Summary

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector

Published: July 25, 2014
doi:

Summary

Trace explosive vapors of TNT and RDX collected on sorbent-filled thermal desorption tubes were analyzed using a programmed temperature desorption system coupled to GC with an electron capture detector. The instrumental analysis is combined with direct liquid deposition method to reduce sample variability and account for instrumentation drift and losses.

Abstract

The direct liquid deposition of solution standards onto sorbent-filled thermal desorption tubes is used for the quantitative analysis of trace explosive vapor samples. The direct liquid deposition method yields a higher fidelity between the analysis of vapor samples and the analysis of solution standards than using separate injection methods for vapors and solutions, i.e., samples collected on vapor collection tubes and standards prepared in solution vials. Additionally, the method can account for instrumentation losses, which makes it ideal for minimizing variability and quantitative trace chemical detection. Gas chromatography with an electron capture detector is an instrumentation configuration sensitive to nitro-energetics, such as TNT and RDX, due to their relatively high electron affinity. However, vapor quantitation of these compounds is difficult without viable vapor standards. Thus, we eliminate the requirement for vapor standards by combining the sensitivity of the instrumentation with a direct liquid deposition protocol to analyze trace explosive vapor samples.

Introduction

Gas Chromatography (GC) is a core instrumental analysis technique of Analytical Chemistry and is arguably as ubiquitous as a hot plate or balance in a chemistry laboratory. GC instrumentation can be used for the preparation, identification, and quantitation of a multitude of chemical compounds and can be coupled to a variety of detectors, such as flame ionization detectors (FIDs), photo-ionization detectors (PIDs), thermal conductivity detectors (TCDs), electron capture detectors (ECDs), and mass spectrometers (MS), depending on the analytes, methodology, and application. Samples can be introduced through a standard split/splitless inlet when working with small sample solutions, specialized headspace analysis inlets, solid phase micro-extraction (SPME) syringes, or thermal desorption systems. GC-MS is often the standard technique used in validation and verification applications of alternative or emerging, detection techniques because of its utility, flexibility, and identification power with established chemical databases and libraries.17 GC and its related sampling and detecting components is ideal for routine chemical analysis and more specialized, challenging analytical applications.

An analytical application of increasing interest to military, homeland security, and commercial enterprises is trace explosive vapor detection, with detection including identification and quantitation. Trace explosive vapor detection is an unique analytical chemistry challenge because the analytes, such as 2,4,6-trinitrotoluene (TNT) and cyclotrimethylenetrinitramine (RDX) have physical properties that make them especially difficult to handle and separate using broader, more generic chemical analysis methodologies. The relatively low vapor pressures and sub parts-per-million by volume (ppmv) saturated vapor concentration, combined with relatively high sticking coefficients, necessitate special sampling protocols, instrumentation, and quantitation methods.812 A GC coupled to an Electron Capture Detector (ECD) or mass spectrometer (MS) is an effective method for quantitating explosive analytes, specifically dinitrotoluene (DNT), TNT, and RDX.6,1317 GC-ECD is particularly useful for nitro-energetic compounds because of their relatively high electron affinity. The U.S. Environmental Protection Agency (EPA) has created standard methods for explosive analyte detection using GC-ECD and GC-MS, but these methods have focused on samples in solution, such as ground water, and not samples collected in the vapor phase.2,1823 In order to detect explosive vapors, alternative sampling protocols must be used, such as vapor collection with sorbent-filled thermal desorption sample tubes, but quantitative detection remains difficult due to lack of vapor standards and calibration methods that do not account for sample tube and instrumentation losses.

Recently, quantitation methods using thermal desorption systems with a cooled inlet system (TDS-CIS), coupled to a GC-ECD have been developed for TNT and RDX vapors.24,25 The losses associated with the TDS-CIS-GC-ECD instrumentation for trace explosive vapors were characterized and accounted for in example calibration curves using a direct liquid deposition method onto sorbent-filled thermal desorption sample tubes. However, the literature focused on instrumentation characterization and method development but never actually sampled, analyzed, or quantitated explosive vapors, only solution standards. Herein, the focus is on the protocol for sampling and quantitating explosive vapors. The protocol and methodology can be expanded to other analytes and trace explosive vapors, such as Pentaerythritol tetranitrate (PETN).

Protocol

1. Instrument Preparation

  1. Ensure the instrument, oven, and detector are at RT. Turn off gas flow to the inlet and detector.
  2. Remove the TDS from the GC. Consult the manufacturer’s user manual for the instrument-specific procedure.
  3. Remove the TDS adaptor from the CIS inlet and remove the liner from the CIS.
  4. Inspect the CIS inlet for particles and debris while the liner is removed. Clean any visible debris with compressed air, or preferably nitrogen.
  5. Attach a new graphite ferrule to a new CIS liner using the manufacturer provided tool and instructions for ferrule-to-liner binding.
  6. Insert the liner with the attached graphite ferrule into the CIS. Replace the TDS adaptor and re-mount the TDS.
  7. Remove a new column from its packaging and remove the silicone protection from the ends of the column.
  8. Insert a nut and ferrule onto each end of the column. Use an ECD detector nut and ferrule for one end of the column and a CIS ferrule for the opposite end of the column.
  9. Using a ceramic column cutting tool, remove approximately 10 cm from each end of the column. Ensure the nuts and ferrules remain on the column but away from the end of the column to avoid clogging and debris.
  10. Secure the column into the oven using the instrument manufacturer guidelines. Insert the column into the inlet. Connect the other end of the column to the detector port. The depth of insertion is specific to instrument, inlet, and detector manufacturer. See the user manual and specifications for the exact column insertion depth.
    NOTE: A pre-bake may be required for the column before connecting the opposite end of the column to the detector ports. Consult the column and instrument manufacturer documentation to determine if a pre-bake is required.
  11. Gently hand-tighten nuts and ferrules onto their respective ports for the inlet and detector. Using a wrench, tighten with approximately a quarter turn of rotation the nuts and ferrules. Too much force or over-tightening will damage the ferrules causing leaks or the column to break and clog.
  12. Bake out the TDS, inlet, column, and detector. A typical bake out consists of setting the temperature for all zones to just below the maximum operating temperature (typically 300 °C) while flowing carrier gas for at least 2 hr.
  13. Cool all zones and retighten all nuts and ferrules to ensure leak-free operation. Heating and cooling during the bake out will cause the nuts and ferrules to loosen, which can introduce leaks.
  14. Load, or reload, the instrument method using the software interface. Verify correct temperatures and flow rates have been achieved. Instrumentation is ready for analysis.

2. Preparation of Standards

  1. Remove 1,000 ng µl-1 3,4-DNT, 10,000 ng µl-1 TNT, and 10,000 ng µl-1 RDX from the freezer or refrigerator and allow the three stock solutions to reach RT.
  2. Dispense 100 μl of stock 1,000 ng µl-1 3,4-DNT and add 900 µl of acetonitrile into an amber sample vial.
  3. Dispense 100 μl of the 100 ng µl-1 3,4-DNT solution from Step 2.2 and add 900 µl of acetonitrile into an amber sample vial.
  4. Dispense 150 µl of the 10 ng µl-1 3,4-DNT solution from Step 2.3 and 4,850 µl of acetonitrile into an amber sample vial. This is the internal standard for direct liquid deposition.
  5. Dispense 100 μl of stock 10,000 ng µl-1 TNT solution, 100 μl of stock 10,000 ng µl-1 RDX solution, and 800 µl of acetonitrile into an amber sample vial.
  6. Dispense 100 μl of the 1,000 ng µl-1 TNT and RDX solution in Step 2.5 and 900 µl of acetonitrile into an amber sample vial.
  7. Dispense 100 μl of the 100 ng µl-1 TNT and RDX solution from Step 2.6 and 900 µl of acetonitrile into an amber sample vial.
  8. Dispense 100 μl of the 10 ng µl-1 TNT and RDX solution from Step 2.7 and 900 µl of acetonitrile into an amber sample vial. This creates the 1.0 TNT/1.0 RDX ng μl-1 solution standard ready for direct liquid deposition onto sample tubes.
  9. Dispense 60 µl of the 10 ng µl-1 solution in Step 2.7 and 940 µl of acetonitrile into an amber sample vial. This creates the 0.6 TNT/0.6 RDX ng μl-1 solution standard ready for direct liquid deposition onto sample tubes.
  10. Dispense 40 µl of the 10 ng µl-1 solution in Step 2.7 and 960 µl of acetonitrile into an amber sample vial. This creates the 0.4 TNT/0.4 RDX ng μl-1 solution standard ready for direct liquid deposition onto sample tubes.
  11. Dispense 20 µl of the 10 ng µl-1 solution in Step 2.7 and 980 µl of acetonitrile into an amber sample vial. This creates the 0.2 TNT/0.2 RDX ng μl-1 solution standard ready for direct liquid deposition onto sample tubes.
  12. Dispense 100 µl of the 1.0 ng µl-1 solution in Step 2.8 and 900 µl of acetonitrile into an amber sample vial. This creates the 0.1 TNT/0.1 RDX ng μl-1 solution standard ready for direct liquid deposition onto sample tubes.

3. Sample Collection

  1. Connect one sorbent-filled thermal desorption sample tube to a sample pump or similar equipment using a small piece of flexible silicone tubing. A red arrow is provided on the sample tubes indicating the air flow direction for sample adsorption, and it should be pointing in the direction of the silicone tubing and sample pump.
  2. Attach a piston flow meter to the sample tube at the opposite end from the sample pump attached in Step 3.1. Adjust the flow rate on the sample pump, or similar equipment, such that the flow rate is approximately 100 ml min-1 through the sample tube according to the readings from the piston flow meter. The flow rate should be set to ±5.0 ml min-1 of the 100 ml min-1 desired set point.
  3. Disconnect the piston flow meter from the sample tube and temporarily shut off the sample pump but leave the sample tube connected to the pump. The sample pump will be reactivated to begin sample collection. The sample tube is ready for collection.
  4. Place the sample tube with the still connected sample pump in the explosives vapor stream. The vapor source could be the headspace above a solid sample, an open environment, or a variety of analyte vaporization systems.
  5. Set a timer based on the approximate sampling times listed in Table 2. The sampling times are listed as a general guideline based on the suspected concentration of material in the vapor phase. These sampling times, with a flow rate of 100 ml min-1, will generally yield a mass in the center of the calibration curve, which is ideal for quantitation.
  6. Activate the sample pump and start the timer. Wait until the timer has stopped and shut off the sample pump. Disconnect the sample tube from the pump and place it in the packaging provided with the sample tube. Cap the tube and store for analysis.
  7. Record the unique serial number stamped onto each sample tube, the sample time, and the flow rate for the sample tube in a laboratory notebook. These values will be important for quantitation.

4. Calibration Curve Generation

  1. Pipet 5.0 µl of the solution standard directly on the glass frit of an unused, conditioned sample tube. Hold the sample tube and pipet upright with a gloved hand during deposition.
  2. Repeat Step 4.1 for each of the six calibration standards onto three different sample tubes.
  3. Deposit 5 µl of the 0.3 ng μl-1 3,4-DNT on each of the tubes as well.
  4. Allow the eighteen sample tubes (three per solution concentration, six solution concentrations) to sit at RT for at least 30 min to evaporate the solvent.
  5. Use the twenty tube autosampler and the previously described TNT and RDX TDS-CIS-GC-ECD method to run and analyze all eighteen tubes O/N.24,25 A summary of the TDS-CIS-GC-ECD parameters for the method is provided in Table 1.
  6. Integrate the peaks associated with 3,4-DNT, TNT, and RDX in the chromatogram for each of the eighteen sample tubes. The 3,4-DNT, TNT and RDX peaks will occur at approximately 4.16, 4.49 and 4.95 min, respectively.
  7. Note the 3,4-DNT, TNT and RDX peak areas for each of the eighteen tubes along with the corresponding mass of TNT and RDX that was deposited on the sample tube in a spreadsheet and laboratory notebook.
  8. Normalize the peak areas for both TNT and RDX by dividing each peak area by the peak area for 3,4-DNT. Do this for all eighteen tubes.
  9. Calculate the average and standard deviation of the normalized TNT and RDX peak areas for the six standard concentrations.
  10. Plot the average normalized peak area versus mass of analyte present on the tubes for both TNT and RDX.
  11. Add a linear trend line for both the TNT and RDX data points. Identify the slope and y-intercept for each analyte. Record the slope, intercept, and R2 value in a spreadsheet and laboratory notebook.
  12. Place used sample tubes in a tube conditioner for 3 hr at 300 ºC and 500 ml min-1 nitrogen flow.

5. Sample Analysis

  1. Deposit 5.0 µl of the 0.3 ng µl-1 3,4-DNT on each of the sample tubes.
  2. Allow the tubes to sit at RT for at least 30 min to evaporate the solvent from the internal standard.
  3. Use the twenty tube autosampler and the previously described TNT and RDX method to run the tubes O/N on the TDS-CIS-GC-ECD.24,25 A summary of the instrumentation parameters for the analysis method is provided in Table 1.
  4. Integrate the peaks associated with 3,4-DNT, TNT, and RDX in the chromatogram for each of the eighteen sample tubes. The 3,4-DNT, TNT and RDX peaks will occur at approximately 4.16, 4.49 and 4.95 min, respectively.
  5. Note the 3,4-DNT, TNT and RDX peak areas for each of the sample tubes in a spreadsheet and laboratory notebook.
  6. Use the peak areas and calibration curve to calculate the vapor concentration in parts-per-billion by volume (ppbv) for each analyte. See Equations 1-4.
  7. Place used sample tubes in a tube conditioner for 3 hr at 300 ºC and 500 ml min-1 nitrogen air flow.

Representative Results

Obtaining quantitative results for trace explosive vapor samples begins with establishing a calibration curve for the TDS-CIS-GC-ECD instrumentation using the direct liquid deposition method of solution standards onto sample tubes to account for instrument losses and differences between solution standards and vapor samples. The TDS-CIS-GC-ECD instrumentation and method for TNT and RDX trace analysis has been previously described in detail elsewhere, but the instrument parameters are summarized in Table 1.24,25 Here, Figure 1 shows a series of chromatograms obtained using the published method and parameters in Table 1. Peaks for 3,4-DNT, TNT, and RDX are observed at 4.16, 4.49, and 4.95 min, respectively. The peak height and area for the internal standard is constant for all masses of TNT and RDX, while the peak height and area increases with mass of the analyte. The peak areas for TNT and RDX for each mass are normalized by the peak area for 3,4-DNT to account for irreproducibility and losses associated with sample tube injection. The normalized peak areas for each analyte are then plotted versus mass on sample tube to establish a calibration curve. A linear regression is conducted to obtain the slope, intercept, and coefficient of determination (R2). The slope and intercept are used for converting the normalized peak area for vapor sample to mass, or ultimately concentration. Figure 2 shows an example calibration curve generated from the chromatograms shown in Figure 1. The error bars indicate one standard deviation with three replicate measurements per mass of analyte (N=3). An ideal calibration curve with no instrument or sampling losses and linear detector response would have an R2 value near unity. An R2 value that significantly deviates from unity, approximately less than 0.98, is typically an indicator the instrument needs servicing, the solution standards were not properly prepared, or solution standards and the internal standard were not properly deposited onto the glass frit of the sample tubes.

The calibration curve, the plot, and related raw data, are saved in the same spreadsheet as the sample information so the calibration used for quantitation is easily accessed and tracked with the analyzed samples. The calibration curve and peak areas from a sample can be used to calculate a vapor concentration using the following set of equations:

Equation 1 (1)

Equation 2 (2)

Equation 3 (3)

Equation 4 (4)

where Aa is the analyte peak area, As is the internal standard peak area, b is the calibration curve Y-intercept for the analyte, S is the calibration curve slope for the analyte, C is a conversion factor to parts-per-billion by volume (ppbv, 109), M is the molecular weight for the analyte (g mol-1), Qs is the sample flow rate (ml min-1), L is a conversion factor from milliliters to liters (103), R is the ideal gas constant (8.314 L kPa K-1 mol-1), T is the temperature (K), t is the sample time (min), and P is the pressure (kPa). These series of equations can be embedded into a spreadsheet for automatic calculation of quantitation values. Importantly, these equations assume an ideal gas, so the concentrations have a reduction in accuracy because none of the analytes are ideal gases.

Figure 3 shows an example of a chromatogram that indicates the instrument is in need of service or new solution standards should be prepared. Additional peaks other than those identified as 3,4-DNT, TNT, and RDX appear in the chromatogram. Additional peaks are always present when using sorbent-filled thermal desorption sample tubes because the sorbent material degrades over time with repeated usage and does not selectively adsorb just DNT, TNT, and RDX. However, the degradation products do not co-elute with 3,4-DNT, TNT, and RDX with a properly maintained instrument.26 A blank tube should be run before and after each calibration series to identify peaks that are present from either sorbent material degradation or impurities captured during the vapor sampling collection. This is easily achieved with the use of a twenty sample tube autosampler, where eighteen calibration standard sample tubes are used for the calibration curve and two additional positions are free for blank tubes at the start and end of the sequence. Additional peaks not observed in the blank, but observed in sample tubes deposited with solution standards to generate a calibration curve, typically indicates solution analyte degradation and new solution standards should be prepared and deposited on a new set of sample tubes. Additional peaks have also been observed if sample tubes are left in the tube conditioner for greater than 3 hr.

Furthermore, the peak shapes deviate greatly from a Gaussian shape, specifically for the peaks at approximately 4.6 and 4.825 min. Some instrumentation and data analysis software packages provide a “Symmetry” calculation for each peak in a chromatogram that attempts to quantitate the deviation from a Gaussian shape. This value can be used as an indicator to replace the column and inlet liner of the instrument when it significantly deviates from unity, where unity indicates a perfect Gaussian peak shape. The ECD is very sensitive to nitro-aromatics such as DNT and TNT, but has a limited dynamic range. This results in peaks becoming clipped at the upper bounds of the dynamic range, as seen for the peak at approximately 4.825 min in Figure 3. If peaks become clipped, then it might be necessary to reduce the sampling time for vapor samples during sample collection. Running a new calibration curve before each sample collection series or on a repeated schedule, such as every other night, is a good way to catalog instrument performance and determine when an instrument requires maintenance or service before analyzing valuable samples.

Figure 1
Figure 1. An example chromatogram of the separation of 3,4-DNT (internal standard), TNT, and RDX using the TDS-CIS-GC-ECD instrumentation with the direct liquid deposition method for generating calibration curves of vapor samples. The chromatogram has been trimmed to the relevant portion, but the total run is 8 min long. The 3,4-DNT peak area is relatively constant (1.5 ng) while the TNT and RDX peak areas, and heights, increase with mass of analyte on sample tube: (black) 0.1 ng, (red) 0.5 ng, (green) 1.0 ng, (blue) 2.0 ng, and (orange) 3.0 ng. Please click here to view a larger version of this figure.

Figure 2
Figure 2. An example calibration curve for (■) TNT and (dot) RDX using the direct liquid deposition method with solution standards and a TDS-CIS-GC-ECD instrumentation. The normalized peak area on the Y-axis is obtained from dividing the TNT and RDX peak areas in a chromatogram by the peak area for 3,4-DNT, the internal standard. The error bars represent one standard deviation for the average of three replicate sample tubes per solution standard, or analyte mass.

Figure 3

Figure 3. A chromatogram resulting from poor instrument maintenance, column degradation, and sample tube sorbent material degradation. Additional peaks other than the 3,4-DNT, TNT, and RDX peaks are observed. The peak at approximately 4.825 min is clipped because the analyte mass is at the upper limit of the dynamic range of the detector. A shoulder appears at 4.850 min, indicating poor separation. The baseline, or lower limit, response is elevated causing baseline drift and increase in noise.

Figure 4

Figure 4. A conceptual diagram illustrating the correct procedure for depositing solution onto a sorbent-filled thermal desorption sample tube for the direct liquid deposition method. The micropipette tip should be touching the glass frit and not the side walls of the sample tube. A new tip should also be used for each deposition between analytes and sample tubes.

Figure 5

Figure 5. An alternative to using personal sample pumps for collecting explosive vapors on sorbent-filled thermal desorption sample tubes. Flexible tubing is used to connect the sample tubes to a mass flow controller (MFC) that allows for electronic input of a desired flow rate. The mass flow controllers, when combined with a pump, automatically adjust for flow rate through the sample tubes to a desired set point regardless of variations between sample tubes. A six MFC configuration is shown with a manifold to connect all MFCs to a common pump, but configurations with different numbers of MFCs are possible.

Parameter Name Value Units
TDS Initial Temperature 25 ºC
TDS Final Temperature 250 ºC
TDS Temperature Ramp 40 ºC min-1
TDS Hold Time 2 min
TDS Flow Rate 455 ml min-1
TDS Mode PTV Solvent Vent N/A
TDS Transfer Line Temperature 300 °C
CIS Initial Temperature 0 °C
CIS Final Temperature 250 °C
CIS Temperature Ramp 12 °C sec-1
CIS Hold Time 3 min
CIS Flow Rate 108 ml min-1
CIS Mode PTV Solvent Vent N/A
Oven Initial Temperature 30 °C
Oven Initial Hold Time 0.5 min
Oven Final Temperature 250 °C
Oven Temperature Ramp 1 40 °C min-1
Oven Temperature Hold 1 210 °C
Oven Temperature Ramp 2 40 °C min-1
Oven Temperature Hold 1 250 °C
Oven Hold Time 1 min
Column Carrier Gas Helium N/A
Column Flow Rate 5.6 ml min-1 
Column Pressure 23.642 psi
Column Coating 5% polysilioxane (DB5-MS) N/A
Column Length 15 m
Column Inner Diameter (ID) 0.25 mm
Column Outer Diameter (OD) 250 mm
ECD Temperature 275 °C
ECD Flow Rate 60 ml min-1
ECD Carrier Gas Nitrogen N/A

Table 1. The TDS-CIS-GC-ECD instrumentation parameters for quantitation of TNT and RDX vapors using the direct liquid deposition method.

Solution TNT and RDX Concentration
(ng μl-1)
Approximate Vapor Concentration
(ppbv)
Sampling Time
(min)
0.1 TNT/0.25 RDX 0.050 TNT/0.125 RDX 120
0.4 TNT/1.0 RDX 0.200 TNT/0.500 RDX 30
2.0 TNT/5.0 RDX 1.00 TNT/2.50 RDX 6

Table 2. The approximate sampling time for collecting explosive vapors for three solution concentrations of TNT and RDX. Actual sampling times may need to be adjusted to yield peaks in a chromatogram suitable for quantitation.

Discussion

Reproducibility is a critical attribute for the quantitation of trace explosive vapors using the direct liquid deposition method with TDS-CIS-GC-ECD instrumentation, and Relative Standard Deviation (RSD) is often used as a metric for reproducibility. We have experienced RSDs for inter- and intra-sample reproducibility of approximately 5% for TNT and 10% for RDX. Any RSD above 15% is used as an indicator to check common sources of variation that reduce the effectiveness of the protocol. Sources of variation that have led to unacceptable RSDs in the past are highlighted in the following discussion.

A common source of variation that can lead to relatively large standard deviations for replicate measurements of solution standards and significant deviation from unity for R2 is consistent deposition of solution standards and the internal standard onto sample tubes. We have found an electronic micropipette is ideal for minimizing variation during deposition, as opposed to a manual micropipette. During several recent projects, where multiple personnel were involved in quantitating explosive vapors over several days of sample collections, the source of variation in the results was largely dependent on the individual and his/her usage of the manual micropipette. At first glance, the usage of a manual micropipette appears relatively straightforward, but small variations in plunger depression and release between users yielded a significant source of variation in the quantitative analysis of explosive vapors. When the manual micropipette was switched for an electronic micropipette, the variation between users could not be distinguished between instrument variation and sampling noise. It is also important to hold the micropipette upright during solution uptake and deposition. The solution should be deposited directly onto the glass frit of the sample tube, i.e., the micropipette tip should contact the glass frit and not the glass side walls. A new micropipette tip should also be used for each deposition and sample tube. Figure 4 shows a conceptual drawing of the procedure for depositing the internal standard or a solution standard onto the glass frit of a sample tube.

Another source of variation that can diminish reproducibility with the quantitation of explosive vapors is the vapor sampling procedure. In the protocol, a commercial sample pump is connected to the sample tube and calibrated with a piston flow meter using a small set screw and screwdriver. The flow rate must be adjusted for every sample tube to account for packing differences of the sorbent material between sample tubes and pump performance. This procedure can be cumbersome and error prone, especially when attempting to collect multiple sample tubes in parallel. Similar to the replacement of a manual micropipette with an electronic micropipette to reduce variation, we have also implemented an electronic sample tube system that uses a vacuum pump and mass flow controllers (MFC). Figure 5 shows a conceptual diagram of a six sample tube vapor collection manifold. The MFCs automatically correct for variations in the packing between sample tubes and automatically adjust the flow rate to 100 ml min-1 without user intervention. The flow rates should still be routinely checked and calibrated with a piston flow meter, but the flow rate can be adjusted electronically rather than manually with a screwdriver. It is possible to create a single MFC sample tube collection manifold, but the MFC-based configuration seen in Figure 5 is meant to demonstrate the scalability of the alternative method. Notably, the individual commercial sample pumps are less expensive than the MFC-based configuration and the MFC-based configuration is a custom assembly, but the MFC-based configuration can reduce variation, improve reproducibility, and be easier to use.

Variation is also present in the TDS-CIS-GC-ECD instrumentation. Over time, as the various internal components of the instrument heat and cool during analysis, parts will expand and contract causing fatigue of consumables, such as ferrules, nuts, columns, and liners. The gradual fatigue of components is unavoidable and a source of variation over time. When performing quantitation of trace (sub parts-per-million by volume, ppmv) explosive vapors, the gradual variation in instrument performance becomes amplified. Thus, it is important to establish calibration curves for quantitation in a timely fashion, typically before analysis of samples. If possible, calibration curves should be generated the same day as the sample analysis to be conducted. This is not always possible due to time constraints and instrument workload. Furthermore, typically at least five replicates are used per mass, or concentration, for a calibration curve because more replicates yield a more robust calibration curve for quantitation. However, the calibration curve in Figure 2 was established with three replicates. The number of replicates was reduced so a full calibration curve over the entire dynamic range of the detector could be established O/N in a single autosampler tray (two blanks, eighteen solution standard sample tubes, and twenty sample tube capacity). To compensate for the reduced number of replicates per analyte mass, a new calibration curve was established O/N with samples run immediately the following day to account for variation associated with instrument drift and prevent a back log of sample tubes for quantitation of explosives vapors with TDS-CIS-GC-ECD, which is approximately 20 min per sample tube.

Determining the peak areas for 3,4-DNT, TNT, and RDX in a chromatogram, such as the example seen in Figure 1, can be a subjective process that can introduce irreproducibility to the quantitation of explosive vapors with direct liquid deposition method and TDS-CIS-GC-ECD instrumentation. Many data analysis software packages supplied with GC-ECD instrumentation include manual and automatic peak detection and integration methods. The field of chromatographic data analysis and its related techniques for automatic peak detection and integration is long and extensive,2731 with many of the algorithms provided in data analysis software. A complete review of the various characteristics and procedures for integrating peak areas is beyond the scope of this work. It is more important for a research group to standardize, document, and use the same procedure for the calibration curve as the samples to minimize variation in quantitation of explosive vapors from more subjective peak area integration processes.

Finally, degradation of sample tubes and solution standards can affect quantitation of trace explosive vapors. Similar to component fatigue from use and thermal cycling of the TDS-CIS-GC-ECD instrumentation, the sorbent material in the sample tubes can degrade over time with repeated sampling and thermal desorption. New sample tubes are tightly packed and white in color. Over time as the sample air is flowed in one direction and carrier gas, typically helium, flowed in the opposite direction during thermal desorption, the packing of the sample tubes becomes loose and yellow in color. The yellow color indicates degradation of the sorbent material from repeated thermal cycling within the instrumentation and tube conditioner. Additional peaks in blanks and sample tubes are also indicators of sorbent material degradation products.26 After each analysis, sample vapors are conditioned in a tube conditioner for a maximum of 3 hr. This is to desorb any remaining material from the sorbent material and effectively clean the sample tubes. However, leaving a sample tube in the conditioner at relatively high temperature (300 °C) longer than 3 hr can significantly shorten the lifetime of the sample tube and introduce variation in quantitation. Similarly, solution standards will degrade over time, which will artificially reduce the mass, or concentration, of analyte in each standard for a calibration curve. To minimize solution degradation, the solution standards should be stored in an amber glass vial in a freezer or refrigerator and the solution standards should be periodically analyzed using the GC-ECD without the TDS-CIS and a standard split/splitless inlet to identify additional peaks or degradation products. A complementary quantitation method, such as gas chromatography mass spectrometry or high performance liquid chromatography, can also be used to ensure the solution standards have not degraded and are suitable for the direct liquid deposition method for quantitating trace explosive vapors.

Divulgaciones

The authors have nothing to disclose.

Acknowledgements

Financial support was provided by the Department of Homeland Security Science and Technology Directorate.

Materials

Name of Material/ Equipment Company Catalog Number Comments/Description
2,4,6-Trinitrotoluene (TNT) Accu-Standard M-8330-11-A-10X 10,000 ng μL-1
Cyclotrimethylenetrinitramine (RDX) Accu-Standard M-8330-05-A-10X 10,000 ng μL-1
3,4-Dinitrotoluene (3,4-DNT) Accu-Standard S-22988-01 1000 ng μL-1
Tenax® TA Vapor Sample Tubes Gerstel 009947-000-00 Tenax® 60/80
CIS4 Liner Gerstel 014652-005-00
Transfer Line Ferrule Gerstel 001805-008-00
Inlet Liner Ferrule Gerstel 001805-040-00
CIS4 Ferrule Gerstel 007541-010-00
ECD Detector Ferrule Aglient 5181-3323
DB5-MS Column Res-Tek 12620

Referencias

  1. McLafferty, F. W., Stauffer, D. B., Twiss-Brooks, A. B., Loh, S. Y. An enlarged data base of electron-ionization mass spectra. Journal of the American Society for Mass Spectrometry. 2 (5), 432-437 (1991).
  2. Psillakis, E., Kalogerakis, N. Application of solvent microextraction to the analysis of nitroaromatic explosives in water samples. Journal of Chromatography A. 907 (1-2), 211-219 (2001).
  3. Babushok, V. I., Linstrom, P. J., et al. Development of a database of gas chromatographic retention properties of organic compounds. Journal of Chromatography A. 1157 (1-2), 414-421 (2007).
  4. National Institute of Standards and Technology. . NIST/EPA/MSDC Mass Spectral Database, Standard Reference Database 1 (NIST 08). , (2008).
  5. Stein, S. E., Pierre, A., Lias, S. G. Comparative evaluations of mass spectral databases. Journal of the American Society for Mass Spectrometry. 2 (5), 441-443 (1991).
  6. Sigman, M. E., Ma, C. -. Y., Ilgner, R. H. Performance Evaluation of an In-Injection Port Thermal Desorption/Gas Chromatographic/Negative Ion Chemical Ionization Mass Spectrometric Method for Trace Explosive Vapor Analysis. Analytical Chemistry. 73 (4), 792-798 (2001).
  7. Ausloos, P., Clifton, C., et al. The critical evaluation of a comprehensive mass spectral library. Journal of the American Society for Mass Spectrometry. 10 (4), 287-299 (1999).
  8. Dionne, B. C., Rounbehler, D. P., Achter, E. K., Hobbs, J. R., Fine, D. H. Vapor Pressure of Explosives. Journal of Energetic Materials. 4 (1), 447-472 (1986).
  9. Ewing, R. G., Waltman, M. J., Atkinson, D. A., Grate, J. W., Hotchkiss, P. J. The vapor pressures of explosives. TrAC Trends in Analytical Chemistry. 42, 35-48 (2013).
  10. Wallin, S., Ang, H. G. Vapor Pressure of Explosives: A Critical Review. Propellants, Explosives, Pyrotechnics. 37 (1), 12-23 (2012).
  11. Pinnaduwage, L. A., Yi, D., Tian, F., Thundat, T., Lareau, R. T. Adsorption of Trinitrotoluene on Uncoated Silicon Microcantilever Surfaces. Langmuir. 20 (7), 2690-2694 (2004).
  12. Moore, D. S. Instrumentation for trace detection of high explosives. Review of Scientific Instruments. 75 (8), 2499-2512 (2004).
  13. Douse, J. M. F. Trace analysis of explosives at the low picogram level by silica capillary column gas–liquid chromatography with electron-capture detection. Journal of Chromatography A. 208 (1), 83-88 (1981).
  14. Douse, J. M. F. Trace analysis of explosives in handswab extracts using amberlite XAD-7 porous polymer beads, silica capillary column gas-chromatography with electron-capture detection and thin-layer chromatography. Journal of Chromatography. 234, 415-425 (1982).
  15. Sigman, M. E., Ma, C. -. Y. In-Injection Port Thermal Desorption for Explosives Trace Evidence Analysis. Analytical Chemistry. 71 (19), 4119-4124 (1999).
  16. Yinon, J., Zitrin, S. . Modern Methods and Applications in Analysis of Explosives. , (1993).
  17. Waddell, R., Dale, D. E., Monagle, M., Smith, S. A. Determination of nitroaromatic and nitramine explosives from a PTFE wipe using thermal desorption-gas chromatography with electron-capture detection. Journal of Chromatography A. 1062 (1), 125-131 (2005).
  18. Hable, M., Stern, C., Asowata, C., Williams, K. The determination of nitroaromatics and nitramines in ground and drinking water by wide-bore capillary gas chromatography. Journal of Chromatographic Science. 29 (4), 131-135 (1991).
  19. Yinon, J. Trace analysis of explosives in water by gas chromatography–mass spectrometry with a temperature-programmed injector. Journal of Chromatography A. 742 (1-2), 205-209 (1996).
  20. Walsh, M. E. Determination of nitroaromatic, nitramine, and nitrate ester explosives in soil by gas chromatography and an electron capture detector. Talanta. 54 (3), 427-438 (2001).
  21. Field, C. R., Lubrano, A. L., Rogers, D. A., Giordano, B. C., Collins, G. E. Direct Liquid Deposition Calibration Method for Trace Cyclotrimethylenetrinitramine Using Thermal Desorption Instrumentation. Journal of Chromatography A. 1282, 178-182 (2013).
  22. Field, C. R., Giordano, B. C., Rogers, D. A., Lubrano, A. L., Rose-Pehrsson, S. L. Characterization of Thermal Desorption Instrumentation with a Direct Liquid Deposition Calibration Method for Trace 2,4,6-Trinitrotoluene Quantitation. Journal of Chromatography A. 1227, 10-18 (2012).
  23. Excoffier, J. L., Guiochon, G. Automatic peak detection in chromatography. Chromatographia. 15 (9), 543-545 (1982).
  24. Vivó-Truyols, G., Torres-Lapasió, J. R., van Nederkassel, A. M., Vander Heyden, Y., Massart, D. L. Automatic program for peak detection and deconvolution of multi-overlapped chromatographic signals: Part I: Peak detection. Journal of Chromatography A. 1096 (1-2), 133-145 (2005).
  25. Vivó-Truyols, G., Torres-Lapasió, J. R., van Nederkassel, A. M., Vander Heyden, Y., Massart, D. L. Automatic program for peak detection and deconvolution of multi-overlapped chromatographic signals: Part II: Peak model and deconvolution algorithms. Journal of Chromatography A. 1096 (1-2), 146-155 (2005).
  26. Fong, S. S., Rearden, P., Kanchagar, C., Sassetti, C., Trevejo, J., Brereton, R. G. Automated Peak Detection and Matching Algorithm for Gas Chromatography−Differential Mobility Spectrometry. Analytical Chemistry. 83 (5), 1537-1546 (2011).
  27. Hargrove, W. F., Rosenthal, D., Cooley, P. C. Improvement of algorithm for peak detection in automatic gas chromatography-mass spectrometry data processing. Analytical Chemistry. 53 (3), 538-539 (1981).
  28. Middleditch, B. S. . Analytical Artifacts GC, MS, HPLC, TLC and PC. 44, (1989).

Play Video

Citar este artículo
Field, C. R., Lubrano, A., Woytowitz, M., Giordano, B. C., Rose-Pehrsson, S. L. Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector. J. Vis. Exp. (89), e51938, doi:10.3791/51938 (2014).

View Video