This study provides an updated home-based tDCS protocol that enables subjects to receive the beneficial effects of tDCS at home with an easy to use device with settings to control the use and dosage, enhancing the feasibility for long-term use at home.
Transcranial direct current stimulation (tDCS) is a non-invasive brain stimulation (NIBS) method, which modulates the membrane potential of neurons in the cerebral cortex by a low-intensity direct current. tDCS is a low-cost technique with minimal adverse effects and easy application. This neurostimulation method has a promising future to improve pain therapy, treatment of neuropsychiatric disorders, and physical rehabilitation.
Current studies demonstrate the benefits of using tDCS over consecutive multiple sessions. However, the daily displacement to the specialized centers, travel costs, and disruptions to daily activities are some of the difficulties faced by patients. Thus, to be more comfortable, easy-to-use, and not disrupt daily commitments, a home-based tDCS was designed. Therefore, the objective of this study was to evaluate the feasibility of a portable tDCS device for home use in healthy subjects and fibromyalgia patients.
Despite increased tDCS use and a reasonably large body of research on the effects across a range of clinical conditions, there is a limited amount of research on developing secure devices that guarantee the dose and contain a block system to avoid excessive use. Therefore, we used a tDCS device with a security system to permit daily use for 20 minutes with a minimal interval of 12 hours between sessions. A programmer preconfigures the equipment, which has a neoprene cap that allows the electrode positions in any assembly, according to individualized protocols for treatments or research. After, researchers can assess the effectiveness of treatment, and its adherence using information kept in the device software.
Results suggest that the device is feasible for home use, with proper monitoring of adherence and contact impedance. There were reports of a few adverse effects, which do not differ from those reported in the literature in studies with the treatment under direct supervision.
Non-invasive brain stimulation (NIBS), such as transcranial magnetic stimulation (TMS) and tDCS, is emerging as a therapeutic option for various conditions, such as neuropsychiatric disorders1,2, pain syndromes3,4,5,6,7, and rehabilitation8,9. tDCS is a form of neurostimulation that modulates the membrane potential of neurons in the cerebral cortex by a low-intensity direct current (1-2 mA). The electrical current is applied directly to the scalp at the targeted brain area (i.e., primary motor cortex, prefrontal dorsolateral, etc.) through sponge electrodes soaked with saline solution or rubber electrodes with conductive gel10. It is a low-cost technique, with easy application and practically no adverse effects.
However, the use of long-term tDCS becomes impracticable for many patients with chronic disease. When we analyze the current evidence about tDCS, we can observe that most protocols used fewer than 20 sessions and also, the treatment was administered in a specialized center6. Nonetheless, it is a frequent complaint of patients that they spend time and money to go the center.
In the real world, there are several everyday difficulties of patients receiving this kind of therapy because frequently they are suffering from refractory pain and/or they have severe psychiatric or neurological disorders. Many of them have difficulties walking or are wheelchair users. An ideal device should guarantee consecutive sessions for long-term therapies as well as be able to run more extensive clinical trials with larger sample sizes. Furthermore, the device should have a block system to guarantee the dose of stimulation while avoiding any abusive or excessive use. Also, the development of a device to be used without direct supervision can test its effectiveness in different samples, either as a unique method or combined with pharmacological treatments, cognitive tasks or physical functions11,12. Thus, we need to take into account that the repeated sessions are an intrinsic characteristic of this kind of treatment according to a vast set of evidence. A clinical response usually requires repeated sessions of stimulation during extended periods5,13,14. In this context, there exists a gap to develop a portable tDCS device for home use, at the lowest cost, that is easy to handle and guarantees the safety and benefits observed in studies under direct supervision at care centers.
Presently the most advanced system available to use tDCS at home requires remote assistance, which is a considerable barrier to large scale treatment, especially in the long-term. In fact, in real life, it is not feasible to maintain a long-term intervention if the patient and the operator controlling the device remotely need to be connected to a video conferencing platform simultaneously. The remote support is a barrier to maintain the treatment because both the patients and the operator controlling the remote need to be available at a pre-determined time of day for the treatment to occur. This is not practical, and it withdraws the patient's autonomy to decide what is the best moment for the treatment session according to their daily life routine15,16. Taking this into account, a tDCS device to be used at home should be practical and friendly for users. The tDCS device presented in this study is simple and with a short training can give the ability for the participants' to use the apparatus at home.
An essential characteristic of our device is that it was designed for home use. For this reason, we include a block system to avoid any indiscriminate use. Thus, an untrained individual can use the device, and the technique can be reproduced across studies to make comparison of results possible between clinical trials11,15,17. Aiming to reduce difficulties related to long-term treatment, we developed a protocol using a home-based tDCS device; here we present a protocol for a safe and straightforward portable tDCS device, preventing its improper use16. We validated this protocol of the home-based tDCS in both healthy subjects and fibromyalgia patients, as well as showed its feasibility as attested by electrophysiological parameters (i.e., motor evoked potential), and clinical measures.
We included healthy subjects and fibromyalgia patients in this study. Fibromyalgia is a syndrome that comprises chronic widespread musculoskeletal pain accompanied by other symptoms such as depressive symptoms, fatigue, sleep disturbance, and morning stiffness. Many patients experience continued symptoms despite initial pharmacological treatment at the maximum tolerated dose. Hence, we advise studying different treatment modalities, including non-pharmacological measures such as the NIBS techniques, in this context, the home-based tDCS. tDCS is a potentially viable option in the treatment of chronic pain, as demonstrated in several studies4,6,7. The development of a home-based tDCS allows the creation of long-term clinical trials, and a better understanding of the cumulative effect of tDCS for chronic disorders (i.e., chronic pain, depression, rehabilitation after stroke, etc.).
Despite the fact that tDCS use has increased and there is a reasonably large body of research on its effects and impact across a range of clinical conditions, a surprisingly limited amount of research has been conducted on developing a device that can be feasibly and securely used at home. Therefore, the objective of this study was to evaluate the feasibility of the tDCS device for home use (monitoring of adherence and contact impedance) in healthy subjects and fibromyalgia patients.
This study was approved by the institutional ethics committee at the HCPA and obtained written informed consent from all subjects. Details of the study with a sample of healthy subjects and fibromyalgia patients are in Figure 1.
1. Participant Recruitment and Screening
2. Proposed Equipment
Note: The Biomedical Engineering department at the Hospital de Clínicas de Porto Alegre in partnership with the Laboratory of Pain & Neuromodulation developed the home-based tDCS. To solve the difficulty of correct placement of the electrodes without the help of another person, we created a tDCS device and a cap that allows the use of two electrodes in several assemblies.
3. Baseline Protocol Visit – tDCS Training
4. Home-based tDCS Sessions
A total of n=20 healthy subjects were recruited (8 men / 12 women) between March 2015 and April 2016. One participant was excluded, due to having performed only 50% of the sessions. Thus, a total of 19 subjects completed the experiment. The mean age was 26.31 years (±4.89).
A total of n=8 fibromyalgia patients were recruited between November 2016 and April 2017 and randomized to two groups: FM active and FM sham. The mean age was 49.5 years (±8.48). Patients who experienced the sham condition received a questionnaire of "blinding validation" asking what stimulation they believe they received. All of the patients answered that they had received active stimulation.
We measured adherence by counting the number of completed sessions, as verified by the records on the software. In the study with HS, there was a 90% adherence to the proposed sessions, totaling 19 training sessions and 171 valid sessions (total: 190). In the study with FM patients, there was a 94% adherence to the proposed stimulation sessions, totaling eight training sessions and 227 valid sessions for FM active and 225 valid sessions for FM sham (total: 460). The total number of valid sessions in the healthy and fibromyalgia sample was 650 sessions, with overall adherence of 93%.
The contact impedance of the sample of HS had a mean of 2.93 (±1.04) and the study with FM patients (only active tDCS) was 2.82 (±1.15). The mean of contact impedance taking into consideration both samples (n = 416) was 2.87 (±1.13) (Figure 9).
The adverse effects were similar between the HS and FM active groups (although the FM group had some items more frequently than the HS group). The sham group had adverse effects, but these were smaller than those found in the other groups (Figure 10).
A total of n=47 sessions were not performed by the participants (HS=19 and FM=28). The justifications were: prescheduled social activities, forgetfulness, and technical problems in the cable and battery. Three participants needed to replace the battery, which did not recharge properly. None of the participants interrupted a session during stimulation. When the device presented malfunctions, it was changed, or the participant was oriented to solve the problem. No cases remained unresolved.
Figure 1: Flow of study
Details of study design – healthy subjects and fibromyalgia patients.
Figure 2: Schematic drawing of the electrode construction
(A) Right side view.(B) Left side view.(C) Upper view.(D) Bottom view.
Figure 3: Stimulation area
(A) Area of stimulation healthy subjects: anodal stimulation over left M1 and cathodal stimulation over contra-lateral supra-orbital area. (B) Area of stimulation FM patients: DLPFC area, with anodal stimulation over left side.
Figure 4: Screen sequence
(A) Opening screen. (B) Battery status. (C) Date and time. (D) Command "wear the cap". (E) Command "inject saline". (F) Command to start the session. (G) Beginning of the stimulation.
Figure 5: Saving the session when stimulation ceases
After the end of the stimulation, the tDCS device automatically powers off and stimulation ceases, saving the session.
Figure 6: Alarm warning screen
(A) High resistance. (B) Alarm warning: "adjust the cap and inject extra saline".
Figure 7: Shut down screen if the session needs to be stopped during stimulation
If the session needs to be stopped during stimulation, the central button can be pressed for a few seconds, followed by pressing the switch off button and a message "Shut down?" will appear.
Figure 8: Typical curves of current intensity versus contact impedance during a tDCS session
An example of current intensity and contact impedance curves (healthy subject). Current intensity (mA): Red line. Resistance (kΩ): Blue line. Please click here to view a larger version of this figure.
Figure 9: Contact Impedance
Mean of contact impedance – healthy subjects and the fibromyalgia patients active tDCS (n= 416).
Figure 10: Results of adverse effects
Comparison between groups of participants: healthy subjects (HS), active tDCS fibromyalgia patients (FM active) and sham tDCS fibromyalgia patients (FM sham). Please click here to view a larger version of this figure.
The device presented an adequate user interface via the LCD (liquid crystal display) and was user-friendly and easy to use. Our home-based tDCS equipment proved to be suitable for long-term use since adherence was higher than 90% even when participants were instructed to perform 60 home-based sessions. This enables the reproducibility of research conditions13,21 necessary for the idealization and confidence in studies with this device.
It is challenging to maintain adherence when research involves consecutive sessions without intervals on weekends. This fact is one of the hypotheses raised to justify why healthy participants had slightly lower adherence than fibromyalgia patients. Also, a rationale to explain the current results may be that patients with chronic pain have the possibility of pain relief as a motivator, which does not occur in healthy subjects.
The use of home-based tDCS allows participants to choose what time of day to receive the treatment, therefore accommodating patients' schedules. Also, other research protocols comparing different times of tDCS application can be devised using this device, linking to other areas of study, such as chronobiology of pain, for example. Besides, the device proved to be safe for use with no real-time supervision of the sessions due to its security system. The device can be preconfigured to allow for an interval between sessions, duration of stimulation, intensity, maximum resistance, number of courses and active or sham stimulation predefined by a programmer. Therefore, it is possible to program for daily use of 20 min with a minimal interval of 12 hours between sessions (or other configurations). Thus, researchers can monitor adherence using information kept in the device software. Therefore, our results proved that device features ensure reliability and reproducibility of the stimulation, which is a crucial factor in the development of clinical trials.
Electrode-skin contact impedance is an important variable to ensure the safety of tDCS use16,21. High impedance may be related to poor coupling between the electrode and the skin or dryness of the electrode's sponge. These situations may increase current density and lead to skin lesions22. In our study, the use of 6 mL of saline solution was sufficient to guarantee low contact impedance and was adequate and similar in both groups.
The adverse effects produced by our device do not differ from those reported in ambulatory use. Thus, our home-based tDCS resembles traditional equipment, and it is feasible, reliable and safe for use at home in consecutive sessions for long-term therapies, allowing the possibility of running more extensive clinical trials with larger samples.
The authors have nothing to disclose.
The authors would like to thank the collaborators from the department of Biomedical Engineering at the HCPA for their assistance and support. Acknowledgments to Mário Eugênio Saretta Poglia, Francisco Mauro and Cássia Garzela for the photos and filming and Emillie Pinheiro and Ketlin Bazzo for their participation in the video and photos. We also thank the União das Américas Faculty, which provided the place for video recording. This research was supported by grants and material support from the following Brazilian agencies: Committee for the Development of Higher Education Personnel (CAPES), National Council for Scientific and Technological Development (CNPq), Postgraduate Program in Medical Sciences at the School of Medicine of the Federal University of Rio Grande do Sul, Postgraduate Research Group at the Hospital de Clínicas de Porto Alegre – FIPE HCPA and FINEP (Research in Biomedical Engineering) – 02/2013: 04.13.0465.00.
Fisiologic solution 0.9% | Eurofarma | Sterile and pyrogenic solution, trilaminated packaging, 500 mL, transparent, two equal nozzles for the introdution of the equipment and the needle, protective sealing, registration at the ministery of health – 1.0043.1047.004-6 | |
Vegetable sponge | Scotch-Brite 3M | Vegetable sponge cloth, package with a unit of 18×22 cm. The vegetable cloth is a composite material of regenerated cellulose, cotton, dimethyl benzylammonium chloride and magnesium chloride (fungicidal substance). It has great absorption power of liquids in general. Available in colors: blue, pink and yellow. | |
Serynge | Advantine | 10 mL Plastic Serynge | |
Electrodes | Three layers electrode: Flexible Vinyl material, conductive rubber and vegetal sponge | ||
Cap | Neoprene of 4 mm thickness – Manufactured by Biomedical Engineering Department of Hospital de Clínicas de Porto Alegre | ||
tDCS device | Manucfated by Biomedical Engineering Department of Hospital de Clínicas de Porto Alegre |