An additive manufacturing strategy for processing UV-crosslinkable hydrogels has been developed. This strategy allows for the layer-by-layer assembly of microfabricated hydrogel structures as well as the assembly of independent components, yielding integrated devices containing moving components that are responsive to magnetic actuation.
Polyethylene glycol (PEG)-based hydrogels are biocompatible hydrogels that have been approved for use in humans by the FDA. Typical PEG-based hydrogels have simple monolithic architectures and often function as scaffolding materials for tissue engineering applications. More sophisticated structures typically take a long time to fabricate and do not contain moving components. This protocol describes a photolithography method that allows for facile and rapid microfabrication of PEG structures and devices. This strategy involves an in-house developed fabrication stage that allows for the rapid fabrication of 3D structures by building upwards in a layer-by-layer fashion. Independent moving components can also be aligned and assembled onto support structures to form integrated devices. These independent components are doped with superparamagnetic iron oxide nanoparticles that are sensitive to magnetic actuation. In this manner, the fabricated devices can be actuated using external magnets to yield movement of the components within. Hence, this technique allows for the fabrication of sophisticated MEMS-like devices (micromachines) that are composed entirely out of a biocompatible hydrogel, able to function without an onboard power source, and respond to a contact-less method of actuation. This manuscript describes the fabrication of both the fabrication set-up as well as the step-by-step method for the microfabrication of these hydrogels-based MEMS-like devices.
MEMS devices have found a multitude of applications especially in the field of medical devices. Although they lend a lot of added functionalities and the miniaturized nature of these devices make them attractive for use as implantables1,2,3, these devices often have inherent safety and biocompatibility issues, as they are composed of materials that could be harmful to the human body (e.g., metals, batteries, etc.)4,5,6. PEG-based hydrogels are liquid swollen polymer networks and have been frequently used for applications such as tissue engineering scaffolds largely in part due to their high biocompatibility7,8. PEG-based hydrogels have also been FDA-approved for use in humans9,10,11. However, due to the material properties of hydrogel, they do not easily withstand normal manufacturing processes such as techniques used in typical silicon-based microfabrication. Thus, hydrogel-based constructs are typically limited to simple monolithic architectures. Current efforts at microfabrication of hydrogels have resulted in structures with micron-sized features; however, these structures are often of a single layer and a single material12,13 and lack moving components14,15,16.
In a previous work, we describe a strategy for fabricating micromachines that are composed entirely of a biocompatible PEG-based hydrogel material17. Micron-sized features can be fabricated easily using a photolithography method and these structures can be built upwards using a layer-by-layer method, enabled by the precise z-axis movement of the substrate on which the hydrogels are polymerized. Hydrogels of different compositions can be fabricated adjacent to each other. Additionally, these devices have moving components that can be actuated using an external magnet. This versatile technique is also suitable for processing any soft material or hydrogel that is photo-polymerizable. Thus, this technique is well-suited for fabricating sophisticated MEMS-like devices composed entirely of hydrogels.
1. Fabrication Stage
2. Fabrication of PDMS Chamber and Determining its "Zero" Level
3. Photomask Design for Photopolymerization of Hydrogel Microstructures
4. Treatment of Glass Coverslips to Prevent Adhesion of Hydrogels
5. Layer-by-layer Fabrication of Hydrogels: Top Sealing Layer and Bottom Support Structures
6. Assembling and Sealing the Hydrogel-based Device
7. Iron Oxide Doping of Hydrogel Components
8. Actuation of the Assembled Device
NOTE: The iron-doped components within the assembled device can be actuated to move using a strong permanent magnet such as neodymium (N52 strength). Be careful to avoid pinching hazards as these magnets are very strongly attracted to ferromagnetic materials.
Figure 3B shows images of the layers of hydrogels polymerized using the fabrication set-up. Figure 3B(i) shows a fabricated 400 µm thick base layer with a 600 µm aperture. Figure 3B(ii) shows a further two layers that were layered on top of the base layer; a 500 µm tall perimeter and an 800 µm tall axle in the middle. The total fabrication time for these three layers was less than 3 minutes taking into account 4 seconds of exposure for each layer and time taken to adjust the height of the bottom substrate and alignment of photomasks. Previous work performed on the same fabrication set-up demonstrates that a variety of designs can be fabricated with resolutions as high as 100 µm.
The hydrogel components could also be easily doped with iron oxide nanoparticles. The exposure times were optimized to ensure thin layers (200 µm) of PEGDA prepolymers doped with iron oxide nanoparticles could be fully polymerized. Figure 5A shows the photomask used to define the shape of the iron oxide segment to be polymerized. The un-doped PEGDA prepolymer can be fully polymerized within 4 seconds of UV exposure. However, when the iron oxide doped prepolymer was exposed for 4 seconds to UV, the resultant hydrogel was not fully polymerized, as can be seen in Figure 5C. The segment generated was thinner (as compared to a fully cross-linked segment shown in Figure 5B), and the edges were uneven with compromised fidelity as compared to the shape defined by the photomask. UV exposure of 10 seconds was required to fully cross link the iron oxide segment and Figure 5B shows the iron oxide segment that was generated; the polymerized iron oxide segment is of full thickness (200 µm) with straight edges, and shape fidelity is closely maintained as compared to the photomask (Figure 5A). Conversely, over exposure (>15 seconds) to UV light generated iron oxide segments that were over polymerized. Figure 5D shows an over polymerized segment that has poor shape fidelity and is larger than the shape defined by the photomask.
Figure 6A shows a complete device after sealing with proper alignment by utilizing photomasks with alignment marks. The gear within the device is entirely within the central void of the device and is thus responsive to magnetic actuation. Figure 6B shows a device with a misaligned sealing layer. Figure 6C shows the bottom layers of hydrogel and the gear itself elucidated with black outlines and Figure 6D shows the misaligned sealing of the top hydrogel layer elucidated in white outlines. As can be seen from Figure 6D, portions of the gear that fall within regions where polymerization would take place during sealing (shown in red fill) results in portions of the gear being anchored to the bulk of the hydrogel material. This prevents the gear from moving during actuation.
Figure 7 shows a functional single gear device that was fabricated (total fabrication time ~15 minutes). The total thickness of the device is 2 mm and the longest dimension of the device is 13 mm. The top and bottom layers of the device is 400 µm thick and the gear has a height of 1 mm. This design allows for a 100 µm clearance on the top and bottom surface of the gear to allow for movement. The top most layer of the device has a 600 µm aperture and the axle for the gear is 400 µm in diameter. Figure 5B shows images of the device when it is actuated with a magnet such that the gear performs a full rotation as can be observed from the change in position of the iron oxide segment from (i) through (vi).
Figure 1. Fabrication set-up for hydrogel-based micromachines. A) Schematic of fabrication stage. This schematic shows the various components of the fabrication set-up including the PDMS chamber in which the hydrogels are formed within the fabrication area, a vacuum-enabled stage which holds down the PDMS chamber as well as attaches the flexible membrane to a micrometer head for height control, and top substrate consisting of a glass coverslip that is either untreated or coated with PDMS. B) Schematic of the top view of the fabrication stage (without PDMS chamber). The UV light source is then positioned such that the incident angle of the light is perpendicular to the horizontal plane of the fabrication stage (not shown in figure). Please click here to view a larger version of this figure.
Figure 2. Schematic of single-gear hydrogel-based device and photomasks used for each layer. A) Schematic of the top- and oblique-view of a typical hydrogel-based device that can be fabricated using this strategy. This device consists of a single gear that contains an iron-doped segment which allows for magnetic control. B) Schematic of the individual layers and components within the device. This single-gear device consists of a top sealing layer (i), support structures such as the post for the iron-doped gear and the walls of the device (ii) as well as a bottom layer (iii). C) Photomask designs used to fabricate the single-gear device. The photomasks are designed dark field; desired features are left transparent while the background is dark. This panel shows the photomask designs corresponding to the top sealing layer (i), support structures (ii) and bottom layer (iii). Please click here to view a larger version of this figure.
Figure 3. Layer-by-layer photolithography of hydrogel-based micromachines. A) Schematic of step-by-step process for device fabrication within the fabrication area of the PDMS chamber. 1: A small volume of PEGDA prepolymer is pipetted onto the glass coverslip bonded to the flexible membrane of the PDMS chamber (bottom substrate). A piece of untreated glass coverslip is used as the top substrate and a photomask is placed on top of this top substrate. The height of the bottom substrate is brought to the desired height (Z1) using the micrometer head. The hydrogel prepolymer is then exposed to UV light through the photomask. The top substrate can then be lifted off the PDMS chamber and the hydrogel remains adhered to the top substrate (inset). This layer is then reserved for later use. 2: Step 1 is repeated but the top substrate is now replaced with PDMS-coated glass. The polymerized hydrogel will remain adhered to the bottom substrate. 3: The height of the bottom substrate is lowered (Z2> Z1) and more prepolymer can be added to the fabrication area. A second photomask is used and the prepolymer is exposed to UV light once again. 4: Step 3 can be repeated (Z3 >Z2) until the desired support structures are created. (i) Once the support structures are completed, the top substrate can be removed to allow for access to fabrication area for introduction of any preformed hydrogel components (e.g., iron-doped gear). (ii) Once the preformed components have been placed and properly aligned, the hydrogel layer from Step 1 can be placed on top of the fabricated structure and aligned. 5: All the layers are then exposed to UV light through a photomask that seals the edges of the device. (i) The sealing step seals the entire device while the interior components are shielded from further UV exposure. (ii) The sealed device can then be lifted off the fabrication chamber as it would preferentially adhere to the top substrate. Please click here to view a larger version of this figure.
Figure 4. Steps for iron-oxide nanoparticle doping of hydrogel components. (i) UV light is exposed through a photomask defining the iron oxide-doped segment within the hydrogel gear. (ii) Thin (200 µm) layers of iron oxide-doped hydrogel is polymerized each time and stacked on top of each other. (iii) The layer of thin layers creates a segment with total height of 1 mm. This segment is left in the fabrication layer. (iv) Un-doped prepolymer is then deposited into the fabrication area and a photomask that defines the complete shape of the gear is then used during cross-linking. This enables the formation of the complete gear with an iron oxide-doped segment. Please click here to view a larger version of this figure.
Figure 5. Photopolymerization of iron oxide-doped hydrogel components. A) Photomask of gear segment to be doped with iron oxide nanoparticles. B) Iron oxide-doped hydrogel that has been optimally polymerized (10 s exposure). C) Iron oxide-doped hydrogel that has been under-polymerized (4 s exposure). D) Iron oxide-doped hydrogel that has been over-polymerized (20 s exposure). Please click here to view a larger version of this figure.
Figure 6. Alignment of hydrogel layers during sealing of device. A) Image showing the correct alignment of hydrogel layers with free-moving gear that is entirely within the void of the device. B) Image showing device with misaligned hydrogel layers (B, C, and D are images of the same device but with different layers highlighted). C) Same image as in (B) but with black outlines elucidating bottom layers which are correctly aligned. The gear is correctly placed within the bottom layers. D) Same image as in (B) but with white outlines showing the misaligned top layer of hydrogel. The gear has been partially polymerized during the sealing step and portions of the gear (red fill) has been anchored to the bulk material of the device. This renders the device non-functional. Please click here to view a larger version of this figure.
Figure 7. Actuation of a single-gear hydrogel-based micromachine. A) Image showing the fabricated device. B) Images showing the different orientations of the gear upon actuation. (i) From its initial orientation (0°), the gear is rotated by (ii) 60°, (iii) 120°, (iv) 180°, (v) 240°, and 300°. Scale bar is 1 mm Please click here to view a larger version of this figure.
Figure 8. Versatile fabrication of various designs for hydrogel-based micromachines. A) A simple gate valve which controls the release of drugs from a single reservoir. The linear movement of the iron oxide-doped hydrogel component gates the diffusion of a hypothetical drug out through and outlet. B) A gated linear manifold which controls the release of drugs from multiple reservoirs. Each reservoir contains hypothetical drugs and the movement of the iron oxide-doped component gates the movement of drugs out of these reservoirs through a window of hydrogel that allows for the diffusion of these drugs out to the exterior. C) A simple rotor that can be actuated to spin about an axis. D) A sophisticated design based on the Geneva drive. A driving gear with a pin is able to engage a larger driven gear and produce intermittent movement; a full rotation of the driving gear rotates the driven gear by 60°. All scale bars are 1 mm. From Chin, S. Y. et al. Additive manufacturing of hydrogel-based materials for next-generation implantable medical devices. Science Robotics. 2 (2), (2017). Reprinted with permission from AAAS17. Please click here to view a larger version of this figure.
This technique is a facile and rapid method for layer-by-layer photolithography of hydrogel microstructures. Using an additive manufacturing approach, we can easily build a variety of 3D structures out of biocompatible materials and even incorporate moving parts. This would thus enable the formation of entirely biocompatible microdevices. The technique is based on simple repetition of lithography steps, which is enabled by the precise control of the height of the bottom substrate via a micrometer head. Traditional fabrication techniques used in the MEMS industry, involving harsh processing techniques and sacrificial materials, is often not compatible with the processing of soft hydrogels. Other methods for 3D printing hydrogels, such as extrusion-based methods, are limited to spatial resolutions above 200 µm and print speeds of mm/s for simple structures that do not include moving parts18,19. Stereolithography (SLA) and digital light project (DLP) based bioprinters maybe be able to achieve better resolutions but are also a lot costlier to setup. These fabrication strategies are also not able to easily print overhangs without supporting substrate materials, which could be difficult to introduce and remove from the completed device. We circumvent this by aligning and polymerizing a pre-formed sealing layer to the fabricated support structure as a final step to form the completed device. The design of the fabrication set-up gives the user easy access to the fabricated structures and allows for the easy alignment of various components with the use of alignment marks.
The strategy presented here is also a lot faster than other techniques of similar resolutions; the total time taken for the demonstrated fabrication of the device with a rotating is about 15 minutes. Another added advantage of this fabrication strategy, though not demonstrated in this protocol but shown in our previous work17, is the ability for the user to quickly and easily change the type of polymer used between steps which can be done in small volumes. In this way, one can create devices that are a composite of different types of hydrogels. The device fabricated using this strategy also has the added advantage of contactless actuation as the gear contains a segment that is doped with iron oxide nanoparticles, rendering the gear sensitive to magnetic actuation and can thus be actuated using an external magnet. Additionally, the device is entirely biocompatible and hence may be safely implanted in vivo.
An important feature of this technique is the treatment of the different glass substrates, which enables the user to preferentially adhere or repel the polymerized hydrogel to either the bottom or top glass substrate. When a combination of untreated glass is used with a PFOTS-treated glass surface (bottom substrate), formed hydrogels will preferentially adhere to untreated glass, as they are repelled from the fluorinated surface of PFOTS-treated glass. Conversely, when PDMS-coated glass is used with the PFOTS-treated bottom substrate, hydrogels will tend to remain on the PFOTS-treated surface as PDMS surfaces more strongly repel the formed hydrogels. This feature allows one to build upwards, adhere hydrogels such that they are immobilized on glass substrates and can be reserved for alignment to other structures at a later point in time, or even build downwards. This adds to the flexibility of the technique and types of designs that can be fabricated as well as enables the incorporation and sealing in of independent, free-moving hydrogel components.
During the layer-by-layer fabrication, it is important to optimize the polymerization time used. Hydrogels should be optimally cross-linked such that they form at full thickness as well as at high fidelity as compared to the shapes defined by the photomask. This is dependent on the power of the lamp and the type of hydrogel used. Although not shown in this protocol, the polymerization time decreases with increasing lamp power and increases with increasing PEG chain length and decreasing concentrations of PEGDA used. Other factors that affect the amount of energy available for photopolymerization, such as the change in opacity of the prepolymer due to the addition of iron oxide nanoparticles (Figure 4), will also affect the polymerization time. Optimization for cross-linking conditions for different hydrogel compositions is thus required before the start of the fabrication process of devices.
The use of alignment marks on the photomasks and the proper alignment of the hydrogel layers, especially the final sealing layer, are important to ensure that proper sealing is performed, and the interior components are not inadvertently cross-linked to the surrounding support structures during the fabrication process. This would prevent these components from freely moving during magnetic actuation. As shown in Figure 5, a misaligned top sealing layer and photomask results in the crosslinking and anchoring of a portion of the gear to the bulk material of the device itself. As a result, this gear does not rotate when actuated with a magnet.
The devices can be actuated using strong permanent magnets such as neodymium magnets. These magnets generate strong magnetic forces when in close range to ferromagnetic materials and care should be taken to prevent injury. The device can be actuated to move without the magnet coming in contact with the device; the magnet can be held or placed ~1cm away from the device. The movement of the iron-doped components should mirror the movement of the magnet and can be actuated to move continuously or oriented intermittently as desired. The device can be manually actuated or an actuation set-up can be used. The magnet can be attached to any actuator (e.g., servo motor) to rotational movement. The speed of rotation of the magnet, and hence the speed of rotation of the iron-doped component, can be controlled using a microcontroller. This provides for a more precise method of actuation.
Figure 8 shows schematics and images of various designs from previous work that were fabricated using this same technique and demonstrate the versatility of this method. These designs range from simple devices that resemble valves (Figure 8A) to more complicated and sophisticated designs that draw inspiration from the Geneva drive design (Figure 8D) that comprise of 2 engaged gears that produce intermittent movement. The smallest features that can be generated using this technique were typically about 100 µm and each design is composed of multiple layers (3 to 6 layers). Different types of hydrogel compositions (with differing mechanical strengths and porosity) can also be polymerized and bonded to each other. Hence, one can easily combine the types of hydrogels to be used within a device depending on the required function of the different components within the device.
The authors have nothing to disclose.
This work was supported by an NSF CAREER award, NIH R01 grant (HL095477-05), and NSF ECCS-1509748 grant. S.Y.C. was supported by the National Science Scholarship (PhD), which was awarded by the Agency for Science, Technology and Research (Singapore). We thank Keith Yeager for help with building the fabrication set-up, and Cyrus W. Beh for photographs of the set-up and devices.
Poly(ethylene glycol) (n) diacrylate [MW 400Da] | Polysciences, Inc | 01871-250 | PEGDA reagent for prepolymer |
Darocur 1173 | Ciba Specialty Chemicals, Inc | – | Photoinitiator |
Iron oxide (II, III) | Sigma Aldrich | 637106-25G | Iron oxide nanoparticles |
Trichloro(1H,1H,2H,2H-perfluorooctyl)silane | Sigma Aldrich | 448931 | Fluorinated compound that is used to vapor silanize the PDMS chamber to prevent adhesion of hydrogel to the glass coverslip that is bonded to the flexible PDMS membrane with prolonged use of the PDMS chamber |
Petri dish, glass | Sigma Aldrich | BR455743 | Glass petri dishes for casting PDMS layers for forming PDMS chamber |
Sylgard 184 Silicone Elastomer Kit (PDMS) | Dow Corning | 240-4019862 | PDMS for fabrication chamber |
Glass coverslips (No. 2), 50 x 45 mm | Fisher Scientific | FIS#12-543F | Glass substrates that cover the fabrication chamber |
Fisherbrand Straight Flat Tip Forceps 4.75in | Fisher Scientific | FIS#16-100-112 | Tweezers for handling polymerized hydrogel layers/devices |
Omnicure S2000 | Cadence Technologies Pte Ltd | 010-00148R | UV lamp |
5 mm Adjustable Collimating Adaptor | Cadence Technologies Pte Ltd | 810-00042 | Collimator for UV lightsource |
Photomasks | CAD/Art Services Inc | – | Photomasks used to define hydrogel microstructures |
Adobe Illustrator | Adobe | – | Designing of photomasks |