Science Education
>

Simultaneous Transcranial Alternating Current Stimulation and Functional Magnetic Resonance Imaging

Instructor Prep
concepts
Student Protocol
JoVE Journal
Behavior
This content is Free Access.
JoVE Journal Behavior
Simultaneous Transcranial Alternating Current Stimulation and Functional Magnetic Resonance Imaging

Conduct all experiments in accordance with institutional ethics committee guidelines. For all studies mentioned in this manuscript, all procedures were performed according to the declaration of Helsinki and approved by the local Ethics Committee of the University Medical Center Göttingen.

1. Stimulation and Computer Setup Prior to Experiment

  1. Stimulator setup
    NOTE: The stimulator used for this fMRI experiment is a specially designed magnetic resonance (MR)-compatible system equipped with an MR-safe inner filter box, an outer filter box, safety resistors, coupled cables, and MR-safe materials. Some instructions pertain specifically to the manufacturer's directions, and these may vary when using another stimulator, so take care to follow equipment instructions provided by the manufacturer that may constitute exceptions to this setup. Figure 1A shows the stimulator components used in this experimental setup.
    1. Navigate through the menu of the stimulator to program the desired experimental parameters (refer to user's manual for details). For example, for a stimulation frequency of 10 Hz, program 10 cycles for ramp-up/down time of 1 s, 300 sinusoidal cycles for 30 s of stimulation, current strength equal to 1,000 µA, and repetitive trigger mode, as conducted for our experiments unless otherwise noted. Save the program to load for each time the experiment is run thereafter.
    2. Connect the stimulus presentation computer trigger output to the stimulator using a BNC cable.
    3. Place a nonmagnetic, shielded local area network (LAN) cable through the radio frequency (RF) waveguide tube from the inside of the scanner room. To avoid resonant capacitive coupling, ensure that the cable is free from loops and placed along the wall of the room, leading to the rear of the magnet bore and along the right-side scanner bed railing inside the bore, leading to the position of the inner filter box (see Figure 1C and safety note in Step 2.4 regarding cable position). Secure the cable with tape placed intermittently along the length of it.
  2. Load the visual stimulus program on a designated presentation computer that is separate from the scanner control computer. As depicted in Figure 1C, connect the presentation computer to the scanner trigger output via an optical-to-electrical converter and to an output device (i.e., projector) that is placed in a shielded case or outside the magnet room. Use nonmagnetic mirrors to direct the projection onto a screen inside the scanner bore.

2. Subject Arrival and Preparation

  1. Pre-screen recruited subjects for any contraindications for MR scanning (e.g., no metal implants, no claustrophobia, experiment-specific subject prerequisites) as well as for tACS (e.g., history of seizures, chronic headaches, pregnancy)26,27.
  2. When the subject arrives, instruct the subject about the fMRI experiment details and describe the experience to expect (e.g., visual stimulus, tingling or phosphenes from tACS, special task instructions).
  3. Place electrodes according to the 10-20 EEG system and stimulator preparation.
    1. Using a tape measure, measure the distance on the subject's head from the nasion to the inion, and from ear-to-ear, over the top of the head. The intersection of both lengths gives the position on the head for Cz, according to the 10-20 EEG system. Mark the spot for Cz on the scalp using a marker.
    2. Place an EEG cap without electrodes on the subject's head, with Cz aligned to the mark on the subject's scalp, determine the desired location of the electrodes and mark them.
      NOTE: It is important that all experimenters use the same placement system to ensure consistency through all experiments; the 10-20 EEG system, which is commonly used in transcranial stimulation experiments, has specific guidelines to maintain accurate electrode placement26,28.
    3. Using alcohol and cotton pads, clean the hair and skin on and around the marked spots on the subject's scalp; remove oils and hair products.
    4. Spread some gel on the rubber electrodes and press each electrode firmly on the marked and cleaned locations on the subject's scalp, ensuring full contact from electrode to conductive gel to the scalp with minimal impedance.
    5. Using a spare shielded LAN cable, connect the filter boxes and MR-safe cables to the stimulator and to the rubber electrodes as depicted in Figure 1A.
    6. Turn on the stimulator and test the impedance (refer to user's manual for details). If the impedance is not below 20 kΩ, press the electrodes on the scalp or add electrode gel as necessary until this impedance guideline is met.
    7. When the impedance is below 20 kΩ, allow the stimulator to output current for a few seconds to familiarize the subject with the sensory experience. Ask the subject about sensory perception during this test, including whether tingling sensation exists and can be withstood, and extent or location of phosphenes during stimulation.
    8. At this point, the subject is prepared to move to the scanner bed. Leaving the electrode cable plugged into the rubber electrodes on the subject, disconnect the stimulator, the spare LAN cable, and the outer and inner filter boxes.
    9. Connect the outer filter box to the LAN cable that runs through the waveguide to the MR scanner, leaving as little exposed LAN cable outside of the waveguide as possible (see Figure 1B). Connect the stimulator to the outer filter box using the stimulator cable and double-check that the stimulator is connected to the presentation computer trigger output.
  4. Prepare the subject inside the MR scanner.
    NOTE: Figure 1C shows the full tACS-fMRI setup during the experiment. It is critical to arrange the cables and inner filter box as specified, with the electrode cable arranged at an approximately 90° angle to the plane of the scanner bed and the inner filter box resting on the scanner bed railing on the right side of the scanner bore. Neglecting to do so can damage the safety circuit of the electrode cable; this configuration applies for both open and closed RF coils.
    1. After ensuring that the subject is free of magnetic materials and ready for the MRI experiment, lead the subject into the scanner room.
    2. Give ear plugs for hearing protection to the subject, and instruct the subject to lie on the scanner bed, placing pillows around and under the head and under the legs for comfort and to reduce movement. When placing the pillows behind the subject's head, pay special attention to lay the electrode cable flat and in a position that is comfortable for the subject to lie on for the duration of the experiment.
    3. Give the alarm ball and MR-safe response button box to the subject to hold such that minimal movement is required to push a button to respond in the experiment.
    4. Secure the RF head coil over the subject's head with a mirror attached such that the subject can see the projection screen reflected in the correct orientation.
    5. Temporarily secure the free end of the electrode cable coming from the rubber electrodes to a place in the head coil such that it does not catch when the bed is moving. Figure 1D shows the subject's head positioned in the head coil with pillows, mirror, and tACS cable in place before moving the bed to the center head coil for imaging. The filter box is also shown placed on the scanner bed railing, as an example of where it must sit relative to the head coil when the scanner bed is in measurement position.
    6. Move the scanner bed into measurement position. From the back end of the scanner bore, connect the electrode cable from the rubber electrodes to the inner filter box that connects to the LAN cable, as depicted in Figure 1C. To prevent excess motion during scanning, secure the cables and filter box along the scanner bed railing to the right of the bore with tape and sand bags. Place projector screen into the rear end of the scanner bore.
    7. Test the impedance on the stimulator once more to ensure that all connections between cables, filter boxes, and the stimulator are made properly.

3. MR Scanning and Experiment

  1. Before the scan begins, test that the presentation computer registers when the subject pushes response buttons.
  2. Acquire high-resolution T1-weighted anatomical data (e.g., three-dimensional turbo fast low angle shot, echo time (TE): 3.26 ms, repetition time (TR): 2,250 ms, inversion time: 900 ms, flip angle 9°, isotropic resolution of 1 x 1 x 1 mm3).
    1. After the acquisition, adjust contrast and windowing on the anatomical MRI to low and high extremes to visually detect noise during scanning that may result from the stimulator setup. Continue this visual monitoring of noise concurrently with functional image acquisition.
  3. Start the experiment on the presentation computer, ready to begin with the scanner trigger, and start the stimulator to wait for the presentation computer output trigger. Leave the stimulator on and connected throughout the fMRI experiment to avoid differences in the temporal signal-to-noise ratio (tSNR) between stimulator on and off conditions22.
  4. Start the fMRI scan (e.g., two-dimensional T2*-weighted gradient-echo echo-planar imaging; TE: 30 ms, TR: 2,000 ms, flip angle 70°, 33 slices of 3-mm thickness, no gap between slices at an in-plane resolution of 3 x 3 mm2, 210 volumes for seven minutes of scanning), which triggers the start of the experiment on the presentation computer. Monitor the stimulator display to guarantee that current is sent at desired times throughout the experimental runs.

4. Experiment Conclusion

  1. After the experiment has run and scanning is finished, unplug the inner filter box from the cable connected to the rubber electrodes before moving the scanner bed, remove the subject from the scanner, and remove the electrodes, leaving the subject free to wash their hair.
  2. Turn off the stimulator, and plug it in to recharge. Clean the rubber electrodes with water for their next use.

Simultaneous Transcranial Alternating Current Stimulation and Functional Magnetic Resonance Imaging

Learning Objectives

Figure 2 and Figure 3 show representative images acquired for equipment noise tests in a phantom and in a human subject, respectively. In every row, Figure 2 and Figure 3 show representative axial slices from an acquired volume or calculated map, labeled accordingly above the row. The rightmost image on each row is a sagittal representation of the corresponding volume or calculated map, indicating axial slice locations with blue lines. Aside from the first row, which illustrates electrode placement in white, the volume is overlaid on a T1-weighted image in each figure. Notice that there is no distortion or signal dropout from the electrodes in the T1-weighted images. The second row of Figure 2 shows representative functional MRI data acquired with the tACS setup in place and turned on. In the phantom in Figure 2, notice there is some signal dropout and distortion due to the electrodes, however, row 2 of Figure 3 shows that these distortions do not extend beyond the scalp in a subject. Rows three and four of Figure 2 show noise measurements in the volume, which are acquired using the same parameters as the fMRI data, but without an RF excitation pulse. The images show the noise level in the scanner room and of the MR hardware during the scan. Row three is a noise measurement with tACS off, and row four is one with tACS on. In the fifth and sixth row of Figure 2 are tSNR maps for functional runs with the tACS setup and the stimulator off and on, respectively. TSNR maps calculated from data acquired in the human subject appear in Figure 3 rows three, with tACS off, and four, with tACS on. Notice there is no visible difference in intensity when comparing between stimulation conditions. As we demonstrated in a previous study, the tACS equipment produces around 5% drop in tSNR in images compared to those acquired without the tACS setup, however the tSNR should remain stable across stimulation on and off conditions22.

Figure 4 represents a series of images that demonstrates signal dropout that can occur when non-MR-compatible electrodes are used. Slices from an fMRI volume acquired of a subject with electrodes that may have some metal contaminations show signal dropout below the electrode placed roughly over primary motor cortex, as indicated with red circles.

Figure 5 shows results of an experiment testing the effects of current strength of 16 Hz Cz-Oz tACS on the BOLD signal in subjects whose only task is central cross fixation. Throughout the experiment, 12-second periods of tACS were interleaved with non-stimulation periods varying from 24 – 32 seconds. In a pseudorandomized order, tACS was applied with a different current strength (500 µA, 750 µA, 1,000 µA, 1,500 µA) in each of four runs. Figure 5A shows event-related averages of the BOLD signal for statistically significant clusters, with increasing effect on the BOLD signal with increased current strength. Additionally, Figure 5B shows current-strength specific T-score maps illustrating regional specificity of effects as well as increasing spatial effect with increased current strength. It is also worthwhile to note that BOLD activity in frontal regions was changed significantly, showing that modulations were not always directly below the electrodes. For details, refer to Cabral-Calderin and colleagues22.

Figure 6 shows representative results of an experiment testing the frequency dependence of tACS effects during a visual perception task. Subjects reported the perceived direction of a bistable rotating sphere. At the same time, tACS was applied with electrodes placed at Cz and Oz at one of three stimulation frequencies (10 Hz, 60 Hz, or 80 Hz) in each of three separate sessions. Figure 6A illustrates the experiment timing with visual presentation and tACS periods between blocks of central cross fixation. TACS condition and frequency effect interaction maps and cluster post-hoc tests show frequency-specific effects in the parietal cortex, with 10 Hz tACS decreasing and 60 Hz increasing signal (Figure 6B). Figure 6C shows T-score maps of specific effects of 60 Hz tACS extending beyond the parietal cortex to include some occipital and frontal regions. For experiment and analysis details, refer to Cabral-Calderin, et al.22.

Figure 1
Figure 1: TACS Setup in the Scanner. (A) TACS Setup with all Necessary Elements. The stimulator and cables are connected outside of the MR shielded room. Also shown are the EEG cap, tape measure, and conductive gel used for electrode placement. (B) Outer Filter Box and Stimulator Placed Outside the Scanner Room. The LAN cable (not visible in the figure) comes from the scanner room through the RF waveguide tube and connects to the outer filter box, with as little LAN cable exposed outside of the scanner room as possible. The stimulator should be connected to the outer filter box as well as to the presentation computer trigger output cable. (C) Scanner Environment with Experimental Setup. Depiction of tACS setup, including presentation computer, scanner computer and trigger output, and projector. (D) Subject Positioning for Experiment. Important elements include pillows, cable placement, viewing mirror, and head coil. Filter box is placed on scanner bed railing as an example of placement inside the bore. Please click here to view a larger version of this figure.

Figure 2
Figure 2: Quality Assessment MR Images Acquired of a Phantom. Row 1: High-resolution anatomical T1-weighted image axial slices with their positions indicated by blue lines on a sagittal slice to the right (also seen in each following row). On the sagittal plane, electrode positions are illustrated in white. Row 2: T2*-weighted echo-planar image slices, with magenta arrows indicating signal dropout and distortion due to electrodes and/or electrode gel. On the sagittal plane, the positioning of the corresponding volume is shown as an overlay (also seen in each following row). Row 3: Noise image slices acquired with fMRI experimental parameters and no RF excitation pulse while the tACS setup is in place and turned on, but not stimulating. Row 4: No-RF-excitation image acquired with tACS setup in place and stimulator on and stimulating at 16 Hz. Row 5: TSNR map calculated from data acquired with the tACS setup in place and turned on, but not stimulating. Row 6: TSNR map calculated from data acquired with the tACS setup in place and stimulating at 16 Hz. Please click here to view a larger version of this figure.

Figure 3
Figure 3: Quality Assessment MR Images Acquired of a Subject. Row 1: High-resolution anatomical image axial slices with their positions indicated by blue lines on a sagittal slice to the right (as seen in each row). Electrode positions are illustrated in white on the sagittal view. Row 2: T2*-weighted echo-planar image slices showing no signal dropout due to electrodes and/ or electrode gel. On the sagittal plane, the positioning of the corresponding volume is shown as an overlay (also seen in each following row). Row 3: TSNR map calculated from data acquired with the tACS setup in place and turned on, but not stimulating. Row 4: TSNR map calculated from data acquired with the tACS setup in place and stimulating at 16 Hz.  Please click here to view a larger version of this figure.

Figure 4
Figure 4: Signal Dropout Due to a Contaminated Electrode. Slices from an fMRI volume acquired of a subject using a contaminated electrode placed roughly over the hand knob of the motor cortex. Red circles indicate regions below the electrode with signal dropout. Please click here to view a larger version of this figure.

Figure 5
Figure 5: Effect of Current Strength on tACS Modulation of the BOLD Signal. (A) F-score Maps Showing the Main Effect of Current Strength on the Effect of 16 Hz tACS. A significant main effect of current strength in a one-way rANOVA [within factor: current strength (500, 750, 1,000, 1,500 µA)] is apparent. The plots show the event-related average time course of the BOLD signal for the tACS-on periods for each current strength. Shaded regions indicate standard error of the mean across subjects. MedialFG = medial frontal gyrus, IPS = intraparietal sulcus, IFG = inferior frontal gyrus, PrC = precentral gyrus, L = left, R = right, *cluster not corrected for multiple comparisons. (B) T-score Maps Showing BOLD Activity Changes during 16 Hz tACS for Each Current Strength. No significant effect was found with 500 µA tACS. LH = left hemisphere; RH = right hemisphere. This picture has been modified from Cabral-Calderin et al.29. Please click here to view a larger version of this figure.

Figure 6
Figure 6: Effect of tACS on the BOLD Signal in a Visual Perception Task. (A) Schematic Representation of the Experiment. Visual stimulus and tACS were applied in a block design, with 30 s on-off tACS blocks occurring during 120 sec blocks of visual stimulus presentation. Each frequency was tested in a different session. SfM = structure-from-motion. (B) TACS Condition and Frequency Interaction Effect. F-statistic maps showing significance in two-way rANOVA [within factors: tACS (on, off), frequency (10 Hz, 60 Hz, 80 Hz)] and beta estimates for two representative clusters in the post-central gyrus. Continuous lines and black asterisks mark significant differences for post-hoc comparisons for tACS on-off interaction effects of 10 Hz versus 60 Hz and 10 Hz versus 80 Hz, and red asterisks imply a significant difference for tACS on versus off post-hoc tests. PoC = postcentral gyrus, IPS = intraparietal sulcus. (C)T-score Map of 60 Hz tACS. Significant differences comparing 60 Hz tACS on versus off. This picture has been reprinted from Cabral-Calderin et al.29. Please click here to view a larger version of this figure.

List of Materials

None
DC-Stimulator MR NeuroConn, Ilmenau, Germany includes: inner filter box, outer filter box, MR-safe electrode and stimulator cables (1 each), stimulator, 2 surface electrodes, and one shielded LAN cable; NOTE: This manuscript describes tACS-fMRI setup with NeuroConn's MR-safe stimulator, but such a stimulator from another manufacturer would be acceptable, with adaptations made based on manufacturer specifications.
3 tesla Tim Trio MR scanner Siemens, Erlangen, Germany
presentation computer
presentation software (e.g., Matlab) The Mathworks, Natick, USA
shielded LAN cable
projector InFocus Corporation, Wilsonville, USA IN-5108
Ten20 Electrode Paste Weaver and Co., Aurora, USA
EEG cap – EASYCAP 32-channel system Brain Products GmbH, Germany
tape measure
marker
pillows
button response box Current Designs, Philadelphia, USA
isopropyl alcohol
cotton pads
tape
MR-safe sand bags Siemens, Erlangen, Germany
MR-safe mirrors Siemens, Erlangen, Germany
MR-safe screen can be built in local machine shop to fit site-specific parameters
E-A-Rsoft ear plugs 3M, Bracknell, UK

Lab Prep

Transcranial alternating current stimulation (tACS) is a promising tool for noninvasive investigation of brain oscillations. TACS employs frequency-specific stimulation of the human brain through current applied to the scalp with surface electrodes. Most current knowledge of the technique is based on behavioral studies; thus, combining the method with brain imaging holds potential to better understand the mechanisms of tACS. Because of electrical and susceptibility artifacts, combining tACS with brain imaging can be challenging, however, one brain imaging technique that is well suited to be applied simultaneously with tACS is functional magnetic resonance imaging (fMRI). In our lab, we have successfully combined tACS with simultaneous fMRI measurements to show that tACS effects are state, current, and frequency dependent, and that modulation of brain activity is not limited to the area directly below the electrodes. This article describes a safe and reliable setup for applying tACS simultaneously with visual task fMRI studies, which can lend to understanding oscillatory brain function as well as the effects of tACS on the brain.

Transcranial alternating current stimulation (tACS) is a promising tool for noninvasive investigation of brain oscillations. TACS employs frequency-specific stimulation of the human brain through current applied to the scalp with surface electrodes. Most current knowledge of the technique is based on behavioral studies; thus, combining the method with brain imaging holds potential to better understand the mechanisms of tACS. Because of electrical and susceptibility artifacts, combining tACS with brain imaging can be challenging, however, one brain imaging technique that is well suited to be applied simultaneously with tACS is functional magnetic resonance imaging (fMRI). In our lab, we have successfully combined tACS with simultaneous fMRI measurements to show that tACS effects are state, current, and frequency dependent, and that modulation of brain activity is not limited to the area directly below the electrodes. This article describes a safe and reliable setup for applying tACS simultaneously with visual task fMRI studies, which can lend to understanding oscillatory brain function as well as the effects of tACS on the brain.

Procedure

Transcranial alternating current stimulation (tACS) is a promising tool for noninvasive investigation of brain oscillations. TACS employs frequency-specific stimulation of the human brain through current applied to the scalp with surface electrodes. Most current knowledge of the technique is based on behavioral studies; thus, combining the method with brain imaging holds potential to better understand the mechanisms of tACS. Because of electrical and susceptibility artifacts, combining tACS with brain imaging can be challenging, however, one brain imaging technique that is well suited to be applied simultaneously with tACS is functional magnetic resonance imaging (fMRI). In our lab, we have successfully combined tACS with simultaneous fMRI measurements to show that tACS effects are state, current, and frequency dependent, and that modulation of brain activity is not limited to the area directly below the electrodes. This article describes a safe and reliable setup for applying tACS simultaneously with visual task fMRI studies, which can lend to understanding oscillatory brain function as well as the effects of tACS on the brain.

Tags