16, and 60 nm adipic acid particles41. The resulting deposition efficiencies are generally between 0.05% and 11%. On a number basis, these systems have been characterized using polystyrene particles12,15, carbon nanoparticles12, and silica particles42, yielding efficiencies between 0.2% and 11%. The PIVEC provides, similar or increased, deposition in comparison to available cellular exposure devices.
Cell exposures were performed using copper nanoparticles of 40 nm, 100 nm, and 800 nm. Cell viability was defined using the LDH assay with no significant decrease in viability within four hours post-exposure. The LDH assay was used with a commercial exposure system for 74 ng/cm2 after 2 hours and 148 ng/cm2 after 4 hours of 9.2 nm copper oxide particles44 and 1 µg/cm2 deposited after a sequential exposure 4 hours, incubation for 2 hours, then another exposure for 4 hours of 25 nm copper oxide particles40, both measuring significant decreases in cell viability. Cytotoxicity was observed in another system for 1.6-7.6 µg/cm2 of 180 nm particles copper nanoparticles16 measured by trypan blue dye exclusion. Exposures of 15 minutes to 40-80 nm copper oxide nanoparticles decreased viability, measured 24 hours post exposure. The lower toxicity observed using the PIVEC was likely due to the shorter exposure time of 10 minutes and shorter post exposure measurement times. After four hours post-exposure, viability of cells exposed in the PIVEC decreased. Cells exposed to humid air controls observed no significant cytotoxicity, in agreement with other studies 9,40,44,45,46. The oxidative stress was determined using the DCFH-DA assay. The production of reactive oxygen species, such as hydrogen peroxide or oxygen radicals, generated an amount of stress that can lead to growth arrest or cell death. After the cell exposures, there was a minimal increase in oxidative stress for cells kept in the incubator as control and for cells exposed to humid air. Elevated oxidative stress occurred for all particle sizes, increasing within 30 minutes post-exposure. Within one study, 9.2 nm copper oxide particles44 and 25 nm copper oxide particles40 also induced oxidative stress measured via the carboxy-DCFH-DA assay. Increased exposure time from 2 hours to 4 hours elevated oxidative stress for 9.2 nm copper oxide particles44. After four hours of exposure, the 9.2 nm copper oxide particles produced a similar oxidative response as the sequential exposure of 25 nm copper oxide particles40, suggesting that exposure duration has a higher influence on oxidative stress response than particle size. Cells exposed within the PIVEC produced elevated oxidative stress compared to that study, however, the particles exposed in the alternate system are copper oxide and will dissolve less quickly than the copper exposed within the PIVEC. Studies performed on the dissolution of copper based on the composition and size of particles agree that larger particles and the metal oxides release ions slower than metal particles or their nanoparticle counterparts16,47,48,49 As the humid air controls did not produce significant amounts of oxidative stress compared to the incubator control, the influence of copper particles to induce oxidative stress is consistent within the PIVEC to similar in vitro exposure systems. The biological responses observed using the PIVEC suggest that the PIVEC is an appropriate system for cellular exposure.
While the characterization and cellular studies of the PIVEC agree well with literature,9,16,40,44,46 the device has limitations. The small design decreases the number of samples that can be exposed simultaneously within a single device in comparison to other exposure systems. Other systems allow for at least three cellular exposures to the same aerosol9,12 facilitating replicate measurements. Although the PIVEC only allows for one insert per system, the small size allows for multiple systems to easily be used, helping to mitigate this issue. While other personal monitoring systems do not use cells, the PIVEC must be kept near vertical to reduce spilling the cell culture media necessary to preserve cellular viability. Although the PIVEC has not yet been optimized for specific particle ranges (e.g. PM10, PM2.5, PM0.1), the PIVEC has been characterized for a range of particle sizes. Similar to other perpendicular flow systems, the PIVEC shows a decreased ability to deposit particles near 100 nm in diameter, while 40 nm and 800 nm particles deposited with similar efficiency.
The analysis of cells post-exposure can be expedited by performing a parallel collection of aerosols within the PIVEC and an SKC 37 mm filter cassette. A high correlation of gravimetric based deposition allows for the estimation of the particle mass collected on the cells from data collected using 37 mm filters. Comparison of the insert to the filter cassette reduces the need to collect additional samples and additional measurements to determine the dose. Work in progress includes the integration of a real-time monitoring system for cytotoxicity, oxidative stress, or other biological endpoints. The small size of the PIVEC allows it to be used in a variety of settings, such as on the body as a personal monitor, on a drone above a chemical plant, or outside in the environment for spatial resolution. This method has shown the use of the PIVEC for the collection of aerosol particles onto cell cultures grown at the ALI. By conditioning the aerosol to 37 ± 1 °C and > 80% relative humidity, cellular viability can be maintained during acute exposures. This method is appropriate for both liquid droplet and solid particle-based aerosols and has been shown to deposit particles between 40 nm and 800 nm in cell culture inserts. The versatility of the PIVEC allows this method to be used in multiple settings with a variety of biological endpoints.
The authors have nothing to disclose.
The authors would like to thank Boris Solomonov and the Virginia Commonwealth Innovation Machine Shop for help with rapid prototyping the device. The authors would also like to thank Cristian Romero-Fuentes of the Lewinski Group, Dr. Vitaliy Avrutin, Dr. Dmitry Pestov, and the Virginia Commonwealth Nanomaterials Core Characterization Facility for their help with particle characterization. This work was supported by startup funds provided to Dr. Lewinski by the College of Engineering at Virginia Commonwealth University.
Scanning mobility particle sizer (SMPS) | TSI, Inc. | 3910 | NanoSMPS |
Optical particle sizer (OPS) | TSI, Inc. | 3330 | |
Stainless Steel Pipe, 4" Long | McMaster-Carr | 4830K116 | Standard-Wall 304/304L, Threaded on Both Ends, 1/8 Pipe Size |
Brass Ball Valve with Lever Handle | McMaster-Carr | 4112T12 | Compact High-Pressure Rating, 1/8 NPT Female |
Steel Pipe, 2" Long | McMaster-Carr | 7753K121 | Standard Wall, Threaded on One End, 1/8 Pipe Size |
HEPA filter | GE Healthcare | 09-744-12 | HEPA-Cap Disposable Air Filtration Capsule |
Vacuum Generator | PISCO USA | VCH10-018C | |
PIVEC | VCU | For design please contact authors | |
Resistive heater | |||
1/4" barbed connectors | Zefon International, Inc. | 459743 | |
Porous tubing | Scientific Commodities, Inc. | BB2062-1814A | Hydrophilic 10 um pores |
Battery power bank | |||
Cell culture insert | Fisherbrand | 353095 | 24 well plate insert |
Filter Forceps | Fisherbrand | 09-753-50 | |
Transfer Pipette | ThermoScientific | 13-711-27 | |
Glass Fiber Filters | SKC | 225-7 | Binder-Free Type AE Filter 37 MM 1.00 um pore |
Ultra Micro Balance | A&D | BM-22 | Housed in environmental chamber |
37 mm filter cassette | SKC | 225-3250 | Filter Cassette Blank, 37 mm, Clear Styrene |
Variable flow vacuum pump | SKC | 220-5000TC | AirChek TOUCH, 5 to 5000 mL/min |
Copper Particles | U.S. Research Materials, Inc. | US1090 | 40 nm |
Copper Particles | U.S. Research Materials, Inc. | US1088 | 100 nm |
Copper Particles | U.S. Research Materials, Inc. | US1117M | 800 nm |