Source: Murphy, C. A., et al. Optogenetic activation of afferent pathways in brain slices and modulation of responses by volatile anesthetics. J. Vis. Exp. (2020)
This video outlines a method for simultaneously imaging and recording electrical activity from transgenic brain slices that contain neuron-specific fluorescent proteins. The procedure involves maintaining the brain slice in a perfusion chamber with aerated artificial cerebrospinal fluid, using a light source to excite fluorescent-labeled cells, and recording the electrical signals with a multichannel probe and patch clamp.
All procedures involving animal samples have been reviewed and approved by the appropriate animal ethical review committee.
1. Preparation of hardware and software for multi-channel recordings
2. Configuration of light stimulation protocols
3. Placing multi-channel probe in ex vivo brain tissue slice
4. Patch clamping targeted neurons and obtaining whole-cell configuration
Table 1: Composition of artificial cerebral spinal fluid and intracellular solution. Reagents and concentrations for eACSF, and intracellular pipette solution for patch clamp recordings are listed.
Experiment ACSF, eACSF (in mM) | |
NaCl | 111 |
NaHCO3 | 35 |
HEPES | 20 |
KCl | 1.8 |
CaCl2 | 2.1 |
MgSO4 | 1.4 |
KH2PO4 | 1.2 |
glucose | 10 |
Internal Solution | |
K-gluconate | 140 |
NaCl | 10 |
HEPES | 10 |
EGTA | 0.1 |
MgATP | 4 |
NaGTP | 0.3 |
pH = 7.2 |
The authors have nothing to disclose.
2.5x broadfield objective lens | Olympus | MPLFLN2.5X | |
40x water immersion objective lens | Olympus | LUMPLFLN40XW | |
95% O2/5% CO2 mixture | Airgas | Z02OX95R2003045 | |
A16 probe | NeuroNexus | A16x1-2mm-100-177-A16 | 16-channel probe |
Anesthetic gas monitor (POET II) | Criticare | 602-3A | |
Calcium Chloride (CaCl2) | Dot Scientific | DSC20010 | ACSF |
Capillary glass (patch clamp recordings) | King Precision Glass, Inc. | KG-33 | Borosilicate, ID: 1.1mm, OD: 1.7mm, Length: 90.0mm |
Capillary glass (viral injections) | Drummond Scientific Company | 3-000-203-G/X | 3.5" |
Control of junior micromanipulator | Luigs and Neumann | SM8 | for control of junior micromanipulator |
Control of manipulators and shifting table | Luigs and Neumann | SM7 | for control of multichannel electrode and shifting table |
Digidata 1440A + Clampex 10 | Molecular Devices | 1440A | Digitizer and software |
E-3603 tubing | Fisher Scientific | 14171208 | for delivery of 95% O2/5% CO2 gas mixture to incubation chamber + application of pressure during patch clamping |
EGTA | Dot Scientific | DSE57060 | intracellular solution |
ERP-27 EEG Reference/Patch Panel | Neuralynx | Retired | |
Filling needle | World Precision Instruments | 50821912 | for filling patch clamp pipettes |
Filter cube for imaging EYFP | Olympus | U-MRFPHQ | |
Filter paper | Fisher Scientific | 09801E | lay over slice template during preparation of tissue block |
Flaming/Brown micropipette puller | Sutter Instrument | P-1000 | 2.5×2.5 Box filament |
Gas dispersion tube | Sigma Aldrich | CLS3953312C | |
Glass syringe (100 mL) | Sigma Aldrich | Z314390 | for filling gas-sealed bags |
Gluconic Acid, Potassium Salt (K-gluconate) | Dot Scientific | DSG37020 | intracellular solution |
Glucose | Dot Scientific | DSG32040 | ACSF |
GTP, Sodium Salt | Sigma Aldrich | G8877 | intracellular solution |
Headstage-probe adaptor | NeuroNexus | A16-OM16 | adaptor to connect 16-channel probe to headstage input |
HEPES | Dot Scientific | DSH75030 | ACSF,intracellular solution |
HS-16 Headstage | Neuralynx | Retired | |
Isoflurane | Patterson Veterinary | 07-893-1389 | |
Isopropyl alcohol (70%) | VWR International | 101223-746 | |
Junior micromanipulator | Luigs and Neumann | 210-100 000 0090-R | for manipulation of patch clamp electrode |
LED Light Source Control Module | Mightex | BLS-PL02_US | optogenetic light source control |
Lynx-8 Amplifier | Neuralynx | Retired | |
Lynx-8 Power Supply | Neuralynx | Retired | |
Magnesium Sulfate (MgSO4) | Dot Scientific | DSM24300 | ACSF |
mCherry, Texas Red filter cube | Chroma | 49008 | for imaging tdTomato fluorescent reporter |
Meloxicam | |||
Micropipette holder | Fisher Scientific | NC9044962 | |
Microsyringe pump | World Precision Instruments | UMP3-4 | |
MultiClamp 700A | Molecular Devices/Axon Instruments | 700A | Amplifier |
Nitrogen (for air table) | Airgas | NI200 | |
Nylon mesh | Fisher Scientific | 501460083 | stretched over horseshoe of flattened platinum wire, slice rest on top of this during recordings |
Nylon, cut from pantyhose | Generic brand | small piece to create slice platform in incubation chamber, single fibers to create platinum harp | |
Pipette | Dot Scientific | 307 | For transferring tissue to rig |
Platinum wire | VWR International | BT124000 | 2 cm, flattened, to make platinum harp |
Polygon400 | Mightex | DSI-E-0470-0617-000 | optogenetic light delivery system, comes with PolyScan2 software |
Potassium Chloride (KCl) | Dot Scientific | DSP41000 | ACSF |
Potassium Phosphate (KH2PO4) | Dot Scientific | DSP41200 | ACSF |
Sealed gas bag | Fisher Scientific | 109236 | |
Shifting table for microscope | Luigs and Neumann | 380FMU | |
Sodium Bicarbonate (HCO3-) | Dot Scientific | DSS22060 | ACSF |
Sodium Chloride (NaCl) | Dot Scientific | DSS23020 | ACSF, intracellular solution |
Syringe (1 mL) with LuerLock tip | Fisher Scientific | 309628 | for application of pressure during patch clamping |
Syringe (1 mL) with slip tip | WW Grainger, Inc. | 19G384 | for filling patch clamp pipettes |
Syringe Filters | VWR International | 66064-414 | |
Upright microscope | Olympus | BX51 | |