The Controlled Odor Mimic Permeation System is a simple, field-portable, low-cost method of odor delivery for olfactory testing and training. It is constructed of an odorant retained on an adsorbent material and contained inside of a permeable polymer bag allowing controlled release of the odorant vapor over time.
The Controlled Odor Mimic Permeation System (COMPS) was developed to provide a convenient field testing method of odor delivery at controlled and reproducible rates. COMPS are composed of an odorant of interest on an absorbent material sealed inside of a permeable polymer bag. The permeable layer allows for a constant release of the odorant over a given amount of time. The permeable bag is further stored in a secondary, impermeable bag. The double-containment procedure allows for equilibration of the odorant from the permeable bag but within the impermeable outer layer, resulting in an instant and reproducible source of odorant vapor upon removal from the outer packaging. COMPS are used in both olfactory testing for experimental scenarios and for olfactory detection training, such as with detection canines. COMPS can be used to contain a wide range of odorants (e.g., narcotics powders) and provide a controlled release of the associated odorants. Odor availability from COMPS is expressed in terms of permeation rate (i.e., the rate of the odorant vapor released from a COMPS per unit time) and is typically measured by gravimetric means. The permeation rate for a given mass or volume of odorant can be adjusted as needed by varying the bag thickness, surface area, and/or polymer type. The available odor concentration from a COMPS can also be measured by headspace analysis techniques such as solid phase microextraction with gas chromatography/mass spectrometry (SPME-GC/MS).
Olfaction is a crucial, yet often overlooked, sensing mechanism used by most animals. For many it is the main mechanism for locating food, finding a mate, or sensing danger1. Furthermore, the olfactory capabilities some animals, most notably canines, are regularly exploited by humans for the detection of contraband (e.g., narcotics or explosives), or other objects of interest, such as missing persons, invasive species, or diseases2,3. For canine detection research or other olfaction research topics, investigators often study the process of olfaction and the strengths and limitations of the olfactory system. As such, it is generally desirable to control the release of an odorant vapor into the environment to reproducibly deliver known quantities of odorant during testing. Failure to account for variations in odor availability due to factors such as vapor pressure or environmental effects often complicates data interpretation and applicability4. It is similarly desirable to provide an established quantity of odor during training scenarios for detection canines. For example, studies by Hallowell et al.5 and by Papet6 have indicated the importance of odor intensity in odor perception, and that altering the intensity of an odorant can affect how it is perceived alone or in a mixture.
In laboratory settings, the use of analytical equipment such as permeation tubes with controllable ovens, vapor generators, or olfactometers may be used to control odor delivery. However, this type of equipment is impractical for use during field testing and training scenarios4. The Controlled Odor Mimic Permeation System (COMPS) was developed as a simple, low-cost, and disposable method for controlled odor delivery requiring no external power. Therefore, they can easily be incorporated into a variety of different testing and training scenarios7. COMPS units are simply composed of an odorant of interest on an absorbent material sealed inside of a permeable polymer bag, stored in a secondary containment system. The utilization of COMPS reduces variability between tests and improves consistency during training exercises8.
Odor delivery or availability from COMPS is measured in terms of permeation rate, as determined by gravimetric analysis in terms of mass of vapor released over time. Permeation rates can be controlled by a number of factors, including the thickness of the polymer bag, its available surface area, the type of absorbent material (substrate) used, and the amount of the odorant. Permeation rate is constant for a given period of time (hours or days) depending on the odorant being used. This allows for minimal variability in odor delivery during testing or training. During storage, COMPS come to equilibrium within the impermeable outer container, resulting in an instant source of odorant vapor at a known permeation rate.
COMPS were initially designed to contain odorants associated with explosive materials and to be used as odor mimics7. As defined by Macias et al., an odor mimic simulates a material of interest, such as an explosive, by providing the dominant volatile compounds, or odorants, found in the headspace of that material without the presence of the parent material itself8. To create an odor mimic, the active odorants of the parent material must be determined. An active odorant, in this scenario, is described as a volatile compound that a trained explosive-detection canine detects, believing that there is an actual explosive material present. Having identified dominant volatile compounds in the headspace of several explosive materials, COMPS were prepared to release these individual odorants at a controlled rate for the duration of canine olfactory detection field trials and determine the active odorant associated with several explosive materials. COMPS were successfully used for this purpose7,9 and have since been used as odor mimics for further explosive detection training.
Macias et al. utilized COMPS containing piperonal, a pure chemical solid at room temperature that, in the vapor phase, has been shown to be the active odorant for MDMA (3,4-methylenedioxymethamphetamine), the psychoactive drug known as ecstasy. The researchers used varying thicknesses and surface areas of low-density polyethylene bags to adjust the permeation rate of piperonal vapor. This series of COMPS was then used to estimate piperonal detection threshold for trained narcotics-detection canines8. Conversely, in a separate study, COMPS bag thicknesses were adjusted to minimize the deviation of permeation rates between each compound in a homologous series though they possessed drastically varying vapor pressures. If a single bag thickness had been used in this study, those compounds with higher vapor pressures would have yielded much higher permeation rates. By increasing the bag thickness for the higher volatility compounds, the permeations rates were adjusted so that they were similar for all compounds4. Both studies demonstrate the utility and adaptability of the COMPS to control vapor release. Similar studies optimizing polymer bag thickness as well as absorbent material have been carried out in the creation of odor mimics for synthetic cathinones (i.e., bath salts)10, other narcotics (including heroin and marijuana11), and human odor compounds12,13. In a final example, Simon et al. investigated the active odorants associated with an invasive fungus species14. Whole pieces of infected tree bark, instead of the extracted odorants, were placed directly into the polymer bag to control release during canine olfaction testing14. COMPS can be utilized for a variety of scenarios, and the protocols discussed herein were chosen to demonstrate the diversity of this tool.
1. Assembly of COMPS (Figure 1)
2. Gravimetric analysis to determine COMPS permeation rate
NOTE: A constant ambient temperature is important for accurate and reproducible measurements, both gravimetric and headspace. A constant temperature must be maintained during all testing. It is recommended to carry out all analytical measurements at desired temperatures during testing.
3. Headspace analysis by solid phase microextraction with gas chromatography/mass spectrometry (SPME-GC/MS) (optional)
4. COMPS storage
5. Field olfactory testing
NOTE: Olfactory testing can be carried out in many different ways depending on the animal being tested, the goal of the experiment, and the environmental conditions. The protocol below describes one such manner of testing. All animal testing should first be reviewed and approved by an Institutional Animal Care and Use Committee (IACUC).
The primary objective of using COMPS in olfactory testing/training is to control the release of the chosen odorants and deliver a controlled amount of the odorant over the duration of the test or training session. Odorant release is measured by gravimetric analysis in terms of mass loss per unit time. Figure 2 gives an example of gravimetric results from the permeation of three identical COMPS prepared from 5 µL of pentanoic acid on cotton gauze through a 3 MIL LDPE bag. A line of regression was added to the graph, and the slope of the line represents the permeation rate of 37 µg/min for this set of COMPS.
It is often desirable to be able to adjust the amount of odorant released for a given test. This can be done in several ways, including adjusting the amount of material in the bag, the surface area of the permeable bag material, or the bag thickness. Figure 3 shows how all three of these factors were used to control the release of piperonal. Figure 3A indicates a logarithmic relationship between the mass in piperonal in the permeable bag (3 x 3 inch, 2 MIL LDPE), where the permeation rate increased quickly at the lower masses, then slowed after 500 mg due to physical restriction in the amount of odorant that could be released from the given bag at a time. The data in Figure 3B depict a linear relationship between permeation rate and surface area of the permeable bag for 2 g of piperonal in a 2 MIL LDPE bag. Finally, permeation rate decreased linearly with increased bag thickness (2 g of piperonal in a 3 x 3 inch bag), as shown in Figure 3C, because the thicker bag restricts and slows emission.
In another example of the utility of controlling permeation rates, Simon et al.4 used bag thickness to standardize permeation rates for compounds of varying vapor pressures in order to present canines with similar odorant availability for each analyte during field testing. A volume of 5 μLs of each neat analyte was pipetted onto separate cotton gauze pads and placed into 2 x 3 inch LDPE permeable bags. The permeation rates were measured by gravimetric analysis. Figure 4 shows the variation in vapor pressures (Figure 4A) across the groups of analytes (RSD = 138%) compared to the variation in permeation rate after adjusting the bag thickness (Figure 4B) to control the rate and make them as similar as possible (RSD = 31.8%). Furthermore, adjusting the bag thickness allowed permeation rates to vary by three orders of magnitude (Table 1).
Headspace measurements can be used to better measure the amount of odorant available during a given testing or training scenario. Macias18 measured the amount of piperonal in the headspace of three COMPS with permeation rates of 1,000, 100, and 10 ng/min (Figure 5). The COMPS were placed in a 1 quart sampling can, and the headspace was extracted for 30 min using SPME. The resulting chromatograph in Figure 5 shows the piperonal peak areas increasing with increasing permeation rate18.
Macias then used these three sets of piperonal COMPS in canine trials. Trained narcotics detection canines were tested on the 0 (blank), 10, 100, and 1,000 ng/s piperonal COMPS in a scent cage (Table 2). The results showed that as the permeation rate, and thus odor availability, increased the number of canines alerting to the appropriate COMPS increased18.
Figure 1: Examples of COMPS. (A) A COMPS constructed from a cotton gauze pad in permeable polymer bag. Reproduced from Simon et al.4 (B) A COMPS inserted into an outer impermeable bag. (C) A COMPS containing infected wood as an odor source in a polymer bag. Figures B and C were reproduced with permission from Simon et al.19. Please click here to view a larger version of this figure.
Figure 2: Example measurement of permeation rate by gravimetric analysis. The mass loss of the analyte (i.e., pentanoic acid) on gauze through a 3 MIL LDPE bag measured over time. A, B, and C indicate replicates of the same material, while "Average" is the averaged value of the three at each time point. The given equation describes the linear fit to the average data. Please click here to view a larger version of this figure.
Figure 3: Examples of factors adjusting the permeation rate. Graphs of piperonal dissipation indicate experimentally measured permeation rates when changing (A) the mass of piperonal (3 x 3 inch, 2 MIL LDPE bag), (B) the surface area of the permeable bag (2 g of piperonal, 2 MIL LDPE), and (C) the bag thickness (2 g of piperonal, 3 x 3 inch bag). All error bars represent one standard deviation of the mean (some bars are within the size of the marker). These figures have been reproduced with permission from Macias et al.18. Please click here to view a larger version of this figure.
Figure 4: Comparison of vapor pressure variation across a group of compounds versus permeation rate variation. (A) Vapor pressures for a selection of 12 compounds (RSD = 138%) compared to (B) permeation rates for the same compounds with selected COMPS thicknesses (RSD = 31.8%). Numbers in parentheses represent LDPE bag thickness in MIL. These figures have been reproduced with permission from Simon et al.4. Please click here to view a larger version of this figure.
Figure 5: Headspace analysis of piperonal COMPS at three permeation rates. Overlaid chromatograms of the headspace components of piperonal COMPS adjusted to permeation rates of 1,000, 100, and 10 ng/s. Reproduced with permission from Macias et al.18. Please click here to view a larger version of this figure.
Bag thickness | Permeation rate (mg/min) | R-squared |
1 MIL | 0.108 | 0.974 |
4 MIL | 0.042 | 0.991 |
8 MIL | 0.00499 | 0.99 |
4 MIL in metal bag w/ 1/8" hole | 0.000179 | 0.972 |
Table 1: Permeation rate versus bag thickness. Permeation rate of methyl benzoate on cotton gauze in COMPS of varying bag thicknesses. Note that the lowest permeation rate was achieved by placing a 4 MIL COMPS inside of a metalized bag with a 1/8 in hole. The R2 value indicates the fit of the line to the gravimetric plot.
Piperonal COMPS permeation rate | Number of alerts | % alert |
0 ng/s (blank) | 0 | 0% |
10 ng/s | 4 | 25% |
100 ng/s | 7 | 44% |
1000 ng/s | 12 | 75% |
Table 2: Example of canine field trial results. Canine responses to piperonal COMPS with permeation rates ranging from 0–1,000 ng/s. Reproduced with permission from Macias et al.18.
Controlled Odor Mimic Permeation Systems (COMPS) are easily created by sealing an odorant of interest into a permeable bag. This may be done by pipetting a neat liquid compound onto an absorbent material and then placing the absorbent material into the bag; by placing a pure, solid compound directly into the bag4, as was done in the case of piperonal8; or by placing the target material containing multiple or unknown odorants into a permeable bag, as was done with fungus-infected wood14. The permeable bag controls the release of the odorant so a known and reproducible amount can be delivered over a given training or testing period. Permeation rate is typically measured by gravimetric analysis, plotting mass loss over time, and can be adjusted by altering a number of parameters, including the absorbent material, the mass/volume of odorant, or parameters of the permeation bag (i.e. thickness, surface area, or polymer type). COMPS are stored in an outer nonpermeable envelope, which allows the COMPS to equilibrate prior to use, thus providing a known amount of odorant immediately upon use.
The greater the permeation rate of a COMPS, the greater concentration of odorant available during a training or testing scenario. To quantitate or compare the odor concentration emitted from a COMPS, headspace analysis of the COMPS in the testing/training container should be completed. This is most often done by extracting the odor using SPME with analysis by GC/MS. For quantitation or comparison purposes, it is recommended to use an internal standard and/or an external calibration curve.
COMPS serve as a low-cost, field-amenable devices for controlling the release of an odorant for olfactory training or testing, such as with canine detectors. COMPS can be used repeatedly until depletion, each time delivering the same odorant emission rate, although the length of time the emission rate is constant will change for each analyte and should be tested in the laboratory prior to use. This overcomes a widely recognized limitation of controlling odor delivery for field use and advances olfaction research and detection animal training.
The authors have nothing to disclose.
This work was funded in part by the Office of Naval Research and the National Institute of Justice (2006-DN-BX-K027). The authors wish to thank the many “Furton Group” students that have participated in this project, as well as collaborators from the U.S. Naval Research Laboratory and the Naval Surface Warfare Center (Indian Head EOD Technology Division). Finally, the authors thank Peter Nunez of U.S. K-9 Academy, Tony Guzman of Metro-Dade K9 Services, and Miami-Dade area law enforcement canine teams.
16 oz economy jars (70-450 finish) | Fillmore container | A16-08C-Case 12 | |
7890A gas chromatograph / 5975 mass selective detector | Agilent | ||
Analytical balance | Mettler Toledo | 01-911-005 | |
Ball regualr bands and dome lids | Fillmore container | J30000 | |
Cotton gauze (2" x 2") | Dukal | ||
Disposable weighing boats | VWR | 10803-148 | |
Epoxy-lined sample containers, 1 gallon | TriTech Forensics | CANG-E | |
Epoxy-lined sample containers, 1 pint | TriTech Forensics | CANPT-E | |
Low density polyetheylene bag | Uline | S-5373 | |
Rtx-Volatiles (30 m x 0.32 mmID) column | Restek | 10901 | |
Silver metalized mylar barrier bag (3.5" x 4.5") | ESP Packaging | 95509993779 | |
Silver metalized mylar barrier bag (5" x 8.5" x 3") | ESP Packaging | 95509993793 | |
Solid phase microextration fiber assembly (PDMS/DVB/CAR) | Sigma-Aldrich | 57328-U | |
Solid phase microextration holder | Sigma-Aldrich | 57330-U | |
Tabletop Impulse Sealer | Uline | H-190 | Heat sealer |