Técnicas não invasivas de medição da pressão arterial

JoVE Science Education
Biomedical Engineering
A subscription to JoVE is required to view this content.  Sign in or start your free trial.
JoVE Science Education Biomedical Engineering
Noninvasive Blood Pressure Measurement Techniques

11,572 Views

09:31 min

April 30, 2023

Overview

Source: Hamna J. Qureshi and Craig J. Goergen, Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana

Here we will highlight the key similarities and differences of noninvasive blood pressure measurement techniques between humans and rodents and examine the engineering principles that govern blood pressure. The principles that govern current cuff technology to acquire systolic and diastolic pressures will also be discussed.

Commercially available cuffs that connect with mobile devices are typically compact and portable, thereby allowing measurements to be taken virtually anywhere. Noninvasive, portable blood pressure cuffs are especially useful for patients with hypertension and other cardiovascular problems that require careful monitoring and early detection of any changes in blood pressure.

Similarly, noninvasive blood pressure measurement systems are also available for rodents. This technology is used in laboratory settings and is useful for monitoring animal health throughout a study. While radiotelemetry is the gold standard of blood pressure measurement for rodents, this technique is invasive and can lead to animal mortality if done incorrectly. Noninvasive methods, therefore, are convenient for taking measurements in animals as they can provide valuable data without the need for device implantation. A commercially available system will be used to demonstrate how blood pressure can be measured in humans outside of a clinical setting. This technique allows patients to monitor their own blood pressure periodically without having to visit a clinic each time they want these measurements taken.

The methods described here take advantage of blood flow through the tail of the rodent by using pressure sensors and occlusion cuffs. Both mobile blood pressure cuffs for humans and noninvasive tail-cuff methods for rodents take advantage of similar hemodynamic principles to acquire blood pressure measurements that can provide useful data for users, including clinicians, researchers, and patients.

Principles

Blood pressure measurements consist of systolic and diastolic components, which are both important indicators of cardiovascular health. Systolic blood pressure is defined as the maximum intraluminal pressure exerted against artery walls due to heart contraction, whereas diastolic blood pressure is defined as the minimum intraluminal pressure when the aortic valve is closed.

Blood pressure measurements taken with commercially available cuffs use oscillometric methods to take measurements. In clinical settings, auscultatoric methods are commonly used to take blood pressure. Clinicians typically inflate blood pressure cuffs around the brachial artery in the arm until the vessel is completely occluded. Once they deflate the cuff, clinicians listen for the point at which they first hear the beating of the heart through the stethoscope. This point indicates the systolic blood pressure, as the pressure within the artery is large enough to open the vessel. The cuff then continues to deflate until no repetitive sound is heard. This point indicates the diastolic blood pressure, as the vessel remains open. Oscillometric technology is similar to auscultatoric technology except that it is automated. Instead of listening for a sound, an algorithm for the cuff detects oscillations in blood flow once the blood pressure cuff begins to deflate. These oscillations begin when blood starts flowing through the artery again, indicating the systolic blood pressure. The oscillations level out when blood flow becomes steady, indicating diastolic blood pressure.

Noninvasive methods to take blood pressure in rodents also use automated technology. The system described here includes an occlusion cuff that inflates and deflates around the tail artery, analogous to how clinical blood pressure cuffs inflate and deflate around the brachial artery. A volume pressure recording (VPR) sensor is also placed around the tail. This sensor can determine blood pressure using a pressure transducer that can indicate when systolic and diastolic pressures occur within the animal's tail, a close surrogate for true system blood pressure.

Procedure

1. Wireless Blood Pressure Cuff for Humans

  1. Turn on the cuff and pair it with a mobile device onto which results will be outputted. Pairing can be done via Bluetooth.
  2. Strap the cuff onto the upper arm so that the blood pressure monitor is placed against inner bicep. The Ssrap should be secured tightly on the arm.
  3. Open the application for the cuff on the mobile device. Ensure that the cuff is paired with the mobile device.
  4. Position yourself so that you are seated with your arm resting such that the cuff is at the same level as your heart. Make sure you are in a calm environment so that you remain relaxed while the data is being collected.
  5. Once you are positioned correctly, begin the blood pressure measurement. Multiple trials can be run so that an average of these measurements is reflected in the results.
  6. After all measurements have been collected, the results will be outputted onto mobile device screen. Take note of the results. Unstrap the cuff from arm and shut it off.

2. Noninvasive Blood Pressure Measurements for Rodents

  1. Turn on the heated stage, blood pressure system, and desktop computer. Attach a VPR cuff and occlusion cuff to the blood pressure system. Use software to check that the cuffs are properly calibrated and are working as expected. Set-up the software with the desired parameters.
  2. Prepare the animal (mouse or rat) for placement into an appropriately sized holder. Make sure the environment is calm with minimal unexpected noises. Try to calm animal before placing it in its holder to prevent the animal from experiencing unnecessary stress.
  3. Gently coax animal into the holder. Make sure its paws do not get stuck in the attachments of the holder and that the tail is outside of the restrainer.
  4. Once the animal is secured, place the holder on the heated stage. Slide the occlusion cuff onto the base of the tail and then slide the VPR cuff on just behind the occlusion cuff. Leave 1-2 mm of space in between the two cuffs. If using an occlusion cuff and VPR cuff sized for mice, you may secure them using a strip of plastic tubing.
  5. After the cuffs are on, tape them to the stage to prevent the animal's tail from moving too much. Allow the tail to warm to about 32-34 °C to promote vasodilation and subsequent blood flow through the tail.
  6. Once setup is complete and the tail is sufficiently warm, begin data acquisition. While the blood pressure is being measured, make note if the animal moves substantially within restrainer. Try to keep the cuff in the same position throughout the measurement. If the cuffs slide down the tail, or if the animal moves inside the restrainer, press the Pause button and readjust the cuffs/animal once the ongoing run has completed. Maintain a calm environment throughout this process.
  7. Check to make sure blood pressure data are being acquired as expected. Repeat measurements if necessary. Multiple trials are recommended (typically 20-40 measurements).
  8. Once data has been collected, save the measurements for further analysis. Remove the animal from the holder and place it back into its cage. Turn off the heated stage and blood pressure system and clean all surfaces that were in contact with the animal.

Blood pressure measurements are often used to monitor cardiovascular health. Blood pressure is the lateral pressure produced by circulating blood on vessel walls. Close monitoring is important to detect and treat hypotension, low blood pressure, and hypertension, high blood pressure, both of which can impair the function of vital organs, such as the heart, kidneys, and brain.

Blood pressure is measured manually by auscultation, where the clinician occludes the artery using an inflatable cuff and then listens for the sound of blood flow through the brachial artery as it deflates. Blood pressure can also be measured automatically by oscillometry, which allows patients to measure their blood pressure at home. The inflatable cuff contains a pressure transducer that converts the arterial vibrations to systolic and diastolic pressures based on hemodynamic principles.

This video will illustrate the principles of blood pressure, demonstrate the current methods to measure the systolic and diastolic blood pressures of humans and small animals, and discuss some applications.

Blood pressure measurements consist of systolic and diastolic components, both of which are important indicators of cardiovascular health.

The systolic blood pressure is a measure of the maximum intraluminal pressure exerted against the artery during heart contraction when your heart beats, and the diastolic pressure is a measure of the minimum intraluminal pressure, or the pressure in between beats when the heart relaxes and the aortic valve is closed. To measure blood pressure, a cuff is placed on the patient’s upper arm and then inflated to the point where blood flow is completely occluded. As the cuff deflates, the clinician auscultates while observing the manometer readings.

The first audible Korotkoff sound corresponds to the systolic pressure, the pressure at which blood begins to flow again through the open vessel. As the cuff continues to deflate, blood flows more freely as the vessel remains open. Diastolic pressure is recorded at the point when the clinician no longer hears the beating sound or the self-monitoring device detects a leveling-off of oscillations.

A similar method can be used to monitor rodent blood pressure. A cuff is inflated and deflated around the tail artery, while a pressure transducer detects systolic and diastolic pressure. Now that you understand the difference between systolic and diastolic pressure, let’s demonstrate how to measure blood pressure using a wireless cuff for humans and a noninvasive technique for rodents.

Turn on the wireless cuff and the mobile device and open the cuff application. Then, pair the devices via Bluetooth and open the cuff application on the mobile device. Then, strap the cuff onto your upper arm. Make sure that the blood pressure monitor is placed against your inner bicep. The strap should be secured tightly on your arm. Sit in a position so that the cuff is at the same level as your heart. Make sure you are calm and relaxed before starting.

Now, begin the blood pressure measurement by pressing start. Each measurement takes about one to three minutes. Repeat the measurement several times to obtain an average value in the results. The result will be displayed on the screen of the mobile device. After saving the results, unstrap the cuff from your arm and shut off the device.

To measure the blood pressure of a rodent, first turn on the desktop computer, blood pressure system, and heated stage. Then, attach the VPR cuff and occlusion cuff to the blood pressure system. Use software to confirm that the cuffs are properly calibrated and working as expected. Choose the appropriate parameters for data acquisition. Parameters include deflation time, the number of sets per cycle, occlusion pressure, and the delay between sets. Occlusion pressure is typically set to 250 millimeters of mercury, and there are usually 20 sets per cycle. Other parameters can be specified based on user needs.

Calm the animal to prevent it from experiencing unnecessary stress. The environment should be quiet, with minimal noise. Then, place the animal in an appropriately sized holder. Make sure the animal’s paws do not get stuck in the attachments of the holder and that the tail is outside of the restrainer. Now, place the holder on the heated stage. Slide the occlusion cuff, followed by the VPR cuff onto the base of the tail, approximately one to two millimeters apart. Once the cuffs are in place, tape them to the stage. This will restrict tail movement.

Then, heat the tail to 32 to 34 degrees Celsius to promote vasodilation and blood flow through the tail. As the stage is warming, choose the type of animal on the computer whose blood pressure you will be acquiring and then enter its animal ID into the system so you can access its data later. Once the stage is sufficiently heated, start acquiring blood pressure data by pressing start. Try to maintain a calm environment. Keep the position of the cuff constant and make a note if the animal moves substantially. If the cuffs slide down the tail, or if the animal moves inside the restrainer, press the pause button, then make adjustments after the current measurement is completed, and resume testing.

Check to make sure the blood pressure data are being acquired as expected. Collect 20 to 40 measurements. Once all data has been collected, remove the animal from the holder and place it back into a cage. Then turn off the equipment and clean the surfaces that were in contact with the animal.

The results using the blood pressure monitor for humans are outputted on a screen as shown. Using this technology, systolic and diastolic blood pressure measurements are displayed and multiple readings are averaged.

Healthy blood pressures in humans are typically below 120 millimeters of mercury for the systolic pressure and below 80 millimeters of mercury for the diastolic pressure. Blood pressures above these values could indicate that a patient is starting to develop high blood pressure, which can lead to further health complications.

Here is a typical graph that is obtained when taking rodent blood pressure. The two curves on the graph represent data collected from the VPR sensor, the blue line, and the occlusion cuff, the red line. The inflection points of the blue curve are used to calculate the systolic and diastolic pressure. The measurements are listed at the bottom of the graph along with the OK status, which indicates that this measurement is deemed acceptable by the system.

Typical blood pressure measurements for mice are around 120 over 70 millimeters of mercury, whereas rats have slightly higher blood pressures at around 130 over 90 millimeters of mercury.

Now, let’s take a look at how blood pressure is noninvasively measured in the public and scientific community.

Portable noninvasive blood pressure cuffs allow hypertensive patients to monitor their health between visits to the clinic. Pregnant women at risk for developing pre-eclampsia can monitor their risk for this condition in the critical months before the baby is born.

Research on the development of new therapeutic agents requires frequent blood pressure assessment. While invasive techniques such as subcutaneous telemetry and carotid artery cannulation in mice provide long-term blood pressure data, noninvasive tail-cuff systems are an attractive method to monitor the blood pressure in animal models of cardiovascular disease.

You’ve just watched JoVE’s introduction to blood pressure measurement techniques. You should now understand how noninvasive measurement methods are used for both humans and rodents and how the technology is applied to improve health monitoring for the community. Thanks for watching!

Results

A typical graph obtained from taking rodent blood pressure using a noninvasive blood pressure system is shown in Figure 1. There are two curves on the graph: one for the VPR sensor and one for the occlusion cuff. The inflection points of the VPR sensor curve are where systolic and diastolic blood pressures are determined, as depicted in Figure 1. A status is indicated at the bottom of the figure that communicates whether this measurement is deemed acceptable by the system. In order to obtain acceptable readings, blood pressures are taken multiple times per trial.

Figure 1
Figure 1. Systolic and diastolic blood pressures obtained by a noninvasive measurement system.

Figure 2
Figure 2. Screenshot of systolic and diastolic blood pressure.

For a commercially available cuff for humans, the results of the blood pressure measurement are outputted on a screen as shown in Figure 2. Users can see their systolic and diastolic blood pressure measurements using this technology. Blood pressure can be taken multiple times so that the results output an average of the data that was collected.

Applications and Summary

Both commercial cuffs available for at-home use and noninvasive methods for animal research studies take advantage of automated technologies. By tracking changes in blood flow with sensors, algorithms detect the systolic and diastolic pressures and then communicate these results to users. While technologically similar, there are some key differences between the two measurement techniques that should be noted. When taking blood pressures in human subjects, the environment is typically relaxed as the subject is unrestrained and fairly comfortable. Conversely, animals must be restrained or anesthetized. Restraining can make the animal anxious and therefore skew blood pressure measurements to be higher, whereas anesthesia can lead to hypothermia and vasoconstriction, making reliable measurements even harder to acquire. This is why it is preferable to obtain rodent blood pressures in conscious animals that are in a calm, dimly lit environment where they can be kept warm with no startling noises. Vasodilation through heating is also used when rodents are having their blood pressure measured to promote blood flow through their tail. Acclimating animals to the holder and cuffs before acquiring data can help reduce discomfort and allow collection of more accurate results. Also, if a user is collecting measurements from the same animal on multiple days, the results will be more meaningful if acquired at roughly the same time each day.

Being able to take blood pressure measurements with a user-friendly, compact cuff is an extremely useful tool for patients as it allows patients to monitor their blood pressures at their own convenience. Commercially available blood pressure cuffs, however, are not meant to replace clinically obtained measurements; rather these cuffs are meant to help patients monitor their systemic blood pressure between visits. Commercial cuffs are also subject to noise and do not always collect accurate results. Similarly, noninvasive measurements of rodent blood pressure are also noisy, especially when the animal moves. As such, it is important to be judicious when looking at data from these systems.

Portable noninvasive blood pressure measurement methods are useful for a variety of applications. Commercial cuffs allow hypertensive patients to monitor their health between visits to the clinic. This method can help patients become more conscious of their dietary and exercise decisions, as they can see the direct impact that their lifestyle has on their blood pressure. Commercial cuffs can also help pregnant women track their blood pressures in the final months of their pregnancies. Pregnant women are at risk for developing preeclampsia. Therefore, it is important for these women to know if they are at risk for this condition during the latter half of gestation. Using a portable cuff helps these women track their blood pressure in the critical months before the baby is born.

For animal use, noninvasive blood pressure measurements are valuable for research purposes, especially those regarding cardiovascular health. Taking blood pressures of animals after procedures and studies, such as those inducing hypertension in rodents, can show the effects of these actions on an animal's health. While other invasive techniques can provide more consistent data, the noninvasive nature of tail-cuff systems make them an attractive method for those using animal models of cardiovascular disease.

Transcript

Blood pressure measurements are often used to monitor cardiovascular health. Blood pressure is the lateral pressure produced by circulating blood on vessel walls. Close monitoring is important to detect and treat hypotension, low blood pressure, and hypertension, high blood pressure, both of which can impair the function of vital organs, such as the heart, kidneys, and brain.

Blood pressure is measured manually by auscultation, where the clinician occludes the artery using an inflatable cuff and then listens for the sound of blood flow through the brachial artery as it deflates. Blood pressure can also be measured automatically by oscillometry, which allows patients to measure their blood pressure at home. The inflatable cuff contains a pressure transducer that converts the arterial vibrations to systolic and diastolic pressures based on hemodynamic principles.

This video will illustrate the principles of blood pressure, demonstrate the current methods to measure the systolic and diastolic blood pressures of humans and small animals, and discuss some applications.

Blood pressure measurements consist of systolic and diastolic components, both of which are important indicators of cardiovascular health.

The systolic blood pressure is a measure of the maximum intraluminal pressure exerted against the artery during heart contraction when your heart beats, and the diastolic pressure is a measure of the minimum intraluminal pressure, or the pressure in between beats when the heart relaxes and the aortic valve is closed. To measure blood pressure, a cuff is placed on the patient’s upper arm and then inflated to the point where blood flow is completely occluded. As the cuff deflates, the clinician auscultates while observing the manometer readings.

The first audible Korotkoff sound corresponds to the systolic pressure, the pressure at which blood begins to flow again through the open vessel. As the cuff continues to deflate, blood flows more freely as the vessel remains open. Diastolic pressure is recorded at the point when the clinician no longer hears the beating sound or the self-monitoring device detects a leveling-off of oscillations.

A similar method can be used to monitor rodent blood pressure. A cuff is inflated and deflated around the tail artery, while a pressure transducer detects systolic and diastolic pressure. Now that you understand the difference between systolic and diastolic pressure, let’s demonstrate how to measure blood pressure using a wireless cuff for humans and a noninvasive technique for rodents.

Turn on the wireless cuff and the mobile device and open the cuff application. Then, pair the devices via Bluetooth and open the cuff application on the mobile device. Then, strap the cuff onto your upper arm. Make sure that the blood pressure monitor is placed against your inner bicep. The strap should be secured tightly on your arm. Sit in a position so that the cuff is at the same level as your heart. Make sure you are calm and relaxed before starting.

Now, begin the blood pressure measurement by pressing start. Each measurement takes about one to three minutes. Repeat the measurement several times to obtain an average value in the results. The result will be displayed on the screen of the mobile device. After saving the results, unstrap the cuff from your arm and shut off the device.

To measure the blood pressure of a rodent, first turn on the desktop computer, blood pressure system, and heated stage. Then, attach the VPR cuff and occlusion cuff to the blood pressure system. Use software to confirm that the cuffs are properly calibrated and working as expected. Choose the appropriate parameters for data acquisition. Parameters include deflation time, the number of sets per cycle, occlusion pressure, and the delay between sets. Occlusion pressure is typically set to 250 millimeters of mercury, and there are usually 20 sets per cycle. Other parameters can be specified based on user needs.

Calm the animal to prevent it from experiencing unnecessary stress. The environment should be quiet, with minimal noise. Then, place the animal in an appropriately sized holder. Make sure the animal’s paws do not get stuck in the attachments of the holder and that the tail is outside of the restrainer. Now, place the holder on the heated stage. Slide the occlusion cuff, followed by the VPR cuff onto the base of the tail, approximately one to two millimeters apart. Once the cuffs are in place, tape them to the stage. This will restrict tail movement.

Then, heat the tail to 32 to 34 degrees Celsius to promote vasodilation and blood flow through the tail. As the stage is warming, choose the type of animal on the computer whose blood pressure you will be acquiring and then enter its animal ID into the system so you can access its data later. Once the stage is sufficiently heated, start acquiring blood pressure data by pressing start. Try to maintain a calm environment. Keep the position of the cuff constant and make a note if the animal moves substantially. If the cuffs slide down the tail, or if the animal moves inside the restrainer, press the pause button, then make adjustments after the current measurement is completed, and resume testing.

Check to make sure the blood pressure data are being acquired as expected. Collect 20 to 40 measurements. Once all data has been collected, remove the animal from the holder and place it back into a cage. Then turn off the equipment and clean the surfaces that were in contact with the animal.

The results using the blood pressure monitor for humans are outputted on a screen as shown. Using this technology, systolic and diastolic blood pressure measurements are displayed and multiple readings are averaged.

Healthy blood pressures in humans are typically below 120 millimeters of mercury for the systolic pressure and below 80 millimeters of mercury for the diastolic pressure. Blood pressures above these values could indicate that a patient is starting to develop high blood pressure, which can lead to further health complications.

Here is a typical graph that is obtained when taking rodent blood pressure. The two curves on the graph represent data collected from the VPR sensor, the blue line, and the occlusion cuff, the red line. The inflection points of the blue curve are used to calculate the systolic and diastolic pressure. The measurements are listed at the bottom of the graph along with the OK status, which indicates that this measurement is deemed acceptable by the system.

Typical blood pressure measurements for mice are around 120 over 70 millimeters of mercury, whereas rats have slightly higher blood pressures at around 130 over 90 millimeters of mercury.

Now, let’s take a look at how blood pressure is noninvasively measured in the public and scientific community.

Portable noninvasive blood pressure cuffs allow hypertensive patients to monitor their health between visits to the clinic. Pregnant women at risk for developing pre-eclampsia can monitor their risk for this condition in the critical months before the baby is born.

Research on the development of new therapeutic agents requires frequent blood pressure assessment. While invasive techniques such as subcutaneous telemetry and carotid artery cannulation in mice provide long-term blood pressure data, noninvasive tail-cuff systems are an attractive method to monitor the blood pressure in animal models of cardiovascular disease.

You’ve just watched JoVE’s introduction to blood pressure measurement techniques. You should now understand how noninvasive measurement methods are used for both humans and rodents and how the technology is applied to improve health monitoring for the community. Thanks for watching!