Cavitation microbubbles are imaged using a high-speed camera attached to a zoom lens. The experimental setup is explained, and image analysis is used to calculate the area of the cavitation. Image analysis is done using ImageJ.
An experimental and image analysis technique is presented for imaging cavitation bubbles and calculating their area. The high-speed imaging experimental technique and image analysis protocol presented here can also be applied for imaging microscopic bubbles in other fields of research; therefore, it has a wide range of applications. We apply this to image cavitation around dental ultrasonic scalers. It is important to image cavitation to characterize it and to understand how it can be exploited for various applications. Cavitation occurring around dental ultrasonic scalers can be used as a novel method of dental plaque removal, which would be more effective and cause less damage than current periodontal therapy techniques. We present a method for imaging the cavitation bubble clouds occurring around dental ultrasonic scaler tips using a high-speed camera and a zoom lens. We also calculate the area of cavitation using machine learning image analysis. Open source software is used for image analysis. The image analysis presented is easy to replicate, does not require programming experience, and can be modified easily to suit the application of the user.
Imaging the motion of bubbles is important for various applications because it controls the hydrodynamics of a system. There are many applications where this can be useful: in fluidized bed reactors1,2, or for cleaning with cavitation bubbles3,4. The purpose of imaging bubbles is to understand more about the bubble dynamics or about the direction and motion of a cloud of bubbles. This can be done through observing structures imaged and also by using image analysis to obtain quantitative information, such as the size of the bubbles.
Cavitation bubbles are gas or vapor entities that occur in a fluid when the pressure drops below the saturated pressure value5. They can occur when an acoustic field is applied to a fluid at ultrasonic frequencies. They repeatedly grow and collapse, and upon collapse can release energy in the form of high-speed micro-jets and shockwaves6,7. These can dislodge particles on a surface through shear forces and cause surface cleaning8. Cavitation bubbles are being investigated for surface cleaning in different industries, such as for semiconductors, food, and wound cleaning9,10,11,12. They could also be used to clean dental plaque from teeth and biomaterials such as dental implants12,13. Cavitation occurs around currently used dental instruments such as ultrasonic scalers and endodontic files and shows potential as an additional cleaning process with these instruments14.
The oscillation of cavitation bubbles occurs over a few microseconds and therefore a high-speed camera is required to capture their motion by imaging at thousands of frames per second8. We demonstrate a method of imaging microbubble cavitation around dental ultrasonic scalers. The aim is to understand how cavitation varies around different ultrasonic scalers, so it can be optimised as a novel way to clean dental plaque.
Previous methods used to investigate the cavitation include sonochemiluminesence, which uses luminol to detect where cavitation has occurred15,16. However, this is an indirect technique and it is not able to visualize the cavitation bubbles in real time. Therefore, it is not able to accurately determine exactly where it happens on the instrument, and no information can be gained on the bubble dynamics, unless it is combined with other imaging techniques17. High-speed imaging can image not only the cavitation bubbles growing and collapsing but also the type of cavitation occurring: cavitation clouds, microstreamers and micro-jets6,7,18. These give more information about how the cavitation can clean surfaces.
We present a method of imaging cavitation microbubbles using a high-speed camera and calculating the mean area of cavitation occurring. This method is demonstrated using an example of cavitation occurring around different dental ultrasonic scaler tips, although the experimental and image analysis steps can be used for other applications, such as for imaging other macro and microbubbles.
1. Instrument setup
2. High-speed camera setup
3. Calibration
4. High-speed video recording
5. Image processing
6. ImageJ macro
The image analysis steps can be seen in Figure 1 for one of the ultrasonic scaler tips tested. A FSI 1000 tip and a 10P tip were imaged inside a water tank with the cooling water turned off (Figure 2). Cavitation occurred near the bend of tip FSI 1000 at maximum power, and near the free end in tip 10P (Figure 3 and Figure 4). The mean area of cavitation was 0.1 ± 0.07 mm2 for the FSI 1000 tip and 0.50 ± 0.25 mm2 for the 10P tip (Figure 5).
Figure 1: High-speed imaging setup and image analysis steps (a) Schematic of the high-speed imaging setup used in the study. (b) Schematic of the image analysis steps used in the study, showing the raw images on the left of the scaler tip only and with cavitation, which were then binarized and subtracted from each other to calculate the area of the cavitation clouds. Please click here to view a larger version of this figure.
Figure 2: Comparison between different tips High-speed image stills showing cavitation occurring around the two ultrasonic scaler tips tested (a) FSI 1000 (b) 10P. Please click here to view a larger version of this figure.
Figure 3: Tip 10P high-speed images: High-speed image stills of tip 10P, from a video taken at 6400 frames per second. Cavitation can be seen around the free end of the tip. Please click here to view a larger version of this figure.
Figure 4: Tip FSI1000 high-speed images: High-speed image stills of tip FSI 1000, from a video taken at 6400 frames per second. Cavitation can be seen around the middle of the tip. Please click here to view a larger version of this figure.
Figure 5: Cavitation area image analysis results. The mean area of cavitation occurring around the FSI 1000 and 10P ultrasonic scaler tips calculated using the image analysis technique described. The error bars represent the standard deviation. Please click here to view a larger version of this figure.
Supplementary File. Please click here to download this file.
The technique described in this paper enables imaging of fast-moving microbubbles with high spatial and temporal resolution. It can potentially benefit a wide range of scientific disciplines such as chemical engineering, dentistry and medicine. Engineering applications include imaging cavitation bubbles for cleaning surfaces, or for imaging bubbles in fluidized bed reactors. Biomedical applications include imaging cavitation around medical and dental instruments and imaging biofilm debridement from hard and soft tissue using cavitation bubbles. In this study we demonstrated the technique by imaging cavitation around two different dental ultrasonic scaler tips. The amount of cavitation varies between the two tips tested in this study, with more cavitation clouds observed around the free end of tip 10P. This has previously been linked to vibration amplitude20. The high-speed videos show that the FSI 1000 tip has less vibration, which is likely to be why there is less cavitation around this tip.
One limitation of the image analysis method is that the image subtraction technique to remove the area of the scaler is not completely accurate because the scaler is oscillating and therefore the subtraction may leave some areas of the scaler falsely segmented as bubbles. However, this has been accounted for by averaging the area from a large number of frames (n=2000). This would not be a problem for applications where the object to be subtracted is stationary. For studies where the moving object to be subtracted has a much higher variance, we recommend synchronizing the movements in both videos before subtracting for accurate results. In the current study, we did not synchronize the oscillations but since the vibration was low, we can assume that the oscillations correspond well to each other in these two measurements.
The image thresholding is accurate because the brightfield illumination provides a uniform background with good contrast. It is critical to ensure that the background is uniform and does not contain any other objects which could be falsely segmented. The thresholding method can be modified by using other automatic thresholds to suit the application. Manual thresholding, where the user sets the threshold value, is also possible but is not recommended as it reduces the reproducibility of the results, since different users will select different threshold values.
Image analysis has been used for many other bubble imaging studies. These also use a similar method of backlighting to get optimum contrast between the bubbles and the background, and thresholding to segment the bubbles21,22,23,24. The method shown in the current study can also be generalized to be used for many different bubble imaging applications, which are not limited to only high-speed imaging. High-speed imaging has been used for cavitation bubbles generated in water and also around instruments such as endodontic files and ultrasonic scalers12,25,26,27,28. For example Rivas et al. and Macedo et al. used a high-speed camera attached to a microscope, with illumination provided by a cold light source to image cleaning with cavitation, and to image cavitation around an endodontic file17,29. Bright field illumination provides more contrast between the background and the bubbles, making it possible to use simple segmentation techniques such as thresholding, as demonstrated by Rivas et al. for imaging and quantifying cavitation erosion and cleaning over time29. Dark field illumination makes thresholding more difficult due to the higher variation in grey scales4,30. Image analysis has been used in other studies to gather more information about bubbles1,2. Vyas et al. used a machine learning approach to segment cavitation bubbles around an ultrasonic scaler20. The method described in the current paper is quicker because it uses simple thresholding so it is less computationally intensive, and bubbles occurring above and below the scaler can be analysed. However, the thresholding method used in the current paper is only accurate if the background is uniform. If it is not possible to obtain a uniform background during imaging, other image processing techniques can be used such as the use of background subtraction using a rolling ball radius to correct for uneven illumination, filtering using median or Gaussian filters to remove noise, or also using machine learning based techniques20,31.
In conclusion, we present a high-speed imaging and analysis protocol to image and calculate the area of a microscopic moving object. We have demonstrated this method by imaging cavitation bubbles around an ultrasonic scaler. It can be used for imaging cavitation around other dental instruments such as endodontic files and it can be easily adapted for other non-dental bubble imaging applications.
The authors have nothing to disclose.
The authors are grateful for funding from the Engineering and Physical Sciences Research Council EP/P015743/1.
0.25x attachment | Navitar | 1-50011 | |
12x with 12mm fine focus Long distance microscope zoom lens |
Navitar | 1-50486 | |
2x adaptor with f mount | Navitar | 1-62922 | |
Cavitron Plus Ultrasonic Scaler | Dentsply Sirona | 8184003 | |
Cavitron Ultrasonic Insert FSI 1000FSI 1000 | Dentsply Sirona | UCAFTHD | |
Fibre light guide. 8mm fibre bundle 1500mm length. Focussing lens assembly for Hayashi light, 1/4"-20 tripod thread for mounting. |
Hayashi | LGC1- 8L1500 |
|
Geared head | Manfrotto | MN405 | 7.5kg load capacity |
HDF7010 High-Power LED Endoscope light source. 150W LED provides cold output equivalent to 250W Xenon. |
Hayashi | LA-HDF710 | |
Heavy weight Tripod | Manfrotto | MN475B | Geared centre column, 12kg load capacity |
High Speed Camera | Photron | 103526 | FASTCAM Mini AX200 900K M3 (16GB memory) |
High-Precision Rotation Stage | Thorlabs | PR01/M | |
Laboratory jacks | Camlab | 1194083 | |
Micropositioning sliding plate | Manfrotto | SKU 454 | |
Micropositioning stage 3D | Thorlabs | PT3/M | |
Micropositioning stage rotation | Thorlabs | OCT-XYR1/M | OCT-XYR1/M – XY Stage with Solid Top Plate |
NEWTRON P5 XS Ultrasonic Scaler | Acteon | F62118 | |
Ultrasonic Insert 10P | Acteon | F00253 |