Özet

Stimulation Location Determination using a 3D Digitizer with High-Definition Transcranial Direct Current Stimulation

Published: December 20, 2019
doi:

Özet

Presented here is a protocol to achieve higher accuracy in determination of stimulation location combining a 3D digitizer with high-definition transcranial direct current stimulation.

Abstract

The abundance of neuroimaging data and rapid development of machine learning has made it possible to investigate brain activation patterns. However, causal evidence of brain area activation leading to a behavior is often left missing. Transcranial direct current stimulation (tDCS), which can temporarily alter brain cortical excitability and activity, is a noninvasive neurophysiological tool used to study causal relationships in the human brain. High-definition transcranial direct current stimulation (HD-tDCS) is a noninvasive brain stimulation (NIBS) technique that produces a more focal current compared to conventional tDCS. Traditionally, the stimulation location has been roughly determined through the 10-20 EEG system, because determining precise stimulation points can be difficult. This protocol uses a 3D digitizer with HD-tDCS to increase accuracy in determination of stimulation points. The method is demonstrated using a 3D digitizer for more accurate localization of stimulation points in the right temporo-parietal junction (rTPJ).

Introduction

Transcranial direct current stimulation (tDCS) is a noninvasive technique that modulates cortical excitability with weak direct currents over the scalp. It aims to establish causality between neural excitability and behavior in healthy humans1,2,3. In addition, as a motor neurorehabilitation tool, tDCS is widely used in the treatment of Parkinson’s disease, stroke, and cerebral palsy4. Existing evidence suggests that traditional pad-based tDCS produces current flow through a relatively larger brain region5,6,7. High-definition transcranial direct current stimulation (HD-tDCS), with the center ring electrode sitting over a target cortical region surrounded by four return electrodes8,9, increases focality by circumscribing four ring areas5,10. In addition, changes in excitability of the brain induced by HD-tDCS have significantly larger magnitudes and longer durations than those generated by traditional tDCS7,11. Therefore, HD-tDCS is widely used in research7,11.

Noninvasive brain stimulation (NIBS) requires specialized methods to ensure that a stimulation site is present in the standard MNI and Talairach systems12. Neuronavigation is a technique that allows for mapping interactions between transcranial stimuli and the human brain. Its visualization and 3D image data are used for precise stimulation. In both tDCS and HD-tDCS, a common assessment of stimulation sites on the scalp is typically the EEG 10-20 system13,14. This measurement is widely used for placing the tDCS pads and optode holders for functional near infrared spectroscopy (fNIRS) in the initial stage13,14,15.

Determining the precise stimulation points when using the 10-20 system can be difficult (e.g., in the temporo-parietal junction [TPJ]). The best way to solve this is to obtain structural images from participants using magnetic resonance imaging (MRI), then obtain the exact probe position by matching target points to their structural images using digitizing products15. MRI provides good spatial resolution but is expensive to use15,16,17. Moreover, some participants (e.g., those with metal implants, claustrophobic people, pregnant women, etc.) cannot be subjected to MRI scanners. Therefore, there is a strong need for a convenient and efficient way to overcome the abovementioned limitations and increase accuracy in determining stimulation points.

This protocol uses a 3D digitizer to overcome these limitations. Compared to MRI, key advantages of a 3D digitizer are low costs, simple application, and portability. It combines five reference points (i.e., Cz, Fpz, Oz, left preauricular point, and right preauricular point) of individuals with location information of the target stimulation points. Then, it produces a 3D position of electrodes on the subject’s head and estimates their cortical positions by fitting with the vast data from the structural image12,15. This probabilistic registration method enables the presentation of transcranial mapping data in the MNI coordinate system without recording a subject’s magnetic resonance images. The approach generates anatomical automatic labels and Brodmann areas11.

The 3D digitizer, used to mark space coordinates based on the data from structural images, was first used to determine the position of optodes in fNIRS research18. For those who use HD-tDCS, a 3D digitizer breaks the finite stimulation points of the EEG 10-20 system. The distance of the four return electrodes and center electrode is flexible and can be adjusted as needed. When using the 3D digitizer with this protocol, the coordinates of the rTPJ were obtained, which is beyond the 10-20 system. Also shown are the procedures for targeting and stimulating the right temporo-parietal junction (rTPJ) of the human brain.

Protocol

The protocol meets the guidelines of the Institutional Review Board of Southwest University.

1. Determination of Stimulation Location

  1. Review the literature and confirm the stimulation location (here, the rTPJ)19,20,21.

2. Preparation of Electrode Holding Cap

NOTE: The following steps are shown in Figure 1.

  1. Ensure that all necessary materials are readily available: the 3D digitizer (Figure 2), standard measurement tape, a marking pen, the headform, and a swimming cap.
  2. Put the cap on the headform and mark the points on the cap.
    1. Localize the Vertex (Cz). To do so, first mark the midpoint of the distance between the nasion and inion using a skin marker13,14,22. Then, measure the distance between the pre-auricular points and mark the midpoint. The point at which both points intersect is the Cz.
    2. Check the location of the center electrode and the return electrodes. Here, the stimulation was applied on rTPJ. The rTPJ roughly corresponds to the midpoint between CP6 and P6 in the 10-10 EEG system19,20,21.
    3. Find CP6 and P622,23,24,25. According to the proportional requirements of the 10-10 system, locate the approximate location of the rTPJ on the scalp and mark it on the cap.
    4. Adjust the radius of the four return electrodes based on the objectives11,14,26. After this decision, mark the center electrode and return electrode locations on the cap.

3. 3D Digitizer Measurement

  1. Scan with the metal scanner to ensure that the environment for 3D digitizer is metal-free.
  2. Placement of the cap on the subject's head
    1. Make sure that the references (Cz, Fpz, Oz, left preauricular point, and right preauricular point) on the cap align with the international 10-10 system for scalp location22. For example, localize the Vertex (Cz) on the scalp and place the cap onto the subject's head, aligning the cap's Cz to the subjects.
  3. Arranging the 3D digitizer equipment
    1. Connect the 3D digitizer to the computer using the Universal Serial Bus (USB) interface and make sure that the digitizer software is available and ready27.
    2. Put the source in front of the subject and fasten the elastic rope of the sensor around the head. Importantly, make sure that neither the source nor sensor moves during 3D digitizer measurement.
      NOTE: The source is a magnetic transmitter that emits an electromagnetic dipole field. The sensor is a receiver that detects the field.
    3. Open the digitizer software on the computer and make sure that the 3D digitizer system communicates with the software.
    4. Test the accuracy of the stylus. Find a length of 10 cm on the ruler and record the zero graduation and ten graduation, respectively, using the stylus.
      NOTE: The measurement distance between the two recording points of the 3D digitizer should be captured. Compare the error with the reading from the 3D tracker.
    5. Select the YENİ icon and create a new subject file. Select the Sessions box, then Reference.
      NOTE: Using the 3D digitizer stylus, the reference position data (Cz, inion, nasion, left ear, right ear) of the subject are collected according to the software prompts.
    6. To cater to the requirement of fNIRS experiments, use the Transmitter, Detector, and Channel options. Collect the position data of the center electrode and four return electrodes 3x for the transmitter, detector, and channel, in order to reduce error. Ensure that five electrodes are numbered and localize in turn.
    7. Save the three files that are generated.

4. Data Conversion and Spatial Registration

  1. Select the three files into the NIRS-SPM to achieve the real coordinates registration into MNI space28. Affine transform the reference points and five electrode points in participants to the corresponding points in each entry according to the MRI database in MNI space.
  2. Register the data to anatomical automatic labels and Brodmann areas and register the spatial information of the five electrode points to both of these.
  3. Compare the coordinates of stimulation in previous research with the obtained coordinates20,29.
  4. Make a small cut aligned to the five points marked on the cap, such that the plastic casing is embedded snugly into the cap.

5. Stimulation

  1. Make sure that participant have no contraindications (i.e., history of neurological or psychiatric disorders) for HD-tDCS1,3 and that they provided written informed consent prior to the study (including HD-tDCS stimulation).
  2. For device installation, ensure that all the necessary materials are available (Figure 3). Install the device as detailed in published literature14. A brief description is provided below.
    1. Install the batteries and check that they are charged.
    2. Connect the conventional tDCS and 4×1 Stimulation Adapter.
    3. Connect the cables of five Ag/AgCI sintered ring electrodes to the matching receivers on the 4×1 adapter output cable.
    4. Check that all materials are connected correctly.
  3. Measure the head of the participant and place the cap on the head.
    1. Embed the five plastic HD casings in the swimming cap.
    2. Localize the Cz, Fpz, and Oz of the subject13,14. Adjust the reference on the cap to align with the international 10-10 system for scalp locations22. Once the cap is in position, ensure that it does not move.
    3. Collect the position data of the stimulated brain areas using the 3D digitizer. Make the corresponding adjustments according to the generated data.
  4. Cover the scalp surface with electrically conductive gel. First, carefully separate the hair through the opening of the plastic casing using the end of a plastic syringe, until the scalp is exposed. Then, cover the exposed scalp with the electrically conductive gel through the plastic casing opening on the scalp surface.
  5. Set the parameters of the tDCS device: quality value, stimulus duration, intensity, and condition setting.
    1. Turn on the 4×1 Multichannel Stimulation Adapter.
    2. Ensure that the default setting is SCAN, which shows the impedance of one electrode at a time in the display window by scanning the electrodes14,30,31. Here, the impedance is described as "quality value". Values below 1.5 indicate sufficient quality14,30,31. In this case, the values were lower than 1.
      NOTE: If the impedance value exceeds these required limits, open the cap of the plastic casing with high impedance and adjust the hair and electrode to obtain the desired impedance value.
    3. Press the "MODE SELECT" button and switch from "SCAN" to "PASS", after the impedance values are acceptable.
    4. Select the center-anode or center-cathode by pressing the "POLARITY" button. "CENTRAL ANODE" is the default setting.
    5. Adjust the settings on the conventional tDCS device to include stimulus duration (min), intensity (mA) and sham condition setting. In this case, anodal active stimulation was 1.5 mA, and the stimulus lasted 20 min. Next, push the "RELAX" lever to switch to full current.
    6. Once everything is set, initiate the stimulation. Press the "START" button, and the DC intensity will ramp up until the target current is reached. The timer will then show the remaining time.
      NOTE: Some participants may feel uncomfortable during periods of increased DC intensity. In such cases, the current may be moderately decreased slightly for a few seconds by pulling down the "RELAX" lever. Then, push the dolly bar to full current, gradually, when participants feel comfortable again.

6. Post-stimulation

  1. When the stimulation is over, turn the lever slowly to adjust the current to zero before turning off the power. Otherwise, participants may perceive stinging sensation or dizziness when turning off the power directly.
  2. After the stimulation, open the plastic cap and remove the Ag/AgCI sintered ring electrodes from the casing.
  3. Remove the swimming cap and clean the materials. Provide participants with tools to clean their hair.
  4. Ask participants to fill out a questionnaire after each stimulation session, if necessary (e.g., to measure adverse effects of screening following HD-tDCS, participant tolerance to brain stimulation, etc.; see Supplementary File).

Representative Results

Using the methods presented, coordinates of the rTPJ were determined, which requires stimulation points beyond the 10-20 system. First, the circumference of the headform should be similar to the actual head. Here, the length of the nasion to inion of the headform was ~36 cm, and the length between the bilateral preauricular was ~37 cm.

The steps for producing the electrode cap guide the measuring positions of the 10-20 system. Here, Nz, Iz, Cz, Fpz, Oz, Pz, T8, T7, C4, P8, O2, P4, C6, P6, and CP6 were determined. The approximate location of the RTPJ (about the midpoint between CP6 and P6) was found on the scalp. The distance between the central and peripheral electrodes should be adjusted based on experimental objectives. Previous research obtained radius values ranging from 3.5–7.5 cm11,14,30. With different radius values, DC intensity and stimulation duration may generate different electric field strengths. In this protocol, the distance between all return electrodes and the central active electrode were fixed to 3.5 cm.

Several important reference points on the swimming cap were kept, including Fpz, Cz, Oz, T8, and C4. The Vertex on the scalp was located before the stimulation, and it is critical that the Cz point on the cap exactly aligns with the Vertex. Once the cap is in position, the cap should not move. One .mat file and two .csv files after digitizing were obtained (i.e., sub01_origin.csv, which included the coordinate information of the reference [with subject number 01]), while sub01_others.csv included the coordinate information of the five targeted points [with subject number 01)].

Three .txt files were obtained after data conversion and spatial registration. In digitizer software, there are transmitter, detector (receiver), and channel options for fulfilling the requirements of fNIRS experiments. The coordinate data of the transmitter, detector, or channel should be the same. However, small operating errors may occur, because of laboratory personnel skills, pen holding gesture, etc.

Using the NIRS-SPM stand-alone registration function, the spatial registration function generates MNI coordinates. The numbers in the first line in Table 1 represent the order in the digitizer. In this protocol, the data from number five is the position information about the center electrode. In Brodmann areas (BA), the anatomical label and its number were obtained. The number after each line indicates the percentage of overlap. In anatomical automatic labels (AAL), the anatomical label and percentage of overlap were obtained. To reduce measurement errors, the average value of three data points from the five electrodes’ final MNI coordinates were calculated. As for AAL and BA, the value represents a percentage of overlap with the cerebral cortex. All possibilities were combined into final data (Table 1).

According to the data from MNI coordinates, AAL, and BA, if the difference between the value and target value is too large, the swimming cap must be adjusted to the relative position of the actual values of X, Y, Z, and the target value, as explained in sections 2–411,14,30,31.

Figure 1
Figure 1: Steps for creating the holding electrode cap. Please click here to view a larger version of this figure.

Figure 2
Figure 2: 3D digitizer. The 3D digitizer is a cost-effective solution for 3D digitizing. It is a dual sensor motion tracker. The source is a magnetic transmitter that emits an electromagnetic dipole field. The sensor is a receiver that detects the field. The stylus allows accurate pinpointing of X, Y, and Z data points. The control box connects to the computer and transfer data. Please click here to view a larger version of this figure.

Figure 3
Figure 3: Necessary materials for stimulation. These materials include a tDCS device, 4×1 Multichannel Stimulation Adapter, four 9 V batteries, five Ag/AgCI sodium ring electrodes, five HD plastic casings and their respective caps, electrically conductive gel, a syringe, a standard tape measure, and a swimming cap. Please click here to view a larger version of this figure.

<!–
1 2 3 4 5
MNI X Y Z X Y Z X Y Z X Y Z X Y Z
Channel 43 -89 13 46 -64 54 71 -29 25 64 -56 -16 60 -66 24
Transmit 42 -89 18 42 -67 55 71 -32 27 64 -57 -16 60 -66 24
Receiver 43 -89 16 45 -67 54 71 -31 27 65 -58 -12 58 -69 22
Mean 42.7 -89 15.7 44.3 -66 54.3 71 -30.7 26.3 64.3 -57 -14.7 59.3 -67 23.3
BA Channel 18 – Visual Association Cortex (V2), 0.27823 7-Somatosensory Association Cortex, 0.27876 2 –Primary Somatosensory Cortex, 0.41667 20 – Inferior Temporal gyrus, 0.089606 21 – Middle Temporal gyrus, 0.0072464
19 – V3, 0.72177 39 – Angular gyrus, part of Wernicke's area, 0.53982 22 – Superior Temporal Gyrus, 0.28086 37 – Fusiform gyrus, 0.91039 22 – Superior Temporal Gyrus, 0.17391
40 – Supramarginal gyrus part of Wernicke's area, 0.18142 40 – Supramarginal gyrus part of Wernicke's area, 0.19136 37 – Fusiform gyrus, 0.07971
48 – Retrosubicular area, 0.11111 39 – Angular gyrus, part of Wernicke's area, 0.73913
Transmit 18 – Visual Association Cortex (V2), 0.15936 7 – Somatosensory Association Cortex, 0.57466 2 – Primary Somatosensory Cortex, 0.38871 20 – Inferior Temporal gyrus, 0.035842 21 – Middle Temporal gyrus, 0.0072464
19 – V3, 0.84064 39 – Angular gyrus, part of Wernicke's area, 0.34389 22 – Superior Temporal Gyrus, 0.15674 37 – Fusiform gyrus, 0.96416 22 – Superior Temporal Gyrus, 0.17391
40 – Supramarginal gyrus part of Wernicke's area, 0.081448 40 – Supramarginal gyrus part of Wernicke's area, 0.31034 37 – Fusiform gyrus, 0.07971
48 – Retrosubicular area, 0.1442 39 – Angular gyrus, part of Wernicke's area, 0.73913
Receiver 18 – Visual Association Cortex (V2), 0.21514 7 – Somatosensory Association Cortex, 0.42601 2 – Primary Somatosensory Cortex, 0.44025 20 – Inferior Temporal gyrus, 0.0071429 19 – V3, 0.0036101
19 – V3, 0.78486 39 – Angular gyrus, part of Wernicke's area, 0.51121 22 – Superior Temporal Gyrus, 0.14151 37 – Fusiform gyrus, 0.99286 22 – Superior Temporal Gyrus, 0.054152
40 – Supramarginal gyrus part of Wernicke's area, 0.06278 40 – Supramarginal gyrus part of Wernicke's area, 0.28302 37 – Fusiform gyrus, 0.12274
48 – Retrosubicular area, 0.13522 39 – Angular gyrus, part of Wernicke's area, 0.81949
AAL Channel Occipital_Mid_R, 1 Parietal_Sup_R, 0.030973 SupraMarginal_R, 0.65741 Temporal_Mid_R, 0.039427 Occipital_Mid_R, 0.13406
Parietal_Inf_R, 0.31416 Angular_R, 0.65487 Temporal_Sup_R, 0.34259 Temporal_Inf_R, 0.93907 Angular_R, 0.33696
Cerebelum_Crus1_R,0.021505 Temporal_Sup_R,0.032609
Temporal_Mid_R, 0.49638
Transmit Occipital_Mid_R, 1 Parietal_Sup_R, 0.20814 SupraMarginal_R, 0.74922 Temporal_Mid_R, 0.032258 Occipital_Mid_R, 0.13406
Parietal_Inf_R, 0.20362 Temporal_Sup_R, 0.25078 Temporal_Inf_R, 0.94265 Angular_R, 0.33696
Angular_R, 0.58824 Cerebelum_Crus1_R, 0.02509 Temporal_Sup_R,0.032609
Temporal_Mid_R, 0.49638
Receiver Occipital_Mid_R, 1 Parietal_Sup_R, 0.044843 SupraMarginal_R, 0.7673 Temporal_Mid_R, 0.11429 Occipital_Mid_R, 0.22022
Parietal_Inf_R, 0.20179 Temporal_Sup_R, 0.2327 Temporal_Inf_R, 0.88571 Angular_R, 0.15523
Angular_R, 0.75336 Temporal_Mid_R, 0.62455
–>

Table 1: Localization of stimulations in the brain area. Please click here to view this table (Right click to download).

Supplementary File. Please click here to view this file (Right click to download).

Discussion

Compared to traditional tDCS, HD-tDCS increases the focality of stimulation. Typical sites of stimulation are often based on the 10-20 EEG system. However, determining the precise stimulation points beyond this system can be difficult. This paper combines a 3D digitizer with HD-tDCS to determine stimulation points beyond the 10-20 system. It is important to clearly define the steps and precautions for making and using the electrode cap in such cases.

In general, the position of target stimulation areas is derived from the results of previous brain imaging studies, and the position of the stimulation areas on 10-20 international system or MNI coordinates can be obtained. The steps for creating the electrode cap guide for measuring positions of the 10-20 system are critical. It is key that the reference on the cap aligns with the international 10-20 system for scalp locations when placing the cap on the head. Once the 3D digitizer starts running, the source and sensor should not move, or it will cause data deviation.

In the software, the reference points are on the scalp and not on the cap, unless all the reference points of scalp and cap are matching. If the error between the measured results and target values is out of the acceptable range, the position of the marked points should be slightly adjusted. After adjustment, measurements should then be made again. Once users press the “MODE SELECT” button and switch from “SCAN” to “PASS”, the current will start passing from the conventional tDCS device through the electrodes into the 4×1 Multichannel Stimulation Adapter.

The modular electroencephalogram recording cap provides fixed positions of probes. However, determining the precise stimulation points beyond this system can be difficult. The positions of electrodes beyond the 10-20 system can be determined using the protocol described, as well as the coordinates of stimulation points. The radius setting should be based on the experimental objectives. Using the method described here, the radius of the four return electrodes and center electrode can be flexibly adjusted.

There are many digitizer software packages (e.g., the Brainstorm software for an fNIRS task; here, the Vpen software was used)15. Different data collection software packages emphasize different functions and should be selected according to the research question. Head circumference varies among individuals; hence, using the same cap can produce errors. However, the modular electroencephalogram recording cap also suffers from this problem.

Açıklamalar

The authors have nothing to disclose.

Acknowledgements

This study was supported by the National Natural Science Foundation of China (31972906), Entrepreneurship and Innovation Program for Chongqing Overseas Returned Scholars (cx2017049), Fundamental Research Funds for Central Universities (SWU1809003), Open Research Fund of the Key Laboratory of Mental Health, Institute of Psychology, Chinese Academy of Sciences (KLMH2019K05), Research Innovation Projects of Graduate Student in Chongqing (CYS19117), and the Research Program Funds of the Collaborative Innovation Center of Assessment toward Basic Education Quality at Beijing Normal University (2016-06-014-BZK01, SCSM-2016A2-15003, and JCXQ-C-LA-1). We would like to thank Prof. Ofir Turel for his suggestions on the early draft of this manuscript.

Materials

1X1 Low Intensity transcranial DC Stimulator Soterix Medical 1300A
3-dimensional Polhemus-Patriot Digitizer POLHEMUS 1A0453-001 PATRIOT system component
4X1 Multi-Channel Stimulation Interface Soterix Medical 4X1-C3
Dell desktop computer Dell CRFC4J2 Master computer to run 3D digitizer application

Referanslar

  1. Nitsche, M. A., Paulus, W. Excitability changes induced in the human motor cortex by weak transcranial direct current stimulation. Journal of Physiology. 527, 633-639 (2000).
  2. Sellaro, R., Nitsche, M. A., Colzato, L. S. The stimulated social brain: effects of transcranial direct current stimulation on social cognition. Annals of the New York Academy of Sciences. 1369 (1), 218-239 (2016).
  3. Chen, W., et al. Sex-based differences in right dorsolateral prefrontal cortex roles in fairness norm compliance. Behavioural Brain Research. 361, 104-112 (2019).
  4. Sánchez-Kuhn, A., Pérez-Fernández, C., Cánovas, R., Flores, P., Sánchez-Santed, F. Transcranial direct current stimulation as a motor neurorehabilitation tool: an empirical review. BioMedical Engineering Online. 16 (1), 76 (2017).
  5. Dmochowski, J. P., Datta, A., Bikson, M., Su, Y., Parra, L. C. Optimized multi-electrode stimulation increases focality and intensity at target. Journal of Neural Engineering. 8 (4), 046011 (2011).
  6. Seo, H., Kim, H. I., Jun, S. C. The Effect of a Transcranial Channel as a Skull/Brain Interface in High-Definition Transcranial Direct Current Stimulation-A Computational Study. Science Report. 7, 40612 (2017).
  7. Datta, A., et al. Gyri -precise head model of transcranial DC stimulation: Improved spatial focality using a ring electrode versus conventional rectangular pad. Brain Stimulation. 2, 201-207 (2009).
  8. Turski, C. A., et al. Extended Multiple-Field High-Definition transcranial direct current stimulation (HD-tDCS) is well tolerated and safe in healthy adults. Restorative Neurology and Neuroscience. 35 (6), 631-642 (2017).
  9. Datta, A., Elwassif, M., Battaglia, F., Bikson, M. Transcranial current stimulation focality using disc and ring electrode configurations: FEM analysis. Journal of Neural Engineering. 5 (2), 163-174 (2008).
  10. Edwards, D., et al. Physiological and modeling evidence for focal transcranial electrical brain stimulation in humans: a basis for high definition tDCS. Neuroimage. 74, 266-275 (2013).
  11. Kuo, H. I., et al. Comparing cortical plasticity induced by conventional and high-definition 4 x 1 ring tDCS: a neurophysiological study. Brain Stimulation. 6 (4), 644-648 (2013).
  12. Singh, A. K., Okamoto, M., Dan, H., Jurcak, V., Dan, I. Spatial registration of multichannel multi-subject fNIRS data to MNI space without MRI. Neuroimage. 27 (4), 842-851 (2005).
  13. DaSilva, A. F., Volz, M. S., Bikson, M., Fregni, F. Electrode positioning and montage in transcranial direct current stimulation. Journal of Visualized Experiments. (51), (2011).
  14. Villamar, M. F., et al. Technique and considerations in the use of 4×1 ring high-definition transcranial direct current stimulation (HD-tDCS). Journal of Visualized Experiments. (77), e50309 (2013).
  15. Jasinska, K. K., Guei, S. Neuroimaging Field Methods Using Functional Near Infrared Spectroscopy (NIRS) Neuroimaging to Study Global Child Development: Rural Sub-Saharan Africa. Journal of Visualized Experiments. (132), (2018).
  16. Logothetis, N. K. What we can do and what we cannot do with fMRI. Nature. 453 (7197), 869-878 (2008).
  17. Glover, G. H. Overview of functional magnetic resonance imaging. Neurosurgery Clinics of North America. 22 (2), 133-139 (2011).
  18. Zhu, H. . The easy and stable marking method for registering fNIRS data to MNI space by using 10-20 system. , (2012).
  19. Young, L., Saxe, R. An fMRI Investigation of Spontaneous Mental State Inference for Moral Judgment. Journal of Cognitive Neuroscience. 21, 1396-1405 (2009).
  20. Young, L., Cushman, F., Hause, M., Saxe, R. The neural basis of the interaction between theory of mind and moral judgment. Proceedings of the National Academy of Sciences USA. 104, 8235-8240 (2007).
  21. Jurcak, V., Tsuzuki, D., Dan, I. 10/20, 10/10, and 10/5 systems revisited: their validity as relative head-surface-based positioning systems. Neuroimage. 34 (4), 1600-1611 (2007).
  22. Schestatsky, P., Morales-Quezada, L., Fregni, F. Simultaneous EEG monitoring during transcranial direct current stimulation. Journal of Visualized Experiments. (76), (2013).
  23. Klem, G. H., Lüders, H. O., Jasper, H. H., Elger, C. The ten-twenty electrode system of the International Federation. The International Federation of Clinical Neurophysiology. Cleveland Clinic Foundation. Electroencephalography & Clinical Neurophysiology Supplement. 52, 3 (1999).
  24. Society, A. E. Guideline thirteen: Guidelines for standard electrode position nomenclature. Journal of Clinical Neurophysiology. 1, 111-113 (1994).
  25. Oostenveld, R., Praamstrac, P. The five percent electrode system for high-resolution EEG and ERP measurements. Clinical Neurophysiology. 112, 713-719 (2001).
  26. Saturnino, G. B., Antunes, A., Thielscher, A. On the importance of electrode parameters for shaping electric field patterns generated by tDCS. Neuroimage. 120, 25-35 (2015).
  27. . L. Real-time Recording System of Visual Head 3D Positioning Information (VPen). China patent. , (2014).
  28. Ye, J. C., Tak, S., Jang, K. E., Jung, J., Jang, J. NIRS-SPM: Statistical parametric mapping for near-infrared spectroscopy. Neuroimage. 44 (2), 428-447 (2009).
  29. Decety, J., Lamm, C. The role of the right temporoparietal junction in social interaction: how low-level computational processes contribute to meta-cognition. Neuroscientist. 13 (6), 580-593 (2007).
  30. Villamar, M. F., et al. Focal modulation of the primary motor cortex in fibromyalgia using 4×1-ring high-definition transcranial direct current stimulation (HD-tDCS): immediate and delayed analgesic effects of cathodal and anodal stimulation. The Journal of Pain. 14 (4), 371-383 (2013).
  31. Borckardt, J. J., et al. A pilot study of the tolerability and effects of high-definition transcranial direct current stimulation (HD-tDCS) on pain perception. The Journal of Pain. 13 (2), 112-120 (2012).

Play Video

Bu Makaleden Alıntı Yapın
Chen, W., Chen, R., He, Q. Stimulation Location Determination using a 3D Digitizer with High-Definition Transcranial Direct Current Stimulation. J. Vis. Exp. (154), e60263, doi:10.3791/60263 (2019).

View Video