Summary

Atomic Force Microscopy to Study the Physical Properties of Epidermal Cells of Live Arabidopsis Roots

Published: March 31, 2022
doi:

Summary

The atomic force microscopy indentation protocol offers the possibility to dissect the role of the physical properties of the cell wall of a particular cell of a tissue or organ during normal or constrained growth (i.e., under water deficit).

Abstract

A method is described here to characterize the physical properties of the cell wall of epidermal cells of living Arabidopsis roots through nanoindentations with an atomic force microscope (AFM) coupled with an optical inverted fluorescence microscope. The method consists of applying controlled forces to the sample while measuring its deformation, allowing quantifying parameters such as the apparent Young’s modulus of cell walls at subcellular resolutions. It requires a careful mechanical immobilization of the sample and correct selection of indenters and indentation depths. Although it can be used only in external tissues, this method allows characterizing mechanical changes in plant cell walls during development and enables the correlation of these microscopic changes with the growth of an entire organ.

Introduction

Plant cells are surrounded by a cell wall that is a complex structure composed of interacting networks of polysaccharides, proteins, metabolites, and water that varies in thickness from 0.1 to several µm depending on the cell type and the phase of growth1,2. Cell wall mechanical properties play an essential role in the growth of plants. Low stiffness values of the cell wall have been proposed as a precondition for cell growth and cell-wall expansion, and there is increasing evidence that all cells sense mechanical forces to perform their functions. However, it is still debated whether changes in the physical properties of the cell wall determines cell fate2,3,4. Because plant cells do not move during development, the final shape of an organ depends on how far and in what direction a cell expands. Thus, Arabidopsis root is a good model to study the impact of cell wall physical properties in cell expansion because different types of expansion occur in different regions of the root. For example, anisotropic expansion is evident in the elongation zone and particularly noticeably in the epidermal cells5.

The method described here was used to characterize the physical properties of the cell wall of epidermal cells at the nanoscale of living Arabidopsis roots using an Atomic Force Microscope (AFM) coupled with an inverted fluorescence phase microscope6. For an extensive revision of the AFM technique, read7,8,9.

This protocol outlines a basic sample preparation method and a general method for AFM-based elasticity measurements of plant cell walls.

Figure 1
Figure 1: Schematic overview of force-indentation experiment in Arabidopsis roots using atomic force microscopy (AFM). The scheme gives an overview of the steps of a Force-Indentation experiment from the preparation of the substrate to immobilize the root sample firmly (1-2), root viability confirmation through propidium iodide staining (3), cantilever positioning on the surface of an elongated epidermal cell of the primary root (4-5), force curves measurement (6), and force curve processing to calculate the apparent Young's modulus (7-8). EZ: elongation zone. Please click here to view a larger version of this figure.

Protocol

1. Preparation of the plant material and growth conditions

  1. To generate the needed plant material, sterilize the seeds of Arabidopsis wild type and mutant lines of interest.
    NOTE: In this protocol, we used the following: ttl1: T-DNA insertion lines Salk_063943 (for TTL1; AT1G53300) – Columbia-0 (Col-0) wild-type; the Procuste1 (prc1-1) mutant, which consists of a knock-out mutation (Q720stop) in the CESA6 gene (AT5G64740), as previously described in10,11; and ttl1 x prc1-1 double mutant.
    1. Put an approximate volume of 50 μL of seeds in a microtube. Add 500 μL of 70% ethanol + 0.01% Tween solution at room temperature. Mix and incubate for 7 min.
    2. Pour the ethanol solution and add 500 μL of 20% sodium hypochlorite. Mix and incubate for 7 min.
    3. Pour the sodium hypochlorite solution and wash the seeds three to four times with sterile water.
  2. Plating
    1. Prepare in advance basal Murashige and Skoog (MS) medium12 + 1.5% sucrose + 1% agar (pH 5.7) (see Table of Materials). Autoclave the medium. Prepare a laminar flow hood to work in a sterile environment and label the square Petri dishes needed for the experiment. Pour the medium into the Petri dishes and allow the medium to solidify.
    2. Plate the sterilized seeds using sterile pipette tips onto square Petri dishes containing the medium. Distribute the seeds evenly in two rows. When the seeds are dry, close the lids and seal the plates. Stratify the seeds at 4 °C in the dark for 4 days.
      NOTE: Stratification is needed to synchronize germination.
    3. Place the plates vertically in the growth chamber in a long-day regime (16 h of light/8 h of dark) with 25 μmol m2 s−1 light intensity and 23 °C (day/night). Grow the seedlings for 7 days.

2. Osmotic stress treatment (optional)

NOTE: This section provides details on the growth of Arabidopsis roots in osmotic potential of -1.2 MPa as estimated by cryoscopic osmometer (Table of Materials). This part can be omitted or changed depending on the experimental question at hand.

  1. Prepare basal MS medium + 1.5% sucrose + 400 mM mannitol + 1% agar (pH 5.7). Autoclave the medium. Pour the medium in square Petri dishes in the laminar flow hood and allow the medium to solidify.
  2. With tweezers (Table of Materials), place 5-day old seedlings pre-grown in basal MS + 1.5% sucrose medium in square Petri dishes containing the medium supplemented with 400 mM mannitol. Place the plates vertically in the growth chamber in the same conditions as in step 1.2.3. Grow the seedlings under this osmotic stress condition for 7 days.
    NOTE: To evaluate root growth adaptation of the primary root during severe osmotic stress, it is suggested to use seedlings 5 days after germination grown in controlled conditions to ensure that the root meristem size is already established13. Alternatively, the analysis can be performed at every stage of development to determine the developmental effect of osmotic stress action.

3. Atomic force microscopy (AFM) nanoindentation experiments

  1. Preparation of the sample for nanoindentation experiments
    NOTE: This is a crucial step for applying AFM to the study of live biological samples. The sample must be firmly attached to the substrate to be mechanically stable during indentation measurements. At the same time, sample structural qualities should be preserved. An effective strategy to stabilize a big sample for AFM is to use adhesives. The adhesive must dry quickly and must not be toxic or reactive with the surrounding medium. In this protocol, polystyrene Petri dishes were used as the substrate and non-acidic silicone glue (Table of Materials) was used to bind the Arabidopsis seedlings.
    1. Spread a thin layer of silicone glue into the Petri dish with a coverslip. Leave the glue in the air for 45 s.
    2. With tweezers, place the seedling on the glue, orienting it in a direction that avoids contact between the seedling's protruding parts and the cantilever. Then, press the root gently to the silicone glue layer to bind it firmly. Leave the seedling in contact with the glue for 45 s before continuing with the next steps of the protocol.
    3. Incubate the attached seedlings with 2 μg/mL propidium iodide (PI; Table of Materials) for 5 min in dark and carefully wash them thrice with 1x phosphate-buffered saline (PBS) solution (Table of Materials).
    4. Place the PI-incubated seedlings into an inverted fluorescence microscope coupled with the AFM. To observe fluorescence, set the excitation and emission wavelengths at 572 nm and 617 nm, respectively. The absence of fluorescence inside the epidermal cells confirms the viability of the roots.
    5. Cover the seedling with 1x phosphate-buffered saline (PBS) solution.
  2. Indentation protocol
    NOTE: Perform all AFM measurements within 1 h after the immobilization of the sample on its support at room temperature. The room where the AFM is located should be maintained at 25 °C with an air conditioner. The PBS solution should be at the same temperature.
    1. Mount a standard silicon nitride cantilever with a pyramidal tip (see Table of Materials) into the AFM probe holder for fluid (Table of Materials).
    2. Align the laser on the cantilever close to the position of the tip. Next, move the photodiode to place the laser spot at the center of the detector.
    3. To calibrate the deflection sensitivity, position a polystyrene Petri dish with 1x PBS under the AFM.
      NOTE: This step is needed to calculate the indentation. When a force measurement is made on a hard surface, there is no indentation into the surface, so any change in the z scanner displacement corresponds to the cantilever deflection. Therefore, it is very important to calibrate the deflection sensitivity on a hard surface. Using a soft surface will overestimate the deflection sensitivity, which will result in spring constants that are too low.
    4. Adjust the photodetector such that the non-contact deflection is near 0 V. Calibrate the deflection sensitivity by performing an indentation (force curve), with a ramp size of 3 μm, an indentation-and-retraction rate of 0.6 μm/s, and a trigger threshold of 0.5 V.
    5. To calibrate the spring constant of the cantilever, use the Thermal Tune utility of the AFM software (Table of Materials) recommended for probes with spring constants less than approximately 5 N/m or use the method described in reference14. Before activating Thermal Tune in the software, ensure that the probe does not interact with the sample.
      1. Enter the cantilever temperature. Click on Calibrate > Thermal Tune or on the Thermal Tune icon in the NanoScope toolbar. Select a frequency range (1-100 Hz).
      2. Click on Acquire Data in the Thermal Tune panel. Wait for the AFM to acquire data, and then click on the Simple Harmonic Oscillator (Fluid) button.
      3. Adjust Median Filter Width to 3. Adjust PSD Bin Width to reduce the noise in the acquired data by averaging. Set fit boundaries around the first resonance peak.
      4. Click on Calculate Spring Constant k, and then, click on Yes in the pop-up window asking if the user wants to use this value.
      5. Repeat the steps described above three times and manually take an average of spring constant values. Enter this average value in the Spring Constant box in the Ramp parameter list in the PicoForce view. The calibration ends at this point.
    6. Using the inverted optical microscope at 10x, 20x, and 40x eyepiece magnification, position the AFM probe on the surface of the fourth elongated epidermal cell of the primary root, ensuring to position it in the center of the cell.
    7. With the calculated spring constant value (step 3.2.5.5), obtain force curves with a ramp size of 3 µm, a trigger threshold of 11 nN, and an indentation-and-retraction rate of 0.6 μm/s at selected points.
      ​NOTE: Previous work15,16 found that low-frequency of indentation-and-retraction helps to minimize hysteresis and drag force.
    8. Obtain force curves from three cells per root for each treatment (use three biological replicates per treatment). Capture at least 150 force curves for each root.

4. Measure the apparent Young's modulus

  1. Fit each force curve to the following model for pyramidal indenters17: Equation 1, where E is the apparent Young's modulus of the sample, Equation 2 is the Poisson modulus of the sample, and α is the semi-angle to the vertex. Discard the fits with a correlation coefficient (r2) < 0.99. Consider the Poisson's ratio for totally incompressible material: Equation 2 = 0.5 and the geometry of the indenter given by the manufacturers.
    NOTE: Use the first 100 nm from the contact point of the loading curve for the fitting to be as sensitive as possible to cell wall mechanics and to reduce the impact of turgor pressure on the apparent Young's modulus values8. The fitting has been done using a custom MATLAB program (available upon personal request by writing to J. C. B) that allows setting a specific indentation depth from the contact point of the loading curve.
  2. Create a normalized histogram with the apparent Young's modulus data and fit it with a Gaussian distribution. Discard data points outside the 95% confidence interval and recalculate both the histogram and the Gaussian fit. Calculate and report the mean and standard deviation of the apparent Young's modulus.
  3. Determine the significance of the comparisons between groups by multiple comparison tests like ANOVA.

Representative Results

Force-Indentation experiments
The following text presents some results expected when a force-indentation experiment is conducted to show the typical output to expect when the protocol is well executed.

Force-displacement curves
Representative force indentation plots that were obtained indenting live samples at a position placed in the center of the cell of the root elongation zone are presented in Figure 2. When the AFM tip starts to indent the surface of the cell wall, the force begins to increase (indicated by 1 in Figure 2A) because of the cell wall opposition to the deformation. The increase of force (loading) continues until the maximum force value is reached (indicated by 2 in Figure 2A); after this point, the unloading part of the indentation begins.

Figure 2
Figure 2: Force-displacement curves obtained in indentation experiments made in cell walls of live root elongation zone cells. (A) A typical force-displacement curve. At the position marked by 1, the force begins to increase; the AFM tip starts to indent the cell wall. The increase of force (loading) continues until the maximum force value is reached, as indicated with 2; after this point, the unloading part of the indentation begins. (B) Example of a force-displacement curve in which it is impossible to detect the position of the cell surface before the indentation (indicated by a), and both the loading and unloading curves are noisy. Please click here to view a larger version of this figure.

The expected force-displacement curves will show in the indentation part that the force grows following a parabola, as predicted by the model (explained in step 4.1) (Figure 2A). Another fitting parameter is the contact point position; it should correspond to the cell wall surface before indentation and is considered the origin of the AFM tip displacement. Force curves in which it is impossible to detect the point of contact before the indentation should be discarded (Figure 2B). The loading and unloading curve of the force-indentation experiment must be devoid of noise. Noisy curves such as the one represented in Figure 2B must be discarded.

Normalized histogram of the apparent Young's modulus
The histograms showing the distribution of the frequency of the obtained values of the apparent Young's modulus for a set of 201 successful indentations on nine different cells of three different plants of Col-0 grown in control conditions are presented in Figure 3. The figure shows an example in which the mean and standard deviation of the apparent Young's modulus (88.12 ± 2.79 KPa) were calculated from the histograms fitted with a Gaussian curve6.

Figure 3
Figure 3: Example of a histogram of apparent Young's modulus that allowed a proper statistical analysis of the results. Data were obtained from force curves on nine different cells of three different plants of Col-0 grown in control conditions. The data obtained from 201 Force curves were fitted to a Gaussian distribution. The relative frequency indicates the number of fitted force curves that correspond to each calculated apparent Young's modulus. The mean and standard deviation values of this condition were obtained from the Gaussian fits. RF: Relative frequency; AEM: apparent Young's modulus. This figure was modified from reference6 Please click here to view a larger version of this figure.

Some genotypes could be very difficult to indent due to the morphology of the root. This was the case for prc1-1 mutant and ttl1 prc1-1 double mutant grown in severe osmotic stress. In this condition, these mutants had few elongated cells in the root elongation zone and developed very long root hairs, which interfere with the cantilever producing the sample movement. Figure 4 presents histograms showing the probability distribution of the obtained values of the apparent Young's modulus on nine different cells of the ttl1 prc1-1 double mutant. The figure shows an example in which the obtained force curves did not allow the proper analysis. Only Histogram H that corresponds to one cell could be fitted to a Gaussian distribution6.

Figure 4
Figure 4: Example of histograms of apparent Young's modulus that did not allow a proper analysis. Data were obtained from force curves on nine different cells of three different plants of the ttl1prc1-1 double mutant. The relative frequency indicates the number of fitted force curves that correspond to each calculated apparent Young's modulus. Only histogram H could be fitted to a Gaussian distribution. RF: Relative frequency; AEM: the apparent Young's modulus. This figure was modified from reference6. Please click here to view a larger version of this figure.

Discussion

Cell and cell-wall mechanics are increasingly becoming relevant to gain insight into how mechanics affects growth processes. As physical forces propagate over considerable distances in solid tissues, the study of changes in the physical properties of the cell wall and how they are sensed, controlled, tuned, and impact the plant's growth are becoming an important field of study2,3,8.

A method is presented here for studying the cell wall physical properties of cells in the elongation zone of the Arabidopsis root. The AFM technique may be readily combined with other techniques, such as growth rate studies, to understand the effect of the cell wall physical properties in root growth.

One of the most critical experimental steps is the mechanical immobilization of the sample; the sample must be firmly attached to avoid the sample movement while it is indented. The layer of the silicone glue must be thin enough to allow it to dry quickly. The manipulation of the silicone glue and the sample should be careful to ensure that the silicone does not cover the areas of the sample that will be measured. Furthermore, the preparation should be completed quickly (within a minute) to maintain the silicone glue's adhesive properties and prevent roots dehydration. It is crucial to prevent tissue damage; dehydration and damage of the epidermal tissue will produce irreversible changes in the mechanical properties of the cell wall. Other methods used immobilization in 1% agarose on a glass slide; however, this method also has the challenge of needing quick manipulation and an additional step of preparation of the glass slides to prevent the samples from moving in the X and Y axes during the analysis. For more information, see references8,18. Another relevant point of this protocol is to count with an optical microscope with epifluorescence coupled to the AFM that has different objectives (10x, 20x, and 40x). This allows selecting the indentation area accurately.

One challenging aspect of the protocol is the selection of the AFM indentations settings. Different indenters and indentation depths render different resolutions and can be used to answer different research questions8. Several studies have shown that the tip's size and geometry are important considerations to discern the desired results. Different information is obtained with different tip shapes7. For example, the conical tip can discriminate mechanical properties at the sub-cellular level, the spherical tip can better represent the mechanical properties of the whole cell19, and pyramidal tips allow to obtain high-resolution images and mechanical properties at the same time7,20. The selection of the pyramidal tip radius (20-60 nm) is important to prevent the tip from sinking, allowing a proper indentation process. For each application, the researcher must choose the type of cantilever (and tip) better suited for the sample and the kind of measurements to perform8.

The method has some limitations: the mechanical properties of only cells at the surface of organs can be measured, and tissues with variable topographies are difficult to work with. For example, root hairs could make it difficult to access the epidermal cells with the AFM tip8,18. Finally, while the method presented measures elasticity, nanoindentation has also been used to report viscosity and turgor pressure in plants and animals8,18.

Offenlegungen

The authors have nothing to disclose.

Acknowledgements

This research was funded by CSIC I+D 2018, grant No. 95 (Mariana Sotelo Silveira).; CSIC Grupos (Omar Borsani) and PEDECIBA.

Materials

1 x Phosphate-Buffered Saline (PBS) Include sodium chloride and phosphate buffer and is formulated to prevent osmotic shock and maintain water balance in living cells.
AFM software Bruker, Billerica, MA, USA
Atomic force microscopy (AFM) BioScope Catalyst, Bruker, Billerica, MA, USA
Catalyst Probe holder-fluid Bruker, Billerica, MA, USA CAT-FCH A probe holder for the Bioscope Catalyst, designed for fluid operation in contact or Tapping Mode.  Also compatible with air operation.
Cryoscopic osmometer; model OSMOMAT 030 Gonotech, Berlin, Germany
Murashige & Skoog Medium Duchess Biochemie M0221 Original concentration, (1962)
Optical inverted microscope coupled to the AFM Olympus IX81, Miami, FL, USA
PEGAMIL ANAEROBICOS S.R.L., Buenos Aires, Argentina 100429 Neutral, non acidic silicone glue
Petri dishes Deltalab 200201.B Polystyrene, 55 x 14 mm, radiation sterile.
Propidium iodide Sigma P4170 For root viability test.
Silicon nitride probe, DNP-10, cantilever A Bruker, Billerica, MA, USA DNP-10/A For force modulation microscopy in liquid operation. Probe tip radius of 20-60 nm. 175-μm-long triangular cantilever,  with a spring constant of 0.35 N/m.
Tweezers Sigma T4537

Referenzen

  1. Anderson, C. T., Kieber, J. J. Dynamic construction, perception, and remodeling of plant cell walls. Annual Review of Plant Biology. 71, 39-69 (2020).
  2. Roeder, A. H. K., et al. Fifteen compelling open questions in plant cell biology. The Plant Cell. 34 (1), 72-102 (2022).
  3. Zhang, B., Gao, Y., Zhang, L., Zhou, Y. The plant cell wall: Biosynthesis, construction, and functions. Journal of Integrative Plant Biology. 63 (1), 251-272 (2021).
  4. Hamant, O., Haswell, E. S. Life behind the wall: Sensing mechanical cues in plants. BMC Biology. 15 (59), 1-9 (2017).
  5. Scheres, B., Benfey, P., Dolan, L. Root development. The Arabidopsis Book. 1, 0101 (2002).
  6. Cuadrado-Pedetti, M. B., et al. The arabidopsis tetratricopeptide thioredoxin-like 1 gene is involved in anisotropic root growth during osmotic stress adaptation. Genes. 12 (2), 236 (2021).
  7. Milani, P., Braybrook, S. A., Boudaoud, A. Shrinking the hammer: micromechanical approaches to morphogenesis. Journal of Experimental Botany. 64 (15), 4651-4662 (2013).
  8. Braybrook, S. A. Measuring the elasticity of plant cells with atomic force microscopy. Methods in Cell Biology. 125, 237-254 (2015).
  9. Bidhendi, A. J., Geitmann, A. Methods to quantify primary plant cell wall mechanics. Journal of Experimental Botany. 70 (14), 3615-3648 (2019).
  10. Desnos, T., et al. Procuste1 mutants identify two distinct genetic pathways controlling hypocotyl cell elongation, respectively in dark- and light-grown Arabidopsis seedlings. Development. 122 (2), 683-693 (1996).
  11. Fagard, M., et al. Procuste1 encodes a cellulose synthase required for normal cell elongation specifically in roots and dark-grown hypocotyls of arabidopsis. The Plant Cell. 12 (12), 2409-2423 (2000).
  12. Murashige, T., Skoog, F. A revised medium for rapid growth and bio assays with tobacco tissue cultures. Physiologia Plantarum. 15 (3), 473-497 (1962).
  13. Perilli, S., Sabatini, S. Analysis of root meristem size development. Methods in Molecular Biology. 655, 177-187 (2010).
  14. Sader, J. E., et al. A virtual instrument to standardise the calibration of atomic force microscope cantilevers. Review of Scientific Instruments. 87 (9), 093711 (2016).
  15. Collinsworth, A. M., Zhang, S., Kraus, W. E., Truskey, G. A. Apparent elastic modulus and hysteresis of skeletal muscle cells throughout differentiation. American Journal of Physiology – Cell Physiology. 283 (4), 1219-1227 (2002).
  16. Mathur, A. B., Collinsworth, A. M., Reichert, W. M., Kraus, W. E., Truskey, G. A. Endothelial, cardiac muscle and skeletal muscle exhibit different viscous and elastic properties as determined by atomic force microscopy. Journal of Biomechanics. 34 (12), 1545-1553 (2001).
  17. Sirghi, L., Ponti, J., Broggi, F., Rossi, F. Probing elasticity and adhesion of live cells by atomic force microscopy indentation. European Biophysics Journal. 37 (6), 935-945 (2008).
  18. Peaucelle, A. AFM-based mapping of the elastic properties of cell walls: At tissue, cellular, and subcellular resolutions. Journal of Visualized Experiments: JoVE. (89), e51317 (2014).
  19. Peaucelle, A., et al. Pectin-induced changes in cell wall mechanics underlie organ initiation in Arabidopsis. Current Biology. 21 (20), 1720-1726 (2011).
  20. Fernandes, A. N., et al. Mechanical properties of epidermal cells of whole living roots of Arabidopsis thaliana: An atomic force microscopy study. Physical Review E. 85 (2), 21916 (2012).

Play Video