Summary

Potentiodynamische Korrosionsprüfung

Published: September 04, 2016
doi:

Summary

Here, we present a protocol to set up and run an in vitro potentiodynamic corrosion system to analyze pitting corrosion for small metallic medical devices.

Abstract

Different metallic materials have different polarization characteristics as dictated by the open circuit potential, breakdown potential, and passivation potential of the material. The detection of these electrochemical parameters identifies the corrosion factors of a material. A reliable and well-functioning corrosion system is required to achieve this.

Corrosion of the samples was achieved via a potentiodynamic polarization technique employing a three-electrode configuration, consisting of reference, counter, and working electrodes. Prior to commencement a baseline potential is obtained. Following the stabilization of the corrosion potential (Ecorr), the applied potential is ramped at a slow rate in the positive direction relative to the reference electrode. The working electrode was a stainless steel screw. The reference electrode was a standard Ag/AgCl. The counter electrode used was a platinum mesh. Having a reliable and well-functioning in vitro corrosion system to test biomaterials provides an in-expensive technique that allows for the systematic characterization of the material by determining the breakdown potential, to further understand the material’s response to corrosion. The goal of the protocol is to set up and run an in vitro potentiodynamic corrosion system to analyze pitting corrosion for small metallic medical devices.

Introduction

Elektrochemische Techniken bieten eine schnelle und relativ kostengünstige Verfahren die elektrochemische Eigenschaften eines Materials zu erhalten. Diese Techniken basieren hauptsächlich auf der Fähigkeit , durch Beobachten der Reaktion des Ladungstransferprozesses auf eine kontrollierte elektrochemische Störung 1-5 Korrosion eines Metalls zu erfassen. Die Korrosion von Metallimplantaten in einem Körperumgebung ist entscheidend aufgrund der negativen Auswirkungen auf die Biokompatibilität und Materialintegrität 6. Der Hauptfaktor , um die Korrosion von Implantaten innerhalb des Körpers beiträgt , ist die Auflösung der Oxidoberfläche zu einer erhöhten Freisetzung von Metallionen führen 7-11. Dies führt zu unerwünschten biologischen Reaktionen, die lokal gefunden werden kann, jedoch mit potentiell systemische Wirkung zum vorzeitigen Versagen des Implantats führen 10,12-28.

Die Korrosionseigenschaften eines Prüflings aus dem Polarisations Scan vorhergesagt erzeugtdurch einen Potentiostaten. Ein Polarisations Scan ermöglicht die Extrapolation der kinetischen und Korrosionsparameter eines Metallsubstrats. Während einer Abtastung, die Oxidation oder Reduktion einer elektroaktiven Spezies kann durch Ladungsübertragung und der Bewegung der Reaktanten oder Produkte beschränkt. Diese Faktoren sind alle von der Polarisations Scan verkapselt; daher wichtig, ein System zu haben, das eine zuverlässige und wiederholbare Polarisations Abtastung über mehrere Zyklen erzeugt, ist von großer Bedeutung. Der Schwerpunkt dieses Manuskripts ist ein Protokoll zur Verfügung, aus denen die Grundlagen und Schritte, um eine gut funktionierende potentiodynamischer Korrosionssystem zu erhalten genommen.

Protocol

1. Konstruktion des Probenhalters Konstruieren Sie den Probenhalter aus Edelstahl Abstandhalter und einem M3 Edelstahl Gewindeschraube, anstelle mit einem M3 Sechskantmutter gehalten. Entfernen Sie den Kopf der Gewindeschraube mit einer Zange und den Schnitt Segment polieren den Faden Muster zu halten. Wenn alle einzelnen Komponenten bereit sind, montieren die Elektrodenhalter. Jeder Elektrodenhalter enthält drei Abstandshalter miteinander verbunden durch die M3-Schrauben, was zu einer 11…

Representative Results

Am Ende des Verfahrens ein in vitro – Korrosionssystem ist Setup Korrosionsuntersuchungen durchzuführen. Spezifische Verfahren, wie beispielsweise die Reinigung des Korrosionsbehälter und dem Faraday-Käfig wurden in das Protokoll eingeführt Rauschleistung zu verbessern. Das grundlegende Konzept eines guten Polarisations Scan ist die elektrophysikalischen Bedingungen des Materials wertvolle Informationen zu identifizieren, um die Korrosionsanfälligkeit aus einem metallischen…

Discussion

Polarisations – Scans von den Proben aus rostfreiem Stahl hergestellt zeigte saubere kontinuierliche Plots mit Scans Korrelation in der Literatur gesehen, das einen gut funktionierenden Korrosionssystem , das sowohl zuverlässig und reproduzierbar 29 ist. Schlechte Reproduzierbarkeit von potentiodynamischen Pitting Potentialen mit einer Ausbreitung von einigen hundert Millivolt identifiziert, mit 29 durch einen stochastischen Prozess gekennzeichnet Potential liegen Grübchenbildung. Dies ist häufi…

Disclosures

The authors have nothing to disclose.

Acknowledgements

The authors had no funding provided for this study.

Materials

Potentiostat Metrohm PGSTAT101
Ag/AgCl reference electrode, shielded Metrohm 6.0729.100
Electrode shaft Metrohm 6.1241.060
Polisher Forcipol 1v Metkon 3602
Clindrical flask 700mL SciLabware FR700F
Reaction lid SciLabware MAF2/41
Dichloromethane Sigma-Aldrich MKBR7629V use under a fumehood. Wear protective clothing
Thermo / HAAKE D Series Immersion Circulators Haake

References

  1. Isaacs, H. S. Aspects of corrosion from the ECS Publications. J. Electrochem. Soc. 149 (12), 85-87 (2002).
  2. Fontana, M. G., Greene, N. D. . Corrosion Engineering. , (1978).
  3. Pourbaix, M. Electrochemical corrosion of metallic biomaterials. Biomaterials. 5 (3), 122-134 (1984).
  4. Rechnitz, G. A. . Controlled-Potential Analysis. , (1963).
  5. Silverman, D. C., Revie, R. W. Chapter 68. Uhlig’s Corrosion Handbook. , (2000).
  6. Gurappa, I. Characterization of different materials for corrosion resistance under simulated body fluid conditions. Mater Charact. 49 (1), 73-79 (2002).
  7. Antoniou, J., et al. Metal ion levels in the blood of patients after hip resurfacing: a comparison between twenty-eight and thirty-six-millimeter-head metal-on-metal prostheses. J Bone Joint Surg Am. 90, 142-148 (2008).
  8. Billi, F., Campbell, P. Nanotoxicology of metal wear particles in total joint arthroplasty: a review of current concepts. J Appl Biomater Funct Mater. 8 (1), 1-6 (2010).
  9. Bradberry, S. M., Wilkinson, J. M., Ferner, R. E. Systemic toxicity related to metal hip prostheses. Clin Toxicol (Phila). 52 (8), 837-847 (2014).
  10. Davda, K., Lali, F. V., Sampson, B., Skinner, J. A., Hart, A. J. An analysis of metal ion levels in the joint fluid of symptomatic patients with metal-on-metal hip replacements. J Bone Joint Surg Br. 93 (6), 738-745 (2011).
  11. Clarke, M. T., Lee, P. T., Arora, A., Villar, R. N. Levels of metal ions after small and large diameter metal-on-metal hip arthroplasty. J Bone Joint Surg Br. 85 (6), 913-917 (2003).
  12. Brown, S. A., Hughes, P. J., Merritt, K. In vitro studies of fretting corrosion of orthopaedic materials. J Orthop Res. 6 (4), 572-579 (1988).
  13. Bryant, M., et al. Characterisation of the surface topography, tomography and chemistry of fretting corrosion product found on retrieved polished femoral stems. J Mech Behav Biomed Mater. 32, 321-334 (2014).
  14. Jantzen, C., Jørgensen, H. L., Duus, B. R., Sporring, S. L., Lauritzen, J. B. Chromium and cobalt ion concentrations in blood and serum following various types of metal-on-metal hip arthroplasties. A literature review. Acta Orthopaedica. 84 (3), 229-236 (2013).
  15. Campbell, P., et al. Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips. Clin. Orthop. Relat. Res. 468 (9), 2321-2327 (2010).
  16. Cook, S. D., et al. The in vivo performance of 250 internal fixation devices: a follow-up study. Biomaterials. 8 (3), 177-184 (1987).
  17. Cooper, H. J., Urban, R. M., Wixson, R. L., Meneghini, R. M., Jacobs, J. J. Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Joint Surg Am. 95 (10), 865-872 (2013).
  18. Langton, D. J., Sprowson, A. P., Joyce, T. J., Reed, M., Carluke, I., Partington, P., Nargol, A. V. Blood metal ion concentrations after hip resurfacing arthroplasty. J Bone Joint Surg Br. 91 (10), 1287-1295 (2009).
  19. Langton, D. J., Jameson, S. S., Joyce, T. J., Webb, J., Nargol, A. V. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 90 (9), 1143-1151 (2008).
  20. Daniel, J., Ziaee, H., Pradhan, C., McMinn, D. J. Six-year results of a prospective study of metal ion levels in young patients with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 91 (2), 176-179 (2009).
  21. De Haan, R., et al. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 90 (10), 1291-1297 (2008).
  22. Dijkman, M. A., de Vries, I., Mulder-Spijkerboer, H., Meulenbelt, J. Cobalt poisoning due to metal-on-metal hip implants. Ned Tijdschr Geneeskd. 156 (42), A4983 (2012).
  23. Fisher, J. Bioengineering reasons for the failure of metal-on-metal hip prostheses: an engineer’s perspective. J Bone Joint Surg Br. 93 (8), 1001-1004 (2011).
  24. Goldberg, J. R., et al. A Multicenter Retrieval Study of the Taper Interfaces of Modular Hip Prostheses. Clin. Orthop. Relat. Res. (401), 149-161 (2002).
  25. Ingham, E., Fisher, J. Biological reactions to wear debris in total joint replacement. Proc Inst Mech Eng H. 214 (1), 21-37 (2000).
  26. Gilbert, J. L., Buckley, C. A., Jacobs, J. J., Res, J. .. B. i. o. m. e. d. .. M. a. t. e. r. .. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion and allot coupling. J. Biomed. Mater. Res. 76 (1), 1533-1544 (1993).
  27. Browne, J. A., Bechtold, C. D., Berry, D. J., Hanssen, A. D., Lewallen, D. G. Failed metal-on-metal hip arthroplasties: a spectrum of clinical presentations and operative findings. Clin. Orthop. Relat. Res. 468 (9), 2313-2320 (2010).
  28. Jantzen, C., Jorgensen, H. L., Duus, B. R., Sporring, S. L., Lauritzen, J. B. Chromium and cobalt ion concentrations in blood and serum following various types of metal-on-metal hip arthroplasties: a literature overview. Acta Orthop. 84 (3), 229-236 (2013).
  29. Frangini, S., De Cristofaro, N. Analysis of galvanostatic polarisation method for determining reliable pitting potentials on stainless steels in crevice-free conditions. Corros Sci. 45 (12), 2769-2786 (2002).

Play Video

Cite This Article
Munir, S., Pelletier, M. H., Walsh, W. R. Potentiodynamic Corrosion Testing. J. Vis. Exp. (115), e54351, doi:10.3791/54351 (2016).

View Video