We describe a multi-angle rotational optical imaging (MAROI) system for in vivo quantitation of a fluorescent marker delivered by saposin C (SapC)-dioleoylphosphatidylserine (DOPS) nanovesicles. Employing mouse models of cancer and arthritis, we demonstrate how the MAROI signal curve analysis can be used for the precise mapping and biological characterization of disease processes.
We describe a multi-angle rotational optical imaging (MAROI) system for in vivo monitoring of physiopathological processes labeled with a fluorescent marker. Mouse models (brain tumor and arthritis) were used to evaluate the usefulness of this method. Saposin C (SapC)-dioleoylphosphatidylserine (DOPS) nanovesicles tagged with CellVue Maroon (CVM) fluorophore were administered intravenously. Animals were then placed in the rotational holder (MARS) of the in vivo imaging system. Images were acquired in 10° steps over 380°. A rectangular region of interest (ROI) was placed across the full image width at the model disease site. Within the ROI, and for every image, mean fluorescence intensity was computed after background subtraction. In the mouse models studied, the labeled nanovesicles were taken up in both the orthotopic and transgenic brain tumors, and in the arthritic sites (toes and ankles). Curve analysis of the multi angle image ROIs determined the angle with the highest signal. Thus, the optimal angle for imaging each disease site was characterized. The MAROI method applied to imaging of fluorescent compounds is a noninvasive, economical, and precise tool for in vivo quantitative analysis of the disease states in the described mouse models.
Whole animal imaging has become a powerful tool in the study of animal physiopathology. Among current imaging systems, the MS FX PRO allows researchers to accurately visualize fluorescently labeled (or luminescent) compounds and/or tissues in living mice, and simultaneously obtain X-ray images. With the recently introduced multi modal animal rotation system (MARS) a complete, automated rotation of the mouse is achieved in order to capture both fluorescent/luminescent and X-ray images at specific angles1. Image acquisition can be programmed such that sequential image series can be captured at specific, incremental angles as small as 1°. This permits one to identify the optimal orientation of the animal, i.e. that in which the distance between the internally generated fluorescent/luminescent signal and the system's detection device is the shortest. This, in turn, facilitates precise repositioning of the animal for subsequent imaging sessions during longitudinal studies.
In this report, we describe the implementation of a multi-angle rotational optical imaging (MAROI) system for in vivo quantitation of fluorescent marker intensity. MAROI signal curve analysis can be used in longitudinal studies for direct correlation of fluorescent signal distribution to precisely map diseased sites or biological processes of interest.
This system was used to monitor the absorption of fluorescently labeled SapC-DOPS nanovesicles by orthotopic and spontaneous tumors, as well as by arthritic foci, in living mice; it provided multispectral and multimodal data sets derived from complete rotational coverage of the animals. Among the numerous fluorescent probes currently available for in vivo imaging, those emitting in the near-infrared and far-red spectral regions confer the lowest interference with skin and tissues, and provide the highest penetration and image resolution. We used CellVue Maroon (CVM)2,3, a far-red fluorescent cell linker (Ex 647/Em 667), to label SapC-DOPS (SapC-DOPS-CVM)4-12.
Ethics statement of animal use. All animal studies were approved by the Institutional Animal Care and Use Committee of the University of Cincinnati (IACUC Protocol Number: 11-05-05-02) and the Cincinnati Children's Hospital Research Foundation (Animal Welfare Assurance Number A3108-01). All experiments involving mice followed the animal care guidelines of the University of Cincinnati and the Cincinnati Children's Hospital Research Foundation.
1. Prepare Animal Models
Note: Three different animal models outlined below have been used in our prior studies:
2. Preparation of Fluorescently-labeled SapC-DOPS Nanovesicles
3. Imaging
4. Image Analysis
We demonstrate here that SapC-DOPS nanovesicles labeled with a far-red dye (CVM) specifically accumulate in orthotopic and spontaneous mouse brain tumors, as well as in arthritic joints of K/BxN mice. Serial fluorescence/X-ray images acquired from an ROI placed over each disease site during complete rotations of the mice were subjected to MAROI curve analysis, which revealed the optimal imaging angle with the highest fluorescence intensity.
The primary purpose for using the MARS system is to determine the optimum angle of fluorescence so that the most accurate measurements can be taken. Representative results from three experiments using mice with brain tumors or arthritis are shown. Using SapC-DOPS-CVM and the MARS system (Figure 1), the best possible image angle for observing the tumor or inflammation due to arthritis was determined. Fluorescence images, followed by an X-ray acquisition, were acquired every 10° during a 380° rotation of the mouse. Fluorescence images were overlaid onto the corresponding X-ray images for image display and rotational movie generation.
Results from the orthotopic brain tumor model are demonstrated in Figure 2. The fluorescence image of a representative orthotopic tumor-bearing mouse (Ortho1) is shown in Figure 2A. The optimal image angle for this animal is 10°, the position at which the fluorescence photon intensity is the greatest (Figure 2B). Measurements were taken before injection with SapC-DOPS-CVM (baseline) and 24 hr after injection. Control mice (tumor free) received a similar treatment.
Figure 3 shows comparable data from the genetically engineered brain tumor mouse model. The fluorescence images and photon measurements were taken before injection with SapC-DOPS-CVM (baseline) and 24 hr (Figures 3A and 3B) and 9 days (Figure 3C) after injection. These graphs show that the optimal imaging angle in the tumor-bearing animal (Tumor Mut49) is 20° 24 hr post injection but changes to 10° 9 days post injection. This suggests that fluorescence signal alteration correlated with morphological changes, likely reflecting tumor growth.
As shown in Table 1, the MAROI method clearly demonstrates that the fluorescent signal decreases for projections at increasing rotation away from the optimal imaging angle. In brain tumors, a 7% average decrease in fluorescent signal was obtained if the animal’s physical orientation was ±10° offset from the optimal imaging angle. An average 21% decrease in fluorescent signal was measured at ±20°. Thus relatively small offsets from the optimal angle can result in significant signal attenuation. Utilizing the MAROI technique for image positioning will allow investigators to produce more consistent and reliable data.
The MAROI method was finally used to assess the targeting of arthritic joints by SapC-DOPS-CVM 24 hr after SapC-DOPS-CVM injection. This animal scored 3 with three arthritic joints. Fluorescence images of toe and ankle of the arthritic mouse are shown in Figures 4A and 4B. Corresponding photon measurements at 10° rotation intervals are graphed in Figures 4C and 4D. The optimal imaging angles found for the toe and ankle are 140° and 120°, respectively.
In summary, the combination of the MAROI system with fluorescent SapC-DOPS nanovesicles represents a noninvasive, accurate and highly sensitive strategy for live imaging, which allows for quantitative studies of tumor and arthritis progression in small animals. The possibility of acquiring a 360° multimodal imaging dataset considerably improves data analysis and interpretation, as compared with what is achievable using single angle imaging techniques.
Table 1. Optimum imaging angles for each mouse model. The differences between the angle of maximum photon fluorescence (FLR optimum angle) and the standard anatomic angle (X-ray) can be seen. When these two angles become increasingly different, the measured signal changes significantly. Click here to view larger image.
Figure 1. Multi-angle rotational optical imaging (MAROI) device. Click here to view larger image.
Figure 2. Fluorescence signal vs. image angle in an orthotopic brain tumor mouse model. (A). Image of the peak fluorescent signal at the optimal image angle of 10°. The blue box shows the ROI used to quantify the emitted photons. (B). Graph of the image angle versus photon emission. The representative orthotopic tumor-bearing mouse (Ortho1) is graphed against averaged fluorescence values from identical ROIs in three nontumor mice. Measurements were taken at baseline (before injection) and 24 hr post injection. Error bars represent Standard Deviation. Click here to view larger image.
Figure 3. Fluorescence signal vs. image angle in a spontaneous brain tumor of a genetically engineered mouse model. (A). Image of the peak fluorescent signal of ROI 1 (top blue box) at the optimal image angle of 20°. (B) and (C). Graphs of the image angle versus photon emission from ROI 1. Values from a representative spontaneous brain tumor-bearing mouse, Tumor-Mut 49, are graphed against averaged values from three non tumor mice. Measurements were taken at baseline (before injection) and 24 hr (B) and 9 days (C) post injection. Error bars represent Standard Deviation. Click here to view larger image.
Figure 4. Fluorescence signal vs. image angle in a mouse with arthritis of the toe and ankle joints. (A) and (B). Images showing the peak fluorescent signal for the toes (A) and ankle joints (B), within the ROIs shown in the red box. (C). Graph of angle versus mean photon emission for the toe. Peak photon emission can be seen at an angle of 140°. (D). Graph of angle versus mean photon emission for the ankle; maximum intensity occurs at an angle of 120°. Click here to view larger image.
Accurate determination of the location and magnitude of solid tumors and inflammatory foci in rheumatic conditions is critical to implement adequate treatment and follow up disease progression or remission. While valuable, current imaging strategies (X-rays; MRI; ultrasound; X-ray computed tomography) provide incomplete assessments of disease status. For instance, arthritic joint damage is commonly assessed by X-rays, which provides information on bone structure but not on soft tissue inflammation and destruction, characteristic of early stages of the disease. The MAROI method presented here combines the advantages of both X-rays and sophisticated soft tissue imaging modalities (e.g. MRI or ultrasound) in an integrated, noninvasive and simpler platform that also allows for a full 3D mapping and reconstruction of the diseased tissue or organ in small animals such as mice.
This method takes advantage of the selective affinity of SapC-DOPS nanovesicles for exposed phosphatidylserine residues, which are abundant in the membranes of cancer and inflammatory cells. The determinant of this binding is SapC, a fusogenic lysosomal protein with a strong affinity for anionic phospholipids such as phosphatidylserine7,10,11. When conjugated to a fluorescent probe (CVM), systemically injected SapC-DOPS can be traced to tumor and arthritic sites by fluorescence imaging.
Limitations of our method are related to its sensitivity, which at present restricts its use to imaging of small animals like mice. As with other imaging methods, optimal fluorescent signal to noise ratio is constrained by the size of the tumor or the extent of arthritis, and may be compromised when imaging tissues or organs with high background (autofluorescence) such as ears (brain imaging), intestines/feces (abdominal imaging) and paws (hind limb imaging). In this respect, we found that a far-red dye such as CVM provides better spectral separation and resolution in the in vivo setting than other fluorescent probes in the visible range.
Other pitfalls include potential movement of the animal during imaging, both while anesthetized and post mortem (rigor mortis). Hind limb positioning, particularly, is often difficult to stabilize to avoid movement during rotation. In its present state the technique is also time consuming, with scan times as long as 60 min needed to complete a full rotation and acquire high quality images.
The MAROI method presents a number of advantages over other imaging modalities. The ability to image diseased tissue from 38 (or more) different angles allows visualization of fluorescence that may be hindered when assessing it from a single plane; this is valuable in animal studies because it may help minimize the number of false negatives that result from imaging at inappropriate angles. By overlying X-ray and fluorescence images, a precise anatomical localization of the diseased site can be determined. Finally, the possibility of live (in vivo) imaging allows for longitudinal studies to be performed.
The authors have nothing to disclose.
This work was supported in part by NIH/NCI Grants Number 1R01CA158372-01 (to Qi) and New Drug State Key Project Grant Number 009ZX09102-205 (to Qi). Writing assistance was provided by Dr. Judy Racadio, and was funded by the University of Cincinnati Department of Hematology and Oncology. Vontz Core Imaging Lab (VCIL) in the College of Medicine at the University of Cincinnati.
Dulbecco's modified eagle medium | Gibco (Grand Island, NY) | 11965 | |
Fetal Bovine Serum | Gibco (Grand Island, NY) | 16000077 | |
Penicillin-streptomycin | Hyclone (Logan, Utah) | SV30010 | |
Dioleoylphosphatidylserine | Avanti Polar Lipids (Alabaster, AL) | 840035C | |
CellVue Maroon | Molecular Targeting Technologies, Inc. (Exton, PA) | C-1001 | |
Sephadex G25 column PD-10 | Amersham Pharmacia Biotech, (Piscataway, NJ) | 17-0851-01 | |
New Standard Stereotaxic for Rats and Mice | Harvard Apparatus (Holliston, MA) | 726335 | |
Bransonic Ultrasonic Cleaners Model 1510 | Branson Ultrasonics (Danbury,CT) | CPN-952-118 | |
Multi-spectral FX system | Bruker Corporation (Billerica, MA) | ||
Multi-angle Rotational Optical Imaging Device | Bruker Corporation (Billerica, MA) |