Source: Hanson, C., et al. Fabricating a UV-Vis and Raman Spectroscopy Immunoassay Platform. J. Vis. Exp.(2016).
In this video, we describe the Raman spectroscopy-based direct immunoassay method for detecting specific target antigens. The antibodies, which are adsorbed onto the surface of gold nanoparticles attached to Raman probes, interact with target antigens and produce a specific Raman spectrum, which is subsequently detected.
1. Prepare UV-Vis/Raman Probes
Table 1. DTTC dilution example. Various dilutions of DTTC and the associated volumes of stock DTTC, gold nanoparticle solution, and water.
Volumes to add of each component (ml) | |||
DTTC final concentration (mM) | DTTC working solution (200 mM) | AuNP | Water |
0.2 | 0.1 | 20 | 79.9 |
0.6 | 0.3 | 20 | 79.7 |
1 | 0.5 | 20 | 79.5 |
2 | 1.0 | 20 | 79 |
5 | 2.5 | 20 | 77.5 |
7 | 3.5 | 20 | 76.5 |
10 | 5.0 | 20 | 75 |
2. Immunoassay Plate Preparation
Figure 1. Prepared immunoassay plate. Image of a prepared immunoassay plate. Rows A through E are test samples while rows F through H are control samples. Column 1 contains the undiluted nanoparticles and every subsequent column has half the concentration of AuNP probes.
The authors have nothing to disclose.
60 nm Gold Nanoparticle | Ted Pella, Inc. | 15708-6 | These are citrate-capped gold nanoparticles. Please see Discussion for relationship between Raman reporter and AuNP surface charge and its imporance to proper selection of AuNP and/or Raman reporter. |
Sodium Bicarbonate | Fisher Scientific | S233-500 | |
Methanol | Pharmco-Aaper | 339000000 | |
Tris-Buffered Saline (10x) pH 7.5 | Scy Tek | TBD999 | |
Bottle Top Filtration Unit | VWR | 97066-202 | |
Tween 20 (polysorbate 20) | Scy Tek | TWN500 | Used as an emulsifying agent for washing steps. |
Phosphate-Buffered Saline 10x Concentrate, pH 7.4 | Scy Tek | PBD999 | |
Protein LoBind Tube 2.0 ml | Eppendorf Tubes | 22431102 | LoBind tubes prevent binding of proteins and AuNPs to surfaces of the tubes. |
Protein LoBind Tube 0.5 ml | Eppendorf Tubes | 22431064 | LoBind tubes prevent binding of proteins and AuNPs to surfaces of the tubes. |
Microplate Devices UniSeal | GE Healthcare | 7704-0001 | Used for sealing and storing functionalized plates. |
Assay Plate, With Low Evaporation Lid, 96 Well Flat Bottom | Costar | 3370 | |
HPLC-grade water | Sigma Aldrich | 270733-4L | |
3,3′-Diethylthiatricarbocyanine iodide (DTTC) | Sigma Aldrich | 381306-250MG | Raman reporter |
mPEG-Thiol, MW 5,000 – 1 gram | Laysan Bio, Inc. | MPEG-SH-5000-1g | |
OPSS-PEG-SVA, MW 5,000 – 1 gram | Laysan Bio, Inc. | OPSS-PEG-SVA-5000-1g | OPSS-PEG-SVA has an NHS end. |
Mouse IgG, Whole Molecule Control | Thermo Fisher Scientific | 31903 | Antigen |
Goat anti-Mouse IgG (H+L) Cross Adsorbed Secondary Antibody | Thermo Fisher Scientific | 31164 | Antibody |
Human Serum Albumin Blocking Solution | Sigma Aldrich | A1887-1G | Bovine serum albumin can be used instead. |
UV-Vis Spectrophotometer | Thermo Scientific | Nanodrop 2000c | |
UV-Vis Spectrophotometer | BioTek | Synergy 2 | |
In-house built 785 nm inverted Raman microscope unit | N/A | N/A | An inverted Raman microscope is best for proper focusing onto surface of the well plate. Otherwise a very low magnification will be used due to height of the 96-well plate. An in-house built system was used as it was cheaper than buying from a vendor. However, any commercially available inverted Raman microscope system can be used. |