We present a convenient solid-phase extraction coupled to high-pressure liquid chromatography (HPLC) with electrochemical detection (ECD) for simultaneous determination of three monoamine neurotransmitters and two of their metabolites in infants’ urine. We also identify the metabolite MHPG as a potential biomarker for the early diagnosis of brain damage for infants.
The extraction and analysis of catecholamine neurotransmitters in biological fluids is of great importance in assessing nervous system function and related diseases, but their precise measurement is still a challenge. Many protocols have been described for neurotransmitter measurement by a variety of instruments, including high-pressure liquid chromatography (HPLC). However, there are shortcomings, such as complicated operation or hard-to-detect multiple targets, which cannot be avoided, and presently, the dominant analysis technique is still HPLC coupled with sensitive electrochemical or fluorimetric detection, due to its high sensitivity and good selectivity. Here, a detailed protocol is described for the pretreatment and detection of catecholamines with high pressure liquid chromatography with electrochemical detection (HPLC-ECD) in real urine samples of infants, using electrospun composite nanofibers composed of polymeric crown ether with polystyrene as adsorbent, also known as the packed-fiber solid phase extraction (PFSPE) method. We show how urine samples can be easily precleaned by a nanofiber-packed solid phase column, and how the analytes in the sample can be rapidly enriched, desorbed, and detected on an ECD system. PFSPE greatly simplifies the pretreatment procedures for biological samples, allowing for decreased time, expense, and reduction of the loss of targets.
Overall, this work illustrates a simple and convenient protocol for solid-phase extraction coupled to an HPLC-ECD system for simultaneous determination of three monoamine neurotransmitters (norepinephrine (NE), epinephrine (E), dopamine (DA)) and two of their metabolites (3-methoxy-4-hydroxyphenylglycol (MHPG) and 3,4-dihydroxy-phenylacetic acid (DOPAC)) in infants’ urine. The established protocol was applied to assess the differences of urinary catecholamines and their metabolites between high-risk infants with perinatal brain damage and healthy controls. Comparative analysis revealed a significant difference in urinary MHPG between the two groups, indicating that the catecholamine metabolites may be an important candidate marker for early diagnosis of cases at risk for brain damage in infants.
Catecholamine neurotransmitters and their metabolite contents in body fluids can affect neural function and affect the balance of response-to-stimulus states to a large extent1. Abnormities may cause a variety of diseases, such as pheochromacytoma, ganglioneuroma, neuroblastoma, and neurological disorders1,2. The extraction and determination of catecholamines in body fluids is meaningful to the diagnosis of the relevant diseases. However, catecholamines in biological samples exist in low concentrations and are easily oxidized. Furthermore, they are very difficult to elute because of the large amount of interference in the medium3. Thus, simultaneous detection of catecholamines in biological fluids is still a challenge.
There have been reviews showing that urinary catecholamines can be a measure of stress, and that their levels are important biological markers responding to tactile stimulation processing in newborns5. According to the research, all infants who have suffered from premature incidents are at risk for brain injury4,5,6, and injury may cause abnormal release of catecholamines and related matters into the fluids. There do exist advanced magnetic resonance techniques that can detect brain damage in earlier phases7,8. However, within the first 48 h, an abnormal neurodevelopmental process will cause permanent brain injury that won't be evident in medical images11. Besides, the high cost and scarce instrument resources, along with other factors, makes it impossible for all neonatal units to have access to these specialized neuro-imaging techniques. However, the use of an easily approachable and practical biomarker (such as catecholamines and their metabolites) could overcome these shortcomings, and the screening of a biomarker in human fluids may help in the early diagnosis of brain injury and lead to prompt identification of new-born infants needing neuroprotection9. The catecholamines in urine can be an easy and obvious index, because of the direct correlation between the amount of them released into fluids and neuroactivity function.
Among biological fluids, cerebrospinal fluid (CSF) and plasma samples are not easy to get via existing traumatic procedures, and it is also very difficult to get rid of interference due to adhesive protein and other impurities, leading to a troublesome and time-consuming sampling process that is unsuited for repeated detection. Also, for children, it is almost impossible to get the samples in a traumatic fashion. Therefore, urinary sampling is better than the other forms of sampling, as it is non-invasive, easy to operate, and can be repeatedly done. Urine samples are abundant and easy to store, and show great advantages over the other forms of biological samples.
The main methods to quantify catecholamines in biological fluids include radioenzymic assays10, enzyme-linked immune-sorbent assays11, voltammetry12 and thermal lens spectrometry13. But shortcomings exist, such as complicated operations and hard-to-detect multiple targets. Today, the dominant analysis technique is high-performance liquid chromatography (HPLC)14, coupled with sensitive electrochemical15 or fluorimetric detection16, because of its high sensitivity and good selectivity. With tandem mass spectrometry technology, such as liquid chromatography /mass spectrometry (LC/MS) and liquid chromatography/mass spectrometry/mass spectrometry (LC/MS/MS), the analysis and quantification of the neurotransmitters can achieve high accuracy and specificity17,18. However, the MS technique requires expensive instrumentation as well as substantially qualified manpower, making the method difficult to apply universally in most conventional laboratories. HPLC-ECD systems are commonly equipped in most conventional and clinical laboratories, and have thus become a common and good choice for research groups to use for chemical determination, but they require the sample introduced into the system to be clean and of microscale volume19. Thus, it is of great importance to purify and condense the sample prior to the analysis. The classical method for the purification step is liquid-liquid extraction14,15,20 and off-line solid-phase extraction, including activated alumina column21,22 and diphenylborate (DPBA) complexation23,24,25,26.
Myeongho Lee et al. have been using polymer resin chemically modified with crown ether as the adsorbent to selectively extract catecholamines from human urine since 200727. Also, in 2006, Haibo He et al. demonstrated a facile synthesis approach for boronate affinity extraction sorbent byutilizing a functionalizable nanomagnetic polyhedral oligomeric silsesquioxane (POSS) based nanomagnetic composite, and applying it to the enrichment of catecholamines in human urine (noradrenaline, epinephrine and isoprenaline)28. They also took advantage of the nanomaterials to fulfill the work, using a technology called nano-electrospinning and forming the polymer fibrous material in the nanoscale. The electrospinning process can adjust the diameter, morphology, and spatial alignment of the product by controlling the working voltage and changing the content of the spinning solution along with other parameters29. Compared with the conventional SPE cartridge, electrospun nanofibers are highly suitable to extract and enrich target analytes from a complex matrix, as they are equipped with high surface-area-to-volume ratios to adsorb the analytes with high efficiency, and exhibit more easily-controlled surface chemical properties, allowing handy attachment of the target compounds. These properties make them good choices for SPE adsorbents, greatly reducing the solid phase and desorption solvent amount30,31,32,33. For catecholamines in urine samples, electrospun nanofibers composed of apolymeric crown ether with polystyrene (PCE-PS) were used to selectively extract three catecholamines (NE, E, and DA)34. The paper indicated that the selective crown ether adsorbed the targets of NE, E, and DA, which was based on its correct geometry for binding catecholamines via forming hydrogen bonds. The results displayed the material crown ether effectively, removing other interfering compounds contained in biological samples. Inspired by this report, a novel method was developed for the selective extraction of the catecholamines by use of electrospun composite nanofibers composed of PCE-PS.
In this paper, the method reported previously34 was improved and employed not only to successfully analyze E, NE, and DA, but also their metabolites, MHPG and DOPAC, in urine. We also explore new possibilities for the mechanism of the adsorption process. The method shows satisfying extraction efficiency and selectivity for the five analytes, and the method was verified in the analysis of urine from high-risk infants with perinatal brain damage and healthy controls.
Informed consent from the parents was obtained, and institutional review board approval was obtained for the study. The study was performed in accordance with the code of Ethics of the World Medical Association (Declaration of Helsinki) for experiments involving humans. Caregivers of all participants provided written consent for being enrolled in the study. Ethical committee approval from Zhongda Hospital, affiliate with Southeast University, was also obtained.
1. Preparation of the Columns and Solutions Needed for the Extraction and Determination of Catecholamines
2. Preparation of Real Urine Samples and Mobile Phase
3. PFSPE Extraction and HPLC Analysis
4. Phenylboronic Acid Cartridges (PBA) Extraction
PBA cartridge extraction procedures were similar to the scheme in Kumar et al. (2011)25. All solutions are pushed through the PBA cartridge (100 mg, 1 mL) with air forced by a syringe.
5. Identification and Quantification of Catecholamines
This protocol is a simple and convenient PFSPE method to pretreat urine samples and enrich five catecholamines for detection via an HPLC-ECD system; a diagram of the process is shown in Figure 1. The protocol mainly includes four steps-activating, loading, rinsing, and eluting- coupled with a small quantity of PCE-PS nanofibers and a simple solid-phase extraction device. The morphology of PCE-PS nanofibers was assessed using a surface and porosity analyzer (see Table of Materials). The textural properties-the BET (Brunauer, Emmett, and Teller) surface area, pore volume and pore size-were 2.8297 m2 g-1, 0.009 cm3 g-1, and 12.76 nm, respectively. These data indicate that the material used in the protocol has nanoscale pores on the surface, which may contribute to the high adsorption efficiency and the lowered binding pH in the protocol.
This protocol uses optimized volumes, sample ingredients, leachate, eluant, etc., as well as working pH and time consumed during the procedures. In Chen et al., the eluant was a 12.0 M acetic acid solution because acidic conditions cause the adsorbent to become protonated and positively charged, which is favorable for the elution of CAs34,35. In this study, the eluant recipe was reconditioned to be 30% phosphoric acid, 15% acetonitrile, and 55% distilled water. The final pH value of eluant was adjusted to 3.0. The eluant solvent should be kept in an acidic environment when eluting, but much more moderate than 12.0 M acetic acid, which may lead to poor peak appearance and damage to the HPLC system.
For identification of the catecholamines, chromatogram retention times of the peaks in the samples with standard solution peaks are compared. Figure 2 shows examples of HPLC-ECD chromatograms from the various solutions. If the protocol is followed successfully, the chromatographic profiles of three catecholamines and their metabolites should be obtained by HPLC-ECD with clear symmetrical, well defined peaks and with minimal background noise, as illustrated in Figure 2. For comparison with the conventional method, a commercial PBA cartridge was selected as a control. In Figure 2(a–c), five target peaks in (b) are significantly higher than thaosein (c), indicating that the PFSPE method is more sensitive than the PBA cartridge method. Also, the DOPAC peak did not show in the PBA cartridge extraction result, indicating that the PBA column was not capable of extracting DOPAC. Figure 2(d) depicts the chromatogram of the blank urine sample without any pretreatment, and Figure 2(e) shows the chromatograms of the urine sample extracted using PCE-PS composite nanofibers. Figure 2(f) shows the chromatograms of the urine sample after the extraction with the PBA column, which also shows no DOPAC extraction consistent with the result in Figure 2(c). The diagram indicates that the PFSPE method could not only extract the targets with good effect, but also could get rid of most of the interference in the urine, giving good peak identification for the target compounds.
Statistical analysis revealed that measurements for the three catecholamines and the two metabolites were reliably reproduced (Table 1). All the target compounds showed good linearity between 1.5 and 100 ng/mL (R2>0.99), and the standard curve of each analyte can be found in Supplementary Figure 1. The curves and R2 values demonstrate that the analytes have good linearity and relativity within a certain linear range, suitable for the calculation of concentrations of analytes in urine samples. The limits of detection (LOD) ranged from 0.25 to 0.54 ng/mL, and the limits of quantification (LOQ) were 0.83 to 1.81 ng/mL, respectively. The signal-to-noise (S/N) value equalled 3. The range of the methodological recoveries of the five target compounds was from 97.4% (MHPG) to 124.2% (DOPAC), which was satisfactory for the application to the actual samples. The intra-day precision was from 2.7 to 4.8% (expressed as relative standard deviation) and the inter-day precision was 2-8.1%, displaying good precision and repeatability.
For detection of targets in real urine samples, 28 high-risk infants and 22 healthy infants in the division of child care, Suzhou municipal hospital, were recruited. All 50 infants taken into the study were boys. When they were six months old, they were taken to a routine health check in September 2016. All the urine samples were collected and pretreated following the protocol steps 3.1 and 3.2, and the samples were analyzed using the remainder of step 3. The concentrations were calculated against the calibration curves. Statistical differences were analyzed by analysis of variance (ANOVA). The results can be seen in Table 2. The difference of the catecholamines and metabolites between the two groups were compared and analyzed. The p-values show that the catecholamines were not significantly different between the high risk and healthy groups, while the metabolite MHPG content was different across these groups (p = 0.001). The high-risk infants group had higher amounts of MHPG than the control group (14.8 ± 3.6 ng/mL vs. 1.4 ± 0.2 ng/mL), which means that the level of urinary MHPG may be a potential marker for early identification of high-risk infants.
Figure 3 and Table 3 describe the classic quantification method for determining monoamine materials, and compare with other methods for which the operation process and figures of interest are given. Compared with classic methods, PFSPE method has advantages like a short timespan (5-10 min), simplified operation process, less organic solvent, and more environmental friendliness with satisfactory methodological parameters. The eluant amount needed is low in volume (50 µL) and the target enrichment step requires no evaporation, which greatly promotes the detection sensitivity for the target compounds in the urine. Compared to conventional particle-based SPE, this method enhanced the efficiency, simplified the preparation process, and reduced the time of the analysis with acceptable reliability, selectivity, and sensitivity.
Figure 1: Schematic flow chart of PFSPE procedure in the paper and a representation of its device. (1) Gastight syringe, (2) Pipette tip, and (3) Packed nanofibers. Please click here to view a larger version of this figure.
Figure 2: Chromatograms of different samples. (a) Spiked water sample with targets and IS in (100 ng/mL) without extraction. (b) Spiked water sample extracted by the PFSPE method, and (c) by commercial phenylboronic acid (PBA) cartridge. (d) Real blank urine sample without extraction. (e) Real urine sample extracted by the PFSPE method. (f) Real urine sample extracted by commercial PBA cartridge. Please click here to view a larger version of this figure.
Figure 3: Analytical flow chart of determination methods. (a,b) Previously reported classic extraction methods. (a) Extraction by alumina, (b) by DPBA complex, and (c) by the method proposed in this paper. Please click here to view a larger version of this figure.
Supplementary Figure 1: Standard curves and equations. The standard curves for the five analytes are constructed in order to use the equation of the line to calculate the unknown analytes concentration in the sample. (a) NE, (b) E, (c) DA, (d) MHPG, and (e) DOPAC. Please click here to download this figure.
Supplementary Figure 2: Boronate affinity interaction between boronic acid and cis-diol group or adjacent two hydroxyl groups. (A) Schematic of the interaction between boronic acids and multiple -OH groups to form five- or six-membered cyclic esters. (B) Cis-diol group or adjacent two hydroxyl groups in red circles depict the major reaction sites for the five analytes to boronic acid compound. Please click here to download this figure.
NE | MHPG | E | IS | DOPAC | DA | |
Linear range (ng/mL) | 1.5-400 | 1.5-200 | 1.5-100 | 1.5-100 | 1.5-400 | |
R-squared values | 0.9945 | 0.995 | 0.9976 | 0.9902 | 0.9954 | |
LOD (ng/mL) | 0.323 | 0.322 | 0.313 | 0.657 | 0.249 | 0.543 |
LOQ (ng/mL) | 1.076 | 1.072 | 1.043 | 2.191 | 0.83 | 1.809 |
Recovery ± RSD (%)(n=9) | 110.7±2.9 | 97.4±9.1 | 103.9±5.2 | 86.5±7.3 | 124.2±3.1 | 117.3±5.4 |
Precision(RSD %) (n=9) | ||||||
Intra-day | 4.5 | 4 | 2.7 | 4.8 | 3.3 | 4.7 |
Inter-day | 4.1 | 8.1 | 2 | 6.3 | 3.2 | 5.9 |
NE, norepinephrine; MHPG, 3-Methoxy-4-hydroxyphenylglycol; E, epinephrine; IS, internal standard; DOPAC, 3, 4-Dihydroxyphenylacetic acid; DA, dopamine; |
Table 1: Analytical results of the proposed protocol for the determination of three catecholamines and two metabolites with standard solutions.
Concentration (ng mL-1) | Control group (22) | High Risk Infant group (28) | P value |
NE | 7.52±1.34 | 5.56±1.7 | 0.37 |
MHPG | 1.4±0.2 | 14.8±3.6 | 0.001 |
E | 24±15.8 | 20.9±5.87 | 0.841 |
DOPAC | 106.36±30.1 | 72.12±18.07 | 0.312 |
DA | 55.53±11.9 | 48.12±20.9 | 0.76 |
P<0.05***shows a significant difference between two groups |
Table 2: The comparison of concentrations for three catecholamines and two metabolites in urines between healthy infants and high-risk infants. The results were statistically analyzed using analysis of variance (ANOVA) with a significance level of p = 0.05 for differences between MHPG content. Means ± standard deviations are shown.
Sample | Exhausted solvent (mL) | Sample volume (mL) | Pretreatment time (min) | Linear range | LOD | Relative Recovery (%) | Evaporate and redissolve |
Analytical method |
solvent | volume | time (min) | recovery (%) | solvent and | ||||
(mL) | (mL) | reconstituting | ||||||
residue | ||||||||
Urine | 0.5 | 0.05 | 10 | 2.0–200 ng/mL (NE, E, DA) | 0.2-0.5 ng/mL (NE, E, DA) | 88.5-94.5 (NE, E, DA) | No | HPLC-ECD (Chen et al., 2016) |
Urine | 19 | 0.7 | — | 47–167 µg/L (DA) | 166-500 nM (DA) | 98.3-101.1 (DA) | No | HPLC-UV (Piotr et al., 2016) |
Urine | 1.3 | 0.01 | — | 0.5–1250 ng/mL (NE, E, DA, NMN, MN) | 0.5-2.5 ng/mL (NE, E, DA, NMN, MN) | 74.1–97.3 (NE, E, DA, NMN, MN) | No | LC-MS/MS (Li et al., 2016) |
Urine and plasma | 20 | 0.05 | 10 | 0.04-2.5 ng/mL (NE, E, DA) | 0.01-0.02 ng/mL (NE, E, DA) | 87.0-97.5 (NE, E, DA) | No | HPLC-UV (Mohammad et al., 2016) |
Urine | < 0.5 | 0.1 | 5 | 1.5-400 ng/mL (NE, MHPG, E, DOPAC, DA) | 0.249-0.543 ng/mL (NE, MHPG, E, DOPAC, DA) | 97.4-124.2 (NE, MHPG, E, DOPAC, DA) | No | this work |
—, Undefined. PBA, Phenylboronic acid |
Table 3: Comparison of this work with studies on pretreatment of monoamines or other related topics in recent years.
Supplementary Table 1: Instrument parameters for detection and quantification of analytes in the paper by HPLC-ECD. Please click here to download this table.
Supplementary Table 2: Preparation of standard curve for five analytes. Please click here to download this table.
The proposed PFSPE method in this paper may be significant and meaningful with respect to its rapidity, simplicity, and convenience. The adsorbents used in the protocol are electrospun nanofibers, which have high surface area-to-volume ratios, and adsorb the analytes with high efficiency. The procedure only needs a few milligrams of nanofiber and a small volume of eluant solvent, and does not require an evaporation step to concentrate the analytes. Here, we have presented a detailed overview of an HPLC-ECD based protocol that will permit new users to establish an effective method for detection and quantification of three catecholamines and two of their metabolites (NE, E, DA, and MHPG, DOPAC).
The superiority of this protocol mainly stems from the four critical steps during the procedures, as shown in Figure 1: using PCE-PS nanofibers to capture the target compounds to reduce the size of sorbent to a few milligrams, leading to a rapid adsorption/desorption, and only needing a small volume of eluting fluid (1); adding DPBA into the urine to complex with the analytes to improve their hydrophobicity (2); rinsing the nanofibers with the solution containing DPBA to retain the analytes and remove the impurities (3); and optimization of the extraction and analysis condition to get good sensitivity and selectivity for the analytes (4).
For step (1), the mechanism of superior performance of adsorbent ability for PCE-PS may attributed to a variety of factors. The crown ether polymers doped in the nanofibers can form a host-guest complex with the guest molecules containing amine groups by H-bonds, such as the catecholamines in this paper. In addition, Hongyou Hu et al. and Jishun Chen et al. also proposed that boronic acid can bond with nitrogen atoms in the chemical structure through B-N interactions, and that the hydrophobic skeleton can interact with benzene rings and other aliphatic groups of analytes36,37. Also, during the desorption process, the hydrogen bonds and B-N interactions are easily broken at very low pH; thus, eluants with acidic pH are usually suitable for exploration for desorption. Furthermore, there are also several secondary interactions, including hydrophobic, ionic, and hydrogen bonding, which can occur between boronic chemicals and related compounds36,37,38. All these interactions may contribute to the adsorption of the analytes to the materials.
For step (2), with the help of added DPBA, PCE-PS nanofibers adsorb DPBA-catecholamine complexes rapidly. Zhen Liu et al. showed that the boronate affinity materials have emerged as important media for the selective separation and molecular recognition of organic compounds, such as nucleosides, saccharides, glycans, glycoproteins, and so on38.The complexation mainly occurs from the boronate affinity interaction between boronic acid and cis-diol groups, or two adjacent hydroxyl groups, to form five or six membered cyclic esters, as shown in Supplementary Figure 2A. Supplementary Figure 2B depicts the major reaction sites for the five analytes in this work to the boronic acid compound. When the surroundings become acidic, the boronic acid-cis-diol complex dissociates. The interaction between boronic acid and the cis-diol group or the adjacent two hydroxyl groups exists extensively and relatively strongly.
For step (3), the complex formed between analytes and DPBA can hold them on the adsorbent while other impurities are rinsed from the PCE-PS surface, achieving the best reservation for the targets. For step (4), there has been a report showing that the optimal pH was about 9.0 for the bonding of boric acid and catechol compound39. However, Chen et al. explored the best pH value for the complexing of catecholamine-DPBA and found that the adsorption performance of the composite PS-PCE nanofibers for the CAs was optimal when the pH value was changed to be between 6.0 and 7.034,35. Liu et al. implicated that the structure of the supporting materials for boronate acid compounds, especially with spatial confinement of cavities and nanoscale pores, and the B-N ligands formed during the reaction, can also significantly lower the binding pH and enhance binding affinity38. The PCE-PS structure as well as catecholamine structure are all equipped with -NH- groups. Thus, the properties and nanoscale pores of PCE-PS (data as shown in Representative Results), as well as B-N ligands formed, may contribute to the lowered binding pH in the protocol. Thus, the optimal pH value in this protocol can be neutral, which greatly simplifies the procedures and avoids the degradation of the analytes.
The main limitations of catcholamine determination are that their concentrations in biological samples are very low and that their structures are unstable (easily oxidized); furthermore, it is difficult to get rid of the impurities in the media. Thus, for analysis of catecholamine in biological samples, pretreatment, usually by solid-phase extraction, is necessary. The PFSPE method proposed in this paper provided a simple and convenient pre-concentration for the analytes in the urine samples. But, when doing the experiment, the experiment condition must be strictly controlled, by means such as avoiding light exposure and shortening the pre-treatment time as much as possible.
The applicability of the method was evaluated by its use to determine the concentrations of three catecholamines and two metabolites in the urine of healthy infants and high-risk infants. There was a significant difference between the two groups in urinary MHPG. As summarized previously, the MHPG amount in the human body is now mainly reported as an index to reflect noradrenergic neuronal tone, the catecholamine metabolism activity in the central and peripheral nervous systems40,41,42, and is a useful plasma/urine marker for central NE metabolism43. For high-risk infants, a variety of factors such as hypoxic-ischemic encephalopathy (HIE) causes brain trauma and brain insults, leading to different degrees of childhood neurodevelopmental deficits. This brain damage in turn will lead to the release of neurotransmitters (MHPG included) into body fluids44,45. According to reports by J.W. Maas, there are significant correlations between the brain, CSF, plasma, and urinary concentrations of MHPG46,47. The measurement of total (free + conjugated) MHPG in urine has long been used to assess the metabolism of central NE and peripheral NE metabolism in humans41,48,49. Thus, it may be concluded that high risk infants have noradrenergic neuronal tone damage compared with healthy ones, affecting catecholamine metabolism in the central and peripheral nervous systems. This discrepancy indicates that further research should be done on the relationship between the urinary MHPG level and neurodevelopmental deficits. It has been demonstrated that the proposed method could be used for the determination of catecholamine presence in urine, with a promising prospect for use in the clinic for evaluating diseases relevant to these neurotransmitters.
The authors have nothing to disclose.
This study was supported by the National Science Foundation of China (No.81172720, No. 81673230), the Social Development Research Program of Jiangsu Province Science and Technology department (No. BE2016741), Science & Technology Project of China General Administration of Quality Supervision, Inspection and Quarantine (2015QK055), The Open Project Program of Key Laboratory of Child Development and Learning Science of Ministry of Education, Southeast University (CDLS-2016-04). We sincerely acknowledge Yuan Song and Ping Liu who assisted us in samples collection.
200 µL pipette tip | column to contain nanofibers | ||
PCE-PS nanofibers | material for PFSPE extraction | ||
steel rod (about 0.5 mm diameter) | fill the nanofibres into the column | ||
gastight plastic syringe (5 ml) | compress solution into the end of the tip | ||
methanol | Sinopharm Chemical ReagentCo., Ltd | 67-56-1 | |
diphenylborinic acid 2-aminoethyl ester(DPBA) | Sigma-Aldrich.Inc | A-106408 | complex reagent |
norepinephrine(NE) | Sigma-Aldrich.Inc | A-9512 | analyte |
3-Methoxy-4-hydroxyphenylglycol(MHPG) | Sigma-Aldrich.Inc | H1377 | analyte |
epinephrine(E) | Sigma-Aldrich.Inc | 100154-200503 | analyte |
3, 4-Dihydroxyphenylacetic acid(DOPAC) | Sigma-Aldrich.Inc | D-9128 | analyte |
dopamine(DA) | Sigma-Aldrich.Inc | H-8502 | analyte |
3, 4-dihydroxybenzylamine hydrobromide(DHBA) | Sigma-Aldrich.Inc | 858781 | interior label |
acetonitrile | Sigma-Aldrich.Inc | 75-05-8 | eluriant and mobile phase |
phosphoric acid | Sinopharm Chemical ReagentCo., Ltd | 7664-38-2 | eluriant |
uric acid | Sinopharm Chemical ReagentCo., Ltd | 69-93-2 | artifical urine |
creatinine | Sinopharm Chemical ReagentCo., Ltd | 60-27-5 | artifical urine |
trisodium citrate | Sinopharm Chemical ReagentCo., Ltd | 6132-04-3 | artifical urine |
KCl | Sinopharm Chemical ReagentCo., Ltd | 7447-40-7 | artifical urine |
NH4Cl | Sinopharm Chemical ReagentCo., Ltd | 12125-02-9 | artifical urine |
NaHCO3 | Sinopharm Chemical ReagentCo., Ltd | SWC0140326 | artifical urine |
C2Na2O4 | Sinopharm Chemical ReagentCo., Ltd | 62-76-0 | artifical urine |
NaSO4 | Sinopharm Chemical ReagentCo., Ltd | 7757-82-6 | artifical urine |
disodium hydrogen phosphate | Sinopharm Chemical ReagentCo., Ltd | 10039-32-4 | artifical urine |
urea | Sinopharm Chemical ReagentCo., Ltd | 57-13-6 | artifical urine |
NaCl | Sinopharm Chemical ReagentCo., Ltd | 7647-14-5 | artifical urine |
MgSO4.7H2O | Sinopharm Chemical ReagentCo., Ltd | 10034-99-8 | artifical urine |
CaCl2 | Sinopharm Chemical ReagentCo., Ltd | 10035-04-8 | artifical urine |
HCl | Sinopharm Chemical ReagentCo., Ltd | 7647-01-0 | artifical urine |
citric acid | Sinopharm Chemical ReagentCo., Ltd | 77-92-9 | artifical urine and mobile phase |
EDTA disodium salt | Sinopharm Chemical ReagentCo., Ltd | 34124-14-6 | mobile phase |
monometallic sodium orthophosphate | Sinopharm Chemical ReagentCo., Ltd | 7558-80-7 | artifical urine and mobile phase |
1-heptanesulfonic acid sodium salt | Sinopharm Chemical ReagentCo., Ltd | 22767-50-6 | mobile phase |
sodium hydroxide | Sinopharm Chemical ReagentCo., Ltd | 1310-73-2 | mobile phase |
phenylboronic acid column(PBA column) | Aglilent | 12102018 | PBA extraction |
Inertsil® ODS-3 5 µm 4.6×150 mm column | Dikma | 5020-06731 | HPLC column for seperation |
SHIMADZU SIL-20AC prominence AUTO SAMPLER | Shimadzu Corporation, Japan | SIL-20AC | auto injection for eluriant |
SHIMADZU LC-20AD High Performance Liquid Chromatography | Shimadzu Corporation, Japan | LC-20AD | HPLC pump |
SHIMADZU L-ECD-60A electrochemical detector | Shimadzu Corporation, Japan | L-ECD-60A | detector for the analytes |
ASAP 2020 Accelerated Surface Area and Porosimetry System | Micromeritics, USA | surface and porosity analyzer |