The Rotating Exercise Quantification System (REQS) can induce exercise in Drosophila melanogaster through rotation while simultaneously measuring the amount of activity performed by the animals. Here, we present a point-by-point protocol detailing how to measure activity levels of animals experiencing rotational exercise treatments using the REQS.
Drosophila melanogaster is a new model organism for studies in exercise biology. To date, two main exercise systems, the Power Tower and the Treadwheel have been described. However, a method to measure the amount of additional animal activity induced through the exercise treatment has been lacking. The Rotating Exercise Quantification System (REQS) fills this need, providing a measure of animal activity for animals experiencing rotational exercise. This protocol details how to use the REQS to assess animal activity during rotational exercise and illustrates the type of data that can be generated. Here, we demonstrate how the REQS is used to measure sex- and strain-specific differences in exercise induced activity. The REQS can also be used to evaluate the impact of various other experimental parameters such as age, diet, or population size on exercise induced activity. In addition, it can be used to compare the efficacy of different exercise training protocols. Importantly, it provides an opportunity to standardize exercise treatments between strains, allowing the researcher to achieve equal amounts of activity between groups if needed. Thus, the REQS is a notable new resource for exercise biologists working with the Drosophila model system and complements existing exercise systems.
Recently, researchers have begun to use the fruit fly Drosophila melanogaster to study exercise biology. D. melanogaster has been a genetic model system for over 100 years1,2. However, Drosophila research has made contribution to not just genetics, but also to a variety of other disciplines including neurobiology, behavioral biology, and physiology3. In 2009, the Power Tower, the first exercise machine for Drosophila was described4. The Power Tower takes advantage of the animals' negative geotaxis response. When disturbed, Drosophila tend to move to the top of their enclosure. This response is well established and is the basis of the popular "RING" (Rapid Iterative Negative Geotaxis5) assay that is used to estimate climbing ability and/or physical fitness in Drosophila. The Power Tower uses a mechanical arm connected to a motor unit to repeatedly lift a set of animals within their enclosures by several inches and dropping them back to the ground to induce the negative geotaxis response (Tinkerhess et al. 20126 provide a video illustrating the use of the Power Tower). Prolonged treatment on the Power Tower thus increases the amount of physical activity (running or flying) the animals perform compared to untreated control animals and over time leads to improved performance in the RING assay for physical fitness4. Thus, this work demonstrated the feasibility of using Drosophila as a model for exercise biology.
To expand the repertoire of tools available for Drosophila exercise research, in 2016, Mendez and colleagues described a second Drosophila exercise machine, the Treadwheel7. Similar to the Power Tower, the Treadwheel exploits the negative geotaxis response of Drosophila. However, this response is induced by continued rotation of the animal enclosures, rather than by lifting and dropping them as in the Power Tower. This induction method is gentler and allows for a more endurance oriented exercise regime that avoids any physical trauma that might occur during exercise in the Power Tower (see Katzenberger, R. J. et al. 20138 for the impact of repeated physical trauma on Drosophila health). Similar to the Power Tower4, exercise treatment of animals on the Treadwheel leads to a variety of physiological responses, including changes in physical fitness, triglyceride levels, and body weight7. Thus, two complementary methods are available for Drosophila biologists studying exercise.
One limitation of both the Power Tower and the Treadwheel is the inability to measure the amount of activity induced by the exercise treatment. Analysis of video-recordings taken from the Treadwheel demonstrated that there were significant differences among the various Drosophila strains in how they respond to the exercise treatment7. Specifically, the strains studied differed in how much additional activity the animals performed when stimulated7. This observation prompted us to develop a third exercise system, the Rotating Exercise Quantification System (REQS), that allows us to measure animal activity levels during rotation-induced exercise9. The REQS utilizes a commercially available activity monitoring unit that is installed on a rotating arm to stimulate exercise through rotation as in the Treadwheel. Initial work with the REQS confirms that genetically different Drosophila strains — and sexes – can have significantly different responses to the rotational stimulation and thus the amount of exercise induced is not identical among different genotypes9. Thus, the REQS now enables Drosophila biologists to measure the amount of exercise induced by the treatment, opening a variety of new research avenues in the exercise field.
Here we describe in detail how to use the REQS for quantification of rotational exercise. The REQS induces rotational exercise and simultaneously measures the activity levels of the animals being treated. The REQS is able to accommodate a variety of exercise programs, ranging from the simple 2 h continuous exercise regime demonstrated here to more complex interval training methods as described by Mendez and colleagues7, and stimulation can be adjusted via rotation speed (between approximately 1-13 rotations per min). Depending on the activity monitor unit used to produce the REQS, this method is adaptable to the analysis of single flies or large populations of animals. Due to this versatility, the REQS provides Drosophila researchers with an array of opportunities to study, for example, different exercise regimes, diet interventions, or impact of population density.
The REQS consists of a Drosophila Activity Monitor unit (for source information, please see Table of Materials) mounted on a rotating arm that is controlled by a motor unit (Figure 1). The activity monitor determines how often in a given time span the array of laser beams dissecting the middle of the assay tube is disrupted. For detailed drawings and an in-depth characterization, see our previous publication9. While our system uses the LAM25H unit, the REQS can be modified to accommodate other Drosophila Activity Monitor units as well.
1. Testing the Setup of the REQS
2. Preparation of Animals
NOTE: All animals were raised and tested under standard conditions in an incubator (25 °C, 60-70% humidity, 12 h light/dark cycle) on molasses/cornmeal media10.
3. Data Collection with the REQS
4. Data Analysis
The output from an individual run with the REQS is a data table produced by the DAMSystem308 software, which will be labeled "Monitor1.txt" by default (for an example see Supplemental File 1). An excerpt from such a table is shown in Table 1. Each column contains the data from an individual assay tube, while the rows contain the activity measured in each time interval from the beginning of the experiment (top) to the end (bottom). The first three data points should be monitored in the event that there are data cutouts (such as those seen in Table 2), in which case the cable must be adjusted.
Figure 2 shows the results from an experiment comparing four genetically distinct Drosophila lines (DGRP2 lines 371, 703, 810, and 89712,13). For each of the four lines, ten replicate measurements from groups of ten virgin flies of both sexes were collected as described in the Protocol section. From the output file of the DAMSystem308 software, "Average activity per 5 min per 10 flies" was calculated by averaging the activity measures across the entire 2-hour time span. This average from each column thus produces a single measure for each assay chamber. The summary table that Figure 2 is based on is provided in Table 3.
The data provided in Table 3 were analyzed by ANOVA, testing for effects of genotype, sex, and the interaction between sex and genotype. As only the effect of genotype was significant (p <2 x 10-16), the two sexes were combined in the graph shown in Figure 2. The ANOVA detected a strong effect of genotype, which is evident in the graph. The mean exercise activity levels between all four genotypes are significantly different from each other (p <0.05; Tukey's HSD), and thus, Figure 2 illustrates how the REQS can be used to detect an impact of genotype on exercise activity levels.
Figure 1: The REQS. Shown in this photograph is the REQS used in this procedure. The activity monitor unit (A) rotates around its horizontal axis driven by a rotating arm (B). The rotation speed is adjustable via a dial (C), and the operation of the machine is controlled by an on/off switch (D). The inset (E) shows a close-up of the data connection between the activity monitor unit and the rotating phone jack. Please click here to view a larger version of this figure.
Figure 2: Comparison of exercise activity as measured by REQS in four different DGRP strains. X-axis: animal genotype. Y-axis: animal activity measured as beam crossings per 10 animals per 5 min. A: DGRP2 line 897; B: DGRP2 line 810; C: DGRP2 line 703; D: DGRP2 line 371. The graph shown here combines data from males and females, as there was no significant sex effect for these specific lines (ANOVA; p = 0.557). However, there is a strong genotype effect (ANOVA; p <2 x 10-16), and all individual comparisons between the four genotypes are highly significant (Tukey's HSD; p <0.05). Black diamond: genotype mean; black line: +/- one SD (standard deviation). Please click here to view a larger version of this figure.
Time | Activity(1) | Activity(2) | Activity(3) | Activity(4) | Activity(5) |
12:25:00 | 21 | 86 | 48 | 32 | 76 |
12:30:00 | 31 | 55 | 58 | 74 | 119 |
12:35:00 | 27 | 45 | 47 | 80 | 125 |
12:40:00 | 28 | 55 | 34 | 83 | 91 |
12:45:00 | 36 | 56 | 45 | 67 | 103 |
Table 1: Example output of accurate data from the DAMSystem308 software. Each "Activity(#)" column represents one vial of 10 flies, with activity recorded at 5 min intervals.
Time | Activity(1) | Activity(2) | Activity(3) | Activity(4) | Activity(5) | Fault? |
12:00:00 | 0 | 0 | 0 | 0 | 0 | yes |
12:05:00 | 98 | 1 | 36 | 0 | 8 | no |
12:10:00 | 0 | 0 | 0 | 0 | 0 | yes |
12:15:00 | 88 | 24 | 44 | 1 | 9 | no |
12:20:00 | 0 | 0 | 0 | 0 | 0 | yes |
12:25:00 | 106 | 51 | 41 | 0 | 15 | no |
Table 2: Example output of faulty data with packet loss from the DamSystem308 software. Each column represents one vial of 10 flies, with activity recorded at 5 min intervals. Rows 12:00:00, 12:10:00, and 12:20:00 show "0" activity recordings across all columns, indicating a problem with the data connection (Fault? "yes" in the last column).
Table 3: Example data used to generate Figure 2. Please click here to download this file.
Supplemental File 1: Unprocessed output file as produced by the DAMSystem308 software. Column 1 records the time point number. Column 2 records the date of the experiment, and column 3 records the time each time point is recorded. Columns 4-10 are not used, and columns 11-42 represent the recordings from the 32 slots of the activity monitor. Please click here to download this file.
As the representative results illustrate, the REQS is capable of accurately measuring the activity of exercising Drosophila. The REQS is flexible and allows researchers to address a variety of research questions related to exercise biology or exercise interventions. There are two critical steps in the protocol to highlight. First, it is essential to test the setup of the REQS to ensure that data transmission from the REQS to the DAMSystem308 is working properly. If not properly set up, the data cable can become tangled during the rotation, and sometimes the connection between the REQS and the DAMSystem308 is disrupted at the rotating connection due to wear and tear (although, with almost daily use of the REQS, we have had to replace the connector once in two years). It is prudent to keep replacement parts on hand. Second, the consistency of environmental parameters is essential to the success of the experiments. Drosophila are very sensitive to noise or vibration, and thus, it is important that the experiment is not disturbed during data collection. Thus, ideally, experiments are run in a dedicated incubator that is not accessed during an experimental run. Attention to these two critical steps will ensure that high quality data is collected from the REQS.
While we demonstrate here how the REQS can be used to measure animal activity during a 2-hour continuous exercise treatment from groups of animals, the REQS is flexible in that the time and intensity of exercise can be adjusted for a variety of exercise treatments. In addition, depending on the activity monitor unit used, it can be modified to measure activity from single flies or very large populations of animals. Additionally, the REQS can be used to test the design and optimization of exercise regimes. It can also be used to measure the effects of additional variables such as time of day, animal age, diet, population size, or drug treatment on induced activity and exercise responses. Depending on the exact experimental setup, i.e., length of exercise regime and timing, the REQS can also be used to interfere with the natural sleep patterns of Drosophila. These examples illustrate the versatile nature of the REQS and some potential uses in Drosophila research. Other small animal research communities might also be interested in adapting the REQS for their purposes, thus broadening the utility of this new tool.
Currently, one limitation of the REQS is the limited number of samples that can be processed at a given time, which is dictated by the activity monitors used, which assay a maximum of 32 samples. While the use of multiple REQS units is possible, to allow the REQS to be used for large-scale genetic screens or similar applications, development of a higher throughput version of the REQS would be ideal.
Due to its ability to measure activity levels induced in an identical manner to the Treadwheel, the REQS can be used in combination with the Treadwheel, which allows for somewhat higher throughput (48 samples can be processed at a time). Exercise protocols can be optimized using the REQS and then implemented on the Treadwheel for further studies. Thus, the Treadwheel and REQS can be used complementary as needed for specific study designs.
The REQS is an important step forward in Drosophila exercise research as it allows for quantification of the induced activity. Having a machine that can simultaneously induce exercise and measure this exercise was a clear need in the Drosophila exercise field, as a German group of scientists independently developed a very similar device, termed “swing boat,” that also utilizes an activity monitor to measure activity during exercise induced by rotation14. The “swing boat” does not use complete rotations, but instead swings the activity monitor unit back and forth approximately 30 degrees around a rotational axis. Thus, the “swing boat,” like the REQS, uses the rotation to continually induce a negative geotaxis response and increase animal activity. The REQS and “swing boat” complement existing video-tracking methods used to assay Drosophila after stimulation, such as the Drosophila ARousal Tracking system (DART)15. Both the REQS and the “swing boat” improve upon systems like the DART, which track activity only after the cessation of the stimulus. Thus, the REQS and “swing boat” are important new tools for researchers in the Drosophila exercise field, which can be used in conjunction with the existing Treadwheel and Power Tower devices.
The authors have nothing to disclose.
The work was supported by Award Number P30DK056336 from the National Institute Of Diabetes And Digestive And Kidney Diseases through a pilot grant from the Nutrition and Obesity Research Center at the University of Alabama at Birmingham to NCR.
Drosophila Activity Monitor | Trikinetics | LAM25H | REQS component |
Telephone Cord Detangler | Uvital | uv20170719 | REQS component |
Vial closures (flugs) | Genesee Scientific | 49-102 | Drosophila culture supplies |
Vials | Genesee Scientific | 32-120 | Drosophila culture supplies |
Drosophila culture netting | Carolina Biological Supply | 173090 | Drosophila culture supplies |
Cornmeal | Pepsico | 43375 | Drosophila media |
Molasses | Golden Barrel | BLA-GAL | Drosophila media |
Agar | Apex Bioresearch | 66-103 | Drosophila media |
Inactive Dry Yeast | Genesee Scientific | 62-106 | Drosophila media |
Tegosept | Apex Bioresearch | 20-258 | Drosophila media |
Propionic acid | Genesee Scientific | 20-271 | Drosophila media |