Source: Wu, D. et. al., Rapid Determination of Antibody-Antigen Affinity by Mass Photometry. J. Vis. Exp. (2021)
In this video, we demonstrate the use of mass photometry to study antigen-antibody interactions in a flow chamber. The focused light scatters differently based on the mass of the molecules. A detector collects the scattered light and generates the interference pattern as spots of varied intensities, which correlate with the molecular masses.
1. Prepare the flow chambers
2. Prepare the antibody-antigen samples for the affinity measurements
3. Collect the Mass Photometry data
4. Analyze the MP data
Figure 1: MP flow chamber preparation and loading. (A) Coverslip holding position for the cleaning procedure. (B) Alignment of the 24 x 24 mm coverslip (middle layer) and the double-sided tape (top layer) on the surface of aluminum foil (bottom layer, not shown). Blue dashed lines show the location of the cut lines. (C) Top and side view of the assembled flow chamber with two sample channels, and a picture of the assembled flow chamber. (D) Procedure for sample loading into a flow channel previously filled with buffer.
Figure 2: Mass photometry images. (A) Representative native view image of the imaging buffer collected on a clean coverslip and (B) on a coverslip with surface imperfections. (C) Differential ratiometric image of the imaging buffer and (D) the AHT·HT solution.
The authors have nothing to disclose.
AcquireMP | Refeyn | MP data collection software | |
Anti-human thrombin | Haematologic Technologies | AHT-5020 | RRID: AB_2864302 |
Cotton-tipped applicators | Thorlabs | CTA10 | Cotton optical swabs for lens cleaning |
Coverslips 24×24 mm | Globe Scientific | 1405-10 | |
Coverslips 24×50 mm | Fisher Scientific | 12-544-EP | |
DiscoverMP | Refeyn | MP data processing software | |
Forceps | Electron Microscopy Sciences | 78080-CF | Soft-tipped forceps for coverslips handling |
Human α-thrombin | Haematologic Technologies | HCT-0020 | |
Immersion oil | Thorlabs | MOIL-30 | |
Isopropanol | Alfa Aesar | 36644 | |
Microsoft Excel | Microsoft | Spreadsheet | |
OneMP | Refeyn | Mass Photometry instrument | |
Origin | OriginLab | Scientific graphing software | |
PBS | Corning | 46-013-CM | 10x stock |
Syringe filter | Millipore | SLGSR33SS | Buffer and sample filtering |