Effective pulmonary vein isolation utilizing a cryoballoon depends on complete pulmonary vein occlusion. The point of occlusion can be effectively predicted by direct analysis of pulmonary vein pressure waveform analysis during balloon inflation using a simple and reproducible technique.
Cryoballoon ablation (CBA) is an established therapy for atrial fibrillation (AF). Pulmonary vein (PV) occlusion is essential for achieving antral contact and PV isolation and is typically assessed by contrast injection. We present a novel method of direct pressure monitoring for assessment of PV occlusion.
Transcatheter pressure is monitored during balloon advancement to the PV antrum. Pressure is recorded via a single pressure transducer connected to the inner lumen of the cryoballoon. Pressure curve characteristics are used to assess occlusion in conjunction with fluoroscopic or intracardiac echocardiography (ICE) guidance. PV occlusion is confirmed when loss of typical left atrial (LA) pressure waveform is observed with recordings of PA pressure characteristics (no A wave and rapid V wave upstroke). Complete pulmonary vein occlusion as assessed with this technique has been confirmed with concurrent contrast utilization during the initial testing of the technique and has been shown to be highly accurate and readily reproducible.
We evaluated the efficacy of this novel technique in 35 patients. A total of 128 veins were assessed for occlusion with the cryoballoon utilizing the pressure monitoring technique; occlusive pressure was demonstrated in 113 veins with resultant successful pulmonary vein isolation in 111 veins (98.2%). Occlusion was confirmed with subsequent contrast injection during the initial ten procedures, after which contrast utilization was rapidly reduced or eliminated given the highly accurate identification of occlusive pressure waveform with limited initial training.
Verification of PV occlusive pressure during CBA is a novel approach to assessing effective PV occlusion and it accurately predicts electrical isolation. Utilization of this method results in significant decrease in fluoroscopy time and volume of contrast.
Pulmonary vein isolation (PVI) is a widely utilized invasive approach to treatment of atrial fibrillation. Standard PVI techniques include radiofrequency (RF) catheter ablation and, more recently, cryoablation utilizing a cryoballoon (Arctic Front, MDT Inc, MN, USA) 1, 2. Both techniques result in equal long-term efficacy in durable PVI and similar clinical outcomes 3, 4. Given the more rapid vein isolation achieved with the utilization of the cryoballoon, this latter approach has gained widespread acceptance by operators and is currently performed in most centers specializing in atrial fibrillation ablation.
Significant differences between the two techniques include, among others, the requirement of complete pulmonary vein (PV) occlusion with the cryoballoon in order to achieve effective electrical isolation. Apposition of the cryoballoon in the PV ostium is typically assessed fluoroscopically with the utilization of intravenous contrast. As such, the total amount of fluoroscopy utilized during cryoballoon ablation is higher compared to standard catheter ablation, thus increasing the risk of radiation injury. Further, risks associated with contrast infusion, such as renal injury and contrast allergy, are a significant limitation of this technique, given the frequent coexistence of chronic kidney disease in patients with atrial fibrillation 5.
We present a novel method of direct PV pressure monitoring during cryoballoon inflation. This technique was validated in a total of 35 patients and reproducibility and accuracy was assessed.
Methods/Procedure setup
We included 32 patients with drug-refractory paroxysmal AF and three patients with persistent AF. Informed consent was obtained from all patients before the procedure.
Ablation Procedure
All patients are anticoagulated at least 1 month prior to the procedure. Warfarin is stopped the day prior to the procedure aiming for an international normalized ratio between 1.8 and 2.0 on the day of the procedure. Antiarrhythmic drugs are continued periprocedurally and for three months post procedure. All patients undergo cardiac MRI or cardiac CT imaging before the procedure to assess the left atrial and pulmonary vein anatomy and allow for measurement of PV diameter.
The ablation procedure is performed under conscious sedation. Electrograms are recorded in a specialized EP recording system (EP Med, St Jude Medical, MN, US).
1. Procedure Setup
2. PV Occlusion
Postablation Care
After the procedure, warfarin is resumed and subcutaneous low molecular weight heparin is administered if necessary until the international normalized ratio is ≥2.0. Warfarin is continued for at least 3 months. Clinical follow up was performed at six weeks, three months, six months and twelve months and included ECG at each visit and event monitoring at three months. Long term antiarrhythmic and anticoagulation therapy was determined by clinical outcomes.
We assessed the accuracy of pressure monitoring in predicting complete PV occlusion in 35 patients (28 males) with paroxysmal (n=32) or persistent (n=3) AF. Six patients had a prior AF ablation (17%). A total of 128 PVs were assessed with pressure monitoring during balloon inflation. Occlusive pressure was demonstrated with balloon inflation in 113 PVs, of which 111 were electrically isolated with CBA (98.2%). Representative results are shown in Figure 5.
Lack of occlusive pressure waveform was more frequently noted in the RIPV (11/15 PVs). Two veins with demonstrated occlusive pressure waveform were still electrically connected despite concurrent demonstration of PV occlusion by contrast venography. The positive predictive value was 99%. In the RIPV, we observed a requirement for up to three cryoapplications per vein when occlusive pressure was observed in 90% of RIPVs, whereas 10% of veins with occlusive pressure required 4 cryoapplications. In contrast, when occlusive pressure was not observed in the RIPV, 4 or 5 cryoapplications were performed in 70% of RIPVs.
A common PV ostium was observed in nine patients (27.1%). Left common PV ostium was noted in eight patients and a right common PV in one patient. Occlusive pressure was achieved in 16/18 pulmonary veins with common ostium.
We observed a significant effect of the pressure monitoring technique in reducing total fluoroscopy time and radioiodine contrast utilization. Mean fluoroscopy time was 25.2±11.2 min, mean contrast utilization 38±33cc and mean LA time 109±30.8 min during cryoballoon ablation. Contrast and fluoroscopy time decreased significantly after the initial ten procedures utilizing the technique (34.4±10.1 vs. 21.6±9.4 min of fluoroscopy, p=0.001 and 74±28 vs 23.6±23cc of contrast, p=0.00), indicating a very steep learning curve. No complications related to the utilization of the pressure monitoring technique were observed in this series of patients.
Post procedure arrhythmia free survival in one year was 60%. Recurrence of AF or atrial flutter was noted in 40% of patients leading to repeat PVI in 35% of patients. 3/12 patients with recurrent AF had previously documented persistent AF and 6/12 patients had undergone PVI prior to the index procedure.
Figure 1. A 3-way manifold is connected via tubing to the inner lumen of the cryoballoon and a single pressure transducer for continuous pressure recording.
Figure 2. Standard monitoring screen during live recording. Surface ECG, intracardiac CS recording and venous and arterial pressure recording are displayed. Click here to view larger figure.
Figure 3. PV pressure waveform recording during sinus rhythm. Upon initial apposition of the deflated balloon (left panel), characteristic LA waveform is recorded with A and V waves noted. With initial inflation (middle panel), a change in the waveform is noted with more prominent V waves and persistent lower amplitude A waves, suggesting incomplete venous occlusion. Further inflation and with successful circumferential apposition results in disappearance of the A wave and a larger V wave displaced towards mid diastole, confirming complete PV occlusion (right panel). Click here to view larger figure.
Figure 4. During atrial fibrillation, continuous atrial activity (A waves) with small amplitude V waves is observed when the PV is not occluded. With further manipulation of the cryoballoon in the PV ostium, abrupt increase in the V wave amplitude with loss of the small continuous atrial A waves is noted in the pressure waveform, confirming PV occlusion during atrial fibrillation. Click here to view larger figure.
Total PVs, n | 128 |
PVs with occlusive pressure, n | 113/128 |
PVs isolated with CB when occlusive pressure is demonstrated (n) and PPV (%) | 111/113, 99 |
Occlusive pressure in PVs with common ostia , n | 16/18 |
Fluroscopy utilization change (25 procedures vs initial ten procedures), min | 21.6±9.4 vs 34.4±10.1 |
Contrast utilization change (25 procedures vs initial ten procedures), cc | 23.6.6±23 vs 74±28 |
12-month arrhythmia free survival after ablation, % | 60 |
Figure 5. Representative results.
Abbreviations
PV=pulmonary vein, CBA=cryoballoon ablation, AF=atrial fibrillation, ICE=intracardiac echocardiography, TEE=transesophageal echocardiography, LA=left atrium
We present a novel technique of direct pulmonary venous pressure monitoring for assessment of complete pulmonary vein occlusion during cryoballoon ablation for atrial fibrillation. We have demonstrated a steep learning curve during application of the direct pressure monitoring technique and a rapid decrease in radioiodine contrast and fluoroscopy utilization after limited training. The technique can be performed with procedure tools commonly available in most laboratories and the setup requires limited additional preprocedure setup time.
Cryoablation utilizing a cryoballoon has emerged as a novel invasive management strategy for the treatment of paroxysmal atrial fibrillation 2. Procedural success with cryoballoon ablation is highly dependent on correct apposition of the cryoballoon at the pulmonary vein ostium and achievement of complete occlusion of the pulmonary vein 6. Several predictors of PV occlusion and electrical isolation have been studied: 1. the minimum temperature achieved during cryoapplication as well as the slope of rewarming are correlated with electrical isolation 7; however, no precise cut-off values have been assessed and this approach can only be utilized in a retrospective manner and does not allow for confirmation of PV occlusion prior to the initiation of the freezing process 2. Confirmation of complete occlusion is most commonly assessed by intravenous injection of radioiodine contrast after balloon inflation at the pulmonary vein ostium 6. Contrast leak upon balloon inflation is assessed fluoroscopically and confirms incomplete pulmonary vein occlusion with the cryoballoon. Multiple injections are frequently required upon cryoballoon repositioning resulting in repeated patient exposure to radioiodine contrast and radiation. As such, contrast may be contraindicated in patients with chronic kidney disease or known allergic reaction to iodine. Lack of continuous feedback during initial balloon positioning and inflation is an added disadvantage of fluoroscopic assessment for PV occlusion 3. Lastly, TEE-guided balloon positioning has been well described and allows for direct visualization of the antral position of the cryoballoon and continuous reassessment of balloon positioning and PV occlusion with saline infusion 8 as well as evaluation for “leaks” with color flow Doppler; however this technique requires continuous utilization of transesophageal echocardiography under general anesthesia, potentially increasing the risk for periprocedural complications associated with general anesthesia and espophageal intubation.
We observed a steep learning curve allowing for rapid and consistent interpretation of pressure waveforms after limited initial experience. Contrast utilization is directly negatively correlated with acquiring expertise in this technique and utilization of fluoroscopy is also dramatically reduced after initial training on pressure monitoring.
The pressure monitoring described surpasses the pitfalls mentioned above and allows for accurate and continuous assessment of PV occlusion. Currently, one balloon design is available for clinical use (Medtronic Inc., MN). This system includes an over-the-wire balloon catheter, which is cooled using nitrous oxide. Contrast is typically injected via an inner lumen; for the purpose of our described method of pressure monitoring, this inner lumen is utilized for direct assessment of PV pressure with the utilization of a single pressure transducer. As such, other than a single 3-way manifold, no extra catheters are necessary for application of this method. Further, if necessary, contrast may still be utilized if needed for further confirmation of PV occlusion, though we have demonstrated in a single center study that contrast utilization is largely unnecessary once the pressure monitoring technique has been established.
Pressure waveforms are obtained via the pressure transducer connected to the inner lumen and are displayed on the screen. The balloon is further advanced to the pulmonary vein ostium and inflated as done during standard cryoablation; no adjustments in the standard balloon manipulation approach are necessary when the technique of pressure monitoring is applied. Continuous monitoring for acute dislodgement during the initial freezing period is also possible with the pressure monitoring technique, until a measured scavenged gas temperature drop below -10 degrees Celsius. Particularly, balloon dislodgement from the PV ostium is usually observed upon initial refrigerant injection with resultant balloon expansion (“pop-out” phenomenon) and refinement of balloon positioning can be performed safely if the pressure waveform loses the occlusive pressure characteristics. A similar technique has been utilized in patients presenting in sinus rhythm for cryoballoon ablation 9. We have observed accurate determination of occlusive pressure in patients with ongoing atrial fibrillation during the procedure.
Currently, two sizes of the cryoballoon catheter are available, 28 mm and 23 mm diameter 10. Usage of the 23 mm balloon is limited, however the pressure monitoring technique is not affected by balloon size. The pressure monitoring technique is independent of rhythm and is applicable to patients presenting in atrial fibrillation. It is imperative that recording noise is eliminated and accurate zeroing is obtained to allow for accurate detection of the point of occlusive pressure. Displaying systemic arterial pressure as well as coronary sinus (CS) or surface ECG tracings are important to determine “live” the presence of A waves and change in V wave morphology. A common ostium is frequently observed, particularly in the left PVs. We observed occlusive pressure waveforms and complete electrical isolation in 16/18 PV with common ostium. Subselective wiring of the inferior and superior branches can be achieved in the presence of common ostium. Given these results, presence of a common PV ostium is not a limiting factor in the utilization of this technique.
Recurrent AF or atrial flutter was noted in fourteen patients (40%) after the initial 3 month blanking period and repeat PVI was performed in twelve patients (35%). Our results are comparable to published literature considering the inclusion of patients with persistent atrial fibrillation as well as six patients who had previously undergone PVI, implying a more extensive AF substrate. Interestingly, recurrences were more common in the first ten cases (6/10) despite higher utilization of contrast on initial testing. Of the last fifteen patients studied with minimal contrast utilization (mean contrast 10.9cc), four had recurrent AF post procedure indicating that the dramatic reduction in contrast utilization did not adversely influence clinical outcomes.
Limitations
Given the inability to record a pressure waveform once the intraluminal temperature decreases to less than -10 degrees, late balloon dislocations cannot be detected. This however is unlikely since at this point cryoadherence has occurred. In addition, perturbations are introduced in the pressure waveform during phrenic nerve pacing and pressure recording characteristics become less accurate immediately upon inflation in the right PVs. Atrial fibrillation is not a limiting factor for pressure monitoring application, however it may make the waveform interpretation more difficult for the inexperienced operator. Potential risks associated with air trapped in the 3-way system are limited but not insignificant. However, similar risk for air embolism is posed by utilization of the inner catheter lumen for contrast injection.
In this pilot study, the positive predictive value was noted to be 99% indicating a high likelihood of complete electrical isolation upon observation of occlusive pressure prior to cryoapplication. In contrast, the negative predictive value was low (40%) and was driven by the frequency of non occlusive pressure waveforms in the RIPV. In all patients, more than one cryoapplications were performed and as such we cannot exclude the possibility of minimal leaks during each cryoapplication with eventual complete electrical isolation due to refined maneuvering in the vein and apposition of the balloon in previously no ablated regions. This is particularly true for the RIPV in which, due to its relatively steep entrance into the antrum, slight changes in balloon orientation are required to achieve effective cryoablation in a segmental approach. As such, pressure monitoring may not be as helpful in the RIPV given this anatomical variation.
In our series, we observed lack of electrical vein isolation despite demonstration of occlusive pressure in two veins. There are several plausible explanations for this discrepancy, including the presence of epicardial muscle fibers between the antrum and the pulmonary vein or balloon-vein mismatch that results in ineffective cryoablation in part of the tissue. This phenomenon occurred during utilization of the first generation cryoballoon and is highly unlikely to occur with the imminent availability of the cryoballoon Advance, which will allow more equal distribution of cryoenergy at the PV ostium. Finally, the technique was primarily applied to patients undergoing cryoballoon ablation under conscious sedation, however assessment of the pressure waveform at end-expiration in the intubated patient should theoretically provide the same accurate results. We have observed consistent results in a small subset of patients undergoing general anesthesia, though this subgroup was not included in the analysis presented here.
Conclusions
The technique of pressure assisted PV isolation is simple, reproducible and safe. It significantly reduces the contrast burden and radiation exposure. Since the pressure can be recorded after the freezing process starts, the risk of loss of occlusion in the initial several seconds of balloon cooling (due to “pop-out” phenomenon) is minimized. Fine adjustments of balloon positioning can be safely made within the first 10 sec of cooling before the freezing occur.
The authors have nothing to disclose.
Funding was obtained by Medtronic Inc.
Name of Reagent/Material | Company | Catalog Number | Comments |
Standard Angiographic Kit with NAMIC manifold, Three-Valve Classic (Clear) Manifold with Ports on Right and ON Handles | Navylist Medical, USA | 91301303 | Individual components of the kit can be purchased separately |
Touhey Needle | Merit Medical, IRL | MAP 111 | |
EP MED system | St Jude Medical, MN, USA | ||
Pressure Monitoring Kit with TruWave Disposable Pressure Trasnducer | Edwards Lifesciences, CA, USA | PX284 |